1 NMR Instrumentation Modern Pulsed Fourier Transform NMR Spectrometer with superconducting magnet Why? Superconducting Magnet Computer Electronics Console.

Slides:



Advertisements
Similar presentations
Physical Layer: Signals, Capacity, and Coding
Advertisements

MRI Phillip W Patton, Ph.D..
MR TRACKING METHODS Dr. Dan Gamliel, Dept. of Medical Physics,
ECE 4321: Computer Networks Chapter 3 Data Transmission.
Advanced Higher Unit 3 Nuclear Magnetic Resonance Spectroscopy.
Nuclear Magnetic Resonance
427 PHC.  Atomic emission spectroscopy (AES) is based upon emission of electromagnetic radiation by atoms.
NMR Spectroscopy Tuning / Matching Shimming and Lock.
Lecture 2 1 H Nuclear Magnetic Resonance. Gas Chromatograph of Molecular Hydrogen at –100 °C Thermoconductivity Detector 12.
NMR Nuclear Magnetic Resonance Spectroscopy. Over the past fifty years nuclear magnetic resonance spectroscopy, commonly referred to as nmr, has become.
NMR SPECTROSCOPY.
Nuclear Magnetic Resonance Spectroscopy Chemistry 243
ELEG 479 Lecture #9 Magnetic Resonance (MR) Imaging
Chapter 5 How the spectrometer works Magnet Probe Transmitter SynthesizerReceiver ADC Pulse programmer Computer 5.1 The magnet: 1. Strength: 14.1 T for.
1 NMR Practical Aspects (I) Processes involving in a NMR study: 1. Sample Preparation/Considerations 2. Setting up Spectrometer 3. Setting up acquisition.
7- Double Resonance 1. Types of double resonance experiments 2. 1 H-{ 1 H} Homonuclear Decoupling C-{ 1 H} Heteronuclear Decoupling.
Spectrum analyser basics Spectrum analyser basics 1.
Chapter 3 Data and Signals
NMR Spectroscopy Part II. Signals of NMR. Free Induction Decay (FID) FID represents the time-domain response of the spin system following application.
FT-NMR. Fundamentals Nuclear spin Spin quantum number – ½ Nuclei with spin state ½ are like little bar magnets and align with a B field. Can align with.
The NMR spectrometer Magnet Probe Coils Transmitters
Nuclear Magnetic Resonance Spectrometry Chap 19
FT-NMR.
William Stallings Data and Computer Communications 7th Edition (Selected slides used for lectures at Bina Nusantara University) Data, Signal.
Dynamic Effects in NMR. The timescale in nmr is fairly long; processes occurring at frequencies of the order of chemical shift differences will tend to.
Magnetic Resonance Imaging Basic principles of MRI This lecture was taken from “Simply Physics” Click here to link to this site.
Resonance condition. Pulse A coil of wire placed around the X axis will provide a magnetic field along the X axis when a direct current is passed through.
Principles of MRI Physics and Engineering
Principles of Magnetic Resonance
Atomic Emission Spectroscopy
Determination of Spin-Lattice Relaxation Time using 13C NMR
Magnetic Resonance Imaging
1 Business Telecommunications Data and Computer Communications Chapter 3 Data Transmission.
Seminar October, 2008 j. brnjas-kraljević. Imaging (MRI)  tomography technique  tomography technique – the volume image is built up by images of thin.
Nuclear Magnetic Resonance
Nuclear Magnetic Resonance Spectroscopy (NMR) Dr AKM Shafiqul Islam School of Bioprocess Engineering.
Nuclear Magnetic Resonance ANIMATED ILLUSTRATIONS MS Powerpoint Presentation Files Uses Animation Schemes as available in MS XP or MS 2003 versions A class.
05- NMR Instrumentation 1.Overview of the JEOL LA400 2.Superconducting magnet 3.The magnetic field and homogeneity 4.Field frequency lock 5.Pulse program.
CHEM 344 Spectroscopy of Organic Compounds Lecture 1 4th and 5 th September 2007.
Sarah Newton University of Oregon Applied Physics.
Pei-Ann Lin and PJ Velez December 13, 2011
W HAT IS NUCLEAR MAGNETIC RESONANCE ? State that NMR spectroscopy involves interaction of materials with low-energy radio frequency radiation. Describe.
The Physical Layer Lowest layer in Network Hierarchy. Physical transmission of data. –Various flavors Copper wire, fiber optic, etc... –Physical limits.
A NMR pulse is often designed to detect a certain outcome (state) but often unwanted states are present. Thus, we need to device ways to select desired.
10 – Experimental aspects of NMR 1.The NMR tube 2.NMR solvents 3.The sample 4.How much time does an NMR experiment require? 5.Line shape 6.NMR coils 7.Shimming.
There are 2 variables in NMR: an applied magnetic field B 0, and the frequency ( ) of radiation required for resonance, measured in MHz. NMR Theory
NUCLEAR MAGNETIC RESONANCE SPECTROSCOPY … or NMR for short.
Figure 8.3 gives the basic layout of a continuous wave NMR spectrometer. These intruments were the original type of instrument and have largely.
Tuning, Lock and Shimming Claridge- 3.3, 3.4, 3.6.
NMR Nuclear Magnetic Resonance Chapter 13. Proton Nuclear Spin States Two states have the same energy in the absence of a magnetic field Magnetic Field.
1. What is an NMR Spectrum ? 2. What are the Spectral Features? 3. What are the Spectral Parameters? 4. How much should be known about the NMR Phenomena.
Nuclear Magnetic Resonance (NMR) for beginners. Overview NMR is a sensitive, non-destructive method for elucidating the structure of organic molecules.
 This depends on a property of nuclei called spin.  Gyroscope: Principle: As long as its disc remains spinning rapidly the direction of the spin axis.
MOLECULAR STRUCTURE ANALYSIS NMR Spectroscopy VCE Chemistry Unit 3: Chemical Pathways Area of Study 2 – Organic Chemistry.
Guided by: Dimal Shah Prepared by: Shweta Singh IICP NEW V.V. NAGAR....1.
1 Instrumental Analysis Tutorial 10 Nuclear Magnetic Resonance NMR.
Tuning, Lock and Shimming
FMRI data acquisition.
CLARIDGE CHAPTER 2.2; 3.2.
(Instrument part) Thanundon Kongnok M
Department of chemistry Smt. K. R. P. Kanya Mahavidyalaya, Islampur
Imaging Science Department, IC Jo Hajnal
Nuclear Magnetic Resonance
10.5 Fourier Transform NMR Instrumentation
10.6 Fourier Transform Mass Spectrometry
10.4 Continuous Wave NMR Instrumentation coherent detection
Structure determination by NMR
(4)ELECTRONIC SUPPORT SYSTEM
Superconducting Magnets
Presentation transcript:

1 NMR Instrumentation Modern Pulsed Fourier Transform NMR Spectrometer with superconducting magnet Why? Superconducting Magnet Computer Electronics Console

2 Chem3204: NMR Instrumentation A sensitive receiver to amplify the NMR signals (RF Detector) A Digitizer to convert the NMR signals into a form which can be stored in computer memory A “pulse programmer” to produce precisely timed pulses and delays A computer to control everything and to process the data Components of a NMR spectrometer: An intense, homogeneous and stable magnetic field (magnet + shim) A “probe” which enables the coils used to excite and detect the signal to be placed close to the sample High-power RF transmitter/s capable of delivering short pulses (RF source + RF Amplifier)

3 use persistent superconducting magnets to generate the B 0 field; at low temperatures (less than 6 K, typically) the resistance goes to zero – that is the wire(eg.Nb alloy) is superconducting; To maintain the wire in its superconducting state the coil is immersed in a bath of liquid helium (4 K, expensive); “heat shield” kept at 77 K by contact with a bath of liquid nitrogen (cheap) to reduces the amount of liquid helium boils off; vacuum flask so as to further reduce the heat flow. Superconducting Magnet (SCM): A Big Stainless Steel Dewar Advantages of SCM ? 1. Strongest Magnet; 2. Stable & homogeneous magnet field Bo; 3. Low running cost.

4 Sensitivity & Price Sensitivity  B 0 3/2 $ USD Tesla MHz

5 High resolution NMR requires linewidths of 1 Hz or less Magnetic field across the sample must be homogeneous so that the corresponding variation in the Larmor frequency is small Surround the sample with a set of shim coils, each of which produces a tiny magnetic field with a particular spatial profile to canceling out the small residual inhomogeneities in the main magnetic field. Modern spectrometers might have up to 40 different shim coils labeled according to the field profiles they generate, such as x, y, z, z 2, z 3, z 4, z 5, xy, xz, yz, x 2 -y 2, etc… Shimming, the process to optimize the shims, requires skill and experience because various shims will interact with each other. Shim Coils

6 The Probe The key part of the probe is the small coil used to excite and detect the magnetization in radio-frequency. Many types: e.g. TXI, TBI, X-BB, BBI Fixed Nucleus or Broad-Band (tunable for different nucleus) Direct detection or Inverse detection probe To optimize the sensitivity this coil needs to be (1) as close as possible to the sample; (2) tuned to resonant at the Larmor frequency of the nuclei being detected and (3) matched to maximize power transfer between the probe and the transmitter and receiver. Usually multi-coils for different nucleus: e.g. 1 H, 2 H (for locking), 13 C, 15 N, etc… with observe coil at inner-most position.

7 The Probe

8 The transmitter: Channel Synthesizer: RF source which produces a stable frequency which can be set precisely. RF amplifier: boost this small signal to a power of 100 W or more to provide enough energy to excite the NMR active nuclei in the sample. Attenuator: altering the RF power level in units of decibels (dB) (Bruker: 120 to –6 dB) All under computer control Each nucleus type required one set of transmitter channel => usually more than one channels

9 The receiver The NMR signal emanating from the probe is detected by a digitizer receiver at regular time intervals (dwell time). A device known as an analogue to digital converter or ADC is used to convert the NMR signal from a voltage to a binary number which can be stored in computer memory. Dynamic range of ADC digitizer is measured by bits (e.g. 16-bit, i.e. 0 to or 65535). Receiver Gain (rg) should be set to have the maximum signal using up all the bits in ADC. What happen maximum signal << 65535? What happen maximum signal > Clipping of FID? Small quantization artifact noise peak appears in spectrum. Loss weak signals

10 Pulse programmer to produce precisely timed pulses and delays required by the NMR pulse experiment Computer system Control all electronics Date acquisition and processing (Bruker software-XwinNMR) Plotting Spectrum (Bruker software-Xwinplot) Third party software may be used for processing or analysis e.g. nmrPipe, Felix, nmrView.

11 Additional Instrumentation: Sample spinner: Spinning equalize xy magnetic field homogeneity, i.e. better resolution Eject/Insert system: using air stream to eject and insert sample tube along the long bore tube

12 Additional Instrumentation: Locking (separate 2 H channel) Even in the best spectrometers the field strength varies to some extent over time The position of the deuterium peak is monitored To counteract the field drift a lock field is applied to maintain a constant deuterium resonance position Deuteriated solvent is usually used to provide the Deuterium Lock signal e.g. CDCl 3, D 2 O, CD 3 OD

13 Additional Instrumentation: The pulse field gradient Addition coils in the probe to provide a magnetic field gradient along a particular direction. Extensively used in MRI experiments to provide spatial encoding. Also useful for solvent peak suppression, artifact peaks suppression, and coherence selection in NMR experiments.

14 Bruker AV600Bruker AV400Bruker DRX500JEOL 270Bruker DRX300 NMR Facilities in Chemistry Dept. Safety Precautio ns : Very Strong Magnetic Field! 5G10G