1 MCSRN Mechanical Circulatory Support Research Network A Multi-Institutional Comparison of Adverse Events in Contemporary Continuous Flow LVAD Significant Differences Exist? Stulak JM, Davis ME, Haglund N, Dunlay S, Cowger J, Shah P, Pagani FD, Aaronson KD, Maltais S

2 Disclosures Research Funding (FDP, KA, SM) NHLBI HeartWare MCSRN

3 Background INTERMACS 6th annual report
8 years, >10,000 pt, accrual 2,000pt/yr Very few comparisons of 2nd and 3rd generation LVADs Aaronson et al., 2012 (n=499 vs. 140) Lalonde, et al., 2013 (n=13 vs. 33) Pagani, et al., 2015 (n=148 vs. 297) MCSRN

4 Aim Evaluate outcomes in a large, multicenter cohort with an “all comers” approach Stratify outcomes according to contemporary 2nd and 3rd generation LVAD Report overall incidence and time-related cumulative risk of AE’s MCSRN

5 Methods IRB approval Retrospective review of prospectively collected database Outcomes Survival Bleeding, TE events Neurologic events Infection MCSRN

6 Patients May 2004 Sept 2014 n =734 Age 18 82 57 yr 577 157
560 HMII, 174 HVAD Age 18 82 57 yr 577 157 MCSRN

7 Preop Clinical Characteristics
HMII HVAD p Age at implant INTERMACS Creatinine Male % % 0.002 HTN % % 0.013 BTT % % <0.001 Preop IABP 48% % 0.011 MCSRN

8 Intraop Clinical Characteristics
HMII HVAD p Redo sternotomy 33% 31% 0.51 TV surgery 38% 21% <0.001 AV surgery % % 0.22 Temporary RVAD 5% % 0.44 MCSRN

9 Early Postop Outcome HMII HVAD p Early mortality 7.4% 7.5% 0.95
Dialysis % 10% 0.96 Postop inotropes 6 days 5 days 0.3 LOS days 16 days <0.001 MCSRN

10 1,120 patient-years of support
Follow-up n = 680 Max., 10.4 yrs. Median, 1 yrs. 1,120 patient-years of support MCSRN

11 Survival by Device Percent HeartMate II HeartWare HVAD p=0.96 Years

12 GI Bleeding by Device HeartMate II HeartWare HVAD Percent p=0.18
Follow-up time, years

13 Stroke by Device HeartMate II HeartWare HVAD Percent p=0.006
Follow-up time, years

14 Any Infection by Device
HeartMate II HeartWare HVAD Percent p=0.039 Follow-up time, years

15 Driveline Infection by Device
HeartMate II HeartWare HVAD Percent p=0.14 Follow-up time, years

16 Pump Thrombus by Device
HeartMate II HeartWare HVAD Percent p=0.35 Follow-up time, years

17 AE Risk Comparison Thrombus Survival Stroke Infection GI bleed
Driveline Hazard ratio vs. HeartMate II

18 AE Comparison (EPPY) HMII HVAD p DLI 0.12 0.06 0051
Stroke Pump thrombus MCSRN

19 Multivariable Model Survival p=0.27 GI bleeding p=0.63
Any infection p=0.32 DLI p=0.1 Stroke p=0.003 (HVAD, H.R.:1.8) Pump thrombus p=0.64 MCSRN

20 Summary and Conclusions
Large, multicenter collaborative analysis to stratify outcomes between devices with “all comers” strategy HeartWare HVAD ↑ stroke risk EPPY (trends) HeartMate II ↑ DLI HVAD ↑ stroke, pump thrombus MCSRN

21 Perspective Understanding differences enhances preop counseling and postop management Need for more collaborative efforts Ultimate result? Guidelines, “best practices” MCSRN

22 MCSRN Mechanical Circulatory Support Research Network


Download ppt "MCSRN Mechanical Circulatory Support Research Network"

Similar presentations


Ads by Google