Integration of Aberrometry and Topography with the i Trace System

Similar presentations


Presentation on theme: "Integration of Aberrometry and Topography with the i Trace System"— Presentation transcript:

1 Integration of Aberrometry and Topography with the i Trace System
July 2005 Joe S. Wakil, MD - Tracey Technologies, LLC EyeSys Vision, Inc.

2 Founding Technology Developers:
Vasyl Molebny, DSc Kiev, Ukraine Ioannis Pallikaris, MD Crete, Greece Canadian & Swedish Governments

3 Why Aberroscopy? Current Laser Technology Permits One to Go Beyond Correction of Sphere and Astigmatism You Can Now Address Your Patient’s Quality of Vision High Order Aberrations Define Quality of Vision

4 Why Ray Tracing? Because… the Eye is NOT a Telescope!
Pupil Size: Accommodation: Tear Film: Effects Refraction (ex. Night Myopia) Refraction is NOT a FIXED Number! Effects Refraction (Instrument Myopia) Effects Quality of Vision (Aberrations) Where are the Sources of Aberrations? How do they change with Surgery? Cornea: Lens: Astigmatism (Irreg.), Sph & other HOA Astigmatism, Coma & other HOA

5 The Eye is NOT a Telescope
Off-Axis design No magnification Variable aperture Variable detector res. Accommodation Changing fixation Brain image processing Nature-made TELESCOPE On-Axis design High magnification Fixed aperture Constant detector res. No accommodation Fixed alignment Digital image processing Man-made

6 Significant Higher Order Aberrations
Trefoil Coma Spherical Aberration

7 Aberrometer/Wavefront Technologies
Hartmann-Shack Lenslet Array Tscherning Aberrometer Differential Skiascopy Ray Tracing Features: -Rapid, point by point, IR measurement no data confusion -Pupillometry with auto-tracking/capture -Programmable sampling (256 pts.) in any pupil up to 8mm -Open Field Fixation – avoid instrument myopia and measure Accommodation -Corneal Topography integration – able to measure Lens Aberrations

8 Hartmann Shack

9 Hartmann Shack

10 Hartmann-Shack Wavefront Sensor
H/S Photo of patient with tight eye lid courtesy David Williams

11 Tscherning

12 Disadvantages of H-S and Tscherning
Measures All Points at Once Data Confusion, Compromised Resolution Limited Dynamic Range – Cannot Measure Highly Irregular Eyes Highly Sensitive to Noise – Slow, Requires Multiple Scans Expensive Components – High Cost to Purchase and Repair H-S Measures Reverse Aberrations – Not Physiologic with Real Vision especially for High Orders in Accomodation Tscherning Needs 2-D Imaging of Retina Additional noise and errors

13 Differential Skiascopy

14 Disadvantages of Differential Skiascopy
Does NOT Measure Skew Aberrations – Inaccurate WaveFront especially for Trefoil, Quadrafoil, etc. Measures Multiple Points at Once (slit) and only in Perpendicular Direction Limited WF measurement (axial bias) No Open Field Fixation – Problem of instrument myopia in young patients

15 Total Ocular Aberrations
Internal Optics Aberrations Corneal Aberrations Total Ocular Aberrations *Measuring Corneal Aberration without Lens or Total Aberration is of Questionable Value

16 The iTrace

17 Principles of Tracey Programmable thin beam ray tracing measuring forward aberrations of the eye Rapid sequential measurement of data points over entire entrance pupil (<50ms) Localization of each reflected retinal spot Integration of individual retinal spots to form Point Spread Function (PSF) Analysis of PSF for higher order aberrations and other data formats

18 Programmable Data Sample Points

19 Multiplying the Number of Sites

20 Higher Local Density of Sites

21 Overlay of Two Sets of Site Configuration

22 Refractive Error Measurements
Myopia Hyperopia

23 Retinal Spot Diagram/Point Spread Function

24 Data Displays Retinal Spot Diagram Refraction Map Ablation Map
X, µm 20 10 -10 -20 Y, µm 40 30 Refraction Map Retinal Spot Diagram Wavefront Map Ablation Map

25 Tracey’s Key Advantage: NO COMPROMISES!
Rapid, point-by-point analysis of 256 data points avoids data confusion associated with simultaneous data measurements, therefore, all eyes (highly irregular) can be measured. All points in any pupil size (2-8mm) each with full dynamic range (+/- 15 D). NO COMPROMISES!

26 Baylor Clinical Study (100 eyes) by Doug Koch, MD

27 Validation Studies Three independent studies of
Tracey vs. Manifest Refraction Koch et al eyes Slade et al - 42 eyes Schalhorn et al eyes Results Accuracy to manifest <0.12 D Reproducibility <0.12 D

28 The iTrace

29 Normal Eye

30 Irregular Eye

31 UCVA vs BCVA

32 Post LASIK

33 Full Corneal Topography

34 Full Corneal Topography

35 Keratoconus

36 Normal

37 iTrace Measures Accommodation Mechanism

38 Very Spherical Accommodation

39 Variations in Mapping Accommodative Power in the Natural Crystalline Lens as Measured by iTrace
Horizontal Cyl Sphere Vertical Cyl Coma

40 Crystalens Accommodative Arching
73 Year Old Male Overall Refraction change is 0.5D but Central Cylinder 2.5 D adds Depth of Field to Enhance Accommodative Effect

41 MultiFocal IOL Analysis with i Trace
PSF Analysis Modulation Transfer Function (MTF)obust Aberrometer Pupil Dependent Analysis Multi-Zone Refraction Analysis Retinal Spot Diagram Conoid of Sturm Dynamic Analysis Complete Corneal Topography Analysis Separates Corneal from Total Aberrations Resulting in Lenticular (internal ocular) Aberrations Measures Accommodation

42 Multifocal Acrylic IOL
Alcon ReStor Lens

43

44

45

46

47

48

49

50

51

52

53

54 Monofocal Acrylic IOL Alcon SA-60

55

56

57

58

59

60

61

62

63

64

65

66 Normal Eye +0.5 D Hyperope

67

68

69

70

71

72

73

74

75

76

77

78 iTrace Summary Robust Aberrometer
Measures Spatially Resolved Refraction and Aberrations for ALL eyes – including highly irregular Measurement Zone from 2.0 to 8.0 mm (Flexible) Multi-Zone Refraction Analysis Can do Over-Refraction with Contact Lenses or Spectacles Measures Psuedophakic eyes Complete Corneal Topography Analysis Separates Corneal from Total Aberrations Resulting in Lenticular (internal ocular) Aberrations Measures Accommodation Accurate Pupil Size Measurement

79 Thank you for your attention.