Presentation is loading. Please wait.

Presentation is loading. Please wait.

December, 16 20081 Calibration of electromagnetic calorimeter of Hall A DVCS experiment Eric FUCHEY Ph.D Laboratoire de Physique Corpusculaire.

Similar presentations


Presentation on theme: "December, 16 20081 Calibration of electromagnetic calorimeter of Hall A DVCS experiment Eric FUCHEY Ph.D Laboratoire de Physique Corpusculaire."— Presentation transcript:

1 December, 16 20081 Calibration of electromagnetic calorimeter of Hall A DVCS experiment Eric FUCHEY Ph.D Student @ Laboratoire de Physique Corpusculaire UMR 6533 CNRS/IN2P3 Université Blaise Pascal Clermont-Ferrand

2 December, 16 20082 Outline 1- Motivations of calibration 2- Real data calibration –Method –Results 3- Simulated data calibration –Method –Results 4- Conclusion

3 December, 16 20083 Outline 1- Motivations of calibration 2- Real data calibration –Method –Results 3- Simulated data calibration –Method –Results 4- Conclusion

4 December, 16 20084 Experimental layout Calibration for ep->ep  0 Incident beam : E = 5.75 GeV during experiment –Scattered electron measured by HRS –2 photons issued from  0 measured by 132 PbF 2 blocks calorimeter –Recoil proton detected by proton array -> not used

5 December, 16 20085 Looking at real and simulated  0 data sets We measure We can reconstruct for each event a missing mass: with and We can either reconstruct for one block a missing mass distribution for all events with one photon hitting this block.

6 December, 16 20086 Mp2Mp2 Reconstructed missing mass distribution in each block depends on block position in calorimeter. Effects assumed to come only from calorimeter. => necessity to calibrate calorimeter to correct these effects.

7 December, 16 20087 For analysis, we need to fit  0 data set with  0 simulated set to extract cross section unconstistency of resolutions for real and simulated. => necessity to calibrate and to smear simulated.

8 December, 16 20088 Outline 1- Motivations of calibration 2- Real data calibration –Method –Results 3- Simulated data calibration –Method –Results 4- Conclusion

9 December, 16 20089 Method of calibration (real) Goal of calibration: to find for each block of the calorimeter a coefficient in order to reconstruct in each block a missing mass squared in accordance with M p 2. Coefficient of the form: for block#  (  = 0,…,89)

10 December, 16 200810 Method of calibration (real) For an event i, a photon is in block #a, the other is in block #b. Correcting each photon energy: Recomputing M X 2 with these corrected energies, and filling #a and #b blocks distributions.

11 December, 16 200811 Method of calibration (real) Fit distributions by a gaussian -> Mean of the gaussian => New missing mass squared value.  New calibration coefficient Correct 2 photons to reconstruct missing mass -> calibration coefficients for each block are correlated.  necessity to calibrate by steps.  necessity to estimate calibration quality for each step.

12 December, 16 200812 Corrections Before filling distributions, we apply some corrections: –By effects of resolution, missing mass squared and invariant mass ( ) are correlated –-> possibility to improve resolution. with R = 13 GeV –Additional cuts: 0.4 GeV 2 < M X 2 < 1.4 GeV 2 ; 0.105 GeV < m  < 0.165 GeV. k’ = pHRS +/- 4.5% ; r-function > 0.005 Each photon is in a calibrated block, and E  > 1.0 GeV -6.0 cm < z vtx < 7.5 cm

13 December, 16 200813 Quality calibration control Once iteration completed : –Plot reconstructed missing mass squared value for each block over all blocks –Fit by a constant –Check for deviation from reference value, and dispersion (RMS).

14 December, 16 200814 Results (real) Quick convergence from iteration 0 to 5-6 After, averaged M X 2 oscillates, RMS is stable. => Converged

15 December, 16 200815 Results (real) Ex: Iteration #11 Dispersion is low, Calibration has converged Pion mass deviation less than 1 MeV

16 December, 16 200816 Outline 1- Motivations of calibration –Experimental layout 2- Real data calibration –Method –Results 3- Simulated data calibration –Method –Results 4- Conclusion

17 December, 16 200817 Method of calibration (simulated) Same principle, except we try to match simulated with data: –Searching for a calibration coefficient: –Searching to bring each block simulated resolution (  of gaussian fit) to real block resolution. => also need for each block a « smearing » coefficient

18 December, 16 200818 Method of calibration (simulated) For an event i, a photon is in block #a, the other is in block #b. Correcting each photon energy: Smearing each photon energy: picking a random variable in a gaussian distribution.  : corrected energy,  : Recomputing M X 2 with these corrected energies, and filling #a and #b blocks distributions.

19 December, 16 200819 Quality calibration control Once iteration completed : –Plot M X 2 | simu - M X 2 | real for each block over all blocks and  (M X 2 | simu ) -  (M X 2 | real ) for each block over all blocks –Fit each plot by a constant

20 December, 16 200820 Results (simulated) Averaged  M X 2 converges until iteration 15, then stable RMS does not stop decreasing: <0.001 since iteration #30

21 December, 16 200821 Results (simulated) Since ~iteration #20, seems to be converged. Check out for resolution to conclude

22 December, 16 200822 Results (simulated) Quick convergence averaged  (M X 2 ) of from iteration 0 to 6-7, stable after. averaged  (M X 2 ) RMS stable since Iteration #15

23 December, 16 200823 Results (simulated) Invariant mass deviation due to calibration is still less than 1 MeV Since ~iteration #20, dispersion is low, calibration has converged. Only advantage to go further is to improve  M X 2 dispersion.

24 December, 16 200824 Outline 1- Motivations of calibration –Experimental layout –Consequences on measurements 2- Real data calibration –Method –Results 3- Simulated data calibration –Method –Results 4- Conclusion

25 December, 16 200825 Conclusion Our goal was to calibrate the calorimeter with our set of Kin3  0 data and our set of Kin3  0 simulated to be able to make consistent fit for analysis. We have reached regime of stability for both real calibration and simulated calibration iteration processes => Calibration has been succesfully completed.


Download ppt "December, 16 20081 Calibration of electromagnetic calorimeter of Hall A DVCS experiment Eric FUCHEY Ph.D Laboratoire de Physique Corpusculaire."

Similar presentations


Ads by Google