Duke nEDM Collaboration Meeting The status of 3 He Relaxation Time Measurement at ~400mK Q. Ye, D. Dutta, H. Gao, W. Zheng, X. Zhu Duke University R. Golub,

Similar presentations


Presentation on theme: "Duke nEDM Collaboration Meeting The status of 3 He Relaxation Time Measurement at ~400mK Q. Ye, D. Dutta, H. Gao, W. Zheng, X. Zhu Duke University R. Golub,"— Presentation transcript:

1 Duke nEDM Collaboration Meeting The status of 3 He Relaxation Time Measurement at ~400mK Q. Ye, D. Dutta, H. Gao, W. Zheng, X. Zhu Duke University R. Golub, P. Huffman, F. Dubose NC State University May 20, 2008

2 Duke nEDM Collaboration Meeting Outline Experimental Apparatus 2007 Summer Results Modifications to the MC & acrylic cell 2008 April Results Future plan

3 Experimental Schematics Detachable cell Detachable cell mount d-TPB coated acrylic cell RF coils Dilution Refrigerator pickup coil 4 He capillary 3 He transfer Buffer Volume (copper) Mixing Chamber Filled with polarized 3 He Distance between the top polarized 3 He cell and the bottom acrylic cell ~ 218.4cm d-TPB coated acrylic cell 293.4 cm IVC

4 Transfer Polarized 3 He Transfer polarized 3 He onto top of the DR using portable magnetic field –90% polarization left

5 Duke nEDM Collaboration Meeting 2007 Results Cell filled with 1.23mol super-fluid 4 He (~63% full), cooled down to 0.530K Total amount of 3 He in the cell was 0.00092 mol (polarization~5%) 3 He relaxation time (~1263 seconds with an AFP loss of 0.5%) at 7.5G holding field Hard to cool down the cell filled with super-fluid 4 He

6 Duke nEDM Collaboration Meeting After Modification 0.25mm capillary restrict 4 He film flow Grooves for copper wires to go thru Gold- plated 99.999% copper wires

7 2008 Results Temperature sensor holder Small copper piece For the 1.5mm pure copper cooling wire to go through 1/8” copper tubing

8 Leak found at the joint Duke nEDM Collaboration Meeting The leak was found around this region. The hole was filled with stycast 1266.

9 Heat Input Study Lowest MC temp. ~ 44mK with the acrylic cell being empty (~15.3 kOhm) Duke nEDM Collaboration Meeting

10 During the filling process, 1Kpot temperature ~ 2.05K (25.8 kOhm) Still pressure ~ 8.5e-2 mBar Still turbo power ~ 200W IVC pressure ~ 1.2-1.4e-5 mBar Stable operation Condensing 4 He into acrylic cell volume: ~ 54cc (need ~ 5000 torr 7.5 liter 4 He to fill)

11 3 He Relaxation Time Duke nEDM Collaboration Meeting Wall depolarization probability is deduced 3 He relaxation time measurements

12 nEDM experimental cell The neutron EDM experimental cell ~ 7.6cm × 10.2cm × 50.0cm (S/V ~ 0.5) Scaled 3 He relaxation time ~ 4260s (S/V=1.284) or 4910s (S/V = 1.48). Duke nEDM Collaboration Meeting

13 Next steps Depolarization probability comparable (slightly worse) to that from 1.9K (~2e-7, Q. Ye et al. to appear in PRA) Improve 3 He relaxation time, how? –Better coating? How do we quantify coating, using atomic force microscope (AFM)? Put on gradient coils along z-axis to study geometric phase effect under nEDM experimental conditions

14 Duke nEDM Collaboration Meeting Thank you!