NCAR/RAF Aerosol and Cloud Particle Instrumentation for DYCOMS-II Investigator meeting 15-Feb-2001 Jeffco.

Similar presentations


Presentation on theme: "NCAR/RAF Aerosol and Cloud Particle Instrumentation for DYCOMS-II Investigator meeting 15-Feb-2001 Jeffco."— Presentation transcript:

1

2 NCAR/RAF Aerosol and Cloud Particle Instrumentation for DYCOMS-II Investigator meeting 15-Feb-2001 Jeffco

3 Outline PMS probes - size ranges & data rates Calibration & maintenance plans 1 - before field deployment 2 - at field site 3 - after returning to Jeffco Analysis software ncplot, ncpp & xpms2d –available to investigators Solicit questions and suggestions

4 DYCOMS-II Particle Probe Locations aerosol sampling with instruments inside aircraft CVICCNCNRDMA + other

5 Probe Calibrations Probes will be calibrated for size using standard reference particles refractive index of PSL 1.59, glass beads 1.56, water 1.33

6 PSL’s & Glass Spheres

7 Particle Probe Maintenance Plan Jeffco - before and after field project –Optical cleaning & alignment –Electronic baselines –Electronic response times –Calibration particles - full size range (5+ sizes) On-site checks as needed –Point checks between flights easy on King Airnot so easy on C-130, especially at field sites

8 Calibration example (PCASP 220nm PSL) 1 sec per box Calibrations  project archive

9 RAF Analysis Software (Chris Webster) ncplot – time plots, xy, xyz, spectral ncpp – PMS probe size distributions xpms2d – Two-D probe image display available to investigators  ask Chris for a demo

10 Questions and Suggestions ? Probe performance checks –special concerns for pre- or post- flight ? Calibrate with particular kinds of particles –Salt, ammonium sulfate,.. ?


Download ppt "NCAR/RAF Aerosol and Cloud Particle Instrumentation for DYCOMS-II Investigator meeting 15-Feb-2001 Jeffco."