Presentation is loading. Please wait.

Presentation is loading. Please wait.

Anesthesia Ventilators

Similar presentations


Presentation on theme: "Anesthesia Ventilators"— Presentation transcript:

1 Anesthesia Ventilators
Anesthesia Department Anesthesia Ventilators Raafat Abdelazim

2 Intended Learning Outcomes
By the end of this lecture, the student will be able to: Identify the components of the anesthesia ventilator (AV) Understand the physical principles of mechanical ventilation (MV) Operate the AV, monitor the parameters of MV and adjust its settings as required Apply the goals of MV Understand the hazards of MV Raafat Abdel-Azim

3 The motive power Electrical Pneumatical Mechanism of action (Mode of operation): 4 phases Inspiratory phase (I) Cycling from I to E Expiratory phase (E) Cycling from E to I T V P F Raafat Abdel-Azim

4 Respiratory cycle Two phases: I & E Two phase transitions:
From I to E = expiratory cycling Between E and I = inspiratory cycling Inspiration can itself sometimes have two phases: an active ‘flow’ (TI flow) phase during which gas is being delivered to the patient an end-inspiratory pause (TI pause ) The total duration of inspiration is made of the sum of these two: TI = TI flow + TI pause

5 Relationship of the Ventilator to the Breathing System
It replaces the reservoir bag in the BS It may be connected to the breathing system: using a hose that is attached at the bag mount the bag/ventilator selector valve without a hose (newer) During MV, the APL valve in the BS must be closed or isolated Most selector valves when turned for MV isolate the APL valve (no need to close) V B Raafat Abdel-Azim

6 B / V APL Raafat Abdel-Azim

7 B / V Raafat Abdel-Azim

8 The flow arrangement of a basic two-gas anesthesia machine
Raafat Abdel-Azim

9 Low pressure circuit P Raafat Abdel-Azim

10 Raafat Abdel-Azim

11 Raafat Abdel-Azim

12 Raafat Abdel-Azim

13 Pipeline Raafat Abdel-Azim

14 Back O2 Raafat Abdel-Azim

15 O2 O2 Ventilator – Control unit Anesthesia machine Side Back
Raafat Abdel-Azim

16 Bellows assembly Control Unit
Rigid housing Distensible bellows Control Unit Raafat Abdel-Azim

17 Control Unit Bellows assembly Distensible bellows Rigid housing
Raafat Abdel-Azim

18 Driving Gas Flow valve BS Raafat Abdel-Azim

19 Driving Gas BS Spill valve Safety relief valve Exhaust valve
Insp Spill valve Safety relief valve Exhaust valve Driving Gas Flow valve BS Raafat Abdel-Azim

20 BS Spill valve Safety relief valve Exhaust valve Flow valve
Raafat Abdel-Azim

21 BS Spill valve Safety relief valve Exhaust valve Flow valve
Raafat Abdel-Azim

22 Overpressure at any time
To atmosphere Safety relief valve BS Raafat Abdel-Azim

23 Exp To atmosphere Exhaust valve BS Raafat Abdel-Azim

24 Scavenging Spill valve Raafat Abdel-Azim

25 Spill Valve To scavenging system Raafat Abdel-Azim

26 Spill Valve I & beginning E End E (2-4 cmH2O) Driving gas
To scavenging system Patient’s gas I & beginning E End E (2-4 cmH2O) Raafat Abdel-Azim

27 Raafat Abdel-Azim

28 Traditional circle system
Raafat Abdel-Azim

29 Traditional circle system
In a system with no significant leaks, the amount of excess gas vented by the spill valve should approximate the incoming FGF Raafat Abdel-Azim

30 Raafat Abdel-Azim

31 Components of the Ventilator
Driving Gas Supply Injector (some) Controls (F, V, T, P) Alarms Safety-Relief Valve Bellows Assembly Exhaust Valve Spill Valve Ventilator Hose Connection preset cmH2O adjustable Raafat Abdel-Azim

32 Injector Air Driving gas Air
Some ventilators use a device called an injector (Venturi mechanism) to increase the flow of driving gas. As the gas flow meets restriction, its lateral pressure drops (Bernoulli principle). Air will be entrained when the lateral pressure drops below atmospheric. The end result is an increase in the total gas flow leaving the outlet of the injector but no increase in consumption of driving gas Raafat Abdel-Azim

33 Control of Parameters of Ventilation
VT VM f (frequency= rate) I:E ratio IFR (Insp. Flow Rate) Maximum Working Pressure Raafat Abdel-Azim

34 Raafat Abdel-Azim

35 Raafat Abdel-Azim

36 Narkomed - Drager Raafat Abdel-Azim

37 Fabius - Drager Raafat Abdel-Azim

38 Cisero - Drager Raafat Abdel-Azim

39 Dameca Raafat Abdel-Azim

40 Penlon a800/80 Raafat Abdel-Azim

41 Penlon Nuffield 200 Raafat Abdel-Azim

42 Raafat Abdel-Azim

43 IFR (Inspiratory Flow Rate)
VT (ml) f (b/min) Cycle time (s) IFR TI (s) TE (s) I:E L/min L/s ml/s 500 20 60/20 = 3 60 1 1000 0.5 2.5 1:5 30 2 1:2 60 L/min 600 Volume (ml) 30 L/min 500 400 300 200 100 0.5 1 1.5 2 2.5 3 3.5 4 4.5 5 5.5 6 Time (sec) Raafat Abdel-Azim

44 Inspiratory Waveforms
Constant Decelerating Accelerating (ramp) Sinusoidal Inspiration 60 Flow (L/min) 30 Expiration shorter TI TI TI TI shorter 30 60 Raafat Abdel-Azim

45 Raafat Abdel-Azim

46 The Ventilation Cycle Raafat Abdel-Azim

47 t Cycling Limiting Triggering - Volume - Flow - Flow - Time - Pressure
- Patient (assisted) - Machine (controlled) Raafat Abdel-Azim

48 IPPV Paw Pmax Pplat 20 Pause IF E t ZEEP I E Raafat Abdel-Azim

49 T V P F Duration of ventilation cycle (sec) f (60/duration)
I phase (IF period, IP period) E phase VT, VM TI, TE, I:E T V P F Raafat Abdel-Azim

50 R C Inspiratory Phase During the IF period: Paw depends on:
The airway resistance (R) The total thoracic compliance (C) (V/P) R C Raafat Abdel-Azim

51 P P P P P P P P Resistance P P P P Flow Rate: FRP P P
Raafat Abdel-Azim

52 Resistance Paw 20 t N R F R F Raafat Abdel-Azim

53 Raw = (PIP – Pplat) / Flow
Calculation of Airway Resistance (Raw) in a Ventilated Patient Raw = (PIP – Pplat) / Flow Examples: If in a given situation, PIP = 40 cm H2O, Pplat = 38 cmH2O, flow = 60 L/min (i.e., 1 L/s), Raw will be: = (40 − 38)/1 = 2 cmH2O/L/s PIP = 40 cm H2O, Pplat = 25 cmH2O, flow = 60 L/min (i.e., 1 L/s), Raw will be: = (40 − 25)/1 = 15 cmH2O/L/s Raafat Abdel-Azim

54 Compliance P P Volume C C Raafat Abdel-Azim

55 Compliance P P Volume: VP Raafat Abdel-Azim

56 During I pause Pause No gas F into or out of the lungs  Paw depends only on VI & CT Gas redistributes among alveoli This improves gas distribution in the lungs of patients with small AWD (BA, smokers). Raafat Abdel-Azim

57 C R Paw t EB intubation CW rigidity Pulmonary edema Secretions
Bronchospasm Kinked ETT Paw 20 t C R Raafat Abdel-Azim

58 Compliance = V/P Dynamic compliance: = VT/(PIP-PEEP) L/cmH2O
Static compliance: = VT/(Pplat-PEEP) L/cmH2O Raafat Abdel-Azim

59 Compliance represents a volume change per unit change in pressure (200 ml/cm H2O in the normal lung). The total respiratory compliance consists of combined lung and chest wall compliance and is normally 70–80 ml/cm H2O. Raafat Abdel-Azim

60 Raafat Abdel-Azim

61 Raafat Abdel-Azim

62 Time Constants of the Lung
In most diseases, the involvement of the lung is not uniform. Regional differences in C and R occur. Owing to this, alveoli in different parts of the lung behave differently; diseased alveoli take longer to fill and to empty. The rate of filling of an individual lung unit is referred to as its time constant. For a particular lung unit Time Constant = R x C Raafat Abdel-Azim

63 It takes the equivalent of 5 time constants for the lung to completely fill (or to empty).
In the time afforded by one time constant, 63% of the lung will fill (or empty); two time constants allow 86% of the inspiratory or expiratory phase to be completed; three time constants allow for 95%, and four time constants for 98%. 100 5 98 95 4 86 3 63 2 1 Raafat Abdel-Azim

64 A lung unit with a normal Raw of 1 cm H2O/L/s
Example: A lung unit with a normal Raw of 1 cm H2O/L/s and a normal compliance of 0.1 L/cm H2O would have a time constant of: = 1 × 0.1 = 0.1 s Five times this is 0.5 s, which would be the time required for this unit to fill or empty satisfactorily. This information comes useful while setting a ventilator’s TI and TE Since diseased air units take longer to fill, deliberately prolonging the TI may enable such units to participate more meaningfully in gas exchange Raafat Abdel-Azim

65 Compliance, Time Constant and WOB in Anaesthesia UK
MV slide show Raafat Abdel-Azim

66 During the expiratory phase
VT I E FRC FRC PEEP PaO2 FRC IA contents Pulmonary edema ARDS PaO2 Raafat Abdel-Azim

67 Goals of Pulmonary Ventilation
Raafat Abdel-Azim

68 necessary to maintain the desired PaCO2
To provide adequate minute alveolar ventilation and to  side effects necessary to maintain the desired PaCO2 PPV   ITP Raafat Abdel-Azim

69 Adequate Ventilation PaCO2 of 40 mmHg = 5.3% of 760 mmHg
Goals Adequate Ventilation PaCO2 of 40 mmHg = 5.3% of 760 mmHg 40/760 = 0.053 Normal resting VCO2= 200 ml/min= 0.2 L/min This requires VM of 3.8 L 0.2/ ? = 0.053 0.2/ = Add dead space (VD) 3.8 L/min Raafat Abdel-Azim

70  150 ml in a 70 Kg (154 pound) adult = 0.15 x 10 (f) =
Goals, Adequate Ventilation N = 2.2 ml/Kg (1 ml/pound)  150 ml in a 70 Kg (154 pound) adult = 0.15 x 10 (f) = Required VM = = A larger VM is required for patients who have VD or VCO2 1.5 L/min 5.3 L For VCO2 For VD VD phys = VD ana + VD alv VD ana = conducting airways = 150 ml in a 70 Kg adult VD alv is created when non-perfused alveoli are ventilated (negligible in health, expands in disease) This 150 ml of VD ana is reduced by ETT and can be cut down to about 60% by tracheostomy Raafat Abdel-Azim

71 When VCO2 & VD are stable: VM  1/ PaCO2 VM x PaCO2 = constant
Goals, Adequate Ventilation When VCO2 & VD are stable: VM  1/ PaCO2 VM x PaCO2 = constant e.g., PaCO2 = 50 mmHg with VM = 5 L/min  VM to 7 L/min   PaCO2 to 36 mmHg V1 x P1aCO2 = V2 x P2aCO2 Raafat Abdel-Azim

72 VM =VT x f Manipulation of VT has a different effect on the PaCO2 than does altering f. Consider the following: A set VT of 500 ml and f of 10 b/min results in a VM of 500 × 10 = 5,000 ml/min. The same VM can be produced by a VT of 250 ml delivered at f of 20 b/min, i.e., 250 × 20 = 5,000 ml/min. If, however, the VD is taken into consideration, the implications of these two settings are vastly different. Assuming a VD phys of 150 ml, the alveolar ventilation (the effective ventilation or the ventilation that takes part in gas exchange) in the first example would be: (500 – 150) × 10 = 3,500, and in the second example would be: (250 – 150) × 20 = 2,000. PaCO2 is inversely proportional not to all of the VM, but to that part of the ventilation that is independent of VD (i.e., the alveolar ventilation VA) Raafat Abdel-Azim

73  Side effects  CO PPV  ITP  PVR  RVA  VR  EDV Goals
Raafat Abdel-Azim

74 Goals,  Side Effects PPV  ITP VD PA > PAP Raafat Abdel-Azim

75 All these effects  mean Paw
Goals All these effects  mean Paw Therefore, a goal of PPV is to  mean Paw while maintaining adequate ventilation and oxygenation Raafat Abdel-Azim

76 Modes of Pulmonary Ventilation
Raafat Abdel-Azim

77 SB Paw t +ve -ve Anesthesia ICU: demand valve   WOB  CPAP
t -ve Raafat Abdel-Azim

78 CPAP Paw t Raafat Abdel-Azim

79 IPPV Paw t Anesthesia (or sedation) + MR Few days  muscle atrophy
t ZEEP Raafat Abdel-Azim

80 IPPV + PEEP Paw PEEP t Raafat Abdel-Azim

81 Assisted Ventilation (A)
Paw Patient  f and timing Hazard: hypoventilation A No A t S If  or No S Raafat Abdel-Azim

82 Assisted Ventilation (A)
Paw Patient  f and timing Hazard: hypoventilation A No A CPAP t If  or No S Raafat Abdel-Azim

83 Assist-Control (AC) Paw A+C Provides a minimum f below which  C
Raafat Abdel-Azim

84 AC Conscious patients are more comfortable on A & AC > C
Muscle atrophy is still a problem Most AV don’t provide A or AC Raafat Abdel-Azim

85 IMV   PaCO2 < apneic threshold  no SB  IMV = IPPV
Preset VT and f Paw SB is allowed Muscle atrophy is less likely S t IMV   PaCO2 < apneic threshold  no SB  IMV = IPPV Raafat Abdel-Azim

86 SIMV Paw t Preset VT and f Synch. Mandatory Mandatory NO S
t Triggering window Raafat Abdel-Azim

87 Pressure-limited ventilation (PLV) (PCV)
Paw Pmax Pplat t Raafat Abdel-Azim

88 Ventilator Classification
Pressure versus Volume Ventilators Single-Circuit versus Double-Circuit Ventilators Bellows Assembly: Ascending (=upright = standing) versus descending (=hanging) Raafat Abdel-Azim

89 P vs V Ventilators P Ventilators V Ventilators Changes in R & C change
VT P Commonly used in Neonates  Barotrauma  Compressible V ICU Most AV Other names P generators P-limited P-stable Flow generators V-stable Raafat Abdel-Azim

90 ICU Anesthesia Raafat Abdel-Azim

91 Bellows assembly Distensible bellows Rigid housing E Driving gas I
BS to patient Raafat Abdel-Azim

92 VT VT Drager (End-insp.) Ohmeda (End-exp.) Raafat Abdel-Azim

93 E Expands partially Fully expands I Empties completely
NA Drager bellows Ohmeda bellows E Expands partially Fully expands I Empties completely Partially empties Spill valve External (visible) Hidden in housing Raafat Abdel-Azim

94 E Ascending bellows Descending bellows Attachment Movement EEP
VT adjustment Disconnection or leak Raafat Abdel-Azim

95 Interaction between the Ventilator and Breathing Circuits
Raafat Abdel-Azim

96 Preset / delivered VT difference due to:
Interaction Preset / delivered VT difference due to: Effect of circuit compliance for all V ventilators: both single circuit and double circuit Effect of FGFR only for double circuit ventilators Raafat Abdel-Azim

97 Effect of Circuit Compliance
Interaction Effect of Circuit Compliance  PIP   Delivered VT Circuit compliance = Compressible volume (gas compression, tubing distension) CC= the V of gas that must be injected into a closed circuit to cause a unit  in the circuit P It is related to the volume of the circuit  P V Ventilator tubing V C Anesthesia 6 L 9 ml/cmH2O ICU  (1.5 ml/cmH2O) Raafat Abdel-Azim

98 Del VT = Preset VT – [(PIP - PEEP) x CC]
Interaction, Circuit Compliance Preset – Del VT difference  CC and PIP Del VT = Preset VT – [(PIP - PEEP) x CC] (i.e, VT loss = P x CC) Example: Ventilator tubing CC (ml/cmH2O) PIP (cmH2O) V Loss (ml) Anesthesia 9 20 30 180 270 ICU 1.5 45 Raafat Abdel-Azim

99 Volume loss is specially important in:
Interaction, Circuit Compliance Volume loss is specially important in: Patients with  PIP ICU patients Patients with small VT Neonates  PLV Raafat Abdel-Azim

100 Effect of Fresh Gas Flow Rate (from anesthesia machine)
Interaction Effect of Fresh Gas Flow Rate (from anesthesia machine)  FGFR   Delivered VT V FGF from AM Raafat Abdel-Azim

101 Volume added = TI x FGFR At a certain I:E,  f   TI   V added
Interaction, FGFR Volume added = TI x FGFR TI depends on I:E  I fraction of the respiratory cycle  f TI = 60/f x I fraction Example: I:E = 1:2, f = 20, FGFR = 6 L/min TI = 60/20 x 1/3 = 1 s FGFR = 6000 ml/min = 100 ml/s V added = 1 x 100 = 100 ml At a certain I:E,  f   TI   V added I:E I fr 1:1 1:2 1/3 1:3 Raafat Abdel-Azim

102 Interaction, FGFR Raafat Abdel-Azim

103 Volume added is specially important:
Interaction, FGFR Volume added is specially important: When making large changes in FGFR O2 flush  L/min (don’t use during insp.) When delivering small VT For a neonate, changing FGFR from 1 to 5 L/min  double VT Raafat Abdel-Azim

104 Delivered VT = Preset VT - V lost (compression)
Interaction Delivered VT = Preset VT - V lost (compression) + V added (An. machine) Raafat Abdel-Azim

105 FGF Independent of Ventilation
Fresh gas compensation – GE Compensate for FGF Active, feedback controls bellows movement Fresh gas decoupling – Draeger FGF goes to reservoir bag during inspiration Fresh gas accommodation – Maquet Flow-i Fresh gas flows only during inspiration Ventilation by circuit flow interruption – Draeger Zeus, Perseus Raafat Abdel-Azim

106 Fresh Gas Decoupling (FGD)
Raafat Abdel-Azim

107 Fresh Gas Decoupling (FGD)
Circle system with piston ventilator and fresh gas decoupling (FGD) (Dräger Narkomed 6000) During the I phase the FG coming from the AWS through the fresh gas inlet is diverted into a separate reservoir by a “decoupling valve” (located between the fresh gas source & the circle BS). In the case of the NM 6000 series, the reservoir bag serves as an accumulator for FG storage until the E phase begins. Raafat Abdel-Azim

108 NPR valve = negative pressure relief valve
During the E phase, the decoupling valve opens, allowing the accumulated FG in the reservoir bag to be drawn into the circle system to refill the piston ventilator chamber. Because the ventilator exhaust valve also opens during the E phase, excess FG & exhaled patient gases are vented into the scavenging system Raafat Abdel-Azim

109 Most FGD systems are designed with either piston-type or descending bellows-type ventilators.
Because the ventilator piston unit or bellows in either of these type systems refills under slight negative pressure, it allows the accumulated fresh gas from the reservoir to be drawn into the breathing circuit for delivery to the patient during the next ventilator cycle. As a result of this design requirement, it is not possible for FGD to be used with conventional ascending bellows ventilators, which refill under slight +ve pressure Raafat Abdel-Azim

110  risk of barotrauma & volutrauma
Advantage of FGD  risk of barotrauma & volutrauma With a traditional circle system, in FGF from the flow meters, or from inappropriate activation of the oxygen flush valve, may contribute directly to VT  excessive FG  pneumothorax, pneumomediastinum, other serious patient injury, or even death. Because systems with FGD isolate the patient from FG coming into the system while the ventilator is in I phase, delivered VT tends to remain stable over a broad range of FGF rates, and the potential risk of barotrauma to the patient is greatly reduced. Raafat Abdel-Azim

111 Disadvantages of FGD The possibility of entraining room air into the patient gas circuit In a FGD system, the ventilator piston unit (or descending bellows) refills under slight -ve pressure. If the total volume of gas contained in the reservoir bag + that returning as exhaled gas from the patient’s lungs is inadequate to refill the ventilator piston unit, -ve patient Paw could develop. To prevent this, a negative pressure relief valve is placed in the FGD BS. If BS P  < a preset threshold value, then the relief valve opens and ambient air is allowed to enter into the patient gas circuit. If this goes undetected, the entrained atmospheric gases could lead to dilution of either or both the inhaled anesthetic agent(s) or an enriched oxygen mixture (lowering an enriched oxygen concentration toward 21%). If allowed to continue, this could lead to intraoperative awareness and/or hypoxia. High-priority alarms with both audible and visual alerts should notify the user that fresh gas flow is inadequate and room air is being entrained. Raafat Abdel-Azim

112 If the reservoir bag is removed during mechanical ventilation, or if it develops a significant leak (from poor fit on the bag mount or a bag perforation), room air may enter the breathing circuit as the ventilator piston unit refills during the expiratory phase. This could also result in dilution of either the inhaled anesthetic agent(s) concentration and/or the enriched oxygen mixture. This type of a disruption could lead to significant pollution of the operating room with anesthetic escaping into the atmosphere. Raafat Abdel-Azim

113 Hazards of Ventilators
Raafat Abdel-Azim

114 Apnea Erroneous setting of ventilator selector switch
Hazards Apnea Erroneous setting of ventilator selector switch Ventilator switched off Disconnection Obstruction of inspiratory gas pathway Ventilator failure V B Raafat Abdel-Azim

115 Hypoventilation Circuit leak Incompetent spill valve
Hazards Hypoventilation Circuit leak Incompetent spill valve Open APL valve in circuit Driving gas leak Improper settings Ventilator limitation at high airway pressure PL S Raafat Abdel-Azim

116 Raafat Abdel-Azim

117 Hyperventilation & dilution of anesthetic gases
Bellows leak Raafat Abdel-Azim

118 Obstruction of expiratory gas pathway Ventilator cycling failure
Hazards High airway pressure Exhaust obstruction Obstruction of expiratory gas pathway Ventilator cycling failure Improper settings Oxygen flush during inspiration Bellows leak Raafat Abdel-Azim

119 Summary of important points
Driving gas and patient gas don’t mix The ventilation cycle is controlled by F, V, P and T During the I phase, Paw depends on R and CT Goals of MV are to provide adequate VM (necessary to maintain the desired PaCO2) and to side effects Modern AV provide a range of ventilatory modes The difference between the delivered VT and the preset VT depends on CC and FGF FGD decreases the risk of barotrauma and volutrauma Hazards of MV can be avoided Raafat Abdel-Azim

120 Examples of questions to assess the ILOs
Describe causes and ventilatory management of high airway pressure during inspiratory phase of mechanical ventilation Discuss goals of pulmonary ventilation under anesthesia With IPPV: if f (rate) = 10/min, I: E= 1:1 and FGF= 3 LPM What will be the volume added to the preset tidal volume? Raafat Abdel-Azim

121 APL = adjustable pressure limiting BS = breathing system
I = inspiration E = expiration TI = inspiratory time TE = expiratory time f = respiratory rate (frequency) VT = tidal volume VM = VE = minute volume (minute ventilation) Raafat Abdel-Azim

122 Thank you Raafat Abdel-Azim


Download ppt "Anesthesia Ventilators"

Similar presentations


Ads by Google