TUDelft-IRI-ST A new Tissue-Equivalent Proportional Counter design M. Farahmand, A.J.J. Bos, J. Huizenga and C.W.E. van Eijk Radiation Technology Group,

Similar presentations


Presentation on theme: "TUDelft-IRI-ST A new Tissue-Equivalent Proportional Counter design M. Farahmand, A.J.J. Bos, J. Huizenga and C.W.E. van Eijk Radiation Technology Group,"— Presentation transcript:

1 TUDelft-IRI-ST A new Tissue-Equivalent Proportional Counter design M. Farahmand, A.J.J. Bos, J. Huizenga and C.W.E. van Eijk Radiation Technology Group, IRI Delft University of Technology

2 TUDelft-IRI-ST Experimental microdosimetry gamma radiation alpha particles Study of the effects of ionising radiation on biological targets by investigating the statistical distribution of energy deposition at micrometer level. From: Broerse et al. d = 14  m d = 1  m

3 TUDelft-IRI-ST Tissue-Equivalent Proportional Counter (TEPC) - Wall and gas are tissue-equivalent. - Determine energy deposition in the wall by measuring in gas. - Single-event measurements. TEPC 25 1 - Lowering the gas pressure. Anode wire Helix cathode

4 TUDelft-IRI-ST Applications Radiobiology - link between experimental microdosimetry & radiobiology based on site measurements Radiotherapy - FNT, BNCT, Proton Therapy, Heavy ions Therapy Radiation protection - estimating dose equivalent (H) - unknown radiation Air and Space - aircraft crews and manned missions

5 TUDelft-IRI-ST Small counter cavity High flux radiation sources in radiotherapy facilities - Pile-up effect (Mini TEPC) Measuring low dose rate for radiation protection purposes - Increasing sensitivity of the counter (Mini multi-element TEPC) Main target size of few tens of nm (chromatin fibre) - Limitation of decreasing gas pressure (Mini TEPC)

6 TUDelft-IRI-ST Mini TEPC(-GEM) Electric field lines From: CERN Courier Gas Electron Multiplier (GEM) 100  m 200  m 50  m

7 TUDelft-IRI-ST Gain measurements d t = 1  m 0.3 atm < p g < 0.8 atm 3.0 mm > d g > 1.0 mm d t = 1  m 0.1 atm < p g < 1.0 atm 5.0 mm > d g > 0.5 mm

8 TUDelft-IRI-ST TEPC-GEM design 76 mm  = 2 mm 9 mm

9 TUDelft-IRI-ST First calibration measurements

10 TUDelft-IRI-ST Conclusion & Future development Calibration of TEPC-GEM works properly. Comparing microdosimetric measurements of TEPC-GEM with traditional TEPC. Employing in high flux radiation sources (Therapy facilities). Employing in low radiation field intensity (Aircraft). Investigating the minimum possible site simulation size and cavity size.


Download ppt "TUDelft-IRI-ST A new Tissue-Equivalent Proportional Counter design M. Farahmand, A.J.J. Bos, J. Huizenga and C.W.E. van Eijk Radiation Technology Group,"