Field Equipment Calibration PH, Temperature, Conductivity, Turbidity, Dissolved Oxygen, Free & Total Chlorine Gregory Langland CCSF.

Slides:



Advertisements
Similar presentations
Analyze-Immediately Parameters pH Analyze Immediately Within 15 minutes of collection: pH Residual Chlorine Temperature Chlorine Dioxide Dissolved Oxygen.
Advertisements

Analyze Immediately Dissolved Oxygen. Topics of Discussion *Definition & Pointers *Applications:Water Quality & Sewage Treatment *Theory *Partial Pressure.
Water Quality Testing Made Easy GPS Aquameter and Multiparameter Aquaprobe The perfect solution for your water quality testing needs.
IC Controls 855 & 802 ppm Dissolved Oxygen System pH / ORP Conductivity Dissolved Oxygen Chlorine Standards R1.0 © 2004 IC CONTROLS.
Biochemical Oxygen Demand (BOD) CE Lab. Introduction The Biochemical Oxygen Demand (BOD) test measures the oxygen consumed by microorganisms in.
ISE Ion Selective Electrodes Prepared By Michigan Department of Environmental Quality Operator Training and Certification Unit.
PH pH is a unit of measure which describes the degree of acidity or alkalinity (basic) of a solution. It is measured on a scale of 0 to 14. The formal.
ACADs (08-006) Covered Keywords analytical balance, conductivity bridge with flow cell, inductively coupled plasma analyzer, in-line sodium monitor, ion.
IC Controls 865 ppb Dissolved Oxygen System pH / ORP Conductivity Dissolved Oxygen Chlorine Standards ppb Galvanic Dissolved Oxygen.
CHAPTER STUDY GUIDE CHEMISTRY SPRING FINAL.
PH probe
Sanitation and HACCP Monitoring with the LIGHTNING MVP BioControl Systems, Inc.
Introduction to Environmental Engineering
Standard Operating Procedures l At l Call for help at Orlando Office Tallahassee.
Chlorination & Chlorine Demand
Water Quality Analysis Methods and Applications. Water Quality Analysis Physical Factors including suspended materials (called suspended solids) and dissolved.
Water Quality Supplies Waste bottle 1413 solution pH 10 solution pH 4 solution Solution cups Water sample cup Hanna Meter.
Measuring & Analyzing GLOBE pH data Jean Morrill, Martha Conklin, Roger Bales, Jonathon Whittier & Alexander Leonard Hydrology and Water Resources University.
+ 2.2 Measuring Abiotic Components. + Investigation Must Choose 2 of the following factors to study Investigate how these factors vary between ecosystems,
Kuskokwim Watershed Council Water Quality Training Workshop
Water Quality, Temperature and Gloss Measurement Instrumentation.
Verifying the Use of Specific Conductance as a Surrogate for Chloride in Seawater Matrices Rob Mooney Technical Marketing Manager In-Situ ® Inc.
pH and Conductivity Best Practices
Module 4 Stream Ecology Laboratory Solids TSS and Turbidity.
Tel: +44 (0) Turbidity Application and Product Data.
Proprietary & Confidential INDIA WATER WEEK 2013 Water Quality Analyzers : An Overview Mr T.V.Shiva Rao (MSc,MBA,MSc,PGDSM,FOSI)
Overview of Continuous Water-Quality Monitoring. Purpose of Monitoring Define the objectives of the water quality monitoring project 1. Environmental.
PH Levels in Tumwater and Valley Creeks Lauryn Last, North Olympic Peninsula Skills Center Natural Resources Prior Knowledge The pH levels in both Tumwater.
FIELD METHODS IN ENVIRONMENTAL GEOLOGY GEOS 3110.
New Hanover County Landfill Leachate Treatment Facility Trevor Smith Shannon Alexander Kelly Abbott Courtney Jenkins Carol Ann Dulin.
Field Sampling Methods KWWOA Louisville April 16, 2013 Presented by Frank Hall, Laboratory Certification Coordinator.
BIOCHEMICAL OXYGEN DEMAND
2.2 Measuring abiotic components of the system.  List the significant abiotic (physical) factors of an ecosystem.  Describe and evaluate.
THE MEASUREMENT OF pH. I GENERAL PRINCIPLES Convenient way to express hydrogen ion concentration, or acidity pH = - log[ H +
Salinity “In one drop of water are found all the secrets of all the oceans”
Field Analytical Methods. Considerations for Field Analytical Methods Which parameters are anticipated to occur at the site? What media will be analyzed?
Guanzon Kilat.  Cement is a fine, soft, powdery-type substance. It is made from a mixture of elements that are found in natural materials such as limestone,
Temperature Measured with a thermometer Units are ◦C
Turbidity Measurement By Douglas Rittmann, Ph.D., P.E. Water/Wastewater Consultant To UTEP Laboratory Course On February 21, 2006.
Laboratory Certification Update Part 2 Common Findings KWWOA Louisville April 15, 2015 Presented by Frank Hall, Laboratory Certification Coordinator.
Freshman Engineering Clinic
The Islamic University of Gaza- Environmental Engineering Department
Ch. 4. Measurements (on Site) in Field  pH 와 E H  Basic water quality parameters  Definitions  pH = -log 10 a H+ (logarithmic value of the reciprocal.
Free Residual Chlorine Measurement
IC Controls Quality Water Solutions for pH pH / ORP Conductivity Dissolved Oxygen Chlorine Standards R1.0 © 2004 IC CONTROLS Concerns.
Temperature. What is Temperature? Temperature is the measure of the average kinetic energy in a system. May be measured in Celsius (°C) or Fahrenheit.
 Water Quality Variability in a Bioswell and Concrete Drainage Pipe, Southwest Lincoln, Nebraska Jessica Shortino, B.S. University of Nebraska-Lincoln.
Lab 5 Conductivity of Aqueous Solutions 1. Learning Outcomes To calibrate a conductometer. To use a conductometer to determine conductivity of various.
EDGE Outreach Chlorine Generator Lonna Edwards (ECE), Zachary Russo (ECE), Ryan Shamel (ECE) Mark Hogg, Executive Director; Bob Browning, Field Operations;
Chlorine Generator Test Apparatus ECE 599, Fall 2011 Vincent Korfhage, Vladimir Chausenko, Sam Ryan, Ryan Ray.
EDGE Outreach Chlorine Generator Lonna Edwards (ECE), Zachary Russo (ECE), Ryan Shamel (ECE) Mark Hogg, Executive Director; Bob Browning, Field Operations;
EDGE Outreach Chlorine Generator Lonna Edwards (ECE), Zachary Russo (ECE), Ryan Shamel (ECE) Mark Hogg, Executive Director; Bob Browning, Field Operations;
Chlorine Generator Test Apparatus ECE 599, Fall 2011 Vincent Korfhage, Vladimir Chausenko, Sam Ryan, Ryan Ray.
Introduction to the F-900.
Water Quality Testing.
Chlorination & Chlorine Demand
Cedar Lake Water Clarity Secchi Readings at Schmidt’s Pte Sampling Site, Number of Readings When It Was “Clear to Bottom” During a Given Year.
EMSA 22 Lab Module 1 Field Equipment Calibration
Water Testing APES.
Ion Selective Electrode
Water Testing APES.
Conductivity Lecture.
Explore the future HORIBA GROUP
PH pH is a unit of measure which describes the degree of acidity or alkalinity (basic) of a solution. It is measured on a scale of 0 to 14. The formal.
Quality Assurance in the clinical laboratory
Warm Up p65 Which part of the water cycle is MOST responsible for polluting surface water? Municipalities (towns) get drinking water from which two sources?
Water Quality Monitoring
Electrode Training HORIBA Instruments, Inc. Pamela Millett Scientific
Water Analysis.
Getting the YSI Sonde Ready For Use
Presentation transcript:

Field Equipment Calibration PH, Temperature, Conductivity, Turbidity, Dissolved Oxygen, Free & Total Chlorine Gregory Langland CCSF

Objective Understand Equipments and Supply Needed Ensure QA/QC is in place for data collection Testing various water samples -pH/Temp/Specific Conductance/Total Dissolved Solids (TDS) -Turbidity -Dissolved Oxygen - Total & Free Chlorine

Necessary Equipment & Supplies Turbidity -< NTU standards for 4 point calibration - Sample vial - Chlorine Standards A & B - DPD-Colormetric Determinant for Free and total Chlorine -TURBIDITY /Chlorine meter pH/temp/Electric Conductivity meter -pH Electrode Storage Solution -4.0 pH buffer -7.0 pH buffer pH buffer -Conductivity Standard -multiparameter probe and meter Dissolved Oxygen -Dissolved Oxygen/Meter Probe Extra batteries for the meters Extra Field data sheets Beaker/ sample Container Deionized water for cleaning probes, vials

pH/Temperature Calibration Probe must be submerged in the solution and allowed to stabilize Calibration Three point calibrations Use pH 7.0, pH 4 & pH 10 for three point depending on the anticipated sample pH. Slope should be % out side this range readings are not accurate pH is temperature dependent; most meters compensate for the temperature effect (not as strong an effect as for EC) Measure Temperature of stabilized buffers Buffers/standards pH measurements are only as accurate as the buffers used for calibration. Discard buffers after their expiration dates. Be careful not to contaminate stock solutions with used buffer, rinse water or with a different buffer.

Conductivity (EC) Calibration Probe must be submerged in the solution and allowed to stabilize Calibration Standards conductivity standards commonly used depend on the instrument example : mS/cm for Hanna Multiparameter probes A maximum of 20% error regardless of standard used is acceptable. Beyond 20%, readings are not accurate

Turbidity Calibration Sensor calibration Turbidity free water = zero (0 NTU) standard filtering either sample water or deionized water through a 0.2 um or smaller filter to remove particles or use deionized water that is degassed by sparging (bubbling) with helium to minimize air bubbles that give false turbidity readings

Turbidity Standards Formazin -Opacity of Formazin Standards ( 0, 2,10,40 80, 400 NTUs) : Higher opacity= higher turbidity Other Standards available that are more environmentally friendly ! Most Portable meters have their own Turbidity Standards (<0.1, 15, NTUs)

Turbidity Range & Calibration Standards range depends on anticipated sample values Drinking water - typically 0-20 NTU Lakes, streams and wetlands , 0-50 or NTU Wastewater; Storm water run off NTU; sometimes>1000 NTU, require dilution 2 standards typically adequate, but four point calibration is better (response is linear) Most commonly use ready made standards <0.1 NTU, 15 NTU, 100 NTU and 750 NTU.

Total & Free Chlorine-Colorimetric Calibration Calibrate the meter with Calibration solution provided with the meter. Most meters have specific calibration standards. Sample is initially read without addition of colorimetric solution For Total or free Chlorine specific colorimetric test reagent are added Field Meters have an interface that converts Absorbance recorded by instrument into ppm (part per million) or mg/L of Total or Free Chlorine Validated Range 0-5 ppm (instrument can read as low as 0.02 ppm)

Dissolved Oxygen-Sensor Probe Calibration Hanna Instrument The probe is under polarization with a fixed voltage of approximately 800 mV. Probe polarization is essential for stable measurements with the same recurring degree of accuracy. With the probe properly polarized, oxygen is continually "consumed" by passing through the sensitive diaphragm and dissolving in the electrolyte solution contained inside the probe. Calibration should be 100% in the air (at sea level) The Probe can measure Oxygen as % saturation in a solution or mg/ L without recalibration, by simply pressing RANGE function, The mg/L readings make it possible to read the concentration of the dissolved oxygen directly in ppm. Range: 0.00 to mg/L O2 /0.0 to 300 % O2

Dissolved Oxygen-Sensor Probe Salinity, Altitude Compensation Temperature is automatically compensated during the Probe sensor read out. Temperature has an impact on amount of dissolved O2. However, reading need to be adjusted manually for altitude and salinity that will affect DO reading. Adjustment charts are provided to compensate for change in altitude or salinity in reference to sea level. Salinity adjustment is required when measuring salt water. Altitude compensation is required if above sea level.

Stabilization Criteria for each parameter QA/QC Reading must be stabilized within certain range for each parameter calibrated and tested Stabilization Criteria for each parameter is required as part of QA/QC. Temperature +/- 0.2 deg C Conductivity <100 mS/cm +/- 5% >100 mS/cm +/-20% pH +/- 0.1 unit DO +/- 0.3 mg/L Turbidity +/- 10%