3D Localization Microscopy With Ground State Depletion (GSD 3D)

Similar presentations


Presentation on theme: "3D Localization Microscopy With Ground State Depletion (GSD 3D)"— Presentation transcript:

1 3D Localization Microscopy With Ground State Depletion (GSD 3D)

2 Basic Principles of the GSD Super-Resolution Technology for Displaying Highly Resolved Cellular Structures in Three Dimensions The exact position of a fluorochrome is based on the principle of temporal separation; This requires switching of fluorochromes between on and off states followed by a sequential readout;

3 Super-resolution imaging
The crucial step of dSTORM/GSD is to temporarily switch off the majority of fluorophores to allow the precise localization of single fluorophores.

4 Super-resolution as reconstructed z-stack
Fluorophore in the focal plane: the image appears round; Below the focus: elongated in the vertical direction; Higher than the focus position: elongated in the horizontal direction.

5 Section localization scheme

6 Super-resolution as reconstructed z-stack
Coordinates in z are assigned to a specific layer in a z-stack based on a calibration of the deformed PSF by the cylinder lens. Resolution around 70 nm can be achieved.

7 Evolution of resolution
AlexaFluor 488 AlexaFluor 532 Rhodamine 6G AlexaFluor 647 Others.

8