PhD Course Biosensors – fundamentals & applications

Slides:



Advertisements
Similar presentations
OFFLINE COMPOSITION MEASURING SENSORS
Advertisements

15-1 Potentiometry Potential measurements of electrochemical cells Ion selective methods §Reference electrode §Indicator electrode §Potential measuring.
EEE 529 Microsystems Amperometric Biosensors
Molecular Biomimetics Polypeptides to Inorganic structures.
PH 0101 Unit-5 Lecture-61 Introduction A fuel cell configuration Types of fuel cell Principle, construction and working Advantage, disadvantage and application.
Summary of Potentiometry:
Reactions in Aqueous Media
Electrode measuring principle
MLAB 2401: C LINICAL C HEMISTRY K ERI B ROPHY -M ARTINEZ Designs in Instrumentation Electrochemistry.
Dr. Marc Madou Class IV. Microfabrication of electrochemical sensors Winter 2011 BIOMEMS.
Electrodes and Potentiometry
14-6 Cells as chemical probes
Potentiometry. pH is a Unit of Measurement  pH = Power of Hydrogen (H + )  Defined as the Negative Logarithm of Hydrogen Ion Activity  pH = log (1/H.
ELECTROCHEMISTRY. During electrolysis positive ions (cations) move to negatively charged electrode (catode) and negative ions (anions) to positively charged.
Fabrication of a Microelectrode Array Biosensor Based on a Modified Enzyme-Chitosan Biocomposite Lorenzo D’Amico October 1, 2008.
Dr. Marc Madou, UCI, Winter 2012 Class V Potentiometric and Amperometric Sensors (I) Electrochemistry MAE-295.
Fall 2014, Prof. JB Lee Ion Sensitive FET (ISFET) - What and Why?
Department of Electrical and computer Eng.
Potentiometric Methods
Amperometric Biosensors. Introduction Enzyme Catalyzed redox reactions The function of the enzyme is to generate or consume an electroactive species in.
Molecular Fluorescence Spectroscopy
Nanobiosensors Sara Huefner November 6, Outline Biosensor Background –What is a Biosensor? –Components of a Biosensor –Principles of Detection Biosensors.
ANALYTICAL CHEMISTRY CHEM 3811 CHAPTER 15 DR. AUGUSTINE OFORI AGYEMAN Assistant professor of chemistry Department of natural sciences Clayton state university.
Electroanalytical Chemistry
POTENTIOMETRY 8th lecture
Electroanalysis measure the variation of an electrical parameter (potential, current, charge, conductivity) and relate this to a chemical parameter (the.
Membrane Transport Chapter 6.
Chapter 26. An electrochemical cell A device that converts chemical energy into electrical energy. A Daniell cell is a device that could supply a useful.
Chapter 3 Enzymes.
Using and Controlling Reactions Assign oxidation numbers and balance atom whose oxidation number changes 2. Balance oxygen by adding water 3. Balance.
Topics covered Techniques of enzyme immobilization
Potentiometry and potentiometric measurements. potentiometer A device for measuring the potential of an electrochemical cell without drawing a current.
Presented by: Jay Leitch. What are Glucose Biosensor?   Immobilized biological components fixed on ordinary analytical detectors   Selectively detects.
An Introduction to Electroanalytical Chemistry Electrochemistry: The study of the interchange of chemical and electrical energy Oxidation is the loss of.
BioSensors Yang Yang 9/28/2004. Outlines BioMEMS Enzyme-coated carbon nanotubes Microcantilever biosensor with environmentally responsive hydrogel Cantilever.
The Optimum Operating Conditions for Detection of Hydrogen Peroxide for the Carbon Paste Electrode Modified with Ferrocene Pao-Tsai Kuo ( 郭寶財 ), Chung-Min.
6/2/20161 CHAPTER 2 CORROSION PRINCIPLES Chapter Outlines 2.1 Oxidation and Reduction Reactions 2.2 Standard Electrode Half- Cell Potentials 2.3 Standard.
Electroanalysis measure the variation of an electrical parameter (potential, current, charge, conductivity) and relate this to a chemical parameter (the.
Electrochemical Methods Dr M.Afroz Bakht. Potentiometry Potentiometry is a method of analysis used in the determination of concentration of ions or substances.
Polarization.
Matter – anything that takes up space and has weight; composed of elements Elements – composed of chemically identical atoms as of 2002, 114 elements known,
Electrochemistry.
Section 13 Analytical Voltammetry.
Chapter 16.  the chemical principles, half-equations and overall equations of simple electrolytic cells; comparison of electrolytic cells using molten.
Electrodes and Potentiometry Introduction 1.)Potentiometry  Use of Electrodes to Measure Voltages that Provide Chemical Information - Various electrodes.
Electrochemical cells - batteries
Gas sensing Panca Mudji Rahardjo, ST.MT Electrical Engineering - UB.
Commercial Use of Enzymes
Ch. 21 Potentiometry 1. General principles
POTENTIALS AND THERMODYNAMICS OF CELLS (1) POTENTIALS AND THERMODYNAMICS OF CELLS (1)
Biochemical Reactions Chapter 1.3 McGraw-Hill Ryerson Biology 12 (2011)
Biosensors Dr. Bhavesh Patel Principal
In the name of GOD.
SCI3023 ELECTROCHEMISTRY Chapter 8e: Potentiometry
BIOELECTRONICS 1 Lec3: BIOSENSOR Dr. Eng. Hani Kasban Mahmoud 2017
© SSER Ltd..
ELECTRO-ANALYTICAL TECHNIQUES   POTENTIOMETRIC TECHNIQUES Principle Electrochemical techniques/ Electrochemical methods are generally used for determination.
Electrochemistry MAE-212
Potentiometry and potentiometric measurements
Biosensors Definition
Ion Selective Electrode
Potentiometry Potential measurements of electrochemical cells
Chapter 5: Analytic Techniques
BIOELECTRONICS 1 Lec4: Calorimetric Biosensor
Enzymes Regulatory enzymes are usually the enzymes that are the rate-limiting, or committed step, in a pathway, meaning that after this step a particular.
Chapter 23 Potentiometry.
Chapter Fifteen VOLTAMMETRY AND ELECTROCHEMICAL SENSORS
Rayat Shikshan Sanstha’s S. M. Joshi College, Hadapsar
Presentation transcript:

PhD Course Biosensors – fundamentals & applications TOPICS IN (NANO) BIOTECHNOLOGY Biosensors – fundamentals & applications June 30th 2003 PhD Course

DEFINITION A biosensor may be defined as a device incorporating a biologically active component in intimate contact with a physico-chemical transducer and an electronic signal processor.

Signal Analyte of interest Interfering species Biocomponent Transducer Processor

So what is an biosensor?

HISTORY 1962, Clark and Lyons 1967, Updike and Hicks 1969, Guilbault 1972, Reitnauer 1975, Yellow Springs Instrument 1975, Janata 1979, Danielsson

BIOCOMPONENTS Enzymes Antibodies Membranes Organelles Cells Tissues Cofactors

TRANSDUCERS Electrochemical Optical Piezo-electric Calorimetric Acoustic

BIOSENSOR TYPES Enzyme/metabolic biosensors Bioaffinity sensors Enzyme and cell electrodes Bioaffinity sensors Antibodies Nucleic acids Lectin

Enzyme/Metabolic Sensors Enzymes are biological catalysts. There are five main classes of enzymes. Oxidoreductases Transferases Hydrolases Lyases Isomerases

Oxidoreductases Dehydrogenases Oxidases Peroxidases Oxygenases

Enzyme/Metabolic Sensors Substrate + Enzyme Substrate-enzyme complex Product + Enzyme Substrate consumption/product liberation is measured and converted into quantifiable signal.

Bioaffinity Sensors These sensors are based on binding interactions between the immobilised biomolecule and the analyte of interest. These interactions are highly selective. Examples include antibody-antigen interactions, nucleic acid for complementary sequences and lectin for sugar.

Analyte of interest (antigen) Antibody Interfering species Antibody-antigen complex

Transducers Electrochemical Potentiometric Amperometric Conductimetric

POTENTIOMETRIC BIOSENSORS In potentiometric sensors, the zero-current potential (relative to a reference) developed at a selective membrane or electrode surface in contact with a sample solution is related to analyte concentration. The main use of potentiometric transducers in biosensors is as a pH electrode.

POTENTIOMETRIC BIOSENSORS E = Eo + RT/nF ln[analyte] Eo is a constant for the system R is the universal gas constant T is the absolute temperature z is the charge number F is the Faraday number ln[analyte] is the natural logarithm of the analyte activity.

POTENTIOMETRIC BIOSENSORS The best known potentiometric sensor is the Ion Selective Electrode (ISE). Solvent polymeric membrane electrodes are commercially available and routinely used for the selective detection of several ions such as K+, Na+, Ca2+, NH4+, H+, CO32-) in complex biological matrices. The antibiotics nonactin and valinomycin serve as neutral carriers for the determination of NH4+ and K+, respectively.

Ag/AgCl reference electrode Internal aqueous filling solution Liquid ion exchanger Membrane/salt bridge Porous membrane containing ionophore

POTENTIOMETRIC BIOSENSORS ISEs used in conjunction with immobilised enzymes can serve as the basis of electrodes that are selective for specific enzyme substrates. The two main ones are for urea and creatinine. These potentiometric enzyme electrodes are produced by entrapment the enzymes urease and creatinase, on the surface of a cation sensitive (NH4+) ISE.

POTENTIOMETRIC BIOSENSORS urease Urea + H2O + H+ 2NH4+ + HCO3- creatininase Creatinine + H2O N-methylhydantoin + NH4+ penicillinase Penicillin Penicillonic Acid In contact with pH electrode.

AMPEROMETRIC BIOSENSORS With amperometric sensors, the electrode potential is maintained at a constant level sufficient for oxidation or reduction of the species of interest (or a substance electrochemically coupled to it). The current that flows is proportional to the analyte concentration. Id = nFADsC/d

incorporating electrolyte (e.g. KCl, NaCl) Auxiliary Electrode (e.g. Pt wire) Working Electrode Reference Electrode (e.g. Pt, Au, C) (e.g. Ag/AgCl, SCE) Buffer solution (e.g. Tris, DPBS, Citrate) incorporating electrolyte (e.g. KCl, NaCl) e flow Stirbar

Example Glucose + O2 Gluconic Acid + H2O2 Oxidase The product, H2O2, is oxidised at +650mV vs a Ag/AgCl reference electrode. Thus, a potential of +650mV is applied and the oxidation of H2O2 measured. This current is directly proportional to the concentration of glucose.

I (nA) 150 100 50 5 10 15 20 [Glucose], mM

AMPEROMETRIC BIOSENSORS Amperometric enzyme electrodes based on oxidases in combination with hydrogen peroxide indicating electrodes have become most common among biosensors. With these reactions, the consumption of oxygen or the production of hydrogen peroxide may be monitored. The first biosensor developed was based on the use of an oxygen electrode.

Clark Oxygen Electrode - + Electrode body Silver anode KCl soln. Polyethylene membrane Platinum cathode

AMPEROMETRIC BIOSENSORS The drawback of oxygen sensors is that they are very prone to interferences from exogenous oxygen. H2O2 is more commonly monitored. It is oxidised at +650mV vs. a Ag/AgCl reference electrode. At the applied potential of anodic H2O2 oxidation, however, various organic compounds (e.g. ascorbic acid, uric acid, glutathione, acetaminophen ...) are co-oxidised.

AMPEROMETRIC BIOSENSORS Various approaches have been taken to increase the selectivity of the detecting electrode by chemically modifying it by the use of: membranes mediators metallised electrodes polymers

AMPEROMETRIC BIOSENSORS 1. Membranes. Various permselective membranes have been developed which controlled species reaching the electrode on the basis of charge and size. Examples include cellulose acetate (charge and size), Nafion (charge) and polycarbonate (size). The disadvantage of using membranes is, however, their effect on diffusion.

AMPEROMETRIC BIOSENSORS 2. Mediators Many oxidase enzymes can utilise artificial electron acceptor molecules, called mediators. A mediator is a low molecular weight redox couple which can transfer electrons from the active site of the enzyme to the surface of the electrode, thereby establishing electrical contact between the two. These mediators have a wide range of structures and hence properties, including a range of redox potentials.

AMPEROMETRIC BIOSENSORS

AMPEROMETRIC BIOSENSORS Examples of mediators commonly used are: Ferrocene (insoluble) Ferrocene dicarboxylic acid (soluble) Dichloro-indophenol (DCIP) Tetramethylphenylenediamine (TMPD) Ferricyanide Ruthenium chloride Methylene Blue (MB)

AMPEROMETRIC BIOSENSORS 3. Metallised electrodes The purpose of using metallised electrodes is to create conditions in which the oxidation of enzymatically generated H2O2 can be achieved at a lower applied potential, by creating a highly catalytic surface. In addition to reducing the effect of interferents, due to the lower applied potential, the signal-to-noise ratio is increased due to an increased electrochemically active area.

AMPEROMETRIC BIOSENSORS Metallisation is achieved by electrodepositing the relevant noble metal onto a glassy carbon electrode using cyclic voltammetry. Successful results have been obtained from a few noble metals - platinum, palladium, rhodium and ruthenium being the most promising.

Glassy carbon electrode Metallised GCE Response Potential Potential Response Metallised GCE Glassy carbon electrodes do not catalyse the oxidation of hydrogen peroxide. GCEs metallised with ruthenium, rhodium, palladium or platinum do.

AMPEROMETRIC BIOSENSORS 4. Polymers As with membranes, polymers are used to prevent interfering species from reaching the electrode surface. Polymers differentiate on the basis of size and charge. An example is that of polypyrrole. A polypyrrole film has to be in the reduced state to become permeable for anions. If the film is oxidised, no anion can permeate.

AMPEROMETRIC BIOSENSORS Examples of commonly used polymers are: polypyrrole polythiophene polyaniline diaminobenzene polyphenol

Electrochemical Transducers 3. Conductimetric Conductimetric methods use non-Faradaic currents. In conductimetric transducers the two electrodes (working and reference) are separated from the measuring solution by a gas-permeable membrane. The measured signal reflects the migration of all ions in the solution. It is therefore non-specific and may only be used for samples of identical conductivity.

K+ K+ K+ A- A-

OPTICAL BIOSENSORS The area of biosensors using optical detection has developed greatly over the last number of years due mainly to the inherent advantages of optical systems. The basis of these systems is that enzymatic reactions alter the optical properties of some substances allowing them to emit light upon illumination. Means of optical detection include fluorescence, phosphorescence, chemi/bioluminescence...

OPTICAL BIOSENSORS Advantages of optical biosensors include due to fibre optics, miniaturisation is possible in situ measurements are possible in vivo measurements are possible diode arrays allow for multi-analyte detection signal is not prone to electromagnetic interference

OPTICAL BIOSENSORS Disadvantages include: ambient light is a strong interferent fibres are very expensive indicator phases may be washed out with time

OPTICAL BIOSENSORS Fibre optics are a subclass of optical waveguides which operate using the principle of total internal reflection. Light incident on the interface between two dielectric media will be either reflected or refracted according to Snell’s Law.

OPTICAL BIOSENSORS A Cladding Core B B Total Internal Reflection

OPTICAL BIOSENSORS If light is entered into a fibre (surrounded by a medium of lower refractive index) at a shallow enough angle, the light will be confined within. Thus, the optical fibres consist of a core of high refractive index surrounded by a cladding of slightly lower refractive index, with the whole fibre protected by a non-optical jacket.

OPTICAL BIOSENSORS Jacket qmax Cladding Core Cladding Jacket

OPTICAL BIOSENSORS Light input, and hence output, is dependent on the diameter of the fibre. As a very small diameter is required for flexible fibres, this size is a limiting factor in the fabrication of the fibres. For this reason, fibres are made from bundles which have the advantage of efficient light collection and flexibility. Fibre bundles of 8, 16 and more fibre strands are available.

OPTICAL BIOSENSORS Generally, fibre-optic based biosensors employ fluorescence or chemiluminescence as the light medium. This is due to the fact that fluorescence is intrinsically more sensitive than absorbance. It is also more flexible due to the fact that a great variety of analytes and influences are known to change the emission of particular fluorophores.

OPTICAL BIOSENSORS There are basically two different configurations used at the tip of the fibre-optic probe the distal cuvette configuration waveguide binding configuration

OPTICAL BIOSENSORS The distal cuvette configuration involves immobilisation of detection molecules in a porous, transparent medium at the fibre tip. The fluorescence changes when the analyte diffuses and is bound. Excitation comes from out of the fibre and emission is coupled back into the fibre.

OPTICAL BIOSENSORS Fibre-optic probe Distal cuvette Detection molecule Analyte

OPTICAL BIOSENSORS The waveguide binding tip configuration involves the binding of fluorescent-labelled detector molecules (e.g. antibodies) to covalently attached analyte molecules on the fibre surface. As the label is close to the surface it is excited by the evanescent wave emanating from the fibre and the resulting fluorescence is coupled back into the fibre. Free analyte competes for the binding sites on the recognition molecules, permitting them to diffuse away from the surface with a resultant decrease in fluorescence.

OPTICAL BIOSENSORS Fluorescent-labelled recognition molecule Free antigen Immobilised analyte

Piezoelectric Transducers The principle of this sensor type is based on the discovery of there being a linear relationship between the change in the oscillating frequency of a piezoelectric (PZ) crystal and the mass variation on its surface. Sauerbrey discovered in 1959 that the change in mass is inversely proportional to the change in frequency of the resonating crystal (usually at MHz frequencies).

Piezoelectric Transducers DF = -2.3x106F2DM/A (Sauerbrey equation) The change in mass occurs when the analyte interacts specifically with a biospecific agent immobilised on the crystal surface. The crystal may be coated with antibodies, enzymes or organic materials. Frequency changes smaller than 1MHz may be measured providing nanogram sensitivity.

Quartz wafer Gold Electrical contacts

SAW Transducers Surface acoustic wave (SAW) devices operate by the propogation of acoustoelectric waves, either along the surface of the crystal or through a combination of bulk and surface. The oscillation of the crystal in SAW devices is greater, by at least a factor of ten, than the oscillation of the crystal used in PZ devices.

Calorimetric Transducers Enzyme-catalysed reactions exhibit the same enthalpy changes as spontaneous chemical reactions. Considerable heat evolution is noted (5-100kJ/mol). Thus, calorimetric transducers are universally applicable in enzyme sensors.

Calorimetric Transducers The thermal biosensors constructed have been based on: direct attachment of the immobilised enzyme or cell to a thermistor Immobilisation of the enzyme in a column in which the thermistor has been embedded. DT = nDH/cp

Enzyme Substrate -DH (kJ/mol) Catalase Hydrogen peroxide 100.4 Cholesterol oxidase Cholesterol 52.9 Glucose oxidase Glucose 80.0 Hexokinase Glucose 27.6 Lactic dehydrogenase Pyruvate 62.1 b - Lactamase Penicillin G 67.0 Urease Urea 6.6 Uricase Uric acid 49.1

polyurethane insulation Aluminium block Recorder Bridge/Amplifier Sample Thermistor Buffer stream Enzyme reactor Heat exchanger polyurethane insulation Aluminium block