Effect of Room Ventilation Rates in Rodent Rooms with Direct-Exhaust IVC Systems Roger Geertsema DVM, DACLAM, DAVCPM.

Slides:



Advertisements
Similar presentations
Factors that Affect Gas Pressure
Advertisements

The Smart Vivarium Physiologic and Environmental Monitoring of Rodents.
SAINT PAUL SUSTAINABLE BUILDING POLICY Training Module Six Indoor environmental quality requirements Saint Paul Sustainable Building Policy.
PowerPoint ® Presentation Chapter 15 Troubleshooting and Mitigating IAQ Problems Troubleshooting and Mitigating IAQ Problems Overview of Non-IAQ Problems.
Carrier Controls Demand Controlled Ventilation. Comfort = Temperature + Ventilation Comfort Is More Than Just The Right Temperature… Ventilation Control.
Controlling the amount of ventilation to a space based on demand Reset based on occupancy: Use carbon dioxide sensors to track space CO2 levels and raise.
The Role of Controls for Indoor Air Quality Kent W. Peterson, PE, Fellow ASHRAE P2S Engineering, Inc. Mid Columbia ASHRAE Chapter.
Healthy Homes in Rural Alaska
Heating and Air Conditioning I
Innovative Air Distribution for Healthcare Facilities.
Chapter 4 — Control Principles Chapter 4 Control Principles Control System Components Sensors Controllers Controlled Devices Control Agents Control Functions.
Break Out Session 3: Group A: Ventilation and Air Quality Chris Cosgrove, Cosgrove FDS, Inc.
Monitoring Air Quality: Emissions and Odor From A Swine Finishing Livestock Facility.
Lab Energy Management Big Ten Mechanical & Energy Conference October
Assumptions about Indoor Environments Mark Lawton P.Eng Patrick Roppel M.A.Sc.
When Do Variable Flow Fume Hoods Save Energy? Implications for lab design and behavior modification Michael Gevelber, Robert Choate, Kevin Sheehan, Brian.
1 Meeting ASHRAE Fundamentals, Standard 55 & 62.1 with Chilled Beams Displacement Ventilation.
The House as a System 1. Air Temperature and Water Vapor.
NARAGHI HALL 3 rd FLOOR VENTILATION MANAGEMENT IMPROVEMENT BEST PRACTICE AWARD HVAC DESIGN/RETROFIT 2014 CALIFORNIA HIGHER EDUCATION SUSTAINABILITY CONFERENCE.
Chemical Fume Hood  A type of local ventilation device that is designed to limit the user's exposure to hazardous or noxious fumes, vapors or dusts.
Unit 6.2. Mechanical and mixed mode ventilation TB Infection Control Training for Managers at National and Subnational Level.
School of Occupational & Environmental Hygiene Evaluation of a new mask and scavenging system for nitrous oxide used in labour and delivery Emission Control.
B. Hubbard, B. Middaugh, N. Zimmerman and J. McGlothlin Grant Title: “Development and Evaluation of a Personal, Respirable Silica Scavenging System for.
Lecture 3411/30/05. Vapor pressure vs. boiling point?
Part 2 - Dilution Ventilation (General Ventilation)
Quiz: What parameters do we measure every day?  High temperature  Low temperature  High humidity  Low Humidity.
Active beams versus VAV with Reheat Analysis of May 2013 ASHRAE Journal article Ken Loudermilk Vice President, Technology & Developement.
HVACR416 - Design Pressure Measurement Part 2. Pressure Measurement Several pressure measurements exist. They all measure the same things, the force of.
World Health Organization
Safety Meets Sustainability Rob Klein, MS, CIH Environmental Health & Safety Yale University New Haven, CT Making Universities Sustainable Copenhagen,
INDOOR AIR QUALITY ASSESSMENT. Why is preliminary assessment required ? To understand the current indoor air quality problems To develop a database on.
INDUSTRY HYGIENE MONITORING AT JB COCOA SDN BHD JB Cocoa Sdn. Bhd. Is a cocoa processing plant in the Port of Tanjung Pelepas (PTP), Johor that operated.
Indoor Environmental Quality: Investigating the Problem
Healthy Homes Overview Introduction. Learning Outcomes Upon completion of this course you will:  Recall the relationship between health and housing 
Objectives Compare locations in North America Potential for moisture problems Different driving forces Analyze potential for moisture problems in cooling.
Using Belt Air to Ventilate Active Working Areas.
The Town Hall of Zevenhuizen S.H. Liem, A.H.C. van Paassen M.Verwaal, H.F. Broekhuizen Delft, April 1998 Presentation of the building Presentation of the.
ASHRAE A Ventilation Standard for Residential Structures Ventilation and Acceptable Indoor Air Quality in Low- Rise Residential Buildings.
Is My Chemistry Building Safe? AIHCE Orlando, FL F Michael B. Amster P.E., CIH, CSP, CHMM M.B. Amster & Associates, Concord, MA M.B. Amster & Associates,
The Basics of Formaldehyde: What You Need to Know.
SDB Demand-Controlled Ventilation: Preliminary Experiments Jay Taneja Software Defined Buildings Kickoff January 11 th, 2013.
Presented by: Angela Benedict SRMT Air Quality Program Manager B REATHING, LIVING AND LEARNING IN THE AKWESASNE COMMUNITY : T OOLS TO IMPROVE INDOOR AIR.
IACUC TRAINING Update from AAALAC Assessment May 2009.
Climate, Air Quality and Noise Graham Latonas Gartner Lee Limited RWDI Air Inc.
The Coudé Lab Environment Boulder 18 March The Coudé Lab.
Labs21: Improving the Performance of Laboratories Optimizing Air Changes – one of the Big Hits September 21, 2006 Dale Sartor, P.E. Lawrence Berkeley National.
Objectives Propose residential system related final project –Compare VAV systems with DOAS systems.
Developing the GR1800 “Double Decker” Rat Cage By Corporate Marketing Department - Tecniplast, Buguggiate - Product Description.
Technical Analysis. 1500U Floor area 1500 cm P drawer style Floor area 2064 cm 2 LET’S FOCUS ON IT!
Tests performed on By Corporate Marketing Department - Tecniplast, Buguggiate -
White House Summit on Sustainable Buildings January 24, 2006 William Fisk* Sr. Scientist, Department Head Indoor Environment Department Lawrence Berkeley.
Copyright © 2015 Optimum Energy LLC. All Rights Reserved. Proprietary & Confidential Incorporating Energy Conservation Strategies into University Research.
1 ITM 1.2 How IT Decisions Impact Data Center Facilities: The Importance of Collaboration Lars Strong P.E. Upsite Technologies, Inc.
HOUSING Division Corporate Marketing Department. Plastic Materials POLYETHERIMIDE (PEI) POLYSULFONE (PSU) PSU 3 years WARRANTY PEI 5 years WARRANTY.
Parameter's Approach to Relative Humidity Control August 25, 2011 by Clay Hile.
FLEXIBILITY = Tecniplast is ready and willing to satisfy all the customers needs and provide the right solution for a professional outcome STAND ALONE.
Tests performed on By Corporate Marketing Department - Tecniplast, Buguggiate -
Fish respiration lab. Class data What can we conclude after looking at the data? Temperature will affect the respiration rate of goldfish. Goldfish respire.
The Science of IAQ in Schools John Girman Indoor Environments Division U.S. EPA National Schools Symposium August 10, 2001.
Animal welfare and standardization in Tecniplast IVCs: better research outcomes Stefano Gandolfi Junior Product Specialist Housing Division.
Successful Energy Conservation Projects at University Research Labs
Impact of High-Efficiency Filtration Combined with High Ventilation Rates on Indoor Particle Concentrations and Energy Usage in Office Buildings National.
The Indoor Air Quality in a Multi-garage
Part 12 Technician’s Guide & Workbook for Duct Diagnostics and Repair
Ch. 22 – Climate AREAS OF INTEREST: Comfort Stress
Effects of Ventilation Systems on Ammonia Concentration Distributions in Manure-Belt Egg Layer Houses Using Computational Fluid Dynamics Xinjie Tong1.
SMART VENTILATION BEST PRACTICES
Miroslav Variny, Otto Mierka
Control strategies for aeroallergens in an animal facility
A Ventilation Standard for Residential Structures
Presentation transcript:

Effect of Room Ventilation Rates in Rodent Rooms with Direct-Exhaust IVC Systems Roger Geertsema DVM, DACLAM, DAVCPM

Background Vivarium with Individual Ventilated Cages (IVC) for rodents with cage exhaust directly ventilated out of room Tecniplast IVC with positive pressure cages (70% cage exhaust rate)

Specific Aims Can room ventilation rates be safely lowered in rodent rooms utilizing direct exhaust individually-ventilated caging (IVC) Air quality within the room that could have an occupational health or animal wellbeing effect Changes in intracage environmental conditions that could impact animal wellbeing or complicate research results

Study Design 8 rodent rooms 7 mouse rooms 1 rat room 2 ventilation ratesLow: 5 – 6 ACH High: 10 – 12 ACH Density of Room in cages/sq.ft. (# of cages in room) Period Rats (140) 1.2 (280) 1.2 (540) 0.8 (175) 0.6 (190) ) 0.2 (65) 0.1 (25) 1HighLow High Low High 2LowHigh Low High Low 3HighLow High Low High 4LowHigh Low High Low

Air Flow Room Volume2800 cu.ft. Room Pressure DifferentialPositive Cage Pressure DifferentialPositive Room Ventilation RateHighLow CFM Supply ACH CFM from Racks (2)100 CFM for Pressure Offset100 CFM from Room Exhaust30050 Ventilation Cost ($3.50/cfm) $1750/year$875/year

Study Design Compare Low vs. High room ventilation rates for: Room CO 2 (difference between supply - exhaust air) Room Dew Point Temperature (difference between supply - exhaust air) Room Mouse Allergen (Mus m1) Room Endotoxin Intracage Ammonia, CO 2, Temperature, and Humidity Create a controlled spill of EtOH in room Evaluate the peak level and amount of time to return to baseline at Low vs. High ventilation

Demand-Controlled Ventilation (DCV) Computer controlled Phoenix valves in supply and room exhaust Monitoring of room air quality for temperature, dew point temperature, CO 2, dust particles, and Total Volatile Organic Chemicals (TVOC) Sample taken every 15 min. from room exhaust, not the cage exhaust Ability to increase ventilation rate based on monitoring parameters

*

*

Summary of Results Low ventilation rate: Slightly increased level of CO 2 Slightly increased Dew Point Temperature Increased time to clear a VOC spill (demand-controlled ventilation will mitigate this) No difference in: Mus m1 Endotoxin Intracage ammonia, CO2, temperature, and humidity

Conclusions It is safe to lower the room ventilation rate to 5 – 6 ACH both for human workers and animals with a direct exhaust IVC system that is properly designed and maintained - This may not apply to all IVC systems Although some statistically significant effects were observed, air quality still well within acceptable guidelines (ASHRAE limit for CO 2 in room air is 1000 ppm) With a demand-controlled ventilation system, the air is cleared of a spilled VOC faster (assuming the VOC is able to be detected by the system)

Acknowledgements ACLAM Foundation Grant Dr. Lindsell, Matthew Gudorf, Alvin Samala, Scott Smith, & Michael Phelan