Professor Ioannis Pallikaris MD, PhD.  Different mediums (Nd:YAG Crystals vs ArF Gas)  Less optics for the shaping of the pulse  No use of toxic gasses.

Similar presentations


Presentation on theme: "Professor Ioannis Pallikaris MD, PhD.  Different mediums (Nd:YAG Crystals vs ArF Gas)  Less optics for the shaping of the pulse  No use of toxic gasses."— Presentation transcript:

1 Professor Ioannis Pallikaris MD, PhD

2  Different mediums (Nd:YAG Crystals vs ArF Gas)  Less optics for the shaping of the pulse  No use of toxic gasses  Improved beam quality and shape  Reduced thermal effect and collateral damage  Hydration dependence on 193nm

3  Corrosive, toxic and high cost gases  Big spot size  Absorption coefficient suggest 193nm affected more by BSS or NaCl 0.9%

4 Mirror 90% Medium Pumping ArF 193 nm@400hz Mirror 100% Medium Nd:YAG 1064 nm 213 nm@300Hz Non-linear optics

5  All treatments performed by CustomVis Pulzar Z1 & Allegretto 400  Last follow-up to 12 months  PRK Mean SEQ = - 4.50 ± 1.81D  [-6.50D to -1.50D Sph and up to –2.50D Cyl]  160 eyes in Excimer and 160 in Solid  MMC solution was applied to PRK cases in both systems according to given attempted correction LASIK and PRK for treating Myopia and Myopic Astigmatism. A retrospective study.

6  89% between + 1.00/-1.00 D at 12 months postop for Excimer  96% between +1.00/-1.00 D at 12 months postop for Solid

7  35% of eyes gained more than 1 line in Solid  31% of eyes gained more than 1 line in Eximer 157 eyes

8  0% of eyes changed more than 0.50D between 3m and last postop Solid

9 Smooth ablation surface No edema or distortion of the adjacent corneal stroma Normal structure of the endothelial layer in both groups S E S EE S

10  The Descemet’s membrane and the endothelium did not demonstrate any morphological difference with intact tissues S E S E Solid Excimer S E

11  No statistical significant difference found between two groups in any postoperative interval ( p >0.05). Study group, 30 eyes underwent myopic PRK with the CustomVis Pulzar Z1 laser system. Control group, 30 eyes underwent myopic PRK with the a Allegretto Wavelight 400Hz Excimer system

12  Coma Y Decrease 0.04  Coma X Increase 0.05  Spherical aberration Increase 0.04  HO rms Increase 0.08

13 Alcon LADARVision4000 System Wavefront-guided Excimer Laser Ablation Using PRK Mean attSphEq : -4.21D Diopters 6months postop :Pupil diameter 6mm RMS HO: decrease 0.29 Spherical Aberration: decrease 0.06 Third order RMS: increase 0.13 Daniel S. Durrie,, et al JRS 2008 WaveLight ALLEGRETTO Laser Wavefront Optimized Treatments in Surface Ablation Mean attSphEq : -3.64 ±1.92 Diopters 3months postop:Pupil diameter 6mm RMS HO: increase 0.03 Spherical Aberration: increase 0.01 Third order RMS: increase 0.03 Aleksandar Stojanovic, et al JRS 2008

14  Predictability presented similar results between the two systems  The excimer seems to give more overcorrection effect due to the use of MMC compared to the Solid system  Both system are safe for correction with PRK and LASIK  Stability of refraction is the same for both systems with stability level being equal at 12 moths PostOp period

15  Histology has proved to have similar results to the samples of Excimer laser  There are no statistical significant changes in corneal endothelium monolayer in post PRK patients (up to one year) with solid state (213 nm) and excimer (193 nm) laser systems.  Similar changes to high order aberration were found between Solid and Excimer