Heartware HVAD Ventricular Assist System

Slides:



Advertisements
Similar presentations
Managing the LVAD patient in Cardiac Rehabilitation
Advertisements

Community Preparation for Caring for Mechanical Circulatory Device Patients University of Wisconsin Hospital And Clinics Ventricular Assist Device Program.
Left Ventricular Assist Devices “LVAD”
BACKGROUND Ventricular assist devices (VADs) are a proven therapy as bridge-to-cardiac transplantation in Class IIIB and Class IV heart failure patients.
1 CAUTION: Federal Law (USA) restricts this device to sale by or on the order of a physician. Refer to the “Instructions For Use” for complete Indications.
Monitoring and Documentation Principles of IV Therapy BSN336.
BACKGROUND Ventricular assist devices (VADs) are a proven therapy as bridge-to-cardiac transplantation in Class IIIB and Class IV heart failure patients.
EMS Management of the SynCardia Temporary Total Artificial Heart (TAH-t) Patient Clinical-xxx Rev001.
LVAD is a battery-operated, mechanical pump-type device that's surgically implanted. Goal of LVAD: providing the patient with as close to a normal lifestyle.
Central Line Removal Competency Assessment for Registered Professional Nurses in the Critical Care Areas References: AACN Procedure Manual for Critical.
CAUTI Prevention.
5/24/ HEMODYNAMIC MONITORING. OBJECTIVE 5/24/ Describe the three attributes of circulating blood and their relationships. 2. Identify types.
Topics Zoll LifeVest  What it is  Who it treats  How does EMS handle these patients?
CENTRAL LINES AND ARTERIAL LINES
Community Preparation for Caring for Mechanical Circulatory Device Patients University of Wisconsin Hospital And Clinics Ventricular Assist Device Program.
Hemodynamic Monitoring By Nancy Jenkins RN,MSN. What is Hemodynamic Monitoring? It is measuring the pressures in the heart.
A guide for healthcare professionals Measuring Blood Pressure at Home Michigan Department of Community Health Heart Disease and Stroke Prevention Unit.
Mechanical Circulatory Support for Advanced Heart Failure Speakers: Lynne A Benish, ACNP-BC Mechanical Assist Device Nurse Practitioner, Cardiac Surgery.
Ventricular Assist Devices Brian Schwartz, CCP February 25, 2003.
Shannen Whiddon.  Cardiac tamponade is a condition in which cardiac filling is impeded by an external force.
Rejection Normal response Inflammation 25% of pt. will have acute rejection during the first year post transplant Causes: Previous Rejection Noncompliance.
Advances In LVAD Patient Management
Ventricular Assist Devices
HOME AND AMBULATORY BLOOD PRESSURE MONITORING
Transition to Next Level of Care Management of Patient with Ventricular Assist Device Hospital Home Transition.
What is it? What are my responsibilities as baby nurse?
Community Preparation for Caring for Mechanical Circulatory Device Patients University of Wisconsin Hospital And Clinics Ventricular Assist Device Program.
Top Ten things you need to know… About VAD’s Kim Byrum Chappell Mechanical Assist Coordinator.
1 Arterial Lines Set Up & Monitoring Union Hospital Emergency Department.
 Hub (IV catheter)  Maintenance (or primary) solution  TKO (KVO) rate  Parenteral  Mechanical gravity devices  EID: electronic infusion device 
DVT Prevention and Anticoagulant Management
Implanted Ports: Procedure for Access and Care
Ventricular Assist Devices A Ventricular assist device, or VAD, is a mechanical circulatory device that is used to partially or completely replace the.
Risk Factors for Adverse Outcome after HeartMate II Jennifer Cowger, MD, MS St. Vincent Heart Center of Indiana Advanced Heart Failure, Transplant, & Mechanical.
PCI What You Need to Know!. What and Where Radial- advantages  Immediate ambulation  Easily compressible vessel  Less risk of nerve injury  Dual blood.
Ventricular Assist Device: Intervals for dressing changes By. Megan Giska.
Cardiac Cath and Angiocardiography Adult II FINAL 2/2015.
Section VII. Home BP Measurement 2015 Canadian Hypertension Education Program Recommendations.
SNO IMAFIDON E.O (Mrs) July, INTRODUCTION Health care institutions and their patients are familiar with the effects of nosocomial infections (NI).
Hemodynamic Monitoring John Nation RN, MSN Thanks to Nancy Jenkins.
Hemodynamic Monitoring
National Patient Safety Goals (NPSG) Online Orientation -the purpose is to improve patient safety -the goals focus on problems in health care safety and.
Chapter 31 Medication Administration. Injections: Intravenous  Three methods:  As mixtures within large volumes of IV fluids  By injection of a bolus.
Starting an IV Hanging a Primary IV Solution Hanging an IV Piggy Back Programming an IV Pump Lab Skills.
Intravenous Infusion.
Patient Safety Overview 2014
Ventricular Assist Device
Use of High Flow Nasal Cannula and Aerosolized Epoprostenol as a Bridge to Lung Transplantation Sherwin Morgan, RRT, Steve Mosakowski, RRT, Stephanie Ostrawski,
Planned Implementation Date: June 18, 2013
University of Wisconsin Hospital And Clinics
2.13 Copyright UKCS #
Alarm Sound Tutorial.
The Call is Made Rhonda Duggan, BSN, RN, CCTC August 29, 2014
What’s New, UCAT?.
Traditional parenteral antihypertensive treatment
Blood Pressure August 2015 Blood Pressure.
Medication Safety & Anticoagulation
Polmunary edema.
Ventricular Assist Devices (VAD’S)
Hemodynamic effects of partial ventricular support in chronic heart failure: Results of simulation validated with in vivo data  Deborah Morley, PhD, Kenneth.
Henry Ford Health System
Mechanical Circulatory Support Devices HOSEIN PASANDI.
Durable Mechanical Circulatory Support in Advanced Heart Failure
Implantation of the Jarvik 2000 Heart
Advances and Future Directions for Mechanical Circulatory Support
Dominique Shum-Tim, MD, Brian W
Clinical experience with an implantable, intracardiac, continuous flow circulatory support device: physiologic implications and their relationship to.
Pericarditis Inflammation of the pericardium Many causes
URINALYSIS DIPSTICK TESTING
Presentation transcript:

Heartware HVAD Ventricular Assist System Last edited 9/12/13

About the Pump Miniaturized circulatory support system for treatment of advanced heart failure Can achieve up to 10L/min cardiac output Designed to be placed only in pericardium, no abdominal access as in competitors FDA approved Utilized as “Bridge to Transplantation” and as “Destination Therapy” Last edited 11/28/12

Last edited 11/28/12

Last edited 11/28/12

Last edited 11/28/12

Last edited 11/28/12

Last edited 11/28/12

Cannulation The inflow cannula is housed within the Pump Helps decrease pump size Patient is cannulated in LV apex and Ascending AO Last edited 11/28/12

Last edited 11/28/12

Always defaults to AC/DC power supply if that is plugged in Last edited 11/28/12

Last edited 11/28/12

Last edited 11/28/12

Last edited 11/28/12

Last edited 11/28/12

Last edited 11/28/12

Last edited 11/28/12

HeartWare®Back-up Controller Keep a back-up controller and extra fully charged batteries with the patient at all times in case of an emergency Verify that the back-up controller parameters match the primary controller parameters verify by looking at the settings sticker on the backside of the back-up controller Alarm parameters should be documented on sticker and updated when changed MD to update sticker when any changes to settings made The back-up controller should be programmed before the implant procedure, upon any parameter change to the primary controller, and when the primary controller is replaced Done by MD and documented in note Last edited 11/28/12

Last edited 11/28/12

Last edited 11/28/12

Last edited 11/28/12

Last edited 11/28/12

Last edited 11/28/12

Last edited 11/28/12

Last edited 11/28/12

Last edited 11/28/12

Only use bags provided from vendor Last edited 11/28/12

Last edited 11/28/12

Last edited 11/28/12

Last edited 11/28/12

Last edited 11/28/12

Last edited 11/28/12

This will be spot checked on 8E This will be spot checked on 8E. Patient does not go home with this monitor. Last edited 11/28/12

Last edited 11/28/12

This does not shut down the device Last edited 11/28/12

Last edited 11/28/12

Last edited 11/28/12

Last edited 11/28/12

Parameters Monitored Speed = RPMs Work = Watts Flow = Cardiac Output Normal 2400-3200 (capable of 1800-4000) Set rate by MD Measured number Work = Watts Measured Number 2.5-7.0 Alarm should be set 2 greater than what patient is running at Flow = Cardiac Output Calculated Number ( roughly Speed + work + HCT) Responsible that HCT programmed in device is within 5 of patients current HCT Alarm should be set 2L/min lower than patient current CO, but never less than 2 L/min Should be >2 L/min Last edited 11/28/12

Management post-operatively and beyond Control hypertension –maintain MAP < 90mmHg*** Continuous flow pumps are preload dependent and sensitive to increased afterload. ALWAYS : Consider clinical (patient) related changes when assessing changes in pump flow Fluids are given to maintain pump flow index at greater than 2.0 L/min/m2 with central venous pressure and left atrial pressure less than 20 mmHg Vasopressors and/or vasodilators can be used as required to adjust vascular tone Patients may require inotropic assistance of right ventricular function Last edited 11/28/12

Daily Care Essentials Always ensure that the controller is plugged into two power sources (red outlet on wall + battery or 2 batteries) A backup controller, programmed to the patient’s current settings, should always be available. Sticker on back of backup controller with patient information (MR, DOB) and settings Batteries should be changed when the battery shows 1 bar left (~25% battery life). Obtaining a blood pressure Continuous flow device, based on native function, may not have pulsatility BP obtained via art line tracing or with Doppler + manual cuff Doppler at brachial pulse to hear continuous flow Inflate cuff around upper arm Pressure at which you first hear flow = approximate MAP Can attempt BP with automated cuff but must confirm with doppler/manual Attempt pleth measurement if pt has pulsatility RVAD and BiVAD: Never open venous lines directly to air No trips to MRI! Listen & document heart sounds every shift Notify MD if changes noted Device is preload-driven: Consider potential causes for decreased preload: Tamponade Hypovolemia Changes in intrathoracic pressure Increased PVR R heart failure Device is afterload-sensitive, consider potential causes for increased afterload Pulmonary vs. systemic hypertension Consider afterload reduction, pulmonary vasodilators, analgesia/sedation Mobilize pt ASAP  Last edited 11/28/12

Nursing considerations Notify MD of suction events, even if intermittent Check Alarm Limits (on sticker; changed by MD in system) High Power alarm 2 watts above average power Low flow alarm 2L/min below average flow (minimum set 2L/min) Assure correct HCT is programmed for CO calculation Last edited 11/28/12

Arrhythmias/Emergency Procedures Arrhythmias may occur in the post-operative period OK to defibrillate HeartWare® System patients Anti-arrhythmic drugs, pacemakers, and ICDs are compatible with the HeartWare System Institute appropriate PALS/ACLS protocols If chest compressions have been administered, confirm function and positioning of HVAD® pump Last edited 11/28/12

Anticoagulation Anticoagulation should be individualized for each patient In general, begin low-dose heparin at 10 units/kg/hr on postoperative day one target PTT of 50-60 seconds Heparin gtt to be freshly made by Pharmacy and changed Q12 hours As patients tolerate oral medication, they should be started on warfarin and aspirin Warfarin should be titrated to maintain the INR of 1.7 to 3 Discontinue heparin when the INR is in therapeutic range. Use low dose Other Labs: Unfractionated heparin levels (0.35-0.5) PTT TEG ATIII Last edited 11/28/12

Dressing Care Good hand-washing technique Always use sterile technique: hats, masks & sterile gloves Initially daily and prn if saturated After 1 week, QOD and prn After 14 days, 2 x per week and prn Dressing: Cleanse site & surrounding tissue with chlorhexidine 2% wand Place 2 x 2 gauze over site & cover any area on drive line with exposed velour coating with gauze (tegaderm that is placed directly on velour coating is a significant infection risk) Cover with tegaderm & date dressing Securement device needed for drive line Document in Powerchart Last edited 11/28/12

Performs safety checks for Heartware system At bedside Back up controller programmed with current patient’s settings Battery charger with back up batteries charging and labeled Doppler and manual BP cuff In Cardiac Storage room, 3rd floor, OR Surgical Back-up equipment Drive line extension cable Outflow graft for pump Another pump Necessary surgical insertion tools 2 monitors 4 batteries and battery charger 2 controllers 2 AC adapters Another patient pack (Velcro holder) Last edited 11/28/12

Last edited 11/28/12

Last edited 11/28/12

Last edited 11/28/12

Last edited 11/28/12

Last edited 11/28/12

Trouble shooting Page “LVAD” for heart failure team www.heartware.com cs@heartwearinc.com Heartware 205 Newbury Street, Suite 101 Framingham, MA 01701 24- Hour Clinical Support 888.494.6365 Last edited 11/28/12