Biomechanical essentials and studies First 12 years processes The Femoral neck screw - prosthesis type Spiron Mülheim a.d. Ruhr, 12.04.2013 B. Birkenhauer.

Similar presentations


Presentation on theme: "Biomechanical essentials and studies First 12 years processes The Femoral neck screw - prosthesis type Spiron Mülheim a.d. Ruhr, 12.04.2013 B. Birkenhauer."— Presentation transcript:

1 Biomechanical essentials and studies First 12 years processes The Femoral neck screw - prosthesis type Spiron Mülheim a.d. Ruhr, 12.04.2013 B. Birkenhauer Stiftungsklinikum Mittelrhein, Location Boppard

2 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 Content The development of Spiron-Prosthesis Biomechanical backgrounds Clinical results and studies -12-years processes - Figuring of incorrect implatations - Prvention of OP-fails

3 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 Self cutting conical (proximal cylindrical) Screw Rotationsproof design No modular concept No Left- / Right-variants Bonit-coated, rough blasetd titanium alloy for cementless fixation Prosthesis design

4 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 „Father“ „Mother“ DSP nach Huggler u. Jacob, Bild: Sulzer Orthopedics Problem of the DSP: often „long suffering period“ of the pateints, lug pain (???), Suspicion: too low Osseointegrationsurface „Child“ Conceptional development

5 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 Proximal-cortical and distal and distal spongiose anchorage by conical screw body Verified good load distribution even with strength components vertical to the implant axis ill.: Allopro. Allospine posterior, technichal information biomechanical investigations. Sulzer Medica. 1999: 3 Pedicalscrew priciple

6 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 On a conical base body the vertical effective load will transfered consistentto the implant => Prevention of punctual bear maximation ill.: Allopro. Allospine posterior, technichal information biomechanical investigations. Sulzer Medica. 1999: 2 Load distribution at the conical screw

7 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 Thread causes an up to 30% Surface enlargement (Osseointegrationssurfac e) as compared to a conventional cementless stem prosthesis Thread = „bone inductor“ Osteological advantages I

8 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 1. Surface enlargement (Osseointegrationsurface) by Threaded sructure 2. Bone induction on protuding prosthesis- structures using the example of revisionsstem SL on Wagner Histolog. Pictures used from: Schenk, Wehrli: Zur Reaktion des Knochens auf eine zementfreie SL-Femur-Revisionsprothese, Orthopäde (1989) Osteological advantages II

9 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 1 year postoperative Change of the femoral x-raymorphology by the Spiron-prosthesis 1 year postoperative

10 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 Application of force of femoral neck prostehses Extension trajectories Sek. Pressure trajectories Prosthesis as „capstone“ of the trajectories Pictures: Permedica, JUST-Prothese Pressure trajectories

11 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 First implantation 08.02.2001 Harris-Hip-Score (HHS) präop. 51 (24-76), n = 51 and increased after one year to HHS 1 : 94 (86- 100), n = 33 on 30 Patients B. Birkenhauer, H. Kistmacher, J. Ries, Zementfreie Schenkelhalsschraubenprothese Typ Spiron, Konzeption und erste klinische Ergebnisse, Der Orthopäde, 11/2004, 1259-1266 1 year postop. Clinical results

12 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 Followup of the implantations 2001-2003 Running followup of the patients with minimum 10 years-process n = 48 substract out (censored) loss: 3 2 cases of death, 1 periprothetic Fractures after fall Not censored loss: 4 2 aseptic early loosening (Implantation fail) 2 Early infects (1 Immunsuppr. Patient) Precental survival rate after 10(-12) years: 91,2% Running evaluation by „HipDo-patient interview sheet“ for assessment of the result quality after min. 10 years with good to very well long term evaluation of patients For Comparison: SP2-survival rate after10(-13) years: 93 % (Lubinus, P.: Die SP2-Schaft Endoprothese, Frankfurter Orthopädie Symposium, November 2005) * Copyright: Orthopädische Klinik Dortmund, K. Buckup

13 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 The „disillusionment“ Abstract out of the A) Australian Hip- prothesesregister 2011 B) schwedish Hip- prothesesregister 2010 Statement Both register shows that cementless endoprostheses in younger patient have worse survival rates than in older patients Reason: higher stress! Furthermore it‘s interesting that cemented supplies shows a very good success! Ergo: If there is now revision problem, (older patients) a cemented implantation is a very welel working supply

14 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 X-ray processes particularly cases: case 1 (2001) 2002 2004 2009 Statement: Bone under the collar can resorb several milimeters without consequences, this meets also with the calculations

15 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 X-ray processes particularly cases: case 2 (2001) 200420072013 statement: Femoral neck was damaged. Cerclage. No consequences

16 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 X-ray processes standard cases: case 3 (2001) 200520082012 Statement: Picture proves, no negative bone rebuilding like with longer prostheses This meets also with the calculations

17 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 Example of a early loosending at varus and lateral contact 1 Months 5 Months 7 Months 1 year before Revison Statement: Avoid Lateral Contact If it still appears already intraoperative change to standard stem

18 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 Example for Varus without lateral contact of the prosthesis Varus-Position without lateral cortical contact operated permanent 1 y post op 2 year post op

19 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 Example: Septic Early loosening (2002) Pat. under Immunosuppression with rheumatoid arthritis with left sided early infect

20 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 What are the central statements? Contact the lateral Spiron should be avoided: - too shallow angle with lateral contact is always wrong - but too steep angle with lateral contact works (But not recommended!) Don't choose Spiron too thin; Spiron length is not so critical. There Spiron is much to suggest that it is good if the Spiron still has some distance to the opposite cortex. If loosening were observed, this was in the first 2 years. No loosening at 2 years post-op If the implantation failure "lateral" contact out of the study expects to get a survival rate of 96.8% 10-12 years * The HHS is remarkably good at Spiron patients. Bone condition for a revision is excellent. Calculations for bone remodeling are consistent with X-ray processes. Application of Spiron useful for patients who are still experiencing revisions. In the older age standard hips and also cements supplies are meaningful. * For reasons of simplicity, the prostheses were summarized by 10 to 12 years in a group

21 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 B. Ebbecke, Universiy Hanover Institute for constructional mechanic and numerical mechanic Numerical studies of Spiron – Prosthesis Results of bone remodelling by Spiron-prosthesis: Density distribution in biological balance, 2003 Ebbecke, B. Knochenwachstum - Endlich verständlich. Zusammenfassung, University Hanover, Institut für Baumechanik und Numerische Mechanik, 2003. Nackenhorst, U. Ein effizientes Finite Element Verfahren zur Simulation des beanspruchungsstimulierten Knochenwachstums. Tagungsband des Workshops ”Die Methode der Finiten Elemente in der Biomedizin, Biomechanik und angrenzenden Gebieten”, Ulm, 1997. Nackenhorst, U. und Schröder, U. Numerische Simulation des beanspruchungsinduzierten Knochenumbaus am Beispiel endoprothetisch versorgter Femura. Bericht aus dem Institut für Mechanik, Hamburg, 1997. Ziefle, M. Numerische Untersuchungen der Druckscheiben Prothese. Zusammenfassung, Universität Hannover, Institut für Baumechanik und Numerische Mechanik, 2002 Further studies

22 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 Study for comparative primary stability 2005 In collaboration with the animal medical school and the MHH Hannover Medical School (Department of Trauma Surgery), a Investigation of the femurs of shepherds carried out: 3 groups (same age, same size, similar bone quality): 15 femurs without prosthesis 15 femurs with Spiron 15 femurs with cementless standard stem type "Zweymüller” Article title: Less motion in bone/prosthesis interface by the new uncemented short-stem prosthesis Spiron(R) in dog bone MS ID : 1648004516228708 Authors : Ulrich Wiebking, Björn Birkenhauer, Franziska Hühn, Christian Krettek and Thomas Gösling Journal : Acta Veterinaria Scandinavica 2007

23 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 The investigation In laboratory The test machine

24 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 Results I Terminal of the investigation: The fracture SpironZweymüller

25 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 Result Strength – Time-Diagram

26 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 Result blue = Spiron; red = Zweymüller; white = without prosthesis

27 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 Result: Maximal stress at femur fracture Femora with Spiron Femora with Zwey- müller Natural Femur

28 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 Elasticity of the system „Prosthesis in bone“ (Force in Newton vs. distance in millimeter)

29 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 The „Helica“-Prosthesis in the veterinary medicine Due to the good biomechanical results of that investigation was the Spiron prosthesis under a new name directly Introduced in the veterinary medicine. Hach, V., Delfs,G.: Initial experience with newly developed hip endoproethesis; Vet Comp Orthop Traunatol 2009; 22: 153-158

30 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 Result A major factor of the primary stability is the preservation of the elasticity of the bone. But a stem prosthesis increases elasticity and stiffens the proximal femur. A femoral neck prosthesis can maintain the elasticity of the proximal femur.

31 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 Indications Good bone quality Physiological hip joint geometry Biological age <65 BMI? (No previous surgery) good compliance Beware of arthritis with periarticular osteoporosis Approx. 15 - max. 20% of hip replacement patients

32 B. BirkenhauerCoST OP-Kurs Hüftgelenk12.04.2013 1.Are Femoral neck prostheses short stem prostheses? 2.Are shortstem- und femoral neck prostheses biomechanical equivalent? Outstanding questions

33 Thank you for your attention!