Presentation is loading. Please wait.

Presentation is loading. Please wait.

University of Wisconsin Diagnostic Imaging Research.

Similar presentations


Presentation on theme: "University of Wisconsin Diagnostic Imaging Research."— Presentation transcript:

1 University of Wisconsin Diagnostic Imaging Research

2 Lecture 1: Introduction (1/2) – History, basic principles, modalities Class consists of: 1)Deterministic Studies - distortion - impulse response - transfer functions All modalities are non-linear and space variant to some degree. Approximations are made to yield a linear, space-invariant system. 2)Stochastic Studies SNR (signal to noise ratio) of the resultant image - mean and variance

3 Nov. 1895 – Announces X-ray discovery Jan. 13, 1896 – Images needle in patient’s hand – X-ray used presurgically 1901 – Receives first Nobel Prize in Physics – Given for discovery and use of X-rays. Wilhelm Röntgen, Wurtzburg Radiograph of the hand of Röntgen’s wife, 1895.

4 Röntgen detected: No reflection No refraction Unresponsive to mirrors or lenses His conclusions: X-rays are not an EM wave Dominated by corpuscular behavior Röntgen’s Setup

5 Projection X-Ray Disadvantage:Depth information lost Advantage:Cheap, simple attenuation coefficient Measures line integrals of attenuation Film shows intensity as a negative ( dark areas, high x-ray detection)

6 SagittalCoronal

7 Early Developments Intensifying agents, contrast agents all developed within several years. Creativity of physicians resulted in significant improvements to imaging. - found ways to selectively opacify regions of interest - agents administered orally, intraveneously, or via catheter

8 Later Developments More recently, physicists and engineers have initiated new developments in technology, rather than physicians. 1940’s, 1950’s Background laid for ultrasound and nuclear medicine 1960’s Revolution in imaging – ultrasound and nuclear medicine 1970’s CT (Computerized Tomography) - true 3D imaging (instead of three dimensions crammed into two) 1980’s MRI (Magnetic Resonance Imaging) PET ( Positron Emission Tomography)

9 Computerized Tomography (CT) 1972Hounsfield announces findings at British Institute of Radiology 1979 Hounsfield, Cormack receive Nobel Prize in Medicine (CT images computed to actually display attenuation coefficient  x,y  Important Precursors: 1917 Radon: Characterized an image by its projections 1961 Oldendorf: Rotated patient instead of gantry Result:

10 First Generation CT Scanner Acquire a projection (X-ray) Translate x-ray pencil beam and detector across body and record output Rotate to next angle Repeat translation Assemble all the projections.

11 Reconstruction from Back Projection 1.Filter each projection to account for sampling data on polar grid 2. Smear back along the “line integrals” that were calculated by the detector.

12 Modern CT Scanner From Webb, Physics of Medical Imaging

13 Computerized Tomography (CT), continued Early CT Image Current technology

14 Exhalation Inhalation

15 Nuclear Medicine -Grew out of the nuclear reactor research of World War II -Discovery of medically useful radioactive isotopes 1948 Ansell and Rotblat: Point by point imaging of thyroid 1952 Anger: First electronic gamma camera a)Radioactive tracer is selectively taken up by organ of interest b)Source is thus inside body! c)This imaging system measures function (physiology) rather than anatomy.

16 Example in medical imaging: Consider a nuclear study of a liver with a tumor point source at x 1 = , y 1 =  Radiation is detected at the detector plane. To obtain a general result, we need to know all combinations h(x 2, y 2 ; ,  ) By “general result”, we mean that we could calculate the image I(x 2, y 2 ) for any source input S( x 1, y 1 )

17 Nuclear Medicine, continued Very specific in imaging physiological function - metabolism - thyroid function - lung ventilation: inhale agent Advantage:Direct display of disease process. Disadvantage:Poor image quality (~ 1 cm resolution) Why is resolution so poor? Very small concentrations of agent used for safety. - source within body Quantum limited: CT 10 9 photons/pixel Nuclear ~100 photons/pixel Tomographic systems: SPECT: single photon emission computerized tomography PET: positron emission tomography

18 Combined CT / PET Imaging

19 Necessary Probe Properties Probe can be internal or external. Requirements: a)Wavelength must be short enough for adequate resolution. bone fractures, small vessels < 1 mm large lesions < 1 cm b)Body should be semi-transparent to the probe. transmission > 10 -1 - results in contrast problems transmission < 10 -3 - results in SNR problems λ > 10 cm- results in poor resolution λ <.01Å- negligible attenuation Standard X-rays:.01 Å < λ <.5 Å corresponding to ~ 25 kev to 1.2 Mev per photon

20 Necessary Probe Properties: Transmission vs. λ Graph: Medical Imaging Systems Macovski, 1983

21 Probe properties of different modalities NMR Nuclear magnetic moment ( spin) Makes each spatial area produce its own signal Process and decode Ultrasound Not EM energy Diffraction limits resolution resolution proportional to λ

22 Introduction (2/2) – Comparison of Modalities Review: Modalities: X-ray: Measures line integrals of attenuation coefficient CT: Builds images tomographically; i.e. using a set of projections Nuclear: Radioactive isotope attached to metabolic marker. Strength is functional imaging, as opposed to anatomical Ultrasound: Measures reflectivity in the body.

23 Ultrasound Ultrasound uses the transmission and reflection of acoustic energy.  prenatal ultrasound image  clinical ultrasound system

24 Ultrasound A pulse is propagated and its reflection is received, both by the transducer. Key assumption: - Sound waves have a nearly constant velocity of ~1500 m/s in H 2 O. - Sound wave velocity in H 2 O is similar to that in soft tissue. Thus, echo time maps to depth.

25 Ultrasound: Resolution and Transmission Frequency Tradeoff between resolution and attenuation - ↑ higher frequency ↓ shorter wavelength ↑ higher attenuation Power loss: Typical Ultrasound Frequencies: Deep Body 1.5 to 3.0 MHz Superficial Structures 5.0 to 10.0 MHz e.g. 15 cm depth, 2 MHz, 60 dB round trip Why not use a very strong pulse? Ultrasound at high energy can be used to ablate (kill) tissue. Cavitation (bubble formation) Temperature increase is limited to 1º C for safety.

26 Magnetic Resonance Imaging

27 Main Magnetic Field B0B0B0B0

28 Magnetic Resonance Imaging There are 3 magnetic fields of interest in MRI. The first is the static field B o. 1) polarizes the sample: 2) creates the resonant frequency: γ is constant for each nucleus: density of 1 H ω = γB

29 Proton Spin Creates Signal Source B0B0B0B0  =  B 64 MHz for H + at 1.5T B0B0B0B0

30 Second Magnetic Field : RF Field B1B1B1B1 An RF coil around the patient transmits a pulse of power at the resonant frequency ω to create a B field orthogonal to B o. This second magnetic field is termed the B 1 field. B 1 field “excites” nuclei. Excited nuclei precess at ω(x,y,z) = γB o (x,y,z)

31

32 Transmit Coils Preamp DemodulateA/D RF Coil

33 Spin Encoding

34

35 Magnetic Resonance The spatial location is encoded by using gradient field coils around the patient. (3 rd magnetic field) Running current through these coils changes the magnitude of the magnetic field in space and thus the resonant frequency of protons throughout the body. Spatial positions is thus encoded as a frequency. The excited photons return to equilibrium ( relax) at different rates. By altering the timing of our measurements, we can create contrast. Multiparametric excitation – T 1, T 2

36 Brain Glioma

37 Sagittal Carotid Non-contrast-enhanced MRI Coronal

38 Contrast-enhanced Abdominal Imaging

39 Time-resolved Abdominal Imaging

40 Contrast-enhanced MR Cardiac Imaging

41 Fat Coronal Knee ImageWater Coronal Knee Image

42 Comparison of modalities Why do we need multiple modalities? Each modality measures the interaction between energy and biological tissue. - Provides a measurement of physical properties of tissue. - Tissues similar in two physical properties may differ in a third. Note: - Each modality must relate the physical property it measures to normal or abnormal tissue function if possible. - However, anatomical information and knowledge of a large patient base may be enough. - i.e. A shadow on lung or chest X-rays is likely not good. Other considerations for multiple modalities include: - cost- safety - portability/availability

43 Comparison of modalities: X-Ray Measures attenuation coefficient Safety: Uses ionizing radiation - risk is small, however, concern still present. - 2-3 individual lesions per 10 6 - population risk > individual risk i.e. If exam indicated, it is in your interest to get exam Use: Principal imaging modality Used throughout body Distortion: X-Ray transmission is not distorted.

44 Comparison of modalities: Ultrasound Measures acoustic reflectivity Safety: Appears completely safe Use: Used where there is a complete soft tissue and/or fluid path Severe distortions at air or bone interface Distortion: Reflection: Variations in c (speed) affect depth estimate Diffraction: λ ≈ desired resolution (~.5 mm)

45 Comparison of modalities: Magnetic Resonance (MR) Multiparametric M(x,y,z) proportional to ρ(x,y,z) and T 1, T 2. (the relaxation time constants) Velocity sensitive Safety: Appears safe Static field - No problems - Some induced phosphenes Higher levels - Nerve stimulation RF heating: body temperature rise < 1˚C - guideline Use: Distortion: Some RF penetration effects - intensity distortion

46 Clinical Applications - Table ChestAbdomenHead X-Ray/ CT + widely used + CT - excellent – needs contrast + CT - excellent + X-ray - is good for bone – CT - bleeding, trauma Ultrasound– no, except for + heart + excellent – problems with gas – poor Nuclear+ extensive use in heart Merge w/ CT+ PET MR+ growing cardiac applications + minor role+ standard

47 Clinical Applications – Table continued… CardiovascularSkeletal / Muscular X-Ray/ CT + X-ray – Excellent, with catheter-injected contrast + strong for skeletal system Ultrasound+ real-time + non-invasive + cheap – but, poorer images – not used Nuclear+ functional information on perfusion + functional - bone marrow MR+ getting better High resolution Myocardium viability + excellent

48 Economics of modalities: Ultrasound: ~ $100K – $250K CT: $400K – $1.5 million (helical scanner) MR: $350K (knee) - 4.0 million (siting) Service: Annual costs Hospital must keep uptime Staff: Scans performed by technologists Hospital Income: Competitive issues Significant investment and return


Download ppt "University of Wisconsin Diagnostic Imaging Research."

Similar presentations


Ads by Google