CT Scanning: Dosimetry and Artefacts

Slides:



Advertisements
Similar presentations
PRINCIPLES OF CT.
Advertisements

IMAGE GENERATION IN CT.
Advanced Biomedical Imaging Lecture 3
Computed Tomography Principles
Computers and Computed Tomography
Image Reconstruction.
CT Scanning Dr. Craig Moore
Image Quality Bushong Chapter 7.
IMAGE QUALITY.
Computed Tomography Physics, Instrumentation, and Imaging
Digital Radiography.
Parameters and Trade-offs
Computed Tomography II
Seeram Chapter 13: Single Slice Spiral - Helical CT
NC HPS Meeting 10/18-19/2001 Boone, NC Recent Advances in CT Technology and Issues of CT Dosimetry T. Yoshizumi 1,2, M. Sarder 1, R. Reiman 1,2, E. Paulson.
SCANCOMEDICAL Computed Tomography SCANCO User Meeting 2005 Dr. Bruno Koller SCANCO Medical AG
Advanced Biomedical Imaging
CT Multi-Slice CT.
IMAGE QUALITY NOISE LINEARITY CROSS-FIELD UNIFORMITY IMAGE ARTIFACTS.
Computed Tomography III
BMME 560 & BME 590I Medical Imaging: X-ray, CT, and Nuclear Methods X-ray Imaging Part 1.
Optimisation of X-ray micro-tomography to perform low-dose imaging of highly-dosed gels P.M.Jenneson, E.C.Atkinson, P.Wai and S.J.Doran In 1993, Maryanski.
tomos = slice, graphein = to write
Unsharpness Calculations and Resolution By Professor Stelmark.
X-Rays and CT Scans.
Chapter 2 Stewart C. Bushong
Computed Tomography
8.1 PRODUCTION AND CHARACTERISTICS OF X-RAYS
Tissue Contrast intrinsic factors –relative quantity of protons tissue proton density –relaxation properties of tissues T1 & T2 relaxation secondary factors.
Seeram Chapter 11: Image Quality
Basic principles Geometry and historical development
COMPUTED TOMOGRAPHY I – RAD 365 CT - Scan
QIBA CT Volumetrics - Cross-Platform Study (Group 1C) May 6, 2009 Interclinic Comparison of CT Volumetry Quantitative Imaging Biomarker Alliance.
Image Quality Radiographic Resolution.
BMME 560 & BME 590I Medical Imaging: X-ray, CT, and Nuclear Methods Introductory Topics Part 2.
Computed Tomography Q & A
Medical Image Analysis Image Reconstruction Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
CT Instrumentation and X-ray system
Factors affecting CT image RAD
Interaction ionizing radiation with biological tissue. Basic dosimetry.
IAEA International Atomic Energy Agency Optimization of Protection in Computed Tomography (CT)-What can radiographers do? IAEA Regional Training Course.
QIBA CT Volumetrics - Cross-Platform Study (Group 1C) March 18, 2009 Interclinic Comparison of CT Volumetry Quantitative Imaging Biomarker Alliance.
Image Quality The capacity to define, measure, and assess image quality is a primarily responsibility of a CT Technologist.
Advanced CT systems and Their Performance
Module D Computed Tomography Physics, Instrumentation, and Imaging.
CT QUALITY MANAGEMENT. SPATIAL RESOLUTION CONTRAST RESOLUTION NOISE IMAGE ARTIFACTS RADIATION DOSE.
Part No...., Module No....Lesson No
Part No...., Module No....Lesson No
RAD 254 Chapter 27 Digital Fluoroscopy
CT Chapter 4: Principles of Computed Tomography. Radiography vs. CT Both based on differential attenuation of x-rays passing through body Radiography.
Modulation Transfer Function (MTF)
디지털 래디오그라피 디텍터의 성능 -Modulation Transfer Function- 6 Nov 2014 Seungman Yun Radiation Imaging Laboratory, School of Mechanical Engineering, Pusan National.
Computed Tomography Computed Tomography is the most significant development in radiology in the past 40 years. MRI and Ultrasound are also significant.
ACR CT Accreditation Dose Pass/Fail Criteria and Reference Levels ExaminationPass/Fail Criteria CTDIvol (mGy) Reference Levels CTDIvol (mGy) Adult Head8075.
Introduction to Medical Imaging Week 2: X-ray and CT
Computed tomography. Formation of a CT image Data acquisitionImage reconstruction Image display, manipulation Storage, communication And recording.
Basic Principles of CT Scanning. CT CT - Computed Tomography CAT Scan - Computerized Axial Tomography.
IMAGE QUALITY. SPATIAL RESOLUTION CONTRAST RESOLUTION NOISE IMAGE ARTIFACTS RADIATION DOSE.
CT Scanner Detection system. 2 CT Scanner n Generator s High frequency, kW n X-ray tube s Rotating anode, high thermal capacity: 3-7 MHU s Dual.
CT Scanning Dr. Craig Moore
CT Multi-Slice CT.
Role of Adaptive Statistical Iterative Reconstruction (ASIR) in lowering radiation dose for pediatric head CT Electronic Poster-eP 140 Azadeh Tabari, MD.
CT PRINCIPLES AYMAN OSAMA.
Computed Tomography Image Reconstruction
Physics Testing for ACRIN 6678 Protocol version 3.0 QIBA Group 1C
Basic principles Geometry and historical development
Computed Tomography (C.T)
Computed Tomography (C.T)
Computed Tomography (C.T)
Presentation transcript:

CT Scanning: Dosimetry and Artefacts Dr. Craig Moore Medical Physicist & Radiation Protection Adviser Radiation Physics Service CHH Oncology

Operator Controlled Variables and their effects on Image Quality and Patient Dose &

Imaging Performance Spatial Resolution Z-sensitivity Significance Factors affecting resolution Z-sensitivity Factors affecting z-sensitivity

Modulation Transfer Function (MTF): Details contrast in image relative to contrast in object MTF50, MTF10, MTF2, MTF0 are often quoted MTF2 approximates to the limit of visual resolution

In-plane Spatial Resolution Maximum that can be achieved is about 20 lp/cm (but usually less) Typical matrix size is 512 x 512 In the scan plane limited by pixel size Pixel size = FOV/matrix i.e. if FOV = 40cm and matrix = 512 then pixel size = 0.8 mm Pixel size determines limit of resolution but there are other factors that affect resolution on a scanner

In-plane Spatial Resolution Other factors affecting resolution: Filter used for back-projection Size of focal spot, geometry of the scanner and size of detectors Sampling frequency (number of times the x-ray beam is sampled as it rotates around the patient) Spatial resolution the same for axial and helical scanning

z-plane Resolution Spatial resolution in the z-direction (parallel to patient) dependent on pitch The greater the pitch the lower the resolution

Summary of spatial resolution In-plane spatial resolution and z-resolution are usually thought of as different parameters In fact, z-resolution is an extension of spatial resolution in third dimension Multi-slice scanning is moving CT away from slice based medium to a truly 3D modality Structural features are 3D, so resolution should be equal in all dimensions

Image Noise

Sources of noise (1) Quantum Noise: Randomness of photon detection This type of noise is the most dominant in CT

FOV

Sources of Noise (2) Structure noise: Affected by back-projection filters

Sources of Noise (3) Electronic noise – small compared to other sources

Image Contrast & Noise Contrast is equivalent to the difference in CT number between an object and its surrounding tissue When viewing objects which have CT numbers close to background noise can mask detail Increased noise Increased contrast Pixel number

Low Contrast Resolution: Measure of how well a system can differentiate between an object and its background having similar attenuation coefficients Low Medium High

Low contrast exams account for approximately 90% of CT scans Low contrast resolution is important when small contrast differences are crucial for early detection of disease Low contrast exams account for approximately 90% of CT scans Affected by all parameters that influence noise Minimum detectable contrast is <0.5% Low contrast of a few HU

Principles of CT Dosimetry

Dose in CT One of the highest dose techniques used in medical imaging Within the UK it has been shown to contribute to 40% of the total dose attributable to medical exposure But only 4% of the total number of exams

Radiation Units Absorbed dose in CT Radiation risk in CT CT Dose Index (CTDI) (mGy) Radiation risk in CT Dose Length Product (DLP) (mGy cm) Effective dose (mSv)

Dose Distribution in CT Absorbed dose not a single value Dose values vary with position in patient in scan plane and along z axis

Dose Distribution Depends on More uniform for Periphery:centre Filtration Beam shaping Scanner geometry Size of patient More uniform for Higher filtration Optimised beam shaping Smaller patient Periphery:centre Body 2:1 Head 1:1

CT Dose Index (CTDI) Measure of dose in the scan plane from a single rotation CTDI defined as: Dose at position z, Dz is integrated over the complete dose profile and divided by slice thickness T

Dose Profile along z patient

Measurement of CTDI Routinely measured with air filled pencil ion chambers Use PMMA phantoms to simulate patient Single rotation with chamber: At the centre At the edges Calculate CTDIcentre and CTDIedge

Complete cross section of dose

HIGHER DOSE ‘MEDIUM’ DOSE LOWER DOSE

Pitch typically between 1 and 2

HIGHER DOSE HIGHER DOSE LOW DOSE (mA typically 100 to 200)

HIGH DOSE LOW DOSE MUST INCREASE mA and/or s kV typically between 80 and 140 kV

For the same pitch

Effective Dose To estimate the stochastic risk to the patient, must consider scan length (DLP) and anatomical location of scan Effective dose used This is the equivalent whole body dose

x 7

Effective Dose (mSv) CT is one of (if not THE), highest dose x-ray modalities used in the hospital Head – 2 mSv Chest – 8 mSv Abdomen and pelvis – 10 mSv CT KUB – 10-15 mSv Planar chest – 0.015 mSv CT chest is 600 times the dose of planar chest Planar IVU has effective dose 1.5 – 3 mSv CT KUB is up to 10 times the dose of planar IVU

MUST BE CAREFUL TO OPTIMISE DOSE – HAVE SPECIFIC INFANT PROTOCOLS

Overview of dose reduction methods X ray beam filtration Harden the beam so to reduce low energy photons Shape the beam to the relevant anatomical region X ray beam collimation The beam should be limited to the minimum dimensions required Tube potential, current and time Tube current modulation and AECs Tube current is chosen to maintain predetermined level of noise Size or Weight Based Technique Charts CT image never appears overexposed Standardise kVp, mA and scan time Detector efficiency No detector is 100% efficient at converting photon into signal Modern scanner have 90% efficiencies or above Noise reduction algorithms Smooth noise without reducing fine detail

Contrast and Special Uses of CT CT is being used increasingly for interventional work Special sequences can allow 3D images at 5 frames per second for biopsy needle placement Alternatively needle can be moved between scans in a ‘step and shoot’ method Contrast given to enhance the visibility of certain structures (i.e. increase the CNR) Should be given with caution or avoided altogether in patients with poor kidney function eGFR 30 – 60 with caution Below 30 significant risk Doctor must be available on-call in case of anaphylactic reaction (treatment of which includes intramuscular adrenaline, coticosteroids and antihistamines) Modern scanners can scan the heart in less than 0.5s is one rotation, thus avoiding motion artefacts of the beating heart

Image Artefacts What are artefacts? Systematic discrepancies between the CT numbers in the reconstructed image and the true attenuation coefficients of the object Non-random, or structured image noise

Non-random, or structured, image noise Artefact Origins Physics based Beam hardening Partial volume effects Photon starvation Undersampling Patient based Presence of metal Motion Scanner based Detector sensitivity Mechanical instability Non-random, or structured, image noise

Beam Hardening Artefacts As beam passes through patient, low energy photons are filtered and the beam becomes harder This causes the attenuation coefficient and CT number for a given tissue to decrease along the beam path. Reconstruction process assumes a monoenegetic beam CT numbers are lower in the centre than they should be Cupping Streaks

Avoiding beam hardening artefacts Adequate beam filtration Flat filter Bow tie filter Correction factors built in to calibrate Beam hardening correction software Avoidance of bony regions when possible Patient positioning and gantry tilt

Objects smaller than a voxel its density will be averaged across the whole voxel so it will appear bigger and less dense

Can still be a significant problem in images even with software correction

Mechanical Instability X-ray tube rotor wobble rotates at ~10,000 rpm high temperatures high frequency rotor wobble causes beam deviations streak artefacts result With rotor wobble Without rotor wobble

Artefacts in Helical Scanning In general, the same as in conventional scanning Additional artefacts occur due to interpolation Worse at higher pitches Due to changing structure in z-direction (e.g. skull)

Helical artefact of spherical phantom at pitch 2 Bilateral subdural haemorrhage or helical artefact?

Minimization Reduce effects of variation along z-axis Pitch of 1 rather than higher pitch 180° rather than 360° interpolation Thin slices

Conclusions Artefacts originate from a range of sources and can degrade the diagnostic quality of an image Some can be partially corrected for in software Good scanner design, careful positioning of patient and optimum selection of scan parameters can minimise the artefacts present in an image

CT Summary CT scanner generate images in transaxial slices CT number represents average linear attenuation coefficient in the voxel Contrast in the displayed image is enhanced by windowing The scanner gantry carries the X-ray tube and generator and a curved bank of detectors The image is reconstructed by filtered back projection Slip ring technology allows the gantry to rotate continuously In helical scanning, the patient is moved through the gantry while the gantry rotates Helical pitch is defined as ratio of table movement per rotation to slice thickness Multislice scanners have several rows of detectors that collect data simultaneously Multislice scanning can produce 3D images In-plane resolution is at best 20 lp/cm Quantum noise is the main limit ti low contrast resolution Dose is generally higher for CT than other X-ray modalities