High performance liquid chromatography (HPLC)

Slides:



Advertisements
Similar presentations
Gas Chromatography Introduction 1.) Gas Chromatography
Advertisements

Gas Chromatography & Gas-Liquid Chromatography
Gas Chromatography There an be many parts to a gas chromatography system but the basic components include: An injection system. A column (controllable.
Biology and Chemistry Chemical Biology research uses the tools of chemistry and synthesis to understand biology and disease pathways at the molecular level.
HPLC 1. Introduction 1.Introduction CHROMATOGRAPHY Chromatography basically involves the separation of mixtures due to differences in the distribution.
1 1. Introduction H: High P : Performance (Pressure) L : Liquid C : Chromatography GC : Gas chromatography TLC: Thin layer chromatography IC : Ion chromatography.
Chromatography and Instrumentation. Invented by a Russian Botanist Mikhail Tswett in 1903 He used chromatography to separate the colour pigments in plants.
Gas Liquid Chromatography
B IOCHEMICAL INSTRUMENTAL ANALYSIS -11 Dr. Maha Al-Sedik.
Pn. Suryati Bt. Syafri 2009 High Performance Liquid Chromatography (HPLC)
CHAPTER 29 Supercritical Fluid Chromatography The mobile phase is a supercritical fluid (a fluid above its critical T and critical pressure) Supercritical.
Chromatography Year 12.
LECTURE 4: CHROMATOGRAPHY Content: - Chromatographic separation - classifying analytical separations; column chromatography, planar chromatography - gas.
High Performance Liquid Chromatography High Performance Liquid Chromatography Chem. 331.
Prabh Jassal SUPA Che113 Forensics Ms. Tolentino.
Chapter 6 - Chromatography
Introduction to High Performance Liquid Chromatography.
High Performance Liquid Chromatography
Chromatography Chapter 6.
Magnet Analytical Chemistry Unit 4
Organic Chemistry I - Lab
By: Thilag.k & Stephen. What is Hpcl??? Hplc or high performance liquid chromatography is the most widely used analytical separation technique. The difference.
GC Advantages 1. Very Large N (Very Long Columns) 2. No Packing Material (A=0) 3. Simple Mobile Phase (Compressed Gas) 4. Universal Detectors (FID) 5.
Chromatography vObjective  To understand the principles of chromatography and know the specific types of Chromatograph used in the analysis of environmental.
By: Arafath and Nick. What is it  Chromatography is a technique that is used to separate the substances present in a mixture.  It is widely used to.
By- Bhavya, Harsh, Harshvardhan, Namrata, Ronit and Vidhatri
HPLC – High Performance Liquid Chromatography
Introduction  High-performance liquid chromatography (HPLC) is a form of liquid chromatography.liquid chromatography  The main purpose is to separate.
Best Broken into four categories
Chromatography High Performance Liquid Chromatography HPLC Chapter Dr Gihan Gawish.
HPLC.
High Performance Liquid Chromatography
Chapter 28 High Performance Liquid Chromatography.
Biochemical instrumental analysis - 11 Dr. Maha Al-Sedik 2015 CLS 332.
HPLC.
Chromatography and Instrumentation. Chromatography Separate Analyze Identify Purify Quantify Components Mixture Chromatography is used by scientists to:
HPLC 1. Introduction 1.Introduction  INSTUMENTAL ANALYSIS  PRACTICAL 213 PHC  HPLC.
Organic Analysis Basic concepts. Elements and Atoms Fundamental building block of all substances is the element. Fundamental building block of all substances.
Principles of chromatography
Chemical analysis as part of Quality Control Overview Learn how analytical chemistry techniques such as chromatography and volumetric analysis can be.
Biochemical instrumental analysis - 9 Dr. Maha Al-Sedik 2016 CLS 332.
High Performance Liquid Chromatography. What is HPLC ? It is a separation technique that involves: Injection of small volume of liquid sample Into a tube.
High Performance Liquid Chromatography Presented by Dr. Kamal Modi 2 nd Year Resident.
Overview Determining Identity Quantitative Analysis
Chemical Ideas 7.6 Chromatography.
LU 3: Separation Technique (P2)
GAS CHROMATOGRAPH. GAS CHROMATOGRAPH Principle: Which type of compounds can analyse? Volatile Thermostable.
Chromatography- TLC & HPLC
Chem. 133 – 5/11 Lecture.
CHROMATOGRAPHY.
Chem. 31 – 11/1 Lecture.
HPLC.
High Performance Liquid Chromatography (HPLC)
HERBAL DRUG FORMULATION Presented by : Department of Natural Product Research Acube Lifesciences.
Chromatographic separation
Best Broken into four categories
HPLC.
Chapter: Chromatography
Satish Pradhan Dnyanasadhana College, Thane (W)
High Performance Liquid Chromatography
BASICS OF SUPERCRITICAL FLUID CHROMATOGRAPHY
VAPOUR PHASE CHROMATOGRAPHY
Password: Password: INTRODUCTION TO HPLC
What is chromatography?
Chromatography Daheeya Alenazi.
High Performance Liquid Chromatography
Chapter: Chromatography
High-Performance Liquid Chromatography
GAS CHROMATOGRAPHY.
High-Performance Liquid Chromatography
Presentation transcript:

High performance liquid chromatography (HPLC) Separate, identify & quantify components dissolved in a liquid solvent with a high analytical resolution. Sample carried by a moving gas stream of Helium or Nitrogen

High Performance Liquid Chromatography (HPLC) is a form of column chromatography that pumps a sample mixture or analyte in a solvent (known as the mobile phase) at high pressure through a column with chromatographic packing material (stationary phase). The sample is carried by a moving carrier gas stream of helium or nitrogen. HPLC has the ability to separate, and identify compounds that are present in any sample that can be dissolved in a liquid in trace concentrations as low as parts per trillion. Because of this versatility, HPLC is used in a variety of industrial and scientific applications, such as pharmaceutical, environmental, forensics, and chemicals.

Retention Time: The retention time of a solute is taken as the elapsed time between the time of injection of a solute and the time of elution of the maximum peak of that solute. Sample retention time will vary depending on the interaction between the stationary phase, the molecules being analyzed, and the solvent, or solvents used. As the sample passes through the column it interacts between the two phases at different rate, primarily due to different polarities in the analytes. Analytes that have the least amount of interaction with the stationary phase or the most amount of interaction with the mobile phase will exit the column faster.

Instrumentation: Main components in an HPLC system include the solvent reservoir, or multiple reservoirs, a high-pressure pump, a column, injector system, the detector and waste.

The reservoir holds the solvent, which is referred to as the mobile phase because it moves. There are usually a minimum of two reservoirs in a system, with each holding up to 1000 cc of solvent and usually fitted with a gas diffuser through which helium can be bubbled. A pump is used to generate a specified flow of the mobile phase. Although manual injection of samples is still possible, most HPLCs are now fully automated and controlled by computer.

The injector, or auto sampler, introduces the solvent into a phase stream that carries the sample into the high pressure (up to 400 bar) column, which contains specific packing material needed to effect separation. The packing material is referred to as the stationary phase because it is held in place by the column hardware. A detector is needed to see the separated compound bands as they elute from the high pressure column. The information is sent from the detector to a computer which generates the chromatogram. The mobile phase exits the detector and is either sent to a waste, or collected, as desired.

Separation: mAU time Start Injection Injector Chromatogram Mixer Pumps Detector Column Solvents Mixer Pumps Chromatogram Start Injection mAU time

Separations mAU time Start Injection Injector Chromatogram Mixer Pumps Detector Column Solvents Pumps Mixer Chromatogram Start Injection mAU time

Separations mAU time Start Injection Injector Chromatogram Mixer Pumps Detector Column Solvents Pumps Mixer Chromatogram Start Injection mAU time

Separations mAU time Start Injection Injector Chromatogram Mixer Pumps Detector Column Solvents Pumps Mixer Chromatogram Start Injection mAU time

Separations mAU time Start Injection Injector Chromatogram Mixer Pumps Detector Column Solvents Pumps Mixer Chromatogram Start Injection mAU time

Separations mAU time Start Injection Injector Chromatogram Mixer Pumps Detector Column Solvents Pumps Mixer Chromatogram Start Injection mAU time

Separations mAU time Start Injection Injector Chromatogram Mixer Pumps Detector Column Solvents Pumps Mixer Chromatogram Start Injection mAU time

Separations mAU time Start Injection Injector Chromatogram Mixer Pumps Detector Column Solvents Pumps Mixer Chromatogram Start Injection mAU time

Separations mAU time Start Injection Injector Chromatogram Mixer Pumps Detector Column Solvents Pumps Mixer Chromatogram Start Injection mAU time

Separations mAU time Start Injection Injector Chromatogram Mixer Pumps Detector Column Solvents Pumps Mixer Chromatogram Start Injection mAU time

Separations mAU time Start Injection Injector Chromatogram Mixer Pumps Detector Column Solvents Pumps Mixer Chromatogram Start Injection mAU time

Separations mAU time Start Injection Injector Chromatogram Mixer Pumps Detector Column Solvents Pumps Mixer Chromatogram Start Injection mAU time

Separations mAU time Start Injection Injector Chromatogram Mixer Pumps Detector Column Solvents Pumps Mixer Chromatogram Start Injection mAU time

Separations mAU time Start Injection Injector Chromatogram Mixer Pumps Detector Column Solvents Pumps Mixer Chromatogram Start Injection mAU time

Separations mAU time Start Injection Injector Chromatogram Mixer Pumps Detector Column Solvents Pumps Mixer Chromatogram Start Injection mAU time

Types of HPLC: Normal–phase chromatography. Reversed-phase chromatography.

Normal–phase chromatography Normal–phase chromatography was one of the first kinds of HPLC that chemists developed. Also known as normal-phase HPLC (NP-HPLC) this method separates analytes based on their affinity for a polar stationary surface such as silica NP-HPLC uses a non-polar, non-aqueous mobile phase (e.g. Chloroform), and works effectively for separating analytes readily soluble in non-polar solvents. The analyte associates with and is retained by the polar stationary phase

Reversed-phase chromatography: Reversed phase HPLC (RP-HPLC) has a non-polar stationary phase , moderately polar mobile phase. One common stationary phase is a silica which has been surface- modified with RMe2SiCl, where R is a straight chain alkyl group such as C18H37 or C8H17. With such stationary phases, retention time is longer for molecules which are less polar, while polar molecules elute more readily (early in the analysis). An investigator can increase retention times by adding more water to the mobile phase

Stationary Phases: Polar (Normal Phase): Silica, alumina Non-Polar (Reversed Phase): ODS Silica gel, C18, C8

Chromatography Stationary Phases Silica Gel Derivatized Silica Gel Where R = C18H37 hydrocarbon chain (octadecylsilyl deriv. silica or “C18”) relatively polar surface relatively nonpolar surface “normal phase” “reversed phase”

Normal vs. Reversed Phase Chromatography

TLC vs. HPLC:

Application of HPLC To control the drug stability 1. Pharmaceuticals industry To control the drug stability Quantity of drug determination from pharmaceutical dosage forms, ex. Paracetamol determination in panadol tablet Quantity of drug determination from biological fluids, ex: blood glucose level 2. Analysis of natural contamination Phenol & Mercury from sea water 3. Forensic test Determination of steroid in blood, urine & sweat. 4. Clinical test

HPLC Applications Consumer Products Bioscience Chemical proteins Environmental Pharmaceuticals Consumer Products Clinical polystyrenes dyes phthalates tetracyclines corticosteroids antidepressants barbiturates amino acids vitamins homocysteine Bioscience proteins peptides nucleotides lipids antioxidants sugars polyaromatic hydrocarbons Inorganic ions herbicides

The Factors Which Influence The HPLC Performance: Internal diameter of column The smaller in diameter, the higher in sensitivity. Pump pressure The higher in pressure, the higher in separation. Sample size The polarity sample, solvent and column Temperature The higher in temperature, the higher in separation.

Advantages: Needs a small sample with a high accuracy and precis Non-destructed sample during operation compared to GC.

Disadvantages: Need a skill to run the instruments. Solvents consuming.