Computed Tomography II

Slides:



Advertisements
Similar presentations
IMAGE GENERATION IN CT.
Advertisements

CT Scanning: Dosimetry and Artefacts
Image Reconstruction.
IMAGE QUALITY.
Computed Tomography Physics, Instrumentation, and Imaging
Digital Radiography.
Advanced Biomedical Imaging Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University Lecture 6 Basic physical principles of.
Spiral CT Bushong Chapter 5.
Seeram Chapter 13: Single Slice Spiral - Helical CT
RAD309 Patient Dose.
SCANCOMEDICAL Computed Tomography SCANCO User Meeting 2005 Dr. Bruno Koller SCANCO Medical AG
Image reconstruction and analysis for X-ray computed microtomography Lucia Mancini 1, Francesco Montanari 2, Diego Dreossi 3 1 Elettra - Trieste 2 A.R.P.A.
CT Multi-Slice CT.
IMAGE QUALITY NOISE LINEARITY CROSS-FIELD UNIFORMITY IMAGE ARTIFACTS.
CT Physics V.G.Wimalasena Principal School of radiography.
Computed Tomography III
Remote sensing in meteorology
Computed Tomography RAD309
tomos = slice, graphein = to write
Seeram Chapter 5: Data Acquisition in CT
Outline (HIBP) diagnostics in the MST-RFP Relationship of equilibrium potential measurements with plasma parameters Simulation with a finite-sized beam.
Chapter 2 Stewart C. Bushong
Planar scintigraphy produces two-dimensional images of three dimensional objects. It is handicapped by the superposition of active and nonactive layers.
Conventional and Computed Tomography
DETECTORS AND DETECTOR ARRAYS
…….CT Physics - Continued V.G.WimalasenaPrincipal School of radiography.
LEC ( 2 ) RAD 323. Reconstruction techniques dates back to (1917), when scientist (Radon) developed mathematical solutions to the problem of reconstructing.
Seeram Chapter 11: Image Quality
Basic principles Geometry and historical development
ADC PRE-PATIENT COLLIMATION POST-PATIENT COLLIMATION.
COMPUTED TOMOGRAPHY I – RAD 365 CT - Scan
Digital Image Characteristic
Medical Image Analysis Medical Imaging Modalities: X-Ray Imaging Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan,
A-LEVEL PHYSICS 15/09/ :49 Medical Physics.
Module B Computed Tomography Physics, Instrumentation, and Imaging.
Image Quality Radiographic Resolution.
Seeram Chapter 7: Image Reconstruction
CT Physics Lecture 3.
Computed Tomography Q & A
PRINCIPLE AND BASIC PHYSICS OF COMPUTED TOMOGRAPHY
CSci 6971: Image Registration Lecture 3: Images and Transformations March 1, 2005 Prof. Charlene Tsai.
Computed Tomography Physics, Instrumentation, and Imaging
CT Instrumentation and X-ray system
Factors affecting CT image RAD
IAEA International Atomic Energy Agency Optimization of Protection in Computed Tomography (CT)-What can radiographers do? IAEA Regional Training Course.
Module C Computed Tomography Physics, Instrumentation, and Imaging.
Image Quality The capacity to define, measure, and assess image quality is a primarily responsibility of a CT Technologist.
The production of X-Rays X-Rays are produced whenever charged particles are slowed down or stopped. X-rays are generated via interactions of the accelerated.
Advanced CT systems and Their Performance
CS TC 22 CT Basics CT Principle Preprocessing. 2 CT Basics CT principle preprocessing CS TC 22 Blockdiagram image processor.
Image Display. But first a review Remember the 3 main steps 1. Data Acquisition 2. Image Reconstruction 3. Image Display.
Part No...., Module No....Lesson No
Part No...., Module No....Lesson No
Quality Assurance.
14FEB2005/KWCAE2-UsersGroup Astro-E2 X-Ray Telescopes XRT Setup & Structure Performance Characteristics –Effective Area –Angular Resolution –Optical Axes.
Computed Tomography Computed Tomography is the most significant development in radiology in the past 40 years. MRI and Ultrasound are also significant.
ACR CT Accreditation Dose Pass/Fail Criteria and Reference Levels ExaminationPass/Fail Criteria CTDIvol (mGy) Reference Levels CTDIvol (mGy) Adult Head8075.
Computed tomography. Formation of a CT image Data acquisitionImage reconstruction Image display, manipulation Storage, communication And recording.
IMAGE QUALITY. SPATIAL RESOLUTION CONTRAST RESOLUTION NOISE IMAGE ARTIFACTS RADIATION DOSE.
Diagnostic Radiology II X-ray Tubes. Anode angle Anode angle defined as the angle of the target surface with respect to the central ray in the x-ray field.
Single Slice Spiral - Helical CT
CT Multi-Slice CT.
Computed Tomography Basics
Computed Tomography Data Acquisition
DETECTORS AND DETECTOR ARRAYS. Xenon Detectors Xenon detectors use high-pressure (about 25 arm) nonradioactive xenon gas, in long thin cells between two.
بسم الله الرحمن الرحیم ..
Basic principles Geometry and historical development
Remote sensing in meteorology
Computed Tomography (C.T)
Computed Tomography (C.T)
Presentation transcript:

Computed Tomography II Detectors and detector arrays Details of acquisition Tomographic reconstruction

Detectors Xenon detectors Solid-state detectors Multiple detector arrays

Xenon detectors Use high-pressure (about 25 atm) nonradioactive xenon gas, in long thin cells between two metal plates Very thick (e.g., 6 cm) to compensate in part for relatively low density Thin metal septa separating individual detectors improves geometric efficiency by reducing dead space between detectors

Xenon detectors (cont.) Long, thin plates are highly directional Must be positioned in a fixed orientation with respect to the x-ray source Cannot be used for 4th generation scanners because those detectors must record x-rays as the source moves over a wide angle

Solid-state detectors Composed of a scintillator coupled tightly to a photodetector (typically a photodiode) Scintillator emits visible light when an x-ray is absorbed, similar to an x-ray intensifying screen Photodetector converts light intensity into an electrical signal proportional to the light intensity

Solid-state detectors (cont.) Detector size typically 1.0 x 15 mm (or 1.0 x 1.5 mm for multiple detector arrays) Scintillators used include CdWO4 and yttrium and gadolinium ceramics Better absorption efficiency than gas detectors because of higher density and higher effective atomic number

Solid-state detectors (cont.) To reduce crosstalk between adjacent detector elements, a small gap between detector elements is necessary, reducing geometric efficiency somewhat Top surface of detector is essentially flat and therefore capable of x-ray detection over a wide range of angles Required for 4th generation scanners and used in most high-tier 3rd generation scanners as well

Multiple detector arrays Set of several linear detector arrays, tightly abutted Use solid-state detector arrays Slice width is determined by the detectors, not by the collimator (although collimator does limit the beam to the total slice thickness)

Multiple detector arrays (cont.) 3rd generation multiple detector array with 16 detectors in the slice thickness dimension and 750 detectors along each array uses 12,000 individual detector elements 4th generation scanner would require roughly 6 times as many detector elements; consequently currently planned systems use 3rd generation geometry

Slice thickness: single detector array scanners Determined by the physical collimation of the incident x-ray beam with two lead jaws Width of the detectors places an upper limit on slice thickness For scans performed at the same kV and mAs, the number of detected x-ray photons increases linearly with slice thickness Larger slice thicknesses yield better contrast resolution (higher SNR), but the spatial resolution in the slice thickness dimension is reduced

Slice sensitivity profile For single detector array scanners, the shape of the slice sensitivity profile is a consequence of: Finite width of the x-ray focal spot Penumbra of the collimator The fact that the image is computed from a number of projection angles encircling the patient Other minor factors Helical scans have a slightly broader slice sensitivity profile due to translation of the patient during the scan

Slice thickness: multiple detector array scanners In axial scanning (i.e., with no table movement) where, for example, four detector arrays are used, the width of the two center detector arrays almost completely dictates the thickness of the slices For the two slices at the edges of the scan, the inner side of the slice is determined by the edge of the detector, but the outer edge is determined either by the outer edge of the detector or by the collimator penumbra, depending on collimator adjustment

Slice thickness: MDA (cont.) In helical mode, each detector array contributes to every reconstructed image Slice sensitivity profile for each detector array needs to be similar to reduce artifacts Typical to adjust the collimation so that the focal spot – collimator blade penumbra falls outside the edge detectors Causes radiation dose to be a bit higher (especially for small slice widths) Reduces artifacts by equalizing the slice sensitivity profiles between the detector arrays

Detector pitch/collimator pitch Pitch is a parameter that comes into play when helical scan protocols are used In a helical scanner with one detector array, the pitch is determined by the collimator Collimator pitch = table movement (mm) per 360-degree rotation of gantry / collimator width (mm) at isocenter Pitch may range from 0.75 (overscanning) to 1.5 (faster scan time, possibly smaller volume of contrast agent)

Pitch (cont.) For scanners with multiple detector arrays, collimator pitch is still valid Detector pitch = table movement (mm) per 360-degree rotation of gantry / detector width (mm) For a multiple detector array scanner with N detector arrays, collimator pitch = detector pitch / N For scanners with four detector arrays, detector pitches running from 3 to 6 are used

Tomographic reconstruction Rays and views: the sinogram Preprocessing the data Interpolation (helical)

Sinogram Display of raw data acquired for one CT slice before reconstruction Rays are plotted horizontally and views are shown on the vertical axis Objects close to the edge of the FOV produce a sinusoid of high amplitude Bad detector in a 3rd generation scanner would show up as a vertical line on the sinogram

Rays and views 1st and 2nd generation scanners used 28,800 and 324,000 data points, respectively State-of-the-art scanner may aquire about 800,000 data points Modern 512 x 512 circular CT image contains about 205,000 image pixels Number of rays affects the radial component of spatial resolution; number of views affects the circumferential component of the resolution

Number of rays CT images of a simulated object reconstructed with differing numbers of rays show that reducing the ray sampling results in low-resolution, blurred images

Number of views CT images of the simulated object reconstructed with differing numbers of views show the effect of using too few angular views (view aliasing) Sharp edges (high spatial frequencies) produce radiating artifacts that become more apparent near the periphery of the image

Preprocessing Calibration data determined from air scans (performed by the technologist or service engineer periodically) provide correction data that are used to adjust the electronic gain of each detector Variation in geometric efficiencies caused by imperfect detector alignments is also corrected

Interpolation CT reconstruction algorithms assume that the x-ray source has negotiated a circular, not helical, path around the patient Before the actual CT reconstruction, the helical data set is interpolated into a series of planar image sets With helical scanning, CT images can be reconstructed at any position along the length of the scan

Interpolation (cont.) Interleaved reconstruction allows the placement of additional images along the patient, so that the clinical examination is almost uniformly sensitive to subtle abnormalities Adds no additional dose to the patient, but additional time is required to reconstruct the images Actual spatial resolution along the long axis of the patient still dictated by slice thickness