Elemental Scientific | | PC 3 Advantages Thermal stabilization of spray chamber improves long-term stability Reduces effect.

Slides:



Advertisements
Similar presentations
Mid-Atlantic Soil Testing & Plant Analysis Work Group
Advertisements

Adventures in Sample Introduction for ICP-OES and ICP-MS
A Modern Approach to Controlling ICP Spray Chamber Temperature Jerry Dulude, Glass Expansion,
FLAME SPECTROSCOPY The concentration of an element in a solution is determined by measuring the absorption, emission or fluorescence of electromagnetic.
Elemental Spectroscopy ICP-OES
Ion Sources Some characteristics of ion sources (especially in high precision work): It should have high efficiency in generating ions of the element of.
1 Atomic Absorption Spectroscopy Atomic Emission Spectroscopy Lecture 18.
AAS and FES (Ch 10, 7th e, WMDS)
Inductively Coupled Plasma Mass Spectrometry or ICP-MS
Flame photometry.
Temperature-Related Stability in Plasma Spectrometry Jerry Dulude, Ron Stux, and Vesna Dolic, Glass Expansion,
AOSC 634 Air Sampling and Analysis Lecture 1 Measurement Theory Performance Characteristics of instruments Nomenclature and static response Copyright Brock.
FLIGHT: Clock Calibration Using Fluorescent Lighting Zhenjiang Li, Wenwei Chen, Cheng Li, Mo Li, Xiang-Yang Li, Yunhao Liu Nanyang Technological University,
Photoacoustic Multi-gas Monitor , page 2 Photoacoustic Multi-gas Monitor 1312 Portable monitor for quantitative analysis of up to 5 gases and.
Dual Wavelength Isotope Ratio FS-CRDS Thinh Q. Bui California Institute of Technology ISMS 2014.
Synthesis and characterisation of thin film MAX phase alloys Mathew Guenette, Mark Tucker, Yongbai Yin, Marcela Bilek, David McKenzie Applied and Plasma.
34 S Analysis of Solids by EA-IRMS. Sample Preparation Material needs to be dry and homogenous Organisms can be analyzed whole if small enough Sample.
Atomic Absorption Spectroscopy (AAS)
Developments on the Front End of Your ICP Optimizing Performance and Reliability of ICP-AES and ICP-MS.
Chapter 5: Signals and Noise
AC Analytical Controls January 2001
Isotope Ratio Performance of an Axial Time of Flight ICP-MS Stuart Georgitis 1, Lloyd Allen 1, and Janos Fucsko 1, Frank Vanhaecke 2 1 LECO Corporation.
EPA Precursor Gas Training Workshop Kevin A Cavender EPA-Office of Air Quality, Planning and Standards Precursor Gas Monitoring NO y Monitoring Training.
QUESTIONS WELCOME ANYTIME!!!
Agilent 7500 ICP-MS Date: 28JUL2005.
Atomic Emission Spectrometry
Mass spectrometry Ions are analyzed on the basis of their m/z Chlorine has 2 isotopes, 35 Cl and 37 Cl, in the approximate ratio of 3 :1. Electrons are.
AAS Atomic Absorption Spectrophotometry. AAS – Widely in clinical laboratories to measure elements such as aluminum, calcium, copper, lead, lithium, magnesium,
Define Problem Select Appropriate Methods Obtain and store sample Pre-treat sample Perform required measurements Compare results with standards Apply necessary.
Advantages of a Syringe Driven Sample Introduction Accessory for ICP Spectrometry Jerry Dulude and Scott Bridger Glass Expansion.
Comparison of Competitors’ Evaporative Light Scattering Detectors
1 Atomic Emission Spectroscopy Lecture Advantages of Plasma Sources 1.No oxide formation as a result of two factors including Very high temperature.
Effect of Sample Introduction on ICP-MS Detection of Nonmetal Analytes in a Simulated HPLC Matrix Adam E. Dill and Jon W. Carnahan Department of Chemistry.
Discovering a New Choice in ICP- MS Low Volume Handling MVX-7100 µL Workstation.
1 Introduction to Atomic Spectroscopy Atomic Absorption Spectroscopy Lecture 12.
A UNIVERSAL PROGRAMMABLE TEMPERATURE SPRAY CHAMBER Jerry Dulude, Glass Expansion,
Adv. Inst. Techs.  flame emission (eg flame photometer) known as low temperature emission ( K) ◦ first form of spectroscopy ◦ used in commercial.
Printed by Techniques for Increasing Productivity in ICP Spectrometry Jerry Dulude, Glass Expansion, Inc., 4 Barlows Landing Rd.,
ATOMIC SPECTROMETRY 1. Flames 2. Electrothermal Atomizers 3. Plasmas.
High Precision Pb Isotope Ratio Measurements of Zircon Zenon Palacz, Simon Meffan-Main, Patrick Turner, & Fadi Abou-Shakra.
SPRAY CHAMBER TEMPERATURE: A CRITICAL (YET OFTEN MISSING) PARAMETER IN ICP EXPERIMENTS Jerry Dulude, Ron Stux, and Vesna Dolic, Glass Expansion,
PL-ELS 2100 Ice. PL-ELS 2100 Ice Evaporation at Sub Ambient Temperatures The PL-ELS 2100 Ice comprises an Integrated Cooled Evaporator in combination.
Advanced Analytical Chemistry – CHM 6157® Y. CAIFlorida International University Updated on 9/26/2006Chapter 3ICPMS Interference equations Isobaric.
Eliminating Molecular Interferences in ICP-MS for the Simplest Method Development Dr. Torsten Lindemann.
Innovations in ICP Sample Introduction
High Accuracy and High Precision Uranium Isotopic Analysis by DRC-ICP-MS Steve Beres July 2008.
Printed by A New Versatile Programmable Temperature Spray Chamber for ICP Jerry Dulude and Ron Stux (USA), Vesna Dolic (Australia),
ATOMIC ABSORPTION SPECTROSCOPY (AAS) Atomization: It is the conversion of molecules to their component atoms in gaseous state using a source of heat (flame).
An evaluation of the isotope ratio performance of an axial TOF - ICP - mass spectrometer F. Vanhaecke, L. Moens, R. Dams, L. Allen* and S. Georgitis* University.
Plasma A Plasma consists of a collection of free-moving electrons and ions and is very hot. Energy must be continually applied to sustain the plasma.
Atomic Absorption Spectroscopy
For the past several years, one of the primary focuses of our research group has been the development of novel ion-exchange systems for the purpose of.
Chapter 1: Introduction 1. Type of instrumental methods Radiation Electrical method Thermal properties Others 2. Instruments for analysis Non-electrical.
Status of the PSD upgrade - Status of the PSD cooling and temperature stabilization system - MAPD gain monitoring system - PSD readout upgrade F.Guber,
Abstract Analysis of organic solvents by ICP is often seen as problematic and challenging due to the nature of the solvents used. Commonly encountered.
Thermal Conductivity Detector
CMS CO 2 Cooling Test Stand Capacity and Load Disruption Tests Simon Kwan Erik Voirin Simon Kwan, Erik Voirin1CMS CO2 Cooling Test Stand Test Results.
1 In-house Training on ICP-MS. ICP-MS??? Inductively Coupled Plasma Mass Spectrometry or ICP-MS is an analytical technique used for elemental determinations.
The “Monitor to measure the Integral TRAnsmittance” (MITRA)
Satish Pradhan Dnyanasadhana College Department Of Chemistry
Flame Emission Spectrometry
Chapter 5: Signals and Noise
Signals and Noise Signal to Noise Ratio Types of Noise
The Analysis of Soils and Waters in Accordance with U. S
ICP-MS ICP – Inductively Coupled Plasma MS – Mass Spectrometry
Dnyanasadhana College, Thane. Department of Chemistry T. Y. B. Sc
Satish Pradhan Dnyanasadhana College Department Of Chemistry
3a) Load or Verify IonSource & Tune File
FLAME SPECTROSCOPY The concentration of an element in a solution is determined by measuring the absorption, emission or fluorescence of electromagnetic.
Inductively Coupled Plasma
Presentation transcript:

Elemental Scientific | | PC 3 Advantages Thermal stabilization of spray chamber improves long-term stability Reduces effect of laboratory temperature variation on signal stability Reduces solvent loading. Fast rinse-out using PFA-ST nebulizer and O-ring-free quartz cyclonic spray chamber. Reduced oxides Enhances stability with PC3-SSI Chamber: Isotope ratio measurements Other analyses requiring improved precision Additional reduction of oxides vs. standard PC3 Optional oxygen addition port for organic solvents.

Elemental Scientific | | PC 3 Features Default settings of spray chamber temperature (+2 C/ -5 C) for aqueous or organic solvents. Custom temperature settings available upon request. Quartz cyclonic or HF resistant cyclonic spray chambers available. Direct connection to torch injector. Completely o-ring-free for organic solvents analysis. Organic solvent analysis using PFA-50 nebulizer. Aqueous sample analysis using PFA-ST nebulizer. PC3-SSI model for isotope ratio determination or other high precision analyses. Optional mounting tray for x-y-z torch adjustment. Air-cooled.

Elemental Scientific | | ELEMENT2 with Standard Scott Spray Chamber Five Orders of Magnitude Rinse-out > 3 minutes PFA-ST Self-aspirating Nebulizer 10 ppb ~ 0.2 ppt Log intensity (cps) Time (min)

Elemental Scientific | | ELEMENT2 with PC 3 Inlet System Five Orders of Magnitude Rinse-out < 30s PFA-ST Self-aspirating Nebulizer 10 ppb ~ 0.1 ppt Log intensity (cps) Time (min) ppb ~ 0.1 ppt Log intensity (cps)

Elemental Scientific | | PC3 Peltier Cooled Cyclonic Spray Chamber 4 Hour Stability 1 ppb Tuning Solution ThermoFinnigan ELEMENT2 ICPMS Time (hour) Integrated Signal Intensity (cps) 7Li 1.8% RSD 59Co 1.5% RSD 115In 1.0% RSD 138Ba 1.2% RSD 238U 1.0% RSD

Elemental Scientific | | PC3-SSI Peltier-cooled Stable Sample Introduction Low flow PFA-50 nebulizer Homogenizes sample aerosol Reduces interferences Reduces drift Improves isotope ratio precision

Elemental Scientific | | Short-Term (20 min) Signal Stability Scott vs. PC3 vs. PC3-SSI Isotope Standard Inlet PC3PC3-SSI Co %0.92%0.53% In-II51.32%1.08%0.40% Ba %0.90%0.48% U %0.93%0.34% n=10, 1 min read, 1 min delay, LR 10% mass window

Elemental Scientific | | PC3-SSI Short-term Precision

Elemental Scientific | | Long-Term (14 h) Signal Stability Scott vs. PC3-SSI Isotope Standard Inlet PC3-SSI Co %0.95% In-II5 3.87%1.05% Ba %1.03% U %1.08% n=102, 3 min read, 5 min delay, LR 10% mass window

Elemental Scientific | | Pb Isotope Ratio Precision 5 ms dwell time PC3-SSI Sample Inlet System withELEMENT2 ICPMS 0.000% 0.020% 0.040% 0.060% 0.080% 0.100% 0.120% 0.140% 0.160% 0.180% Number of Passes % RSD 204/ / / ppb Pb

Elemental Scientific | | Summary: PC3 and PC3-SSI with Finnigan ELEMENT Rinse-out improved by 2 minutes Scott-type chamber. Sensitivity enhancement up to 2x or more depending upon nebulizer flow rate. Reduced metal-oxide formation. SSI version for improved short-term and long-term signal stability. Convenient system for organic solvent analysis.