The visible-UV spectrometer we use permits the quantitative determination of the absorption and transmission of light in the wavelength range 200 nm*

Slides:



Advertisements
Similar presentations
Spectrophotometer.
Advertisements

Introduction to Spectrophotometry & Beer’s Law
Spectroscopy and Beer’s Law
An introduction to Ultraviolet/Visible Absorption Spectroscopy
What factors affect the intensity of color
Use of the Spectronic 20 Spectrophotometer
SPECTROPHOTOMETRY IN BIOTECHNOLOGY
Introduction to Spectrophotometry
Introduction to Spectrophotometry
Introduction to Spectrophotometry. Why Spectrophotometry? Imagine you are to make a 1μM solution of a specific protein that you believe could have anti-carcinogenic.
Lecture 2b. Electromagnetic Spectrum Visible range: = nm Ultraviolet: = nm Low energyHigh energy.
Ultraviolet – Visible Spectrophotometer. What is UV – Vis spectroscopy? Ultraviolet – visible spectroscopy is a measure of the absorbance of radiation.
Spectrophotometry and Plotting of Calibration Curve
Principles of instrumentation Prepared by: Ibtisam H. AlAswad Reham S. Hammad.
Dilution 2003 Required D. Information Given A student is instructed to determine the concentration of a solution of CoCl 2 based on absorption of light.
Introduction to Spectrochemical Methods
Spectrophotometer Prof.Dr. Moustafa M. Mohamed Vice Dean Faculty of Allied Medical Science Pharos University in Alexandria, EGYPT.
spectrophotoMETER Dr. Beenish Zaki, Instructor
Fluorometric determination of riboflavin
Fluorometric determination of riboflavin. Introduction.
UV-Vis Absorption Spectroscopy
Dilution and Spectroscopy Lab Lab Word Document Lab Word Document.
Spectrophotometry and Plotting of Calibration Curve BIO-2.
Simple Double Beam Spectrometer
Simple Double Beam Spectrometer
Visible Spectrophotometer
SPECTROPHOTOMETRY IN BIOTECHNOLOGY
(glass pipette, centrifuge, spectrophotometry)
Chapter 13 An Introduction to Ultraviolet- Visible Molecular Absorption Spectrometry.
Electromagnetic radiation travels through space in the form of a wave, with the distance between two crests of a wave termed a wavelength. At one end.
Exp. 18: Spectrophotometric Analysis: Concentration of a Solution Using Beer’s Law White light violet (400nm) – red (800nm) “visible spectrum” sample (light.
Spectrophotometry.
Serial Dilution Activity Using a pipet Simulated Acid: Red = nitric acid Yellow = sulfuric acid Blue = hydrochloric acid.
Spectrophotometer(2) Prof.Dr. Moustafa M. Mohamed.
Spectrophotometer.
SPECTROPHOTOMETRY. Principle : there is interaction between the light and sample particle, spectrophotometer is employed to measure the amount of light.
CLS 332 CLINICAL INSTRUMENTAL ANALYSIS. A VISIBLE ABSORPTION SPECTROMETER.
Spectrophotometry Ability of molecules to absorb and transmit light energy is the basis of one of the most widely used procedures for determining the concentration.
Introduction to Spectrophotometry & Beer’s Law
AP Free Response Questions 2003
Fluorometric determination of riboflavin
SPECTROPHOTOMETRY IN BIOTECHNOLOGY. LIGHT IS A TYPE OF ELECTROMAGNETIC RADIATION Imagine electromagnetic radiation like waves on a pond – But instead.
Absorption Spectroscopy CHEM 251 Week of November 1 st, 2010 Alexis Patanarut.
Spectrophotometry at a Glance
COLORIMETRY & SPECTROPHOTOMETR
Lab1 A VISIBLE ABSORPTION SPECTROMETER. -One of the simplest and most widely used methods to determine concentration of a substance in solution -Measures.
Theory of Spectrophotometry
Introduction to Spectrophotometry
Simple Double Beam Spectrometer
Lab1 A VISIBLE ABSORPTION SPECTROMETER
Spectroscopy Techniques
SPECTROPHOTOMETRY White light contains all wavelengths of light in the visible region of the EM. A substance appears colored because it absorbs light.
Principles and practice of Spectrophotometer
Spectrophotometer - Visible Type Spectro - photometer consists of two instruments : Spectrometer   Photometer  Spectrometer   for producing light.
Introduction to Spectrophotometry
Introduction to Spectrophotometry
Spectrophotometric Determination [Co2+]
Spectrophotometer Dr . S. Jayakumar.
SPECTROSCOPY.
SPECTROPHOTOMETRY White light contains all wavelengths of light in the visible region of the EM. A substance appears colored because it absorbs light.
SIDEBAR: PLANTS ABSORB A LOT OF LIGHT ENERGY…
Utilizing Spectrophotometry in Life Science
Lecture 2b Beer’s Lambert Law.
Let’s Talk About Beer’s Law
Beer's- Lambert Law and Standard Curves of concentrations
Let’s Talk About Beer’s Law
Colourimetry The absorption of light by a coloured solution is directly related to the concentration of the solution. Colourimetry determines the concentration.
Spectrophotometry A method to determine concentration of a species exploiting the absorption of EMR.
Spectrophotometry (A type of Spectroscopy)
Clinical instrumental analysis
Presentation transcript:

The visible-UV spectrometer we use permits the quantitative determination of the absorption and transmission of light in the wavelength range 200 nm* nm by a clear liquid sample. The device consists of simple components. * We actually encounter a lower wavelength limit of ~350 nm due to the use of glass sample containers. Glass absorbs significant amounts of radiation below 350 nm.

sample compartment wavelength adjustment zero adjustment wavelength switch 100% transmittance adjustment mode switch The Apparatus display

Variations in Spectrometers do not play a role Some models of spectrometers include other switches and capabilities which do not play a role in our exercises, e.g., mode switches labeled –concentration or factor wavelength switchSome models of spectrometers do not include the wavelength switch shown in the preceding picture. analog digitalA few of the spectrometers have an analog instead of digital output. While the details may differ slightly, the basic procedures described herein are the same for all the models. wavelength switch

Step 1 - Set Wavelength If spectrometer has wavelength switch, be sure is it set properly Adjust wavelength using wavelength adjustment knob wavelength reading

Initializing the Spectrometer Once the desired wavelength is set, we adjust the electronics at the selected wavelength. We must set (or confirm) that spectrometer displays: –0% transmittance when no light reaches the detector When sample compartment cover is closed with no sample in chamber, the light beam is interrupted internally. –100% transmittance when the sample absorbs no light at the given wavelength This latter step requires the preparation of a BLANK, BLANK: a solution containing all the substances in the solution except the one whose absorption we seek to determine.

Step 2 - Set/Check 0% Transmittance EMPTY SAMPLE CHAMBER Sample compartment cover closed Adjust transmittance to zero Mode switch set to transmittance transmittance reading

0 (zero) percent transmittanceno light Having adjusted the electronics so the detector reads 0 (zero) percent transmittance when no light reaches it, 100 percent transmittance blankno sample We next want to insure that the detector reports 100 percent transmittance (at the appropriate wavelength) when the blank, but no sample, is in the beam. The setting of 100% is a critical reference point. sample It does not mean that the blank does not absorb light at this wavelength, but that we will measure only the absorption due to the sample.

Step 3 - Set/Check 100% Transmittance Place blank in sample compartment. Cover closed Adjust transmittance to 100% Mode switch set to transmittance transmittance reading DO NOT TOUCH THIS SETTING

Initializing the Spectrometer Major changes in the 0% or 100% controls may require repeating both settings to be sure that adjusting one of them has no effect on the other –I.e., confirm that transmittance is 0% with no sample after setting 100% with blank If 0% or 100% have changed, repeat the initialization steps In principle, the 0% setting should need no adjustment after the spectrophotometer electronics have stabilized. In practice, small adjustments may be required.

Step 3a – Re-Check 0% Transmittance Remove Blank Sample compartment cover closed Re-Adjust transmittance to zero ONLY if necessary Mode switch set to transmittance transmittance reading

After the Spectrometer is Initialized at the set wavelength Once the 0% and 100% settings are stable, the spectrometer is ready to measure samples at the set wavelength the subsequent steps depend on whether the spectrometer will be used: fixed wavelength to make a series of measurements at the same fixed wavelength (e.g., in a Beers Law determination) change wavelength to change wavelength (e.g., to determine the absorption spectrum of a sample) Only periodic checks of the 0 and 100% settings are necessary The 100% setting must be readjusted each time the wavelength is changed OR

Summary 1.Prepare Blank 2.Turn on spectrophotometer 3.Adjust wavelength setting 4.Adjust zero setting with empty sample compartment 5.Adjust 100% setting with blank in sample compartment 6.If working at fixed wavelength, check both zero and 100% settings periodically 7.If working at varying wavelengths, adjust 100% setting whenever wavelength is changed. Also check zero setting periodically.