Presented at EME 2012 Aug 17, 2012.  The purpose of this paper is to measure the accuracy of a variety of UHF wattmeters at 1296MHz to see what their.

Slides:



Advertisements
Similar presentations
SOME OF THE MOST COMMON TESTING MISTAKES MADE BY TECHNICIANS PERFORMING VERIFICATIONS TO CAN/ULC The reason I know some of them is because I have.
Advertisements

Instrumentation and Measurements Dr. Mohammad Kilani
Measurement of Voltages and Currents
This lesson covers the following outcomes Unit 54 P1, P7, P8 Unit 6 P10, P11.
Doug Millar SBMS March 3,  RF power measurement is made difficult because of the number of factors in measuring high frequency AC signals.  Detecting.
1 Chapter 5 Sensors and Detectors A detector is typically the first stage of a communication system. Noise in this stage may have significant effects on.
Physics and Measurements.
CHAOS IN A DIODE: PERIOD DOUBLING AND CHAOTIC BEHAVIOR IN A DIODE DRIVEN CIRCUIT Maxwell Mikel-Stites University of Rochester, Rochester, NY, Chris.
1 Lab Equipment. 2 TopicSlides DC Power Supply3-4 Digital Multimeter5-8 Function Generator9-12 Scope – basic controls13-20 Scope – cursors21-24 Scope.
EE 198 B Senior Design Project. Spectrum Analyzer.
EET Electronics Survey Chapter 8 - Measuring Voltage, Current, and Resistance.
Dummy loads & RF power measurement
Classification of Instruments :
Checking Antenna Systems
Precision Power Measurement Solutions from Bird Precision Power Measurement Solutions from Bird.
Oscilloscope Tutorial
Lecture 3: Bridge Circuits
Off-Grid Power Using Enphase Micro-Inverters Off-grid inverter systems, using batteries, can be used to provide AC power when the grid is down. It is possible.
1 Chelmsford Amateur Radio Society Advanced Licence Course Anthony Martin M1FDE Slide Set 24: v1.0, 23-Oct-2004 (4) Measurements – RF Power Chelmsford.
Care Use and Feeding of a MFJ-259B Antenna Analyzer.
Engineering 1040: Mechanisms & Electric Circuits Winter 2015 Analog & Digital Signals Analog to Digital Conversion (ADC)
Understanding ADC Specifications September Definition of Terms 000 Analogue Input Voltage Digital Output Code FS1/2.
2.5Gbps jitter generator Part 1 final presentation.
Training Making accurate Electrical Measurements.
12 Volt Marine Electrics Mike Litzkow.
Analog and Digital Instruments
Performance characteristics for measurement and instrumentation system
DIGITAL MULTIMETER A Must Have For Dealing With Electronics.
1 Chelmsford Amateur Radio Society Advanced Licence Course Carl Thomson G3PEM Slide Set 22: v1.0, 1-Oct-2004 (10) Measurements-2 - Frequency Chelmsford.
Uncertainties Using & Calculating Uncertainties for Electrical Measurement.
ELECTRONIC INSTRUMENTATION & PLC DKT Signal Conditioning Circuits.
Technician License Course Chapter 4 Lesson Plan Module 9 – Antenna Fundamentals, Feed Lines & SWR.
Uncertainties for AH Phys. Accuracy and Precision The accuracy of a measurement tells you how close the measurement is to the “true” or accepted value.
Capacitive transducer. We know that : C=kЄ° (A/d) Where : K=dielectric constant Є° =8.854 *10^-12 D=distance between the plates A=the area over lapping.
Chapter 20 Measurement Systems. Objectives Define and describe measurement methods for both continuous and discrete data. Use various analytical methods.
1 ATF2 Q BPM electronics Specification (Y. Honda, ) Design System –Hardware layout –Software –Calibration Testing Production schedule ATF2 electronics.
The Probe: Measurement Accuracy Begins at the Tip  Attenuation –The ratio of the probe’s input signal amplitude to the output signal amplitude..  Bandwidth.
Sanna Taking & Mohamad Nazri Abdul Halif School of Microelectronic Engineering Prepared by ALTERNATING CURRENT METERS Part 2 Syarifah Norfaezah Edited.
نظام المحاضرات الالكترونينظام المحاضرات الالكتروني Instrumentation and measurments Asst. lecture:-ola hussein.
Russell Taylor. Digital Cameras Digital photography has many advantages over traditional film photography. Digital photos are convenient, allow you to.
Lecture 3: Bridge Circuits
Uncertainty2 Types of Uncertainties Random Uncertainties: result from the randomness of measuring instruments. They can be dealt with by making repeated.
BASIC INSTRUMENTS - oscilloscopes
A WIRELESS PASSIVE SENSOR FOR TEMPERATURE COMPENSATED REMOTE PH MONITORING IEEE SENSORS JOURNAL VOLUME 13, NO.6, JUNE 2013 WEN-TSAI SUNG, YAO-CHI HSU Ching-Hong.
Digital Multimeter – Basic Guide
Done by: Khalid Hijjawi Mohammad Massad.  Is an electrical instrument that measures electrical resistance, the opposition to an electric current. The.
Broadcasting devices. Broadcasting devices overview USB Power Monitor, type channel, 4-line display, USB, Digital out, Opto out WEB Power Monitor,
Unit IV Ultrasonic examination of welds Foreword.
Digital Instrumentation Topics Multimeters Digital Multimeters (DMMs) Specifications (Display,Accuracy, Resolution) Meter Applications.
Internal usage only Truck Scale Technology & Accuracy.
FUNCTION GENERATOR.
DC bias circuit effects in CV measurements A.Chilingarov, D.Campbell Lancaster University, UK 9 th RD50 Workshop CERN,
QRP POWER/SWR METER PROJECT – PART 1 Developed By Al Mecozzi WA8LBZ For Presentation At The February 16, 2011 Meeting Of The CRES Amateur Radio Club.
EME Station Back End Db6nt converter DEM 144/28mhz transverter
WEBENCH® Coil Designer
MECH 373 Instrumentation and Measurements
Session - 10 Basics of Power Power in D.C and Power in A.C.
Unit - 5 Analog and Digital Instruments. Digital Voltmeter (DVM) Used to measure the ac and dc voltages and displays the result in digital form. Types:
Investigation of Wireless Power Transfer in Through-wall Applications
USING A DIGITAL VOLT- OHM - METER
McCrometer Nano Board Test Procedure
Graphing with Uncertainties
Microwave Power sensor Limitations
Analog and Digital Instruments
A. To reduce television interference B. To reduce signal loss
RF Power Meters RF Power can be measured using Thermal instruments
USING A DIGITAL VOLT- OHM - METER
724 Temperature Calibrator
Instruments & Electrical Measurements Lecture 3
Signal level Meter THROW DOWN
Presentation transcript:

Presented at EME 2012 Aug 17, 2012

 The purpose of this paper is to measure the accuracy of a variety of UHF wattmeters at 1296MHz to see what their error is.  Most of the meters are not specified at this frequency so this will be new data.  Common knowledge supposes the additional error may be 10%.

 There are very few meters that can do an accurate calibration at 23cm at high power.  I have been using a Bird 6091 calorimeter, one of the few meters that can do this type of measurement. It’s specifications are-  Accuracy 1.25% of reading  Frequency range MHz  Power range watts

 Steve, K1FO emphasizes that both line SWR and element directivity can have a negative effect on accuracy. His paper is in the resource section. In the testing set-up I used a single double female N connector to connect wattmeters to the calibrator.  I compared an HP436/8487A and attenuator in current calibration with the Bird 6091 at 10Watts on 144MHz and found that the two meters agreed to within 1%. This verified that my measurements would be accurate enough to be meaningful.

 The idea was to take a group of randomly selected meters and test them in their intended range and then at 23cm and see how their accuracy changed. All the meters were used and except for some minor repair, all were tested as found, simulating “that great swap-meet deal”, rather than using all new meters.

 Most of the meters tested were specified at 5% accuracy as a percent of a full scale reading (OFS). On a 25 Watt scale a 10 Watt reading could be off by 1 Watt.  Some were specified at 5% of reading. This is a significantly better accuracy. On a 25 Watt scale a 10 Watt reading would be off by no more than 1/2Watt.  The source data shows that some meters looked better than they were at low ends of the scale.

 Wattmeters read power as a voltage. Using E²/r produces a scale with ¼ power at mid range. This scaling helps compensate for the 5% ofs accuracy problem at low power. Somehow the Struthers meters use a linear power scale.

(The avocado was not tested for accuracy, but tasted good.) An M/M N connector was used to couple meters under test to the 6091.

 A San Bernardino Microwave Society meeting testing meters at 432MHz and 23cm.  At this session we found that combinations of elements and meters all read differently, but not necessarily out of spec at 5% OFS. K6JEY,KN6VR and W6DQ

The 6091 uses water for heat transfer and can be calibrated at DC.

 Calorimeter diagram

Bird 43, 4410 and 4304

 Bird 43- Uses elements with a directivity of about 17-20db. Specified at 5% OFS.  Bird special elements with 7 ranges as part of a bridge for thermal stability. 5% of reading accuracy.  Bird Multi range meter with two couplers. Needs calibration graph below 100MHz. Great from 144MHz to above 1000MHz. 5% OFS

Note the F to F QC coupling block. Models 6737 and 6734.

 Sensors are a capacitive voltage divider * and have a broader frequency response than the regular inductive elements. At higher power all the elements seem to have a flatter frequency response. They are more accurate when one end is attached to a load. But work fine as pictured. * *

Reading error vs. frequency

120Watt Heath IM

 Micro-  match

 All of the Micromatch single range boxes tested inconsistently from box to box. Accuracy vs. frequency range seems to be good however.  Several of the multi range box 711 tested did not meet specifications unless calibrated on a single frequency. Scale to scale readings off on the low ends, even after calibration.  The Heathkit IM4190 tested well at 432 and 23cm.

Notice the scale resolution.

 The dial calibration intervals did not give as good a resolution as other meters tested.  The meter was not as accurate as I thought it would be, although it met specifications.  The meter itself is extremely well made with one of the best couplers I have ever seen and very rugged.  The coupler diodes can be easily blown in the low power direction and are custom made and extremely hard to find.

Easily the most sophisticated meter tested. Battery operated or AC with several sensors available. Dual displays that are analog and digital. Auto ranging as well.

 The finest meter tested. Easily met specifications of 5% of reading on all frequencies.  Wide variety of measurements possible. Analogue and digital display.  Sturdy and robust.  Operated from either Ac or rechargeable batteries  A number of sensors available.  Expensive relative to the other meters.

Also made as Douglas Microwave, Sierra, and others

 Relies on mechanical adjustment for calibration. Rough handling and wear easily changes the calibration, but easy to calibrate.  The lid can fall away from the pot as latches are light duty, and coax cord can pull from connector.  Usually comes with a several multi range sensors.  Meets specifications.

MeterModel144MHz432MHz1296MHz Bird (4) 0 (-9) w Notes on Measurements: Numbers in parenthesis are % of reading. All others are % OFS.

ElementModel144MHz432MHz1296MHz Coaxial Dynamics Bird500E E

R&SModel144MHz432MHz1296MHz NAUS NAP1100w element (4) (4.8) (-13) 350w element (4) (-5) uMatchModel144MHZ432MHz1296MHZ MicroMatch Meters

MeterRange144MHz432MHz1296MHz IM

MeterRange144MHz432MHz1296MHz Struthers.2-1GHz element Douglas.2-1GHz element The Struthers meter looked new and tested well on all ranges. The Douglas was calibrated in 1994 but had obviously had rough usage.

 Doing the measurements has been revealing.  The Bird 4410 seemed to be the most accurate at all frequencies and power levels  The NAP/350 exceeded its specifications.  The Bird 4304 more than met specifications and tracked well between ranges. The Bird Termalines were very accurate. The Thruline elements were all bought used except one made by Coaxial Dynamics and met specifications.

 The data clearly shows that if a meter does well on 432MHz, it will do well on 1296MHz and probably read not more than 10% ofs high. If the meter does not do well at 432MHz, it will not do well at 1296MHz. However, it is worth looking in the detailed measurement results to see how the full- scale calculations work out as a percentage of reading for the meters.

 If you have a high power Watt meter that you know is good at 432Mhz, you can use it with some assurance as to its accuracy on 23cm. It will likely be 10% OFS high in reading.

 K1FO’s article on wattmeter accuracy  Bird Power Point on their metrology procedures and set up.  Bird’s manual on Termaline meters  Bird’s metrology procedures  The above are all available at K6JEY Presentations orK6JEY Presentations  ns/

 is