Equalization of Medipix2 imaging detector energy thresholds using measurement of polychromatic X-ray beam attenuation Josef Uher a,b, Jan Jakubek c a CSIRO.

Slides:



Advertisements
Similar presentations
Sezione di Napoli Univ. “Federico II” Experimental study of beam hardening artefacts in photon counting breast computed tomography M.G. Bisogni a, A. Del.
Advertisements

Advanced GAmma Tracking Array
1 Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration Joonki Noh, Jeffrey A. Fessler EECS Department, The University.
First DQE measurment of 500 µm thick Si hybrid pixel sensor with Photon counting readout for X-ray medical imaging M.Caria, J.Chaput, L.Sarry, B.Surre.
Imaging Properties of the Medipix2 System exploiting Single and Dual Energy Thresholds L. Tlustos 1, R. Ballabriga 1, M. Campbell 1, E. Heijne 1, k Kincade.
1 Pixel Readout of GEMs Motivation The Setup Results Next Steps A. Bamberger, K. Desch, J. Ludwig, M. Titov, U. Renz, N. Vlasov, P. Wienemann, A. Zwerger.
ED and WD X-ray Analysis
1 Cluster Quality in Track Fitting for the ATLAS CSC Detector David Primor 1, Nir Amram 1, Erez Etzion 1, Giora Mikenberg 2, Hagit Messer 1 1. Tel Aviv.
tomos = slice, graphein = to write
FPIX0 Electronic Test Marina Artuso Paul Gelling Jianchun Wang  The system works fine with charge injection calibration  Gain curve, threshold, and noise.
Can people meet from 2:40 to 3:30 on Tuesday, September 5?
experimental platform
Derek Liu E. Poon, M. Bazalova, B. Reniers, M. Evans
The Transverse detector is made of an array of 256 scintillating fibers coupled to Avalanche PhotoDiodes (APD). The small size of the fibers (5X5mm) results.
Why silicon detectors? Main characteristics of silicon detectors: Small band gap (E g = 1.12 V)  good resolution in the deposited energy  3.6 eV of deposited.
Measurement of gas gain fluctuations M. Chefdeville, LAPP, Annecy TPC Jamboree, Orsay, 12/05/2009.
ITAM CAS, IEAP CTU IWORID 2004Glasgow, th July IWORID 2004 OPTIMIZATION OF X ‑ RAY DYNAMIC DEFECTOSCOPY USING MEDIPIX-2 FOR HIGH FRAME RATE READ-OUT.
The 6th International Workshop on Radiation Imaging Detectors, July 2004, Glasgow Digital X-ray portable scanner based on monolithic semi-insulating.
Applications of a Pixellated Detection System to Digital Mammography
Imaging and spectroscopic performance studies of pixellated CdTe Timepix detector Dima Maneuski Vytautas Astromskas, Erik Fröjdh, Christer Fröjdh, Eva.
LRT2004 Sudbury, December 2004Igor G. Irastorza, CEA Saclay NOSTOS: a spherical TPC to detect low energy neutrinos Igor G. Irastorza CEA/Saclay NOSTOS.
Digital analysis of scintillator pulses generated by high-energy neutrons. Jan Novák, Mitja Majerle, Pavel Bém, Z. Matěj 1, František Cvachovec 2, 1 Faculty.
Performance of a 128 Channel counting mode ASIC for direct X-ray imaging A 128 channel counting ASIC has been developed for direct X-ray imaging purposes.
Ion Beam Analysis of Gold Flecks in a Foam Lattice F E Gauntlett, A S Clough Physics Department, University of Surrey, Guildford, GU2 7XH, UK.
Performance limits of a 55  m pixel CdTe detector G.Pellegrini, M. Lozano, R. Martinez, M. Ullan Centro Nacional de Microelectronica, Barcelona, 08193,
Munich-Centre for Advanced Photonics A pixel detector system for laser-accelerated ion detection Sabine Reinhardt Fakultät für Physik, Ludwig-Maximilians-Universität.
Advanced semiconductor detectors of neutrons
Energy calibration of the threshold of Medipix for ATLAS Céline Lebel Université de Montréal presenting for the Institut of Experimental.
NA62 Gigatracker Working Group Meeting 23 March 2010 Massimiliano Fiorini CERN.
Mitglied der Helmholtz-Gemeinschaft Calibration of the COSY-TOF STT & pp Elastic Analysis Sedigheh Jowzaee IKP Group Talk 11 July 2013.
Shashlyk FE-DAQ requirements Pavel Semenov IHEP, Protvino on behalf of the IHEP PANDA group PANDA FE-DAQ workshop, Bodenmais April 2009.
Experiment TGV II Multi-detector HPGe telescopic spectrometer for the study of double beta processes of 106 Cd and 48 Ca For TGV collaboration: JINR Dubna,
DEAR SDD --> SIDDHARTA
Geant4 User Workshop, Catania, Italy, 15th October, 2009
The development of the readout ASIC for the pair-monitor with SOI technology ~irradiation test~ Yutaro Sato Tohoku Univ. 29 th Mar  Introduction.
K. Jahoda, 6 Aug 2007 X-ray School, GWU Proportional Counters Some of what you should know in order to use proportional counters for Spectroscopy, Timing,
Jyly 8, 2009, 3rd open meeting of Belle II collaboration, KEK1 Charles University Prague Zdeněk Doležal for the DEPFET beam test group 3rd Open Meeting.
A. Zelenski a, G. Atoian a *, A. Bogdanov b, D.Raparia a, M.Runtso b, D. Steski a, V. Zajic a a Brookhaven National Laboratory, Upton, NY, 11973, USA b.
1 GEM R&D Activity in USTC Yi Zhou University of Science & Technology of China.
Francesca Bisello, Prague , IV ARDENT Workshop IV ARDENT Annual Workshop 22-26/06/2015 Prague Francesca Bisello ESR 10.
Simbol–X workshopMay 14th, 2007 The Simbol-X Detector Payload P. Laurent CEA/Saclay & APC.
X-ray Dynamic Defectoscopy (XRDD) Direct observation of damage development in time dependent gradually stressed materials by means of X-ray transmission.
A Prototype Diamond Detector for the Compton Polarimeter in Jefferson lab, Hall C Medium Energy Physics Group Amrendra.
Progress with GaAs Pixel Detectors K.M.Smith University of Glasgow Acknowledgements: RD8 & RD19 (CERN Detector R.&D. collaboration) XIMAGE (Aixtron, I.M.C.,
Calculation of the Coded Aperture zero-beam-size image (the “image”). The CA fitting procedure: The image is parameterized as a Sum-Of-Gaussians.
MPI Semiconductor Laboratory, The XEUS Instrument Working Group, PNSensor The X-ray Evolving-Universe Spectroscopy (XEUS) mission is under study by the.
PSA: ADAPTIVE GRID SEARCH The Method Experimental Results Optimization aspects Roberto Venturelli (INFN Padova - IPSIA “Giorgi” Verona) SACLAY, 05-may-06.
ICT 1 SINTEF Edge-On Sensor with Active Edge Fabricated by 3D-Technology T. E. Hansen 1), N. Ahmed 1), A. Ferber 2) 1) SINTEF MiNaLab 2) SINTEF Optical.
Based on a paper which will appear on Med. Phys. Issue of Dec. 05 Introduction Contrast cancellation algorithm Experimental setup Experimental images:
September 2002Lukas Tlustos, IWORID2002 Fixed Pattern Signal fluctuations in Si pixel detectors L. Tlustos, D. Davidson, M. Campbell, E. Heijne, B. Mikulec.
Status report Pillar-1: Technology. The “Helmholtz-Cube” Vertically Integrated Detector Technology Replace standard sensor with: 3D and edgeless sensors,
DESCRIPTION OF PIXIRAD  The PIXIRAD Imaging Counter is an INFN-Pisa Spin-off  It works in photon counting  Pulse discrimination with two thresholds.
Basis of Mammography F. Milano Dept. Clinical Physiopathology University of Florence
CdTe prototype detector testing Anja Schubert The University of Melbourne 9 May 2011 Updates.
Medipix3 chip, downscaled feature sizes, noise and timing resolution of the front-end Rafael Ballabriga 17 June 2010.
Study of the Radiation Damage of Hamamatsu Silicon Photo Multipliers Wander Baldini Istituto Nazionale di Fisica Nucleare and Universita’ degli Studi di.
6:th IWORID, Glasgow, Scotland, July 2004 Energy Dependence in Dental Imaging with Medipix 2 Börje Norlin & Christer Fröjdh Mid Sweden University.
DEVELOPMENT OF PIXELLATED SEMICONDUCTOR DETECTORS FOR NEUTRON DETECTION Prof. Christer Fröjdh Mid Sweden University.
Update on 50 um Sensor Calibration using LNLS Data Mathieu Benoit.
Progress with GaAs Pixel Detectors
Particle tracking with semiconductor pixel detector Timepix
Khan, The Physics of Radiation Therapy, Ch7
M. C. Veale1, S. J. Bell1,2, D. D. Duarte1,2, M. J. French1, M
IHEP group Shashlyk activity towards TDR
Performance of the Medipix and Timepix devices for the recognition of electron-gamma radiation fields C. Teyssier1,3, J. Bouchami1, F. Dallaire1, J. Idarraga1,
The response function of semiconductor pixel detectors for highly ionizing particles Jan Jakubek, P. Soukup Institute of Experimental and Applied Physics,
High Rate Photon Irradiation Test with an 8-Plane TRT Sector Prototype
ME instrument and in-orbit performance
Why silicon detectors? Main characteristics of silicon detectors:
Presentation transcript:

Equalization of Medipix2 imaging detector energy thresholds using measurement of polychromatic X-ray beam attenuation Josef Uher a,b, Jan Jakubek c a CSIRO PSE, Lucas Heights, NSW, Australia, b CSIRO MDU, NRF, Australia c IEAP-CTU, Prague, Czech Republic

CSIRO. Outline Medipix detector description Energy threshold equalization using noise floor Energy threshold equalization using fits of X-ray attenuation curves Examples of application Microfocus X-ray tube Sample Imaging detector X-ray transmission was used in measurements

CSIRO. Detectors of Medipix family Images from Medipix detectors: array of 256 x 256 pixels pixel size 55 x 55  m total area 1.5 x 1.5 cm 2 Planar 300  m thick silicon pixel detector (GaAs and CdTe also available) Bump-bonded to a readout chip containing amplifier, discriminator (low and high threshold) and counter for each pixel 3-bit fine threshold adjustment in each pixel

Threshold equalization 3-bit fine threshold adjustment in each pixel values minimizing the threshold level spread has to be found the common technique is based on finding the noise floor in each pixel threshold can be set just above the noise floor CSIRO. Adjustment bits = 0 Adjustment bits = 7 Optimized adjustment bits

Signal-to-thickness calibration (STC) CSIRO. Replacement of the flat-field correction STC converts the measured number of counts into equivalent thickness of the calibrating material (Al) Measured signal Equivalent thickness X-ray attenuation in Al foils is measured

Signal-to-thickness calibration (STC) CSIRO. X-ray tube running at 40kVp, energy threshold set to ~13 keV i.e. above k-edges of Ni and Cu and below k-edges of Ag and Sn STC raw imagecorrected image sample photo Al Ni Cu Ag Sn 13 keV threshold Sn Ag Ni Cu

STC and threshold close to k-edge of Sn CSIRO. Aluminum ~3 mm thick Tin foil 25µm thick STC corrected imageImage histogramDetected X-ray spectrum X-ray tube running at 40kVp, energy threshold set to ~27 keV (Sn k-edge is at 29.2 keV) Can we further reduce the threshold spread?? Note: The spectra do not include the detector response.

Better threshold equalization CSIRO. Equalization using monochromatic X-rays: + reduces the total threshold spread by 30% - requires change of the X-ray setup arrangement - count rates are low => timely procedure Equalization using polychromatic X-rays measurement fit Analytical function calculating the detected signal: Requires: mass attenuation coefficients X-ray spectrum Medipix detector response matrix Attenuation curves at different thresholds Free parameters of the fit: amplitude thresho ld 6 keV 12 keV 20 keV 30 keV

X-ray spectrum CSIRO. Measured using Silicon Drift Detector: AmpTek SDD X-123D sensor thickness 450 µm Be window 12.5 µm thick collimator ∅ 1 mm X-ray tube: Oxford Instruments Monoblock XRS D W target Inherent 1.5 mm Al The measured spectrum was corrected on the SDD detector response Precise knowledge of the spectrum is essential for this technique!

A function that calculates the charge sharing in the detector was developed. Its parameters (charge spread) are obtained from fit of experimental data. Needs to be constructed only once for the given detector bias. Medipix detector response matrix CSIRO. Photon Pixel cells The electric charge deposited in one pixel spreads into adjacent pixels. The hit pixel “sees” less charge! Pd k α Pd k β The charge sharing produces a large tail in the measured spectrum. IEEE NSS 2010, Knoxville, Tennessee, USA

Noise optimized adjustment bits Equalization using the X-ray attenuation curves CSIRO. Adjustment bits = 0 Adjustment bits = 7 STC optimized adjustment bits X-ray attenuation in Al is measured for both extreme values of THL adjustment Measured attenuation curves are fitted in individual pixels and threshold value in keV is determined Optimal THL adjustment value is determined FWHM: 1.4 keV → 1.1 keV FWTM: 2.7 keV → 1.6 keV Note: The upper threshold can be equalized relative to the lower threshold

STC and threshold close to k-edge – new equalization CSIRO. Aluminum ~3 mm thick Tin foil 25µm thick STC corrected imageImage histogramDetected X-ray spectrum X-ray tube running at 40kVp, energy threshold set to ~27 keV (Sn k-edge is at 29.2 keV) Aluminum ~3 mm thick Tin foil 25µm thick STC corrected imageImage histogramDetected X-ray spectrum FWHM: 723 → 634 µm FWTM: 1671 → 1099 µm old new

New equalization – raw images before the STC CSIRO. Aluminum ~3 mm thick Tin foil 25µm thick Raw image with noise equalization Image histogram X-ray tube running at 40kVp, energy threshold set to ~27 keV (Sn k-edge is at 29.2 keV) FWHM: 6647 → 4058µm FWTM: → 7587 µm Raw image with attenuation fit eq. FWHM: 5146 → 3368 µm FWTM: 9220 → 6298 µm

No Medipix response matrix? CSIRO. The attenuation curve shape should be dominated by the material of sample, sensor, the X-ray spectrum and threshold. Can we omit the response matrix? Attenuation curve fit Fit residuals THL adj bit set to 7 pixel 111,87 THL adj bit set to 0 pixel 111,87 measurement fit with no resp. matrix fit with resp. matrix measurement fit with no resp. matrix fit with resp. matrix threshold from fit: 19.1 keV vs 17.5 keVthreshold from fit: 27.0 keV vs 26.2 keV

THL adjustment with and without the response matrix CSIRO. Distributions of STC calibrated values for Sn foil 25 µm thick Can we omit the response matrix? Distribution of differences of the adjustment with and without the response matrix adj. without resp. matrix adj. with resp. matrix

k-edge imaging - examples CSIRO. Sn Ag Pd Mo Sn Ag Pd Zr Cd Sn Ag Cu Ni X-ray tube: 50kVp, 500µA Threshold scan: 7÷40 keV Energy step: ~1.4 keV Total measurement time: 15 min K-edges identified by relative differences of the measured spectrum and a spectrum under Al. Mo 25-75µm k = 20.0keV Pd 25-75µm k = 24.3keV Ag 10-30µm k = 25.5keV Sn 25-75µm k = 29.2keV Ni 20-60µm k = 8.3keV Cu 20-60µm k = 9.0keV Ag 10-30µm k = 25.5keV Sn 25-75µm k = 29.2keV Zr k = 18.0keV Pd k = 24.3keV Ag k = 25.5keV Cd k = 26.7keV Sn k = 29.2keV Thicknesses: 10 – 200µm Mo Sn Pd Ag Ni Sn Cu Ag

Conclusion Advantages of the presented equalization method: better energy threshold equalization compared to the noise based technique comparable performance to the monochromatic source based equalization simple to implement easy to scale to large area detectors easy automatization of this method rough threshold energy calibration signal-to-thickness calibration for raw image correction precise calculation of the detector signal can be used: to calculate signal-to-thickness calibration for different materials other than the one used in the measured calibration better beam-hardening correction in iterative CT reconstruction techniques (the Expectation Maximization method) CSIRO.

Contact Us Phone: or Web: Thank you Process Science and Engineering On-Line Analysis & Control Josef Uher Phone:

Backup slides CSIRO.

Detection of k-edge in the spectrum CSIRO. Spectrum behind Al Spectrum behind Sn ∆E I Al /I Sn =1 I Al /I Sn >1 Spectrum of k-edges is generated by running the window across the measured spectrum. Only values above and below the edge can be measured if the elements in the sample can be anticipated.