HIGH FREQUENCY OSCILLATORY VENTILATION

Slides:



Advertisements
Similar presentations
Neonatal Mechanical Ventilation
Advertisements

Respiratory System Physiology
Mechanical Ventilaton Ramon Garza III, M.D.. Indications Airway instability Most surgical patients or trauma Primary Respirator Failure Mostly medical.
Educational Resources
CPAP Respiratory therapy EMT-B. CPAP Overview  Applies continuous pressure to airways to improve oxygenation.  Bridge device to improve oxygenation.
HFOV high frequency Oscillatory Ventilation
Essentials of Exercise Physiology
Mechanical Ventilation in the Neonate RC 290 CPAP Indications: Refractory Hypoxemia –PaO2 –Many hospitals use 50% as the upper limit before changing.
Neonatal Options for the 3100A. VIASYS Healthcare, Inc. Neonatal Options for the 3100A Early Intervention Pro-Active Rescue.
Dr Tristan GR Dyer RCSEd Fellow in Pre-hospital Emergency Medicine.
Initiating Positive Pressure CMV A generic approach!
3100B Theory of Operation and Controls. VIASYS Healthcare, Inc. 3100B Theory of Operation and Controls Approved for sale outside the US in 1998 for patients.
Initiation of Mechanical Ventilation
3100A Ventilator. VIASYS Healthcare, Inc. 3100A Ventilator Approved in 1991 for Neonatal Application for the treatment of all forms of respiratory failure.
High Frequency Ventilation
Pulmonary Volumes and Capacities—Spirometry A simple method for studying pulmonary ventilation is to record the volume movement of air into and out of.
Topic 6.4 – Gas Exchange.
The Respiratory system Pulmonary ventilation – Chp 16 Respiration.
Objectives Discuss the principles of monitoring the respiratory system
Ventilator.
Respiratory system. Mechanism of lung ventilation.
Respiratory System Chapter 16 Bio 160.
08/14/061 HFOV vs. Conventional Ventilation Latoya Robinson Julie Ordones Joshua Globke Matthew Heaton.
Theory of HFV.
Principles of Mechanical Ventilation
Building a Solid Understanding of Mechanical Ventilation
1 Elsevier items and derived items © 2010 by Saunders, an imprint of Elsevier Inc. Chapter 19 Mechanical Ventilation of the Neonate and Pediatric Patient.
Roberto Fumagalli Ospedale Niguarda Ca ’ Granda Università degli Studi Milano Bicocca Milano Disclosure: none Management of native lung on ECMO.
To what extend human body is similar to a machine ? Human body must have an energy source in both phases, electrical and mechanical Human body consists.
Ventilators All you need to know is….
1 © 2013 Cengage Learning. All Rights Reserved. May not be copied, scanned, or duplicated, in whole or in part, except for use as permitted in a license.
Without reference, identify principles about volume/pressure and high frequency ventilators with at least 70 percent accuracy.
CPAP Murila fv. Respiratory distress syndrome 28% of neonatal deaths are due to prematurity The most common respiratory disorder in the preterm is Respiratory.
HFOV Mechanisms of Gas Transport
Human Anatomy and Physiology Physiology of air breathing The lungs.
Lung Mechanics Lung Compliance (C) Airway Resistance (R)
3100B Theory of Operation and Controls. SensorMedics 3100B u Electrically powered, electronically controlled piston-diaphragm oscillator u Paw of 5 -
Neonatal Ventilation: “The Bivent”
Pulmonary Circulation- THIS IS A REVIEW!!!! ______________ blood enters the lungs from ______ ventricle of heart through the pulmonary ______. Pulmonary.
Pandemic [H1N1] 2009 RT Education Module 2 Lung Protection.
1 Dr.Wahid Helmy pediatric consultant. Basics of Mechanical Ventilation in Neonates.
1 RESPIRATORY ANATOMY. 2 The primary role of the respiratory system is to: 1. deliver oxygenated air to blood 2. remove carbon dioxide from blood The.
Advanced Modes of CMV RC 270. Pressure Support = mode that supports spontaneous breathing A preset pressure is applied to the airway with each spontaneous.
Mechanical Ventilation EMS Professions Temple College.
Mechanical Ventilation Mary P. Martinasek BS, RRT Director of Clinical Education Hillsborough Community College.
Basic Concepts in Adult Mechanical Ventilation
The purpose of the breathing rhythm is to ventilate the lungs to allow delivery of oxygen to the alveoli, and elimination of the waste gas carbon dioxide.
HIGH FREQUENCY VENTILATION (HFV)
Clinical Simulations for the Life Pulse HFJV IMPORTANT: Tap or click on the slide to advance. Do not use the navigation arrows.
Ventilator Management James Eakins, MD FACS Director, Trauma and Surgical Critical Care Hahnemann University Hospital.
Lung Protective Jet Ventilation Basic Lung Protective Strategy for Treating RDS and Air Leaks with HFJV.
BY: NICOLE STEVENS.  Primary objective of mechanical ventilation is to support breathing until neonates own respiratory efforts are sufficient  First.
1 Elsevier items and derived items © 2010 by Saunders, an imprint of Elsevier Inc. Chapter 20 Neonatal and Pediatric High-Frequency Ventilation.
ECAP BIOL The Respiratory System Mrs. Riel.
3 nd LECTURE VENTILATORS Part One. Ventilators One of the major life support systems. Ventilators take over the vital role of the respiratory muscles.
 Understand the dual control concept  Understand the pressure regulation mechanism in PRVC  Demonstration of PRVC  Settings and adjustment with Servo.
Several types of HFV  HFPPV  HFJV  HFOV. Principles of Oscillation Richard F. Kita BS, RRT, RCP Edited by Paula Lussier, CRT, NPS, RCP, BS.
 Be sure to check the absent folder if you have been absent!  Last day to Make up Blood/Cardiovascular System Exam will be Wednesday. After that it will.
PRESSURE CONTROL VENTILATION
11.1 The Function of Respiration. Agenda Lesson 11.1 The Function of Respiration Read p Vocabulary Learning Check SG 166 # 1-3, SG 167 #1-3.
Mechanical Ventilation
Ventilation Strategies in Newborn
Introduction to Basic Waveforms
RESPIRATORY MECHANISM
High Frequency Oscillatory Ventilation
Mechanical Ventilation
Basic Concepts in Adult Mechanical Ventilation
HFOV Mechanisms of Gas Transport
Respiratory System.
Ventilation strategies
Presentation transcript:

HIGH FREQUENCY OSCILLATORY VENTILATION Michael Gordon

What is HFOV? A ventilator that delivers more than 150 breaths per minute Delivers a small tidal volume, usually less than or equal to anatomical dead space Uses diaphragm (acoustic speaker) to produce oscillatory gas flows within the airway Active inspiratory and expiratory phase

Types of high frequency ventilation High Frequency Positive Pressure Ventilation – conventional ventilation at fast rates, not a true HFV. Conventional ventilation provides rates of 1 – 120 and volumes of 4 – 20ml/kg High Frequency Oscillatory Ventilation – delivers rates of 120 – 1200, with volumes equivalent to or less than anatomical dead space (0.1 – 2 ml/kg). Has active inspiration and expiration.

Terminology Dead space –the air in the nose, mouth, larynx, trachea, bronchi and bronchioles. Frequency – measure in Hertz (Hz), 1 Hz = 1 cycle/sec = 60 bpm Mean Airway Pressure – continuous distending pressure (cmH2O) Oscillation Amplitude - peak to trough of pressure waveform = delta P (cmH2O). This is the measure of pressure the ventilator uses to push air into the circuit.

History high frequency ventilation In 1915 Henderson et al found that panting dogs maintained an adequate gas exchange with tidal volumes less than anatomical dead space. In 1964 it was recognised that low frequency, high tidal volume (>15ml/kg) caused lung injury. Low frequency, high tidal volumes could cause loss of surfactant activity and decreased compliance in otherwise normal lungs. Barotrauma (pressures) and Volutrauma (volumes) implicated in lung damage as well. In 1994 Clark found that large tidal volumes caused more lung damage than high airway pressure. Volutrauma causes -  epithelial and endothelial cell injury, pulmonary oedema, surfactant dysfunction, decreased compliance, hyaline membrane formation and impaired gas exchange Therefore High Frequency Ventilation

Method of gas transport It is thought that gas transport in HFOV is accomplished by the following methods. Direct Alveolar ventilation Pendulluft effect Convective streaming Augmented (Taylor) dispersion Cardiogenic mixing Molecular diffusion

Direct alveolar ventilation Small part of proximal alveoli is ventilated directly. Gaseous exchange takes place the same as in CMV.

Pendulluft effect Neighbouring alveoli with different time constraints (different compliance and resistance) are ventilated out of phase, filling and emptying at different rates. Due to this asynchrony these alveoli can mutually exchange gas. By this mechanism even very small fresh gas can reach large numbers of alveoli and regions.

Convective streaming Describes lateral transport of gas in the airways. Gas flows through a tube in layers which travel at different velocity. Velocity of each layer increases towards centre of tube Particles at outer edge of tube are more effected by viscious resistance and travel at slower velocities Assymetric velocity profiles cause bidirectional flow of gas to and from peripheral airways

Augmented (Taylor) dispersion Describe radial transport of gases due to turbulence A plane boundary surface between two gases develops into a pike shaped profile as the velocity of one of the gases increases. Gases exchange occurs at the boundry surface through lateral diffusion(pic)

Augmented (Taylor) dispersion The amount of gas transport depends on the effective longitudinal diffusivity, and is inversely proportional to molecular diffusivity. At airways bends or bifurcations secondary gas movements occur. (pic)

Cardiogenic mixing The heart is a natural oscillator

Molecular diffusion Passive movement of molecules along a concentration gradient due to random thermal oscillation of molecules Diffusion across membrane

HFOV vs CMV waveform

HFOV Waveform

During HFV only approximately 10% of the delta P delivered at the patient wye is transmitted to the alveoli and terminal bronchioles. During CMV >90% of the PIP delivered at the patient wye is transmitted to the alveoli and terminal bronchioles. During HFV alveoli are kept open and not subjected to large pressure and volume swings in order to maintain gas transfer. During CMV alveoli close and open on every breath, large pressure swings are required to move large volumes (in comparison to HFV) to maintain gas transfer.

HFOV Oxygenation and CO2 removal Controlled by – optimal MAP that promotes lung recruitment and FiO2 CO2 removal Controlled by – increasing Delta P (which increases tidal volumes) and decreasing frequency which increases tidal volumes A lower frequency results in larger tidal volumes also because there is diminished effect of flow dependant ETT resistance, and there is increased time available for flow.

Optimal ventilation strategy From all this an optimal ventilation strategy was developed. This entailed the following  limited FiO2  small tidal volumes  minimal PIP  high Frequency  sufficient PEEP to proximate normal functional residual capacity (FRC)

Optimal ventilation strategy Critical opening point/critical closing point The problem is determining WHEN you are at this point.

Indications for HFOV Neonatal Respiratory Distress Syndrome Persistent Pulmonary Hypertension of the Newborn (PPHN) Neonatal Meconium Aspiration Syndrome Congenital Diaphragmatic Hernia Neonatal Air Leak Syndrome with PIE Pneumonia Pulmonary Haemorrhage Lung Hypoplasia

Advantages of HFOV over CMV Does not cause large alveolar pressure and volume swings to produce adequate gas exchange May reduce haemodynamic effects of ventilation – depends on MAP Produces more uniform lung inflation and normalises lung structure Lung recruitment on HFV allows more uniform dispersal of surfactant and Nitric Oxide May reduce lung inflammatory response Patients can breath as if on CPAP

Advantages of HFOV over ECMO Less invasive, does not require anticoagulants Less expensive, can be used in non ECMO centres Can be used as a bridge to ECMO when CMV has failed Can be used on newborns <2kg ECMO centres now using HFOV report 30 – 50% decrease in need for ECMO to treat respiratory failure Easier to transport on HFV than ECMO ( HFJV runs on batteries)

Sensormedics 3100A Type – electromagnetic flow generator Specifications – Paw range 3-45 cm H20 Weight range - up to 35kg Amplitude 0-90 cm H20 Frequency range 5-15 Hz I:E ratio adjustable between 1:2:3 and 1:1 (30% and 50%) Sigh breaths - no

MAP - PEEP ΔP (amplitude) Frequency inflate alveoli oxygenation - distance piston travels - shake alveoli - chest wiggle - pressure above and below MAP - ↑ ΔP = ↑ tidal volume = ↑ CO2 removal Frequency - 1kHz = 60bpm - speed of piston - ↓ frequency = ↑ tidal volume = ↑ CO2 removal

Babylog 8000plus Type – oscillating expiratory valve with venturi assisted expiration Specifications Paw range 3-30cm H20 Weight range – up to 4kg (but realistically 2 kg) Amplitude arbitrary scale 0 – 100% (100% amplitude = 60 Paw) Frequency range 5-20cm Hz I:E ratio – automatically adjusted between 1:5 and 1:1 Sigh breaths- yes

Frequency - ↓ frequency = ↑ tidal volume = ↑ CO2 removal MAP - PEEP - inflate alveoli - oxygenation Frequency - ↓ frequency = ↑ tidal volume = ↑ CO2 removal - (changing freq. has more effect on tidal volume) - (numbers ≤ 10 are better because better flow rates) ΔP (amplitude) - relative scale 0-100% (0% = MAP) - % of difference of 60 minus the MAP - shake alveoli - chest wiggle - pressure above and below MAP - ↑ ΔP = ↑ tidal volume = ↑ CO2 removal - (amplitude ≥ 60% has little effect on tidal volume) DCO2 - gas transport coefficient - represents CO2 likely to be removed - VT2 x f - ↑ DCO2 = ↑ CO2 removal

Set Up Sensormedics - Practical Points Space in the unit and around the bedside Care with hoses and power cords Always ensure brakes are applied Refer to policy and procedures (9W-04-2-11)

Set Up Drager – Practical Points Appropriate tubing (non disposable) Care with power cords

Prior to use adjust settings for patient (off patient) - as ordered by medical officer: Set the ordered bias flow (usually remains set at 25 LPM (centre of ball)). Set the ordered frequency (Hz). Inspiratory time to remain set at 33%.  Set the mean pressure limit: return the MEAN PRESSURE ADJUST control knob to maximum then rotate the MEAN PRESSURE LIMIT knob anti-clockwise until the Paw reading is 10cm above the ordered  (treatment) MAP.

Going onto HFOV Set the MAP 2 – 4 cmH2O over the Map on the CMV Set the amplitude to see a “wiggle” from the nipple line to the umbilicus Set the Hertz to an appropriate rate Set the FiO2 according to CMV FiO2 Ensure appropriate alarm limits set Obtain a chest X-ray Ensure the rib expansion on x-ray is 8 –9 ribs Obtain frequent arterial blood gases to monitor progress

Nursing Considerations Observations - How are they different to CMV? Respiratory assessment Cardiac monitoring ?pneumothorax Care of the ETT Positioning, pressure care Parent – caretaker interactions

Potential complications Hypotension due to impaired cardiac function IVH Lung over distension and Air Leak Syndrome Unplanned extubation

Potential operating issues Alarm settings High and low Paw must be set appropriately Problems? If set to close to operating Paw and patient is spontaneously breathing the oscillator alarms are triggered frequently If alarms set to far from operating Paw circuit may not pressurise and re-start after disconnection

Potential operating issues Unable to decrease Paw as oscillator being weaned Paw is bias flow dependent – set as 25l/min at NICU RWH, therefore may be to high as Paw being weaned Solution – reduce bias flow

Potential operating issues Unable to increase Paw when oscillator in use Check Paw LIMIT knob not set too low

Potential operating issues Unable to calibrate circuit mean pressure 39-43cm Check length of circuit for air leaks Ensure water trap T piece is turned off Ensure all balloon diaphragms are fitted snuggly and not punctured Check temperature probe site for snug fit Increase biasis flow*