Diagnostics and Instrumentation for ERL  Pieces of ERL and their beam specifics  Injector diagnostics  Drive Laser diagnostics  Problems with 2-beam.

Slides:



Advertisements
Similar presentations
Matching Injector To Linac. Caveats This is all loose and fuzzy – sort of religion We dont have real tight control over and knowledge of the machine –
Advertisements

The JLab IR/UV FEL Driver D. Douglas for the JLab FEL anarcho-syndicalist commune.
Sub-ps bunch length measurements at the JLab FEL with coherent transition radiation (CTR) and “Martin-Puplett” interferometer Pavel Evtushenko, JLab 
Beam-based Measurements of HOMs in the HTC Adam Bartnik for ERL Team, Daniel Hall, John Dobbins, Mike Billing, Matthias Liepe, Ivan Bazarov.
J. Rudolph, Helmholtz-Zentrum Berlin EuCARD 2nd ANNUAL MEETING Slice emittance measurements at the ELBE superconducting RF photoinjector.
Halo, SLAC, 2014 Towards wire scanner measurements with Large Dynamic Range (> 10 6 ) Pavel Evtushenko, Jefferson Lab.
1 Bates XFEL Linac and Bunch Compressor Dynamics 1. Linac Layout and General Beam Parameter 2. Bunch Compressor –System Details (RF, Magnet Chicane) –Linear.
Diagnostics and commissioning on ERLP Yuri Saveliev ASTeC CONFORM Project: EMMA Design Review Workshop February 2007, Daresbury Laboratory.
Emittance Measurement: Quadrupole Scan C. Tennant USPAS – January 2011.
TUPD02 BEAM DIAGNOSTICS FOR THE ESS BLM BPM Trans Profile Bunch Shape BCM Preliminary System Count A. Jansson, L. Tchelidze, ESS AB, Lund, Sweden Hybrid.
Generation and Characterization of Magnetized Bunched Electron Beam from DC Photogun for MEIC Cooler Laboratory Directed Research and Development (LDRD)
Beam Diagnostic Needs and Challenges of some Future Light Sources Pavel Evtushenko, JLab This talk to a large extend based on discussion and comments:
Generation and Characterization of Magnetized Bunched Electron Beam from DC Photogun for MEIC Cooler Laboratory Directed Research and Development (LDRD)
Modelling of the ALICE Injector Julian McKenzie ASTeC STFC Daresbury Laboratory IOP Particle Accelerators and Beams Group Status and Challenges of Simulation.
Injector Setup/Mini-phase  Description of injector setup  sources of drift  Mini-phase procedure for injector  Checking the rest of the machine. Stephen.
Accelerator and Detector R&D, August Diagnostics Related to the Unwanted Beam Pavel Evtushenko, Jlab FEL.
Transverse emittance Two different techniques were used to measure the transverse emittance. The multislit mask in the injector 9 MeV Quadrupole scan for.
Longitudinal transfer function a.k.a. (M 55 ) measurements at the JLab FEL Pavel Evtushenko, JLab  Jlab IR/UV upgrade longitudinal phase space evolution.
SPPS, Beam stability and pulse-to-pulse jitter Patrick Krejcik For the SPPS collaboration Zeuthen Workshop on Start-to-End Simulations of X-ray FEL’s August.
05/05/2004Cyrille Thomas DIAMOND Storage Ring Optical and X-ray Diagnostics.
Experience from the Spallation Neutron Source Commissioning Dong-o Jeon Accelerator Physics Group Oak Ridge National Laboratory May 9, 2007.
Recent Experiments at PITZ ICFA Future Light Sources Sub-Panel Mini Workshop on Start-to-End Simulations of X-RAY FELs August 18-22, 2003 at DESY-Zeuthen,
Transverse Profiling of an Intense FEL X-Ray Beam Using a Probe Electron Beam Patrick Krejcik SLAC National Accelerator Laboratory.
Operated by the Southeastern Universities Research Association for the U.S. Dept. Of Energy Thomas Jefferson National Accelerator Facility FEL Bunch length.
Optics considerations for ERL test facilities Bruno Muratori ASTeC Daresbury Laboratory (M. Bowler, C. Gerth, F. Hannon, H. Owen, B. Shepherd, S. Smith,
S. Molloy, P. Emma, J. Frisch, R. Iverson, M. Ross, D. McCormick, M. Woods, SLAC, CA, USA S. Walston, Lawrence Livermore National Laboratory, CA, USA V.
Accelerator Science and Technology Centre Extended ALICE Injector J.W. McKenzie, B.D. Muratori, Y.M. Saveliev STFC Daresbury Laboratory,
ICFA-HB 2004 Commissioning Experience for the SNS Linac A. Aleksandrov, S. Assadi, I. Campisi, P. Chu, S. Cousineau, V. Danilov, G. Dodson, J. Galambos,
October, 2009 Internal Review Beam Instrumentation David Gassner.
LDRD: Magnetized Source JLEIC Meeting November 20, 2015 Riad Suleiman and Matt Poelker.
Y. R. Roblin Hall A beamline and accelerator status.
Beam Halo Monitoring using Optical Diagnostics Hao Zhang University of Maryland/University of Liverpool/Cockcroft Institute.
Christopher Gerth DL/RAL Joint Workshop 28-29/4/04 Modelling of the ERLP injector system Christopher Gerth ASTeC, Daresbury Laboratory.
February 5, 2005D. Rubin - Cornell1 CESR-c Status -Operations/Luminosity -Machine studies -Simulation and modeling -4.1GeV.
July LEReC Review July 2014 Low Energy RHIC electron Cooling Jorg Kewisch, Dmitri Kayran Electron Beam Transport and System specifications.
P. Urschütz - CTF3 Collaboration Meeting 2007 CTF3 commissioning & operation in 2006 P. Urschütz for the CTF3 operations team  Commissioning of the Delay.
DaMon: a resonator to observe bunch charge/length and dark current. > Principle of detecting weakly charged bunches > Setup of resonator and electronics.
MeRHIC Internal Cost Review October, Dmitry Kayran for injector group MeRHIC Internal Cost Review October 7-8, 2009 MeRHIC: Injection System Gun.
Lessons Learned From the First Operation of the LCLS for Users Presented by Josef Frisch For the LCLS March 14, 2010.
Construction, Commissioning, and Operation of Injector Test Facility (ITF) for the PAL-XFEL November 12, 2013 S. J. Park, J. H. Hong, C. K. Min, I. Y.
What did we learn from TTF1 FEL? P. Castro (DESY).
XFEL Beam Dynamics Meeting Bolko Beutner, DESY First results of micro-bunching and COTR experiments at FLASH Bolko Beutner, Winfried Decking,
Seesion1 summary Unwanted Beam Workshop (UBW 2012) Dark Current Issues for Energy Recovery Linacs.
Beam Commissioning Adam Bartnik.
Emittance measurements for LI2FE electron beams
Slice Parameter Measurements at the SwissFEL Injector Test Facility
Andrei Shishlo ORNL SNS Project Florence, Italy November 2015
Have a chance to operate your own beam at CERN
BUNCH LENGTH MEASUREMENT SYSTEM FOR 500 KV PHOTOCATHODE DC GUN AT IHEP
Timing and synchronization at SPARC
Coupling Correction at the Australian Synchrotron
Controls CBETA controls will extend existing EPICS control system
WG2 Summary: Diagnostics, measurements, and commissioning
The Cornell High Brightness Injector
ERL BPM System Mike Billing, Mark Palmer
Diagnostics overview and FB for the XFEL bunch compressors
Z. Huang LCLS Lehman Review May 14, 2009
Linac/BC1 Commissioning P
Test of Booster at UITF Reza Kazimi (12/12/18)
High Level Physics Applications for LCLS Commissioning
Linac Diagnostics Patrick Krejcik, SLAC April 24, 2002
Injector Experimental Results John Schmerge, SSRL/SLAC April 24, 2002
Linac Physics, Diagnostics, and Commissioning Strategy P
Breakout Sessions SC1/SC2 – Accelerator Physics
LCLS Injector/Diagnostics David H. Dowell, SLAC April 24, 2002
First results of micro-bunching and COTR experiments at FLASH
Diagnostics RF and Feedback
Linac Diagnostics Commissioning Experience
Generation of Magnetized Bunched Electron Beam for MEIC Cooler
CLIC luminosity monitoring/re-tuning using beamstrahlung ?
Presentation transcript:

Diagnostics and Instrumentation for ERL  Pieces of ERL and their beam specifics  Injector diagnostics  Drive Laser diagnostics  Problems with 2-beam Linac  Large dynamic range measurements – direction of development  Transversal beam match  Longitudinal match, bunch compression, bunch length measurements Pavel Evtushenko, JLab

ERL regions and their specific 1.Electron gun / injector: low 5-10 MeV beam energy space charge dominated beam transversely large, few ps long 2. Linac: two beams, must preserve beam quality 7. Beam dump put all in there 3. Beam transport: higher beam energy, bunch compression, match transversely 4. FEL wiggler (interaction region) transversely smallest beam, shortest bunch length, precisely set Twiss parameters, no beam loss, precise trajectory to overlap with optical mode 5. More beam transport the same, but used beam 6. More Linac used beam, energy compression 0. Drive Laser / Photocathode: change and affect everything

Gun / Injector diagnostics Beam parameters that need to be measured and monitored in the injector:  the transverse phase space distribution (Twiss parameters and emittance )  the longitudinal phase space distribution (bunch length and energy spread)  be able to measure and control halo  phase of the beam in RF cavities (1. setup 2. monitor)  drive laser parameters (transverse & longitudinal profile, position, phase )  cathode (Q.E.) distribution  make sure that the beam parameters measured with tune-up pulsed beam do not change when going to CW beams and changing the beam average current

Gun / Injector diagnostics Viewers (3) Solenoids (2) Quads (4) SRF cavities (2)Electron gun (1)

Injector / Transverse beam profile  100 um thin YAG:Ce viewers oriented normal to the beam is used  first viewer (on the left) located downstream of the booster in non dispersive location  used when centering beam in injector elements, emittance measurements and beam phase  second viewer (on the right) located at the point close to max. dispersion and small horizontal beta function  used for energy spread (energy distribution) and mean energy measurements also beam phase

Injector / Transverse Phase Space The multislit or a single slit scanning through the beam (or a beam scanning across the slit) does the job very well (pulsed beam only).  well established technique  works for space charge dominated beam  beam profile is measured with YAG  measures not only the emittance but the Twiss parameters as well  enough information to reconstruct the phase space  has been implemented as on-line diagnostics  works with diagnostics mode only (low duty cycle, average current)

Injector / Longitudinal Phase Space  Single cell cavity or multi-cell structure with TM dipole mode impose on the bunch time dependent transverse kick.  The dipole creates dispersion in the transverse direction perpendicular to the cavity kick Problems:  the same as multislit – pulsed beam only  resolution limited by transverse emittance; solution – put a small hole in front of the cavity and do the measurements for small beamlet AND as f(x,y) - 3D charge distribution  usually a special setup placed at the beam line for injector study and removed when it is done.

Injector / Drive Laser Diagnostics  Transversal profile of the D.L. on the cathode (standard technique) imaging reflection from the input window to a CCD placed on the same distance from the window as the cathode  Longitudinal profile: auto-correlator works well for Gaussian pulses for non Gaussian pulses streak camera  Amplitude stability photo diode (at JLab FEL is monitored all the time in control room)  Transversal stability long term slow drifts or misalignment – the same CCD as for profile meas. fast transverse jitter 4-quadrant position sensitive photodiode  Phase stability (phase noise) all the way down to DC fast (few GHz) photodiode to look at FM at very high harmonic number BUT also must do measurements at DC to separate AM from FM + Signal Source Analyzer (SSA) expansive but saves a lots of time

Injector / Drive Laser Diagnostics  all focusing elements are placed upstream of the pick off plate  distance from the pick off to the cathode = distance from pick off to reference screen  the reference screen used for D.L. profile and position measurements

Injector / Drive Laser Diagnostics Typical auto-correlator signal JLab FEL D.L. (observed in the control room permanently) Drive Laser transverse profile dynamic range ~500: 10-bit frame grabber, 60 dB SNR CCD

Injector / Cathode Q.E.  For accurate modeling of the injector beam dynamics the input in to a code must be accurate. This includes the emission transversal profile, which is a product of the Drive Laser profile and Q.E. profile  One technique to measure the Q.E. profile is scanner with a laser (spot on the cathode is much smaller than cathode dimensions) and two scanning mirrors. Usually automated and gives 2D distribution of the Q.E. Can be used only with beam and gun voltage off.  Another technique: run low current emittance dominated beam (even DC), measure the laser trans. profile and image the cathode to a view screen. The ration of the beam profile and the laser profile gives Q.E. profile. If used with the real drive laser (not always possible) gives the emission profile. Laser profile on cathodePhotoemission profile

2-pass viewers  there are two beams in the LINAC  when trying to measure decelerating beam with a viewer the accelerating beam gets also intercepted  ultimately the measurements should be done with non intercepting technique (or intercepting but negligibly )  JLab FEL uses OTR viewers with 5 mm hole (first beam goes in to the hole)  44% transparent mesh 5 micron thin  SRF cavities see the radiation due to the intercepted beam (can trip cavity) JLab FEL LINAC OTR viewer  With the ultra bright beam OTR might be useless (OTR  COTR LCLS)  The best solution might be to use wire scanners (intercepts small part of the beam).  If the scanner measures radiation created by the wire, must take care of the background.  Laser wire scanner: Will the measurements time and cost be acceptable? What is the lowest energy when the technique is practical?  We also use: 1. Diamond-like carbon foils ~ 1 micron thin

2-pass BPM (ideas) Stripline BPM signal measured with scope, the picture limited by the scope BW  A. Lebedev proposed at ERL05 to make a system based on small AM of the beam current (no experiments so far)  Time domain approach might be very different for different machines ✔ long recirculation time vs. short; ✔ every bucket filled vs. not  The phase difference is not always 180 deg, especially when tuning machine  In JLab FEL the ΔT between two beam depends on rep. rate.; smallest is 6 ns; Time domain approach: (will not measure every bunch) 1. make BPM pulses broader to 1.5 ns (LPF or dispersion) 2. grab peak value with S&H (GHz BW) 3. digitize S&H output with ~ 10 MHz ADC Motivation:  To do differential orbit measurements (measure transport matrix) with both beams in the LINAC  The decelerating beam gets adiabatically anti-dumped – small errors corrections in the beginning leads to big orbit change at the end  Orbit stabilization and feedback

Transport / Transverse match  combination of OTR and P-46 phosphor coated viewer is used  set of transverse beam profiles measured through UV FEL beam line (~1/2 of the accelerator)  together with the machine optics model used to understand and adjust the transverse match  iterative process between measurements and fitting/adjusting model and beam optics  anything but Gaussian distributions

Large dynamic range beam profile measurements Measured in JLab FEL injector, local intensity difference of the core and halo is about 300. (500 would measure as well) 10-bit frame grabber & a CCD with 57 dB dynamic range PARMELA simulations of the same setup with 3e5 particles: X and Y phase spaces, beam profile and its projection show the halo around the core of about 3e-3. Even in idealized system (simulation) beam dynamics can lead to formation of halo.

Large dynamic range measurements (example)  Main principal of one of the ways to make large dynamic range measurements is to reduce a measurement to frequency measurements. Then make it work for 1 Hz and for 100 MHz and this is 10 8 dynamic range.  For instance use PMT and keep them working in counting mode.  Can be applied to e - beam measurements, laser, (light), X-rays.  Example: wire scanner measurements: Courtesy of A. Freyberger (measured at CEBAF)

Drive Laser large dynamic range measurements  typically in simulation the drive laser pulse is assumed “ideal” i.e. as we want it  to find out how far is it true and if concerned about 1e-6 details of the phase distribution there is the need to make D.L. pulse measurements with 1e6 dynamic range  transverse and longitudinal profile of the D.L. needs to be measured  AND reflections from D.L. transport, which might be far from D.L. spot (also in time)  for transversal profile measurements: ➟ could use 120 dB dynamic range CCD (2 CCD in one camera each 60 dB) ➟ use a usual ~ 60 dB CCD with an ND filter and accurate cross-calibration  for longitudinal profile measurements: ➟ auto-correlator with dynamic range 120 dB can be built using PMT ➟ auto-correlator works fine for Gaussian pulses ➟ WHAT to do for non Gaussian?

Drive Laser “ghost” pulses  For machine tune up, beam studies, intercepting diagnostics a “diagnostic beam” with very low average current but nominal bunch charge is used (all beam can be lost without damaging machine)  For example, JLab FEL: max rep. rate MHz (CW) diagnostic mode: rep. rate MHz (÷16), 250 μs / 2 Hz (÷2000) average current ~300 nA  Most of the laser pulses are “stopped” by EO cell(s), but the extinction ratio of the an EO cell is about 200 (typical), two in series ~ 4×10 4  Another example: want to reduce 1300 MHz (100 mA) to 300 nA (2 18 ) than for every bunch Q b we want we also get 6.55×Q b of “ghost” pulses (655 % !!!) we do not want  “ghost” pulses overall intensity must be kept much lower than real pulses!!! for “usual” measurements ~ 1% might be fine much bigger problem if want to study halo, let’s say effects, than “ghost” pulses should be kept at (???)

Drive Laser “ghost” pulses  Using a Log-amp is an easy way to diagnose presence of the “ghost” pulses Beam current measurements with a linear circuit current measurements with A Log-amp circuit (60 dB) Problem:  this does not work for CW beam

Drive Laser “ghost” pulses  Using a Log-amp is an easy way to diagnose presence of the “ghost” pulses  Log-amps with dynamic range 100 dB are available 631 uA (100%) 135 pC x MHz 5.7 uA (~0.9 %) MHz “ghost” pulses

Longitudinal transfer function measurements The M 56 and T 566 transport matrix elements are validated via longitudinal transfer function measurements – essential for optimal bunch compression. T 566 is adjusted by sextupoles in dispersive location (ARC1) M 56 is adjusted by quadrupoles in dispersive location (ARC1) Changes in the E 0 and it effect on the bunch compression

Pulsed and CW beam measurements / transition  Most affordable way to measure ps and sub-ps bunches  Works with pulsed beam (tune up) and CW beam, essentially at any average current  Used at JLab FEL to ensure that the bunch length does not change for pulsed or CW and when the average current is increased  Ultimately needs to be setup in vacuum (or N 2 purge) due to atmosphere absorption of THz  Phase information is lost – no direct bunch profile reconstruction  A detector measuring total CTR (CSR) power – a bunch length monitor Power spectrum Interferogram (autocorrelation)

Conclusion and Outlook  Most of the measurements can be made with fairly simple diagnostic tools  For an injector transverse and longitudinal phase spaces can be measured very well with pulsed (diagnostics) beams.  One direction to improve that is to increase the dynamic range of such measurements to (10 6 – 10 8 ?)  Another direction of improvement and development is to monitor injector beam parameters with CW beam (beam size most challenging)  Measurements of D.L. with big dynamic range to understand beam dynamics with the same big dynamic range  LINAC (part of the machine with 2 beams) needs specific solutions  reducing the CW beam to diagnostic beam needs careful consideration this (via “ghost” pulses) affects all diagnostics

Conclusion and Outlook (2)