Analysis of LumiCal data from the 2010 testbeam

Slides:



Advertisements
Similar presentations
Beam test of novel ASICs and sensors Marcin Chrząszcz Cracow University of Technology Itamar Levy Tel Aviv University.
Advertisements

Collection Of Plots for A Testbeam Paper. List of Possible Plots R/Phi resolution, charge sharing, noise etc. Noise performance and few Landau distributions.
Adil Khan Kyungpook National university Korea-Japan Joint ScECAL Group Meeting kobe University Japan 3 rd September 2010.
13/02/20071 Event selection methods & First look at new PCB test Manqi Ruan Support & Discussing: Roman Advisor: Z. ZHANG (LAL) & Y. GAO (Tsinghua))
Bulk Micromegas Our Micromegas detectors are fabricated using the Bulk technology The fabrication consists in the lamination of a steel woven mesh and.
Adding electronic noise and pedestals to the CALICE simulation LCWS 19 – 23 rd April Catherine Fry (working with D Bowerman) Imperial College London.
March, 11, 2006 LCWS06, Bangalore, India Very Forward Calorimeters readout and machine interface Wojciech Wierba Institute of Nuclear Physics Polish Academy.
Standalone VeloPix Simulation Jianchun Wang 4/30/10.
1 Study of the Tail Catcher Muon Tracker (TCMT) Scintillator Strips and Leakage with Simulated Coil Rick Salcido Northern Illinois University For CALICE.
The first testing of the CERC and PCB Version II with cosmic rays Catherine Fry Imperial College London CALICE Meeting, CERN 28 th – 29 th June 2004 Prototype.
GLAST LAT Readout Electronics Marcus ZieglerIEEE SCIPP The Silicon Tracker Readout Electronics of the Gamma-ray Large Area Space Telescope Marcus.
GLAST LAT Readout Electronics Marcus ZieglerIEEE SCIPP The Silicon Tracker Readout Electronics of the Gamma-ray Large Area Space Telescope Marcus.
The Design of MINER  A Howard Budd University of Rochester August, 2004.
Anne-Marie Magnan Imperial College London CALICE Si-W EM calorimeter Preliminary Results of the testbeams st part On behalf of the CALICE Collaboration.
SPiDeR  First beam test results of the FORTIS sensor FORTIS 4T MAPS Deep PWell Testbeam results CHERWELL Summary J.J. Velthuis.
Performance test of STS demonstrators Anton Lymanets 15 th CBM collaboration meeting, April 12 th, 2010.
Timing Jitter and Cross Talk in the SRS + Xray scan of FIT zigzag board Michael Phipps, Bob Azmoun, and Craig Woody 1.
15 Dec 2010 CERN Sept 2010 beam test: Sensor response study Chris Walmsley and Sam Leveridge (presented by Paul Dauncey) 1Paul Dauncey.
Updates on GEMs characterization with APV electronics K. Gnanvo, N. Liyanage, K. Saenboonruang.
VC Feb 2010Slide 1 EMR Construction Status o General Design o Electronics o Cosmics test Jean-Sebastien Graulich, Geneva.
GEM MINIDRIFT DETECTOR WITH CHEVRON READOUT EIC Tracking Meeting 10/6/14 B.Azmoun, BNL.
ATLAS Liquid Argon Calorimeter Monitoring & Data Quality Jessica Levêque Centre de Physique des Particules de Marseille ATLAS Liquid Argon Calorimeter.
Montpellier, November 15, 2003 J. Cvach, TileHCAL and APD readout1 TileHCAL- fibre readout by APD APDs and preamplifiers Energy scan with DESY beam –Energy.
Design and development of micro-strip stacked module prototypes for tracking at S-LHC Motivations Tracking detectors at future hadron colliders will operate.
Fine Pixel CCD for ILC Vertex Detector ‘08 7/31 Y. Takubo (Tohoku U.) for ILC-FPCCD vertex group ILC vertex detector Fine Pixel CCD (FPCCD) Test-sample.
Apollo Go, NCU Taiwan BES III Luminosity Monitor Apollo Go National Central University, Taiwan September 16, 2002.
Progress on the beam tracking instrumentation Position measurement device Tests performed and their resolution Decision on electronics Summary.
CALICE Tungsten HCAL Prototype status Erika Garutti Wolfgang Klempt Erik van der Kraaij CERN LCD International Workshop on Linear Colliders 2010, October.
5-9 June 2006Erika Garutti - CALOR CALICE scintillator HCAL commissioning experience and test beam program Erika Garutti On behalf of the CALICE.
(s)T3B Update – Calibration and Temperature Corrections AHCAL meeting– December 13 th 2011 – Hamburg Christian Soldner Max-Planck-Institute for Physics.
August DESY-HH VFCAL Report W. Lohmann, DESY Infrastructure for sensor diagnostics FE Electronics Development Sensor test facilities Laser Alignment.
Lucia Bortko | Optimisation Studies for the BeamCal Design | | IFJ PAN Krakow | Page 1/16 Optimisation Studies for the BeamCal Design Lucia.
Albuquerque 1 Wolfgang Lohmann DESY On behalf of the FCAL collaboration Forward Region Instrumentation.
February 29, 2016LCWS Arlington Labs involved: Argonne, Vinca Inst, Belgrade, Bukharest IFIN-HH & ISS, CERN, Univ. of Colorado, Cracow AGH-UST & IFJ-PAN,
CALICE, CERN June 29, 2004J. Zálešák, APDs for tileHCAL1 APDs for tileHCAL MiniCal studies with APDs in e-test beam J. Zálešák, Prague with different preamplifiers.
Irradiated 3D sensor testbeam results Alex Krzywda On behalf of CMS 3D collaboration Purdue University March 15, 2012.
Testbeam analysis Lesya Shchutska. 2 beam telescope ECAL trigger  Prototype: short bars (3×7.35×114 mm 3 ), W absorber, 21 layer, 18 X 0  Readout: Signal.
Juan Valls - LECC03 Amsterdam 1 Recent System Test Results from the CMS TOB Detector  Introduction  ROD System Test Setup  ROD Electrical and Optical.
A Brand new neutrino detector 「 SciBar 」 (2) Y. Takubo (Osaka) - Readout Electronics - Introduction Readout electronics Cosmic ray trigger modules Conclusion.
Energy Reconstruction in the CALICE Fe-AHCal in Analog and Digital Mode Fe-AHCal testbeam CERN 2007 Coralie Neubüser CALICE Collaboration meeting Argonne,
Upgrade with Silicon Vertex Tracker Rachid Nouicer Brookhaven National Laboratory (BNL) For the PHENIX Collaboration Stripixel VTX Review October 1, 2008.
Update on works with SiPMs at Pisa Matteo Morrocchi.
Elena Rocco Nikhef, The Netherlands On behalf of the ALICE Utrecht-Nikhef group Jamboree – Utrecht December 2012.
3/06/06 CALOR 06Alexandre Zabi - Imperial College1 CMS ECAL Performance: Test Beam Results Alexandre Zabi on behalf of the CMS ECAL Group CMS ECAL.
0 Characterization studies of the detector modules for the CBM Silicon Tracking System J.Heuser 1, V.Kyva 2, H.Malygina 2,3, I.Panasenko 2 V.Pugatch 2,
SRS TIMING ANOMALY RD51 Mini-Week June 9, 2015 M. Phipps, BNL 1.
Understanding of SKIROC performance T. Frisson (LAL) On behalf of the SiW ECAL team Special thanks to the electronic and DAQ experts: Stéphane Callier,
ICARUS T600: low energy electrons
*Institute of Space Science
Luminosity and Beamtuning Calorimeters in the very Forward Region
EZDC spectra reconstruction and calibration
FCAL R&D towards a prototype of very compact calorimeter
Marcin Chrząszcz Cracow University of Technology Itamar Levy
Huagen Xu IKP: T. Randriamalala, J. Ritman and T. Stockmanns
GLAST LAT tracker signal simulation and trigger timing study
CLAS12 software workshop
ALiBaVa A brief overlook of Liverpool software changes, common issues and questions. A. Affolder, A. Greenall, I. Tsurin, G. Casse, V. Chmill, M. Wormald,
Panagiotis Kokkas Univ. of Ioannina
Detection of muons at 150 GeV/c with a CMS Preshower Prototype
Analysis of FADC single-crystal data
Calice Collaboration Meeting Sep. 2009
CMS ECAL Calibration and Test Beam Results
Felix Sefkow CALICE/EUDET electronics meeting CERN, July 12, 2007
Test Beam Measurements october – november, 2016
SVT detector electronics
BESIII EMC electronics
High Rate Photon Irradiation Test with an 8-Plane TRT Sector Prototype
Rick Salcido Northern Illinois University For CALICE Collaboration
DESY drift chambers efficiency and track reconstruction
Testbeam analysis on module #1
Presentation transcript:

Analysis of LumiCal data from the 2010 testbeam Itamar Levy Tel Aviv University Belgrade, Serbia September 2011

Introduction 2 In the summer of 2010 the FCAL collaboration had the first test beam measurement of the Lumical silicon prototype, and of the BeamCal GaAs. In these measurements the full readout chain including silicon sensors, kapton fan-out and front-end electronics were tested. An external CAEN VME-V1724 ADC was used. It is a 14 bits 8 channels 100Msp/s pipeline ADC controlled by computer via an optical link. 2 area of the silicon sensor have bean measured, 8 upper pads and 8 bottom pads. During the beam test the response of readout chain to electromagnetic showers from tungsten absorber was also studied The sensor board and the electronic board The silicon sensor with 8 upper pads and 8 bottom pads measured. 2

Test Beam Set Up 3 100 mm 163 mm 250 mm ZEUS telescope planes Sensor box Sensor pcb Trigger scintillators 4.5Gev electron beam external ADC Telescope DAQ coinc. unit trigger busy Test beam took place in beam line 22 of DESY II ring in Hamburg with 4.5 GeV electron beam. Measurements are the combination of our sensors, ZEUS telescope for position reconstruction and temperature monitoring. The Telescope composed of 3 modules. Each module have 2 single-sided Perpendicular Si planes (32 x 32 x 0.3 mm) with 640 strips/plane of 25 um pitch. The Readout pitch is 50 um. The DAQs receive a trigger signal from the Telescope 3 scintillators via coinc. unit. The slower DAQ (Telescope), generates a “busy” signal, to level the events between the DAQs. 3

Spectrum analysis : one event 4 In each trigger all 8 channels are recorded, with 128 sample points. In the legend of figures 1semple1sample 4

Spectrum analysis : number of events 5 When looking at number of events: Baseline fluctuation (RMS ~15-20 ADC counts). Noise correlation between channels is observed. 5

Spectrum analysis : S/N 6 Integration window To analyze the spectrum, the complete data set was taken. ~ 1Mevents for upper pads and ~1.2Mevents for bottom pads. Only 1/8 is in signal area of spectrum. For spectrum different methods tried: “spectrum” – integration within the Integration window of signal minus baseline. “spectrumNB” – integration within the Integration window of just signal. “generalspectrum” – integration in the full range (128 points) of just signal. “amplitude” – the max amplitude of the signal minus the baseline. “amplitudeNB” – the max amplitude of the signal. We define S/N as : To complete the measurement, the spectrum was fitted by an Gauss-Landau convolution. Baseline 6

Spectrum analysis : S/N 7 The calculated S/N for all channel : Better not to use baseline subtraction: Please round-up the numbers to at most two digits!!! 7

Spectrum analysis : S/N 8 Only in the bottom pads we can see temperature difference in the data set. The difference in the mean value of the pedestal is ~20ADC counts/1°C for all channels. Y-axis legend ‘padastal’  ‘pedestal’ 8

Noise correlation 9 To check noise correlation we calculate for every 2 channels, the correlation coefficient, in pedestal events, where S is the sample and σ is from the 128 samples in one channel in event. For each pair, we got a distribution between -1 to 1 of the correlation coefficient. The correlation coefficient MPV is presented. X-axis legend, ‘correlation factor (arb. unit)’  correlation coefficent By construction, there are no units for the correlation coefficent !!! 9

Cross-talk Point by Point correlation ch0 and ch1 Profile Y axis 10 Point by Point correlation ch0 and ch1 Profile Y axis To see the noise correlation we make a Point by Point correlation 2D graph for each pair of channels. Each graph contains ~128M samples (1M events times 128 samples each). We try to estimate cross-talk coefficient by using the ROOT profile tool, and study the behavior of the mean value of the pedestal as function of signal size in the other channel. We can also see that the cross-talk coefficient is getting bigger at very large signals > ~6MIP Profile X axis 10

Cross-talk 11 The cross-talk coefficient was normalized by the ratio between the amplitude MPV’s of the signal. We can see that the cross-talk coefficient of the bottom pads is more then twice the coefficient in the upper pads, due to the long transition lines to the ASICs Normalized cross-talk coefficient – bottom pads Normalized cross-talk coefficient – upper pads 11

Cross-talk 12 The cross-talk coefficient was normalized by the ratio between the amplitude MPV’s of the signal. We can see that the cross-talk coefficient of the bottom pads is more then twice the coefficient in the upper pads, due to the long transition lines to the ASICs Normalized cross-talk coefficient – bottom pads Normalized cross-talk coefficient – upper pads 12

Time stamping 13 In this testbeam the time stamping was critical since one needs to match between events in sensor DAQ and in Telescope. We rely on the trigger/busy logic and low rate of event at 4.5 GeV to determine that the two systems have the same number of events. Then by shifts in small number of events one can match, and use all of the data. Position reconstruction was made with track that hit only 1 time per plane of telescope. reconstruction efficiency is ~50% Position reconstruction with correct shifts +1 Position reconstruction with correct shifts Position reconstruction with correct shifts In this plots every color is for one channel over threshold Position reconstruction of bottom pads Position reconstruction of upper pads 13

Charge collection uniformity 14 As a start of the study of charge collection uniformity of the sensor, position reconstruction data combined with spectrum analysis. Each reconstructed event got a weight of his integrated charge normalized to the MPV of the signal in that channel. We averaged the combined charge by the number of hits per area (since the beam profile is not uniform). In this measurement we got around 2000-4000 hits per square mm (half in the upper pads) . One can see that in the pads the charge collection is uniform (plus some edge effects). There is decrease of charge collection of ~10% of 0.2 mm around the 0.1 mm gaps between pads Sensor uniformity upper pads Sensor uniformity bottom pads 14

Summary 15 We present here some new results from the 2010 LumiCal testbeam. S/N – measurement of spectrum using the max amplitude can give S/N of up to 20, by measuring the total signal (without any subtraction) this level of S/N and higher can by achieved. It is difficult to use this kind of measurement since the baseline of the signal has temperature dependence of ~20 ADC count/°C . Noise correlation – in this data set a high noise correlation coefficient is observed. Cross-talk – by using the profile of correlation graph an estimate of cross-talk coefficient can by obtained. The estimate suggests a cross-talk coefficient of the order of 1% for the short lines and 3% for the long lines. Charge collection uniformity – first result shows that there is uniformity in the pads and between similar pads, there is same decrease in the uniformity in the small gap between pads.

Supplement 16 16