© 2013 TriQuint Semiconductor, Inc. Automation of RGAs and Sensors for FDC of PVD Metal Thin Film Processing Arif Choudhury 1, Eric McCormick 1, Guy Takayesu.

Slides:



Advertisements
Similar presentations
X-ray Photoelectron Spectroscopy
Advertisements

High Pressure Plasma with a third electrode James Roberts Physics TSP 2002 Supervised by Dr Kerrie Balla.
Click to Begin. Home Automation is the process of taking something that normally has to be switched or monitored manually and automating that process.
Mass Spectrometer Autosampler Jake Ahrens Stephen Pearson Kyle Sala Seth Yellin Project Mentor: Professor Jane Hill.
The Components There are three main components of inDepth Lite, inDepth and inDepth+ Real Time Component Reporting Package Configuration Tools.
© Copyright 2002 ABB. All rights reserved. - Microelectromechanical Systems for Process Analytics IFPAC 2003 Dr. Berthold Andres ABB Automation Products.
MET ONE 7000 Series Remote Airborne Particle Counters Customer Name Date Presenter Presenter Title.
An Introduction To Helium Leak Detection Technology Darrell R. Morrow, PhD Leak Detection Associates, Inc. Blackwood, New Jersey.
A PRESENTATION ON ELECTRONIC NOSE PRESENTED BY vivek gangwar ELECTRONICS & COMMUNICATION.
Automated Model-Based Testing of Hybrid Systems Michiel van Osch PROSE January 25,
This work is supported by the National Science Foundation under Grant Number DUE Any opinions, findings and conclusions or recommendations expressed.
There are no FREE lunches. Anti-reflective coating A anti-reflective coating is deposited on the top side to help transmit more of the incident sunlight.
DUE Security and Fire Alarm Systems LEARNING OUTCOME 7B Describe design overview and location considerations.

Physical Vapor Deposition
Thin Film Deposition Prof. Dr. Ir. Djoko Hartanto MSc
S. Kugler: Lectures on Amorhous Semiconductors 1 Preparation.
An Introduction to SIMS and the MiniSIMS ToF © Millbrook Instruments Limited Blackburn, UK
Kinetix 3 Component Servo Drive
Windows Server 2008 Chapter 11 Last Update
G4 Control and Management Solution for Data- Centers and Computer Rooms.
Nitride Materials and Devices Project
The ESS vacuum team has overall responsibility for all technical vacuum systems used on the: Accelerator, Target and Neutron Scattering Instruments and.
An Introduction to SIMS and the MiniSIMS alpha
Workshop “Vacuum systems of Synchrotron Light Sources“ organized by MAX-LAB & ALBA Barcelona, th September 2005 Instrumentation and Vacuum Control.
Methods in Surface Physics Experimentation in Ultra-High Vacuum Environments Hasan Khan (University of Rochester), Dr. Meng-Fan Luo (National Central University)
Vacuum science.
Bloodhound™ Logging System Infrared™ Hydrocarbon Analyzer System 0-100% auto ranging Total gas with Chromatograph model- C1-C4 iso and C4 normal Accurate.
VTS Sputter Roll Coater
April 2011 November 2010.
Data Acquisition Dr inż. ZDZISŁAW PÓLKOWSKI Polkowice, 2015 University of Pitesti Dolnośląska Wyższa Szkoła Przedsiębiorczości i Techniki w Polkowicach.
Save-bees’ solution : Consists of standalone devices that are placed in the beehives and offer accurate measurement information for The weight of the.
Component Environmental Test Chamber Jeff Hughes 1, Peter Adam 1, Thong Vo 2 and Phong Ngo 2 1. Department of Mechanical and Material Engineering, 2. Department.
Relationship Between in-situ Information and ex-situ Metrology in Metal Etch Processes Jill Card, An Cao, Wai Chan, Bill Martin, Yi-Min Lai IBEX Process.
Power Systems Specialists Monitoring and Controls System (MCS)
Chemistry Topic: Atomic theory Subtopic : Mass Spectrometer.
Micromechanical Testing of Thin Films WarrenOliver MTS Nano Instruments Oak Ridge, Tennessee.
11/8/ Microplasma Optical Emission Spectrometer (MOES) on a chip SFR Workshop November 8, 2000 Michiel Krüger, David Hsu, Scott Eitapence, K. Poolla,
CONTENTS: 1.Abstract. 2.Objective. 3.Block diagram. 4.Methodology. 5.Advantages and Disadvantages. 6.Applications. 7.Conclusion.
Oxygen Permeation Measurements Across Films using OxySense® luminescence Technology Devinder Saini OxySense Inc S. Eastern Ave Suite 14G Las Vegas,
Secondary Ion Mass Spectrometry A look at SIMS and Surface Analysis.
CONTENTS: 1.Abstract. 2.Objective. 3.Block diagram. 4.Methodology. 5.Advantages and Disadvantages. 6.Applications. 7.Conclusion.
Interactions with Rest Gas – Typical Case Interactions with Rest Gas – ELENA Quantitative analysis for ELENA Evaluations at 100 keV Ejection Energy Evaluations.
T EST T OOLS U NIT VI This unit contains the overview of the test tools. Also prerequisites for applying these tools, tools selection and implementation.
Applications as Residual Gas Analyzer (RGA)
Implementation of the Digital Integrated Public Address System Based on Network Jung-Sook Kim' Division of IT, Kimpo College ' 97, Gimpodaehak-ro, Wolgot-myun,
Quadrupole Mass Spectrometry Concepts Mass Spectrometers for Residual Gas Analysis.
Section 5: Thin Film Deposition part 1 : sputtering and evaporation
Hiden HPR-20 QIC EGA for evolved gas analysis in TGA-MS.
SEMINAR ON P.L.C. SUBMITTED BY HRIDAY MUKHERJEE AMIT DEBNATH SOMNATH SHIL
Upgrading the Data Processing for the Vacuum Furnaces Jennifer Case Northern Illinois University Mechanical Engineering Supervisor: Mayling Wong Technical.
PROPRIETARY INFORMATION - © 2015 WOODWARD, INC. PAGE 1 HighPROTEC-2 MRI4-2 Feeder Relay Sales presentation 2015/08/10.
EE588 : Semiconductor Manufacturing Equipment and Systems Course Introduction Dr. Stephen Daniels Dublin City University.
Roughness and Electrical Resistivity of Thin Films Spencer Twining, Marion Titze, Ozgur Yavuzcetin University of Wisconsin – Whitewater, Department of.
July 11, 2016 Run-to-Run Control of Linewidth and Overlay in Semiconductor Manufacturing Christopher Allen Bode Dr. Thomas F. Edgar, Advisor University.
Energy Management Solution
REMOTE JAMMING DEVICE.
Vacuum At LLO 2016 SURF Udochu, Ogbonnaya LIGO-T
K. H. Lee, H. Y. Lee, Young-Gi Kim, J. Yang, S. M. Yang, K.J. Chung, Y.S. Na and Y. S. Hwang Residual Gas analysis during Glow discharge cleaning, Baking.
How SCADA Systems Work?.
Energy Management Solution
Bluetooth Based Smart Sensor Network
-Honors Thesis Defense- In Situ Ellipsometry of Surfaces in an Ultrahigh Vacuum Thin Film Deposition Chamber Joseph Choi Department of Physics and Astronomy,
Plasma Cleaning Safety Presentation
Phoenix Family Overview Test with Confidence.
Mass Spectroscopy (MS) Applications as Residual Gas Analyzer (RGA)
Simple-to-use, manually-loaded batch RIE system
Easy-to-use Reactive Ion Etching equipment
Residual Gas Analyzer (RGA)
Operation of Target Safety System (TSS)
Presentation transcript:

© 2013 TriQuint Semiconductor, Inc. Automation of RGAs and Sensors for FDC of PVD Metal Thin Film Processing Arif Choudhury 1, Eric McCormick 1, Guy Takayesu 1, Keith Han 3, Alex Lee 3, Young Hwan Lee 3, Justin Wong 2, and John Yoo 2 1 TriQuint Semiconductor Corporation, 500 W Renner Rd, Richardson, Texas, USA BISTel America, 4633 Old Ironsides Dr, Suite 300, Santa Clara, CA BISTel Inc. 2F, Jeongpung Bldg, Seocho-dong, Seocho-gu, Seoul, Korea

Content Introduction –RGA Usage In PVD Process Applications of RGA in Manufacturing Architecture of FDC and Sensors RGA Filament Control Schemes Examples of FDC with RGA Future Work Conclusion © 2013 TriQuint Semiconductor, Inc.

Introduction of RGA and RGA Usage in PVD The RGA is commonly used to identify atomic and molecular species in a high vacuum environment. In the PVD process the RGA is commonly used to monitor contamination before and during the process. –Screens the incoming substrate for resist residue –Monitoring process gases The RGA is very sensitive, it can detect low pressure contaminants in Torr © 2013 TriQuint Semiconductor, Inc.

Pressure Range of PVD Process Various Vacuum Mediums* © 2013 TriQuint Semiconductor, Inc. *

RGA Advantage The RGA is an ideal tool for monitoring of PVD process because it can detect subtle changes in the chamber condition, i.e. contamination The mean free path increases at high vacuum so the probability of detecting a contaminant goes up. Can report the probable species of molecule present in the PVD chamber. Possible real time detection of contaminants or chamber leaks. Not reliant on metrology feedback. © 2013 TriQuint Semiconductor, Inc.

Applications of RGA in Manufacturing Baseline Spectrum –Spectrum scan of a known good chamber or condition Spectrum Analysis (Troubleshooting) –Analyzing the species that exist inside a chamber that is having some unknown issue He Leak Checking –Standard practice after a PVD target change Stand Alone Process Analysis –The RGA manufacturers offer stand alone packages for near real time analysis of spectrum as the tool is running. –Typically not integrated into the house FDC system. –High cost for the software. © 2013 TriQuint Semiconductor, Inc.

Benefits of Integrating RGA with FDC  Having one system that can connect to multiple components and interdict in the event of an incident saves the expense of having multiple FDC systems for each component  Adding RGAs from different manufacturers that are not supported by the tool OEMs. © 2013 TriQuint Semiconductor, Inc.

Architecture of FDC and Sensors Y-TAP PVD Controller FDC RGA Application RF Application Pump Application RGA Control Libraries RF Data Collection Library Pump Data Collection Library Pump PM1 PM2 PM3PM4 PM5 PM6 RF RGA Chamber Processing Status Wafer Processing Status of Each Chamber Trace Data © 2013 TriQuint Semiconductor, Inc.

Component Connections Tool SVIDs  HSMS Ethernet RF Monitor  A to D device to receive the analog signals and pass them to the Ethernet Pump data  RS-232 hub to Ethernet convertor Others? © 2013 TriQuint Semiconductor, Inc.

What is Gained? Context for sensors that are not available on processing chambers. © 2013 TriQuint Semiconductor, Inc.

RGA Filament Control Schemes RGA operation constraint –Open Ion Source mode of RGA operation –RGA operating pressure region is limited –Filament tripped off when P > 9.0 x Torr Customized RGA filament control schemes –P < 9.0 x Torr, filament is always on –P > 9.0 x Torr, filament on only for a short time (10sec.) at end of wafer process, otherwise filament is off © 2013 TriQuint Semiconductor, Inc.

Real time control with integrated sensors like the RGA Filament Status (On) Chamber Pressure AMU 28 (N 2 ) Example of RGA control to sample at the end of a process run. Process is complete. FDC sends the command to ignite the RGA filament. © 2013 TriQuint Semiconductor, Inc.

RGA contaminated bottle controlled test © 2013 TriQuint Semiconductor, Inc. A small Ar bottle that was purposely contaminated with 1600ppm of N2 was connected to a PVD chamber as a controlled test. N2 Contamination Rs Sample Time Pressure * The Ar36 isotope was chosen for the graph instead of Ar40 because of the signal magnitude relative to the magnitude of N 2.

Example 1: Real Time RGA Alarm © 2013 TriQuint Semiconductor, Inc. Ar should be present. O 2 and H 2 are baseline signals. N 2 peak (pulse) detected.

Example 2: Chamber Lid Leak RGA Trace © 2013 TriQuint Semiconductor, Inc. AMU28 (N2)- Red  Chamber lid leak Blue  Normal

Future Work Implementing RGA at house Ar gas supply –Detect problem at time close to t 0 –Minimize recovery Adding Real-time leak test recipes for maintenance and operators –Post chamber recovery –Weekly chamber integrity check © 2013 TriQuint Semiconductor, Inc.

Conclusion There is great value in integrating RGAs and other sensors into an in-house FDC system. –Sensor Control –Real time context with existing tool parameters –Real time detection © 2013 TriQuint Semiconductor, Inc.

Acknowledgement The authors want to express their appreciations for the valuable inputs on this project from Romek Bobkowski and Richard Groom, both of TriQuint Semiconductor Corporation on PVD processing and RGA operation respectively. © 2013 TriQuint Semiconductor, Inc.