1 Hemoccult and Gastroccult Testing Review
June 20, 2011 By Holly A. White, MLS(ASCP)CM, MSHS The Johns Hopkins Hospital Department of Pathology

2 Hemoccult Reagents and Supplies
Beckman Coulter Kit Includes: Hemoccult test slides Hemoccult applicators Hemoccult developer Also needed: Calibrated Timer

3 Hemoccult Procedure Affix label to specimen card
Place specimen in Box A Place specimen from different area in Box B Close flap Set timer for 3-5 minutes Open back flap Add two drops of developer on each test area – wait 60 seconds Blue color on the edges in Positive result Record results on paper and in computer

4 Negative and Positive Results
POSITIVE TEST The developer turns the stool specimen blue - this is an abnormal result and means that blood is present in that stool specimen.     NEGATIVE TEST If the stool specimen does not turn blue, then no hidden blood is present. This is a normal result.

5 Quality Control of Hemoccult
Place one drop of developer between pos and neg circles Set timer for 10 sec. Record results DO NOT record results prior to testing DO NOT compare results of test to patient test results

6 Gastroccult Reagents and Supplies
Beckman Coulter kit includes: Gastroccult test slides Gastroccult applicators Gastroccult developer A plastic storage pouch to protect slides and developer from light and heat Also needed: Calibrated timer

7 Gastroccult Testing Place one drop of specimen in Ph testing area, and one drop in the gastro testing area Set timer for 30 sec. Read PH at 30 seconds If sample is already In the pH testing area pH cannot be completed. Ph needs to be read at 30 seconds

8 Gastroccult Testing 2 drops of developer in Gastro testing area
Test timer for 60 seconds Blue edge is positive QC same as with Hemoccult

9 Quality Control of Gastroccult
Place one drop of developer between pos and neg circles Set timer for 10 sec. Record results DO NOT record results prior to testing DO NOT compare results of test to patient test results

10 Comments Interferences
-Hemoccult- incompletely cooked meat, some raw vegetables may cause false positive -Gastroccult- some red meat, raw vegetables may cause false positive QC must be done and recorded after completion of test No + and -, must be written pos or neg Always set a calibrated timer

11 More Comments and Suggestions
Blue is a positive result Green is a negative result Blue/Green is a positive result Any hint of blue color is a positive result When in doubt, if you can, test it again. If the background of a card is already blue prior to the dropping of developer, do not use the card. It may have been exposed to heat or liquids. Open a new card, preferably a new lot number, and start the test over. Inform your lead tech. Cancel the test if necessary.