1 Finnish Custom Lab., Pekka Ravio
Experiences on developing a screening method for Orbitrap mass spectrometer Esitys käsittelee kemiallista kvalitatiivista orgaanista analytiikkaa tänä päivänä nestekromatografia-massaspektrometrian näkökulmasta. 11. NPRW , Tallinn Finnish Custom Lab., Pekka Ravio

2 Finnish Custom Lab., Pekka Ravio
Orbitrap Elite ETD Hankinta: olemassa olevasta 15 vuotta vanhasta ioniloukusta > ioniloukku-orbiloukkuun >> hinta >> FIMEAn mukaan tulo Laitteen osien kuvaus: neste – interface –ioniloukku – orbiloukku - ETD 11. NPRW , Tallinn Finnish Custom Lab., Pekka Ravio

3 Orbitrap Elite ETD: schematic
Toimintaperiaatteen ero verrattuna moneen muuhun massaspektrometriin: analysoidaan ionipaketteja, ei jatkuvaa ionivirtaa Sähkösumutus interface – kaksiosainen ioniloukku – korkea paine loukussa erilaiset operaatiot – matalapaineloukusta skannataan – omat detektorit – C-trap - Orbitrap toimintaperiaate – HCD cell - ETD

4 Finnish Custom Lab., Pekka Ravio
Pesticide residues screening with Orbitrap Elite Primary instrumental parameters to choose: 1. Mass resolution 15 000, , , or (at m/z 400) 2. Fragmention technique In-source CID, ion trap, HCD all ion fragmentation 3. Collision energy 11. NPRW , Tallinn Finnish Custom Lab., Pekka Ravio

5 Finnish Custom Lab., Pekka Ravio
Measurement Alternating scanning Full scan, mass range 130 – 900 amu All ion fragmentation in-source CID, mass range amu, collision energy 45 V Cycle time around 0.5 s (R = ) > 15 or more scans over UPLC peak External mass calibration 11. NPRW , Tallinn Finnish Custom Lab., Pekka Ravio

6 Finnish Custom Lab., Pekka Ravio
Alternating scan run, full scan – in source CID scan Zoom 11. NPRW , Tallinn Finnish Custom Lab., Pekka Ravio

7 Finnish Custom Lab., Pekka Ravio
EUPT-SM , pepper + around 30 pestisides Furathiocarb, C18H26N2O5S, M ∆ m =1,6 ppm Full scan C10H11O2S , ∆ m = 4.1 ppm In-source CID 11. NPRW , Tallinn Finnish Custom Lab., Pekka Ravio

8 Finnish Custom Lab., Pekka Ravio
Screening tests at the moment against a database and MS/MS library only, no validation with pesticide standards Database bought and collected from literature, 450 compounds Data processing software TraceFinder 3.2 Processing against database and MS/MS spectral library (bought) 11. NPRW , Tallinn Finnish Custom Lab., Pekka Ravio

9 Tullilaboratorio, Pekka Ravio
Details in compound database Tullilaboratorio, Pekka Ravio

10 Tullilaboratorio, Pekka Ravio
TraceFinder processing parameters Tullilaboratorio, Pekka Ravio

11 Finnish Custom Lab., Pekka Ravio
Processing with TraceFinder software 11. NPRW , Tallinn Finnish Custom Lab., Pekka Ravio

12 Finnish Custom Lab., Pekka Ravio
Finding correct fragments ” fragment soup” + library search in-source CID scan 0.019 mg/kg fenhexamid in raspberry Mass resolution 60 000 Fenhexamid MS/MS-library spectra 11. NPRW , Tallinn Finnish Custom Lab., Pekka Ravio

13 Tullilaboratorio, Pekka Ravio
Evaluating isotopic pattern, red=meas., blue=theor. Fenhexamid 0,019 mg/kg Tullilaboratorio, Pekka Ravio

14 Tullilaboratorio, Pekka Ravio
Screening results, EUPT-SM06 Tullilaboratorio, Pekka Ravio

15 Tullilaboratorio, Pekka Ravio
Screening results EUPT-SM07 Tullilaboratorio, Pekka Ravio

16 Finnish Custom Lab., Pekka Ravio
Conclusions Orbitrap Elite Easy to use Stable mass calibration Same mass calibration may be used 1- 3 days (error < 3 ppm) All ion fragmentation with HCD-cell would be nice Data processing with TraceFinder 3.2 Visual outlook very informative Should be able to use several collision energies 11. NPRW , Tallinn Finnish Custom Lab., Pekka Ravio

17 Finnish Custom Lab., Pekka Ravio
Principle how the isotopic pattern is evaluated is unclear Overall conclusions Encouraging results All depends on the database A lot of work to build up the database, ”fine tune” it and ”fine tune” the processing parameters + MS/MS library The more compounds you have in database, the more false positives you get The software needs some upgrading and better manuals 11. NPRW , Tallinn Finnish Custom Lab., Pekka Ravio

18 Finnish Custom Lab., Pekka Ravio
”Our newest weapon” Customs moneydog Massi Thank You 11. NPRW , Tallinn Finnish Custom Lab., Pekka Ravio