MCQs (1) – Questions Ultrasound propagates through the medium as a transverse wave Ultrasound velocity is equal to the product of frequency and wavelength.

Slides:



Advertisements
Similar presentations
Chapter3 Pulse-Echo Ultrasound Instrumentation
Advertisements

* NEAR ZONE is the nearest area to the transducer & it always has the same diameter of the transducer, BUT if the transducer: Linear the near zone length.
Foundations of Medical Ultrasonic Imaging
Handbook of Medical Informatics
Sound Chapter 15.
Ultrasound Dr.mervat mostafa.
Hospital Physics Group
Pegasus Lectures, Inc. Volume II Companion Presentation Frank Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition.
ECHO BASICS PHYSICS AND INSTRUMENTATION
Basic Physics of Ultrasound
ECE 501 Introduction to BME
Ultrasonic Nonlinear Imaging- Tissue Harmonic Imaging.
Ultrasound Imaging Atam Dhawan.
ESAT 3640 Therapeutic Modalities
Welcome.
Sound In Medicine. Sound waves Sound is a mechanical waves travel in air as longitudinal waves in which vibrations in air causes increases (compressions)
GENERATING AND DETECTING OF ULTRASOUND
Ultrasound – Physics & Advances
DOPPLER SONOGRAPHY By Dr/ Dina Metwaly.
THE ULTRASOUND IMAGE: GENERATION AND DISPLAY
Ultrasound Physics Sound is a mechanical, longitudinal wave that travels in a straight line Sound requires a medium through which to travel Ultrasound.
Ultrasonography.
Doppler echocardiography & Magnetic Resonance Imaging.
George David Associate Professor Ultrasound Physics 04: Scanner ‘97.
Resident Categorical Course
High School Physics Day April 3 rd, 2013 Ultrasound Physics & Instrumentation.
Medical Imaging Ultrasound Edwin L. Dove 1412 SC
Display of Motion & Doppler Ultrasound
Review of Ultrasonic Imaging
Sarah Gillies Ultrasound Sarah Gillies
Introduction to the Physics Echocardiography Introduction to the Physics Echocardiography Jose L. Rivera, M.D. January 9, 2010.
© Houghton Mifflin Harcourt Publishing Company Preview Objectives The Production of Sound Waves Frequency of Sound Waves The Doppler Effect Chapter 12.
Ultrasound Physics Reflections & Attenuation ‘97.
Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc.
University of Kurdistan Food Quality Evaluation Methods (FQEM) Lecturer: Kaveh Mollazade, Ph.D. Department of Biosystems Engineering, Faculty of Agriculture,
Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition.
1 Linear Wave Equation The maximum values of the transverse speed and transverse acceleration are v y, max =  A a y, max =  2 A The transverse speed.
Ultrasound Basis Michel Slama Amiens.
Piezoelectric Effect  Sound waves striking a PZ material produce an electrical signal  Can be used to detect sound (and echoes)!
Artifacts Ultrasound Physics George David, M.S.
Ultrasound. History: Available in 19th century. Available in 19th century. Was for sonar (SONAR  Sound Navigation and Ranging) Was for sonar (SONAR 
Chapter 12 Preview Objectives The Production of Sound Waves
Ultrasound Robert L. Metzger, Ph. D.. RAPHEX General Question 2001 D53: In ultrasound, the largest reflections will occur between: A. Liver and muscle.
PSF What is PSF 1way/2way PSFs PSFs for continious & discrete apertures.
Ultrasound Learning Objectives: Describe the properties of ultrasound;
Fundamental Ultrasound Principles Karen Potts Clinical Scientist Review date Jan 2010 Department of Medical Physics Kent & Canterbury.
TISSUE HARMONIC IMAGING (THI)
 Ultrasound waves are longitudinal with high frequencies ( ≈ > 20,000 Hz, though medical Ultrasound is between 1 to 15 MHz.)  When an ultrasound reaches.
TISSUE HARMONIC IMAGING (THI)
TISSUE HARMONIC IMAGING NUR FASHIHA BINTI AZMAN A DIAGNOSTIC IMAGING AND RADIOTHRAPY /2.
NUR INANI BINTI MOHD RUZLAN A Tissue harmonic imaging is a technique in ultrasonography that provides images of better quality as compared to conventional.
The Physics of Diagnostic Ultrasound
Ultrasound Physics Image Formation ‘97.
Lecture on Ultrasonography
Sound in medicine Lect.10.
Reflections & Attenuation
Ultrasound.
Single-element transducers properties
Ultrasound.
Properties of sound.
Food Quality Evaluation Methods (FQEM)
Medical Ultrasound: Fundamental Physics & Safety Implications
Sound COS: #18 Use Models to determine how light and sound waves differ in how they are absorbed, reflected, and transmitted through different media.
Review of Ultrasonic Imaging
PARAMETRIC SUB-BOTTOM PROFILER: A NEW APPROACH FOR AN OLD PROBLEM
Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc.
Traveling Waves Waves and Sound
Fundamentals of Sonographic Wave Propagation and New Technologies
Chapter 12 Image Artifacts
Sound Waves and Ultrasound
Presentation transcript:

MCQs (1) – Questions Ultrasound propagates through the medium as a transverse wave Ultrasound velocity is equal to the product of frequency and wavelength Ultrasound velocity is lower in a stiff medium Acoustic impedance is the product of ultrasound velocity and the density of the medium Acoustic impedance of air is almost zero

MCQs (1) - Answers Ultrasound propagates through the medium as a transverse wave Ultrasound velocity is equal to the product of frequency and wavelength Ultrasound velocity is lower in a stiff medium Acoustic impedance is the product of ultrasound velocity and the density of the medium Acoustic impedance of air is almost zero

MCQs (1) - Corrections Ultrasound propagates through tissue as a longitudinal wave, therefore, the local movement of the tissue particles is in the same direction as the wave propagation Ultrasound travels faster in stiff, non-compressible media

MCQs (2) – Questions Ultrasound is attenuated by energy loss from the main beam caused by absorption, reflection and scattering The attenuation is proportional to frequency Specular reflection occurs when the size of the reflector is much smaller than the ultrasound wavelength The degree of ultrasound reflection at an interface between different soft tissue is up to 25% (Fat-muscle interface) The angle of refraction is governed by Snell’s law

MCQs (2) - Answers Ultrasound is attenuated by energy loss from the main beam caused by absorption, reflection and scattering The attenuation coefficient is proportional to frequency Specular reflection occurs when the size of the reflector is much smaller than the ultrasound wavelength The degree of ultrasound reflection at an interface between different soft tissues is up to 25% (Fat-Muscle interface) The angle of refraction is governed by Snell’s law

MCQs (2) - Corrections c) Specular reflection occurs when the size of the reflector is much larger than the ultrasound wavelength d) Up to 1-2% is reflected.

MCQs (3) - Questions An ultrasound transducer contains a piezoelectric crystal At the resonant frequency the transducer produces ultrasound of a wavelength equal to twice the thickness of the crystal For a single crystal transducer the length of the near field is proportional to the squared radius of the transducer and inversely proportional to wavelength For a single crystal transducer the area where beam starts to diverge is termed Fresnel region For a single crystal transducer the angle of divergence is proportional to ultrasound wavelength

MCQs (3) - Answers An ultrasound transducer contains a piezoelectric crystal At the resonant frequency the transducer produces ultrasound of a wavelength equal to twice the thickness of the crystal For a single crystal transducer the length of the near field is proportional to the squared radius of the transducer and inversely proportional to wavelength For a single crystal transducer the area where beam starts to diverge is termed Fresnel region For a single crystal transducer the angle of divergence is proportional to ultrasound wavelength

MCQs (3) - Corrections d) The Fresnel region is another name for the near field, the Fraunhofer region is another name for region where the beam starts to diverge in far field.

MCQs (4) - Questions Focusing on array probe is achieved by transmitting from the innermost elements of the active group first The longer the time delay between transmitting from the elements of the array, the longer the focal distance A low Q transducer has a wide bandwidth Lateral resolution is the ability to resolve two objects lying side by side at the same depth In the focal area, axial and lateral resolutions are equal

MCQs (4) - Answers Focusing on array probe is achieved by transmitting from the innermost elements of the active group first The longer the time delay between transmitting from the elements of the array, the longer the focal distance A low Q transducer has a wide bandwidth Lateral resolution is the ability to resolve two objects lying side by side at the same depth In the focal area, axial and lateral resolutions are equal

MCQs (4) - Corrections Focusing in a linear array is achieved by first energising the outermost elements of the active group A long focal distance is obtained by using a short delay between energising the outer and innermost elements of the active group Axial resolution is a component of pulse length and lateral resolution is a component of beam width. Beam width is larger than the pulse length, so the lateral resolution is worse

MCQs (5) - Questions In B-Mode to avoid flicker the frame rate should be at least 50 fps The maximum achievable frame rate in B-mode with a line density of 100 and a depth of view of 15 cm is approximately 100 fps THI uses second harmonic frequencies of the fundamental frequency THI requires a wide bandwidth transducer One of the main advantages with THI is that it reduces reverberation and side lobe artefacts

MCQs (5) - Answers In B-Mode to avoid flicker the frame rate should be at least 50 fps The maximum achievable frame rate in B-mode with a line density of 100 and a depth of view of 15 cm is approximately 100 fps THI uses second harmonic frequencies of the fundamental frequency THI requires a wide bandwidth transducer One of the main advantages with THI is that it reduces reverberation and side lobe artefacts

MCQ (5) - Corrections The limit of temporal resolution of a human eye is about 40 ms, so ideally the frame rate should be above 25 fps In one second ultrasound needs to travel double the distance ‘d’ (from the probe to the depth of view and then back) as many times as the number of scan lines multiplied by frame rate. Therefore, it travels a total distance of 2 x d x frame rate x number of scan lines with average velocity of 1540 m/s. Therefore, the max frame rate can be calculated as FR= 1540 / (2 x d x n)= 1540/(2 x 0.15 x 100)= 50f/s

MCQs (6) - Questions The Doppler shift frequency only depends on the original frequency of the ultrasound the velocity of the ultrasound through the medium and the velocity of the reflector Continuous wave Doppler allows you to monitor the depth of a target Continuous wave Doppler is an audible sound To detect slow flow it may be necessary to decrease the Doppler frequency or increase PRF The signal to noise ratio can be improved by using a low pass frequency filter

MCQs (6) - Answers The Doppler shift frequency only depends on the original frequency of the ultrasound the velocity of the ultrasound through the medium and the velocity of the reflector Continuous wave Doppler allows you to monitor the depth of a target Continuous wave Doppler is an audible sound To detect slow flow it may be necessary to decrease the Doppler frequency or increase PRF The signal to noise ratio can be improved by using a low pass frequency filter

MCQs (6) - Corrections The Doppler shift also depends on the angle of isonation There is no depth discrimination An increase in Doppler frequency results in a larger frequency shift which is easier to detect. Decreasing the PRF increases sensitivity to small frequency shifts

MCQs (7) - Questions The sampling volume in Colour Doppler is used to determine depth information The maximum depth depends on changes in the velocity of the medium Aliasing is governed by the Nyquist criterion Colour Doppler is not susceptible to aliasing Colour Doppler is based on a continuous wave

MCQs (7) - Answers The sampling volume in Colour Doppler is used to determine depth information The maximum depth depends on changes in the velocity of the medium Aliasing is governed by the Nyquist criterion Colour Doppler is not susceptible to aliasing Colour Doppler is based on a continuous wave

MCQs (7) - Corrections The maximum depth depends on the PRF. d=c/2PRF Nyquist criterion applies as with spectral Doppler Colour Doppler uses a pulsed Doppler to give depth dependant information

MCQs (8) - Questions TI is the ratio of the power emitted to the power required to increase the temperature by 1• C B-Mode imaging poses greater heating risks than Doppler Heating is only caused by transducer surface warming up There are three TI equations used for tissue, bone an cranial ultrasound The ICE Standard (2007) specifies that the surface temperature rise must not exceed 43• C when in contact with the patient

MCQs (8) - Answers TI is the ratio of the power emitted to the power required to increase the temperature by 1• C B-Mode imaging poses greater heating risks than Doppler Heating is only caused by transducer surface warming up There are three TI equations used for tissue, bone an cranial ultrasound The IEC Standard (2007) specifies that the surface temperature rise must not exceed 43• C when in contact with the patient

MCQs (8) - Corrections b) Doppler techniques pose greatest thermal risk c) Heating is caused by the combination of apsorption of the ultrasound waves and the heat produced at the transducer surface

MCQs (9) - Questions MI equals the peak rarefactional pressure divided by the ultrasound frequency Non-inertial cavitation causes acoustic streaming and shock For MI < 0.7, conditions for inertial cavitation probably do not exist The presence of microbubbles decreases the risk of cavitation The risk of cavitation is increased in a prenatal lung

MCQs (9)- Answers MI equals the peak rarefactional pressure divided by the ultrasound frequency Non-inertial cavitation causes acoustic streaming and shock For MI < 0.7, conditions for inertial cavitation probably do not exist The presence of microbubbles decreases the risk of cavitation The risk of cavitation is increased in a prenatal lung

MCQs (9) - Corrections a) MI equals the peak rarefactional pressure divided by the square root of the ultrasound frequency d) The use of contrast agents (microbubbles) will increase the likely hood of cavitation e) The fetal lung is not aerated, therefore, no increase to the risk of cavitation

MCQs (10) – Questions BMUS Guidelines Up to 10 weeks fetal gestation the operator should monitor TIS There are no scan time restrictions when scanning with TI<0.7 Scanning of embryo or fetus is not recommended, however briefly, with TI>2.0 Consent is required from the patient when the scan is performed by a trainee There are no time restrictions when an ultrasound examination is performed by a trainee

MCQs (10) - Answers Up to 10 weeks fetal gestation the operator should monitor TIS There are no scan time restrictions when scanning with TI<0.7 Scanning of embryo or fetus is not recommended, however briefly, with TI>2.0 Consent is required from the patient when the scan is performed by a trainee There are no time restrictions when an ultrasound examination is performed by a trainee

MCQs (10) - Corrections Scanning of embryo or fetus is not recommended however briefly, with TI>3.0 e) The time should not exceed double that needed for a diagnostic examination

MCQs (11) - Questions An imaging phantom is made using a tissue mimicking material (TMM) RCR guidance states electrical safety testing of the ultrasound scanner should be performed bi-annually RCR guidance states QA testing of the ultrasound scanner should be performed bi-annually Doppler ultrasound could be tested using a flow phantom A radiometer can be used to measure the heat pattern from a transducer

MCQs (11) - Answers An imaging phantom is made using a tissue mimicking material (TMM) RCR guidance states electrical safety testing of the ultrasound scanner should be performed bi-annually RCR guidance states QA testing of the ultrasound scanner should be performed bi-annually Doppler ultrasound could be tested using a flow phantom A radiometer can be used to measure the heat pattern from a transducer

MCQs (11) - Corrections b) RCR guidance states electrical safety testing of the ultrasound scanner should be performed annually c) RCR guidance states QA testing of the ultrasound scanner should be performed annually or bi-annually