Presentation is loading. Please wait.

Presentation is loading. Please wait.

Detectors of HPLC.

Similar presentations


Presentation on theme: "Detectors of HPLC."— Presentation transcript:

1 Detectors of HPLC

2 HPLC Bulk Property Detectors
Refractive Index Detector

3 HPLC Bulk Property Detectors
Refractive Index Detector Plus: 1) Measures a bulk property 2) Nearly Universal (different RI than mobile phase) 3) Comparable response for different analytes 4) Detects species with no chromophores Minus: 1) Temperature dependent 2) Poor sensitivity (LOD ≈ 100 ng) 3) No gradient elution

4 HPLC Bulk Property Detectors
Evaporative Light Scattering Detector Solid state laser

5 HPLC Bulk Property Detectors
Evaporative Light Scattering Detector Plus: 1) Measures a bulk property 2) Nearly Universal (must be non-volatile) 3) Does not detect liquids (gradients are ok) 4) Detects species with no chromophores Minus: 1) Signal not linear with concentration 2) Fair sensitivity (LOD ≈ 1 ng) 3) No salts or buffers in mobile phase

6 HPLC Specific Property Detectors
Conventional UV-Vis Absorption Detector Photodiode 1 Photodiode 2 Hg Pen Lamp 254 nm

7 HPLC Specific Property Detectors
Multi-wavelength UV-Vis Absorption Detector Deuterium Lamp Photodiode Array

8 HPLC Specific Property Detectors
Multi-wavelength UV-Vis Absorption Detector

9 UV-Vis detection of sugars in juice

10 HPLC Specific Property Detectors
Multi-wavelength UV-Vis Absorption Detector Plus: 1) Measures a specific property 2) Nearly Universal (must have chromophore) 3) Most common of all detectors (~75%) 4) Potential to provide qualitative info. 5) Simple, robust Minus: 1) Fair sensitivity (LOD ≈ 1 ng) 2) Expensive with PDA 3) Misses some important analytes

11 HPLC Specific Property Detectors
Fluorescence Detector

12 HPLC Specific Property Detectors
Fluorescence Detector

13

14 HPLC Specific Property Detectors
Fluorescence Detector Plus: 1) Measures a specific property 2) Highly Selective (must fluoresce) 3) Second most common of all detectors (~15%) 4) High sensitivity (LOD ≈ 0.01 ng) 5) Can interrogate very small volumes Minus: 1) Not Universal 2) Limited Applications

15 HPLC Specific Property Detectors
Electrochemical Detector

16 HPLC Specific Property Detectors
Electrochemical Detector Plus: 1) Measures a specific property 2) Highly Selective (depends on reduction potential) 3) High sensitivity (LOD ≈ 0.01 ng) Minus: 1) Not Universal 2) Must have electrolyte in mobile phase 3) Mobile phase must be aqueous 4) Gradients not possible

17 What is Mass Spectrometry?
IUPAC Definition: The branch of science dealing with all aspects of mass spectrometers and the results obtained with these instruments. Applications: identification Quantification Molecular structure higher-order structure (H/D exchange, cross-link) tissue imaging

18 Mass Definitions Molecular masses are measured in Daltons (Da) or mass units (u). One Dalton = 1/12 of the mass of a 12C atom. Monoisotopic mass = sum of the exact masses of the most abundant isotope of each element present, i.e., 1H= , 12C= , 16O= , etc. This is the most accurately defined molecular mass and is preferred if a measurement of it can be determined. Average mass = sum of the abundant averaged masses (“atomic weights”) of the constituent atoms of a given molecule. The result is a weighted average over all of the naturally occurring isotopes present in the compound. This is the common chemical molecular weight that is used for stoichiometric calculations (H=1.0080, C=12.011, O=15.994, etc.). The average mass cannot be determined as accurately as the monoisotopic mass because of variations in natural isotopic abundances. The mass to charge ratio (m/z). A quantity formed by dividing the mass (in u) of an ion by its charge number; unit: Thomson or Th. .

19 Ionization Techniques and Mass Analyzers
Mass Spectrometry Ionization Techniques and Mass Analyzers

20 Mass Spectrometer Schematic
Vacuum envelope Sample in Inlet System Ion Source Mass Analyzer Detector GC HPLC Direct injection Inlet systems: Data System CI EI ESI APCI MALDI Ion sources: Mass spectrum out Mass analyzers: Time-of-flight (TOF) Quadrupole Ion trap FT-ICR Orbitrap

21 Ionization Techniques
Gas-Phase Methods Electron Impact (EI) Chemical Ionization (CI) Desorption Methods Matrix-Assisted Laser Desorption Ionization (MALDI) Fast Atom Bombardment (FAB) Spray Methods Electrospray (ESI) Atmospheric Pressure Chemical Ionization (APCI)

22 Electron Impact

23 Advantages Well-Established Fragmentation Libraries No Supression Insoluble Samples Interface to GC Non-Polar Samples Disadvantages Parent Identification Need Volatile Sample Need Thermal Stability No Interface to LC Low Mass Compounds (<1000 amu) Solids Probe Requires Skilled Operator

24 Chemical Ionization

25 Chemical Ionization Advantages Parent Ion Interface to GC
Insoluble Samples Disadvantages No Fragment Library Need Volatile Sample Need Thermal Stability Quantitation Difficult Low Mass Compounds (<1000 amu) Solids Probe Requires Skilled Operator

26 FAB

27 Advantages Parent Ion High Mass Compounds (10,000 amu) Thermally Labile Compounds (R.T.) Disadvantages No Fragment Library Solubility in Matrix (MNBA, Glycerol) Quantitation Difficult Needs Highly Skilled Operator Relatively Low Sensitivity

28 MALDI

29 Matrix-Assisted Laser Desorption/Ionization (MALDI)
Analyte is dissolved in solution with excess matrix (>104). Sample/matrix mixture is dried on a target and placed in the MS vacuum. Requirements for a satisfactory matrix: It must co-crystallize with typical analyte molecules It must absorb radiation at the wavelength of the laser (usually 337 nm) To transfer protons to the analyte it should be acidic Typical successful matrices for UV MALDI are aromatic carboxylic acids. Sinapinic acid a-cyano-4-hydroxycinnamic acid (CHCA) 2,5-dihydroxybenzoic acid (DHB) HO COOH OH CH3O CH=CH-COOH CH=C-COOH CN

30 Matrix-assisted laser desorption ionization (MALDI)
Pulsed laser (337 nm) 3.5 ns Sample stage Desorbed sample ions and neutrals Mass analyzer Sample and matrix, crystallized on stage ±20 kV

31 MALDI Ionization Mechanism
1. Laser pulse produces matrix neutrals, + and - ions, and sample neutrals: M --> M*, MH+, (M-H)- (M= Matrix) - + 2. Sample molecules are ionized by gas-phase proton transfer: MH+ + A --> AH+ + M (A=Analyte) (M-H)- + A --> (A-H)- + M

32 MALDI Mass Spectrum of Protein Tryptic Digest
m/z

33 MALDI Advantages Parent Ion High Mass Compounds (>100,000 amu)
Thermally Labile Compounds (R.T.) Easy to Operate Disadvantages No Fragment Library Wide variety of matrices Quantitation Difficult

34 Electrospray Ionization
0 kV Sample in solution flows into capillary tube Tube held at kV Nitrogen flowing in outer tube aids nebulization

35 Electrospray ionization

36 Electrospray Ion Formation
Needle at High Voltage MH+ + + + + + + + + + + + + + + + + + + + + + + + + + + [M+3H]3+ + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + [M+2H]2+ Droplets formed in electric field have excess positive ions. Evaporation of neutrals concentrates charge. Droplets break into smaller droplets. Eventually one molecule + n protons is left.

37 Nanospray Online analysis ~ 20 µm tip ID Interface with nanoLC
Flow rate: ~300nL/min Offline analysis (static infusion) ~ 2 µm tip ID Flow rate: ~40nL/min Requires pure sample free from salt New Objective, Inc.

38 ESI

39 ESI Advantages Parent Ion High Mass Compounds (>100,000 amu)
Thermally Labile Compounds (<0º C) Easy to Operate Interface to HPLC Zeptomole sensitivity with nanospray Disadvantages No Fragmentation Need Polar Sample Need Solubility in Polar Solvent (MeOH, ACN, H2O, Acetone are best) Sensitive to Salts Supression

40 Electrospray MALDI Ionization Methods Online LC/MS possible
Poor for mixtures without LC Quantitation possible Good for MW <600 Generate highly charged ions MALDI Very long sample lifetime; repeated measurements possible Good for mixtures Matrix peaks can interfere at MW <600 Salt tolerant Low maintenance Generate ions with few charges

41 APCI

42 APCI Advantages Parent Ion Insensitive to Salts Interface to HPLC
Can use Normal Phase Solvents Handles High Flow Rates Disadvantages Need Volatile Sample Need Thermal Stability

43 Mass Analyzers Double Focusing Magnetic Sector Quadrupole Mass Filter
Quadrupole Ion Trap Linear Time-of-Flight (TOF) Reflectron TOF Fourier Transform Ion Cyclotron Resonance (FT-ICR-MS)

44 Double-Focusing Magnetic Sector
Advantages Very High Resolution (60,000) High Accuracy (<5 ppm) 10,000 Mass Range Disadvantages Very Expensive Requires Skilled Operator Difficult to Interface to ESI Low resolution MS/MS without multiple analyzers

45 Quadrupole Mass Analyzer/Filter
+ U + V cos wt -U - V cos wt Uses a combination of RF and DC voltages to operate as a mass filter. Mass analyzer. Mass selection device Ion transport device (RF-only collision cell). Mass scan and stability diagram Randall E. Pedder Extrel Application Note

46 Quadrupole Mass Filter
Advantages Inexpensive Easily Interfaced to Many Ionization Methods Disadvantages Low Resolution (<4000) Low Accuracy (>100ppm) MS/MS requires multiple analyzers Low Mass Range (<4000) Slow Scanning

47 Quadrupole Ion Trap 3D Trap
End caps Ions out to detector Ring electrode (~V) Insulated spacer He gas 1x10-3 Torr Ions in (from ESI) Uses a combination of DC and RF fields to trap ions Ions are sequentially ejected by scanning the RF voltage Linear Trap Essentially a quadrupole with end-caps Advantage: Larger ion storage capacity, leading to better dynamic range Raymond E. March, JOURNAL OF MASS SPECTROMETRY, VOL. 32, 351È369 (1997)

48 Quadrupole Ion Trap Advantages Inexpensive
Easily Interfaced to Many Ionization Methods MS/MS in one analyzer Disadvantages Low Resolution (<4000) Low Accuracy (>100ppm) Space Charging Causes Mass Shifts Low Mass Range (<4000) Slow Scanning

49 Time-of-Flight (TOF) Mass Analyzer
Source Drift region (flight tube) + + detector + + V Ions formed in pulses. Measures time for ions to reach the detector. or

50 Linear Time-of-Flight (TOF)
Advantages Extremely High Mass Range (>1 MDa) Fast Scanning Disadvantages Low Resolution (4000) Low Accuracy (>200ppm) MS/MS not possible

51 Reflectron Time-of-Flight (TOF)

52 Linear and Reflector TOF Analyzers
Reflector compensates for initial variation in kinetic energy, improving resolving power and mass accuracy.

53 Reflectron Time-of-Flight (TOF)
Advantages High Resolution (>20,000 in some models) High Accuracy (<5ppm) 10,000 Mass Range Fast Scanning Disadvantages Low Resolution for MS/MS (PSD)

54 Fourier Transform Ion Cyclotron Resonance (FT-ICR)
Ions trapped and measured in ultrahigh vacuum inside a superconducting magnet. Excite + R C Detect + B0 A.G. Marshall

55 FT-ICR-MS Advantages Extremely High Resolution (>500,000)
Very Good Accuracy (<1 ppm) MS/MS in one analyzer Disadvantages Expensive Requires Superconducting Magnet Slow MS/MS

56 Electron Multiplier Multi-Channel Plate (MCP)
From Detector Technolgy: B. Brehm et al., Meas. Sci. Technol. 6 (1995)

57 Differential Amplifier
Fourier Transform Ion Detection Differential Amplifier FT 100 150 200 250 Frequency (kHz) 7+ 8+ 10+ 11+ 12+ 9+ 600 1000 1400 1800 m/z Calibration 80 240 400 Time (ms) Image Current Bovine Ubiquitin 1072 1071 A.G. Marshall

58 Orbitrap FTICR Confined ion trajectory Image current detection
Fourier transform data conversion TOF Simultaneous excitation Unique to Orbitrap 3D electric field trapping No need for magnet Easy access Final detection device Alexander Makarov, Anal. Chem. 2000, 72,

59 Image Current Detection in Orbitrap
From Alexander Makarov’s 2008 ASMS Award Address

60 Comparison of Analyzer Types
Ion Trap/ Quadrupole TOF OrbiTrap FT-ICR Sensitivity +++ ++* to +++ ++* +* Mass Accuracy +** ++ +++** Resolving Power ++++** Dynamic Range + to +++** ++** Upper m/z + ++++ *Sensitivity lowered due to losing ions on way to analyzer, rather than inherent sensitivity. **Can be improved by scanning narrower mass range or slower.

61 Hybrid/Tandem Instruments
Combine (1) ion selection, (2) ion dissociation, and (3) mass analyzer devices Quadrupoles and ion traps good for selective isolation of precursor ions and for fragmentation (required for MSMS - Topic of Lecture 2) Reflectron TOF, FT-ICR, and OrbiTrap have higher mass accuracy and resolving power (high mass accuracy is good for identification – Lecture 3)

62 Ion Isolation Quadupole Continuous ion beam Quadrupole ion trap
Pulsed-mode operation; space charge issue SWIFT in FTICR Ultrahigh selectivity; only works well in ICR traps TOF Only implemented on TOF/TOF

63 Ion Dissociation Collision Induced Dissociation (CID or Collision Activated Dissociation (CAD) ion traps: off-resonance excitation rf-only multi-poles: higher kinetic energy (HCD) and cascaded CID TOF/TOF: single collision Electron capture dissociation (ECD) and Electron transfer dissociation (ETD) ECD: FTICR, reagent: electron ETD: ion traps, reagent: free radical anion Other important factors to consider: how product ions are collected and detected

64 MALDI-TOF/TOF (4700 Proteomics Analyzer)
250 l/s turbopump Collision gas 2nd source (accelerates fragment ions) Ion gate Multiplier detectors Reflectron Floating collision cell N2 laser Retarding lens Beam 1 lens & x/y deflectors Beam 2 lens, metastable suppressor & x/y deflectors High performance TOF analysis for MS1 and MS2 give high resolving power and good mass accuracy. High accelerating voltage allows high energy CID, giving a wider range of fragment ions and facilitating side-chain cleavages that distinguish isomeric amino acids Ile and Leu. Marvin L. Vestal and Jennifer M. Campbell, Methods Enzymol ;402 :79-108

65 Hybrid Instrument: QqTOF Mass Spectrometer (QSTAR)
5x10-7 Torr Liner Ion mirror Ion detector Pusher Puller Q1 q2 MS1 mass selection CID Ion focus & transfer q0 N2 ESI spray tip Roughing pump Turbo pump High vacuum Collision gas MS2 mass analysis Ion formation Chernushevich, Loboda and Thomson, J. Mass Spectrom. 2001; 36: 849–865

66 Linear Ion Trap – FT-ICR (LTQ-FT)
Linear Ion Trap (LTQ) Ion Cyclotron John E. P. Syka, et al, Journal of Proteome Research 2004, 3,

67 Data Dependent Acquisition
Data Dependent Scans MSMS based on intensity ranking of precursor ions Dynamic Exclusion Precursor m/z of previous MSMS are memorized and no MSMS done on them during a defined time period Automatic Gain Control (AGC, unique to ion trap) Control how many ions are scanned – to achieve signal/noise ratio and to minimize space charge effect

68 Scan Sequence of LTQFT Thermo Application Note: 30046

69 Linear Ion Trap - Orbitrap - ETD
Alexander Makarov, Stevan Horning, et al, Anal. Chem. 2006, 78, G. C. McAlister, et al, J Proteome Res. 2008, 7,

70 Linear quadrupole ion trap (LTQ) video clip
J. Schwartz, et al, JASMS2002v13p659


Download ppt "Detectors of HPLC."

Similar presentations


Ads by Google