A WEIGHT EXCHANGE SYSTEM FOR AN ABSOLUTE PRESSURE BALANCE Authors: H. Kajastie, J. Manninen, M. Rantanen, K. Riski and S. Semenoja, Centre for Metrology.

Slides:



Advertisements
Similar presentations
Mass Metrology, April 2003Mass Metrology - Susceptometer, June 2007 Sartorius Susceptometer - for Precise Measurement of: Susceptibility and Magnetization.
Advertisements

CO2 CascadeComp_Studi Galileo  1 Semi-Hermetic Compressors for CO 2 Cascade Systems Hermann Renz Bitzer Kuehlmaschinenbau GmbH  Germany.
Sandvik Construction Module 4b CH & CS cone crushers
Cooling System Get the engine up to optimum operating Temperature as quickly as possible and maintains it at that temperature. Controls the heat produced.
Connor O’Leary, Micah Uzuh, Brandon Zimmerman, Matthew Howard Advisor: Dr. Dyer Harris.
Manual Set-Up for Density by Hydrometer Method ASTM D1298, D287, D6822 and IP160 Feb
Mechanical Engineering
Presented by: Piston Pumps PVplus. Model max. Displacement [cm³/U] Speed [min-1] PV PV PV PV PV PV
The Swagelok® SK series and AFS Multipurpose Ball Valves
Mechanical Assembly High Purity-Atmospheric Pressure Glove Box Dean.
Iván Fernández CIEMAT 2 nd EU-US DCLL Workshop, University of California, Los Angeles, Nov th, 2014.
Pedal Purification University of Notre Dame Senior Design Group A6 November 28, 2006 Team Rallye (from left): Nicole Del Rey Eric Sabelhaus Mike McConnell.
MAGUIRE MCF CONCENTRATE FEEDER Metering and Blending Equipment.
Build Basic Knowledge of Internal Combustion Engines (I.C.E)
ENTC-303Dr. ALVARADO1 Fluid Mechanics LAB: PRESSURE MEASUREMENT AND GAGE CALIBRATION ENTC
Hydraulics.
A Neutrino Factory Target Station Design based on Solid Targets J. R. J. Bennett STFC, Rutherford Appleton Laboratory, Harwell Science.
Brake Systems.
Selection of measuring instruments
Senior Design: Validation of Design December 15, 2008.
@ Perkins Proprietary Information of Perkins Engines Company Ltd., © 2000.
Platzhalter Bild Vacuum Transfer System for loading the Sartorius Prototype Mass Comparator CCL1007 Thomas Fehling, NCSLI-Conference, Washington D.C.,
Pneumatic Actuators Brief Overview Kunal Sinha Grad General Engineering EFR : 2 per min.
1 Practical Interpretation of Unbroken Chain in Metrological Traceability as to VIM 3 Center for Measurement Standards (CMS) Industrial Technology Research.
Miruna DOBRE – SMD (Belgium) Stephanie BELL – NPL (United Kingdom) Dolores del CAMPO – CEM (Spain) Martti HEINONEN - MIKES (Finland)
Precipitation Gauge Calibration System RYU-SOO HO, BYUNG SUN KIM, YUN BOK LEE, IN-TAE KIM, SUN KI LEE Meteorological Observation Standardization Division,
Korea Institution of Standard Technology 1 Contents 1. Company Summary 2. Company History 3. Organization 4. R&D Achievements 4-1 R&D Achievements Contents.
Lesson 2 Pressure.
A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY Mr Drago GROSELJ Dr Jérôme DUVERNOY Dr John GORMAN
HVAC523 Circulator Pumps.
1 CPB 5000 Pressure Balance -Product overview-. 2 CPB product line CPB 5000 „Standard“ CPB 3800 „Compact Line“ CPB 5000 „Premium“ HighEnd / Laboratory.
Solution Preparation. is a solution preparation system that automatically calculates and dispenses the exact amount of solvent required to prepare a precise.
1 A Local and Remote Radio Frequency Identification Learning Environment Andrew Shields & David Butcher Wireless and Mobility Research Group, Institute.
Standard Weights and Classes of Balances
DESIGN AND FABRICATION OF RECIPROCATING PUMP
ZPDA /ZPD-KI high accuracy pump dosing system
Unit 4: Mechanical Systems. Topic 2: The Wheel and Axle, Gears, and Pulleys.
CHAPTER 5: Mass and Energy Analysis of Control Volumes
1 / 24 Introduction Eccentricity effects INTI Buenos Aires Oliver Mack: Recommendation For The Revision Of Test Procedures For Load Cells In Legal Metrology.
7.3 ENERGY LOSSES AND ADDITIONS  Objective: to describe general types of devices and components of fluid flow systems.
Preliminary Detailed Design Review P15080 Flow Culture System November 13, 2014.
CMS FPIX Cooling System Studies Joe Howell, Fermilab for the FPIX Upgrade Mechanical Working Group CMS Upgrade Workshop April 27,
INTERCOMPARISON BETWEEN RA VI ( France, ) AND RA II (Australia) REGIONAL INSTRUMENT CENTRES ENVIRONMENTAL AGENGY OF THE REPUBLIC OF SLOVENIA Slovakia,
Manual Density Business Today
THE NEW HOME APPLIANCE CLAMPS – GENERATION G. FIELD OF APPLICATION Carton Clamps or Domestic Appliance Clamps are used throughout the logistic chain to.
Arctic Circle 2015, October 16-18, Reykjavík - Iceland MeteoMet Project: Metrology Activities in Ny-Ålesund (Svalbard) Chiara Musacchio Istituto Nazionale.
A U.S. Department of Energy Office of Science Laboratory Operated by The University of Chicago Argonne National Laboratory Office of Science U.S. Department.
Barometry The art or science of barometric observation.
Pressure.
Topic 2: The Wheel and Axle, Gears, and Pulleys
AT-ECR/C.FabreMay 31, 2007 Cryogenics for Liquid Argon Calorimeters Caroline Fabre on behalf of the ATLAS Liquid Argon Cryogenics Collaboration.
Platzhalter Bild Design and Performance of a Fully Automatic Robot System for Mass Determination Thomas Fehling, NCSLI–Conference, Orlando/Florida, August.
APMP TCFF Workshop Nov. 22, Investigation on calibration of fuel ethanol flow meters Takashi Shimada, Ryouji Doihara, Noriyuki Fuiichi, Yoshiya.
Sartorius Volume Comparator VD1005 A new Method for highly accurate Density Determination Benno Gatzemeier, Market Manager Mass Metrology, Sartorius AG.
Maturing and cold smoking
EEN-E1040 Measurement and Control of Energy Systems Control II: Controlled devices, control in energy technology, complex systems Nov 7th 2016.
KASIMOV INSTRUMENT-MAKING ENTERPRISE – AFFILIATE OF JSCO “RSIE”
Magnetic Impelled Arc Butt Welding
LABVIEW BASED MAGNETIC FIELD MAPPING SYSTEM FOR CYCLOTRON DC-280
HFM Test Station Main Cryostat
Calibration of Weighing Instruments
Forging new generations of engineers
Application Operation principle Advantages Highlights Comparison
Cooling System Get the engine up to optimum operating Temperature as quickly as possible and maintains it at that temperature. Controls the heat produced.
High Pressure screw compressor
Forging new generations of engineers
VISUAL AIDS for instruction in VACUUM TECHNOLOGY AND APPLICATIONS
SCITEQ End closure series
Presentation transcript:

A WEIGHT EXCHANGE SYSTEM FOR AN ABSOLUTE PRESSURE BALANCE Authors: H. Kajastie, J. Manninen, M. Rantanen, K. Riski and S. Semenoja, Centre for Metrology and Accreditation (MIKES), P.O. Box 239, FIN Helsinki, Finland, Telephone: , Speaker: S. Semenoja, Centre for Metrology and Accreditation (MIKES)

The pressure laboratory of MIKES The national laboratory for pressure in Finland Secondary level laboratory The laboratory standards are commercial instruments, covering the pressure range from 5x10 -4 Pa to 500 MPa

Introduction The pressure laboratory of MIKES has designed and built a system for changing a load of a pressure balance while the reference vacuum is maintained The pressure balance is a well known DH Instruments PG7601

Traditional calibrations with PG7601 calibrations are performed using a pressure balance equipped with a bell jar and a vacuum pump each pressure point needs its own weight combination, and the weights can be changed only after removal of the bell jar the reference vacuum must be pumped separately for each point time consuming and not very convenient

DH Instruments PG7601

Automated mass handling system for PG7601 fully automatic full-range (5 kPa – 350 kPa) systems are commercially available an automated system concentrated in most used pressure range (850 hPa – 1050 hPa) can be much simpler, and a self-made weight exchanger is good alternative (= low-cost) for the commercial systems MIKES have some previous experience on developing weight exchangers for mass and density calibrations

General discussion for a weights exchanger centring of the weights the motorised exchange system and its operator generate heat in the environment, which can lead to temperature gradients in the weighing chamber magnetic interaction between the weights, a measurement device and the weight exchange system

…and special discussion for the weights exchanger of the pressure balance a user interface into the vacuum chamber the weighing point (on a piston of the pressure balance) is rotating the piston is floating and it moves vertically during loading/unloading the weights exchange system should be easily removable from the pressure balance

The weights exchange system - The pressure balance the original bell jar was replaced with a larger chamber to fit in a chain of eight weights a lid and the chamber is attached to the balance with help of vacuum like the original bell jar rubber gaskets are used

The chamber with the rubber seals

The weights exchange system - Mechanical construction the aluminium lid a weight set a flexible aluminium KF connection between the weights and a threaded shaft for mechanical feed through a gear box which moves the threaded shaft movement for the gear box by a step motor / a manual crank

The mechanics of the vertical movement

The weights exchange system - The weights the weight set of 8 discs from stainless steel the weights are designed so that rotated weights are stacked and centred with help of chamfer under and top of the weights when the weights are lifted they are chain connected with small hooks The true masses, densities and volumes of the 250 g weights were determined in the mass laboratory of MIKES

The weights

The weights exchange system - The control unit an eurocard rack the step motor a driver for the motor a power source a controlling device PC-connection with an RS 232/485 – interface adapter, option for future

The weight exchange system with the control unit

Testing The operational testing of the new weights exchange system was made by calibrating three barometers with the new system and comparing the results to those obtained by the traditional method on the same day The barometers used in the testing were products of Vaisala, a Finnish company well known for its weather monitoring instruments

Testing The following three instruments were calibrated: –n:o Y of the type PTB 220. This is a new instrument belonging to the MIKES density laboratory –n:o P of the type PTB 200, which is one of the reference standards of the Finnish Meteorological Institute –n:o of the type PTB 200, a working standard of MIKES pressure laboratory since 1994

Testing The instruments P and have been calibrated in the pressure laboratory of MIKES for several times They have been proved to be very stable, the drift is certainly less than 2 Pa / year

The results on Y as averages with uncertainty bars (coverage factor k=2)

The results on P as averages with uncertainty bars (coverage factor k=2)

The results on as averages with uncertainty bars (coverage factor k=2)

The result points for the barometer Y

The result points for the barometer

The temperatures recorded for each measurement point during calibration

The reference vacuum values recorded for each measurement point during calibration

Summary of results Barometer n:oE n –value ranges Y , ,12 P , , , ,17 Obviously the two methods give equivalent results

Conclusions The new weight exchanger system has been used in routine calibrations for barometers The results are not only in a good agreement with the traditional method but they suggest possibilities to decrease the measurement uncertainty The main target, considerable savings in measurement time was achieved

The end. Thank You for Your attention