Download presentation
1
Radiation Dosimetry of the Patient
Robert L. Metzger, Ph.D.
2
1. Dosimetry Radiation dosimetry is primarily of interest because radiation dose quantities serve as indicators of the risk of biologic damage to the patient The biologic effects of radiation can be classified as either deterministic (non-stochastic) or stochastic Deterministic or non-stochastic effects are believed to be caused by cell killing if a sufficient number of cells in an organ or tissue are killed, its function can be impaired
3
1. Dosimetry Deterministic or non-stochastic effects
effects include terratogenic effects to the embryo or fetus, skin damage and cataracts a threshold can be defined below which the effect will not occur for doses greater than the threshold dose, the severity of the effect increases with the dose to assess the likelihood of a deterministic effect on an organ from an imaging procedure, the dose to that organ is estimated
4
1. Dosimetry A stochastic effect is caused by damage to a cell that produces genetically transformed but reproductively viable descendants cancer and hereditary effects of radiation probability of a stochastic effect, instead of its severity increases with dose No dose thresholds below which the effects cannot occur The NRC’s radiation dose limits described in Chapter 23 are intended to limit the risks of stochastic effects and to prevent the non-stochastic effects
5
1. Dosimetry Entrance Skin Exposure
The radiation exposure incident on a patient is the entrance skin exposure Skin doses are easy to measure but they are poor indicators of patient risk They do not take into account the exposed area, penetrating power of the x-ray beam, or the radiosensitivity of the exposed region At diagnostic energies, the f-factor (roentgen-to-rad) conversion is close to 1.0 so that dose is numerically equal to exposure
6
1. Dosimetry Dose-Area Product (DAP)
Product of patient entrance skin exposure and cross-sectional area of the x-ray beam (exposed area) Units are in mGy-cm2 or mrad-cm2 Used in fluoroscopy
7
1. Dosimetry Radiation Dose
Radiation dose is defined as the absorbed energy per unit mass but this says nothing about the total mass of tissue exposed and the distribution of the absorbed energy Would you prefer to receive a dose of 10 mGy to the whole body or 20 mGy to the finger? The 10 mGy whole body dose represents about 1,000 times the ionizing energy absorbed for a 70-kg person with a 35 g finger
8
1. Dosimetry Imparted energy
the total amount of energy deposited in matter is called the imparted energy (Joules), is the product of the dose (Gray) and the mass (Kg) over which the energy is imparted assume each 1-cm slice of a head CT scan delivers a 30 mGy dose to the tissue in the slice If the scan covers 15 cm, the dose is still the same, however the imparted energy is approx. 15 times that of a single slice (you also have to consider scatter from adjacent slices, about 10-25%)
9
1. Dosimetry The disadvantage of imparted energy is that it does not account for the different sensitivities of the exposed tissue to biologic damage Effective dose is used for comparing risk of stochastic effects E (Sv) = wT x HT has shortcomings, wT were developed from epidemiologic data and incorporate significant uncertainties
10
1. Dosimetry Organ Doses It is possible to estimate organ doses from a given entrance skin exposure (ESE) Organ doses are substantially lower than skin dose For AP projections, the embryo dose will be between 1/3rd and 1/4th the ESE (in the direct beam) For PA projections, the embryo dose will be about 1/6th of the ESE (in the direct beam) For LAT projection, the embryo dose will be about 1/20th of the ESE (in the direct beam)
11
c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2nd ed., p. 59.
12
1. Dosimetry Comparing ESE is useful for assessment of equipment performance and calibration, when a comprehensive analysis of effective dose is unnecessary c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2nd ed., p. 797.
13
1. Risk The International Commission on Radiological Protection (ICRP) estimates the risk of fatal cancer for exposures to adults of working age to be deaths per Sv or per rem this translates to 1 cancer death per 2,500 people receiving an effective dose of 10 mSv (1 rem) Because of the linear, no-threshold assumption used in risk estimates, risk is presumed to be proportional to the effective dose
14
1. Risk Risk is proportional to the effective dose
there would be a 1 in 25,000 chance that a fatal cancer would result from an effective dose of 1 mSv (0.1 rem), or a 1 in 500 chance of a fatal cancer from an effective dose of 50 mSv (5 rem) The ICRP estimates the risk to be two or three times higher for infants and children and substantially lower for adults older than 50 years of age
15
1. Typical Absorbed and Effective doses
c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2nd ed., p. 798.
16
1. Risks Procedure Effective Dose (mSv)
Risk of Fatal Cancer (per million) Equivalent to Number of Cigarettes Smoked Equivalent to Number of Highway Miles Driven Chest Radiograph 0.04 1.6 12 29 Skull Exam 0.1 4.0 71 Mammography Thoracic Spine 1.0 40.0 292 714 Pelvis 1.1 44.0 321 786 Abdomen 1.2 48.0 350 857 CT Head 1.8 72.0 526 1286 Lumbar Spine 2.1 84.0 613 1500 Intravenous Urography 4.2 168.0 1226 3000 CT Pelvis 7.1 284.0 2073 5071 CT Abdomen 7.6 304.0 2219 5429 CT Chest 7.8 312.0 2277 5571 Barium Enema (with fluoro) 8.7 348.0 2540 6214
17
1. Interventional Radiologic Procedures
c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2nd ed., p. 799.
18
2. Radiographic Procedures
Geometry for measuring the output free-in-air of a radiographic system c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2nd ed., p. 801.
19
2. Radiographic Procedures
c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2nd ed., p. 802.
20
2. Radiographic Procedures
Geometry for measuring the output free-in-air of a radiographic system when phototiming is used c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2nd ed., p. 804.
21
3. Effective Dose Comparison with Chest PA Exam
Procedures Eff. Dose (mSv) Equivalent no. of chest x-rays Approx. period of background radiation Chest PA 0.02 1 3 days Pelvis 0.7 35 4 months Abdomen 1.0 50 6 months CT Chest 8 400 3.6 years CT Abdomen or Pelvis 10-20 500 4.5 years
22
Question Assuming the skin entrance dose from a single slice CT study is 5 rad, the dose for a 10 slice examination would be approximately _____ rad and the imparted energy would be ____ rad (ignore scatter). A. 5, 15 B. 15, 5 D. 50, 5 E. 5, 50
23
Question The skin entrance exposure from a CT slice is 2.0 R. Ten contiguous slices are taken, then dye is injected and 10 slices are repeated. The total entrance skin exposure is about _____ R. A. 2.0 B. 2.2 D. 5.0 E. 20.0 You have to consider scatter. 25% of 2 R = 0.5. So 2.5 per scan is the rad exp. For two scans, 2.5*2 = 5.0
24
Question The national average ESE for a normal 23 cm thick A/P abdomen film with a 400 speed screen-film system is: A. 13 mR B. 150 mR C. 300 mR D. 850 mR E mR
25
Question Match the exposure or dose with the appropriate item:
A. 15 mR B. 40 mR C. 5 R D. 10 R E. 50 mrem 1. CT head scan ESE 2. Lateral chest ESE 3. 10 min fluoro (thin patient) 4. Monthly limit for a pregnant technologist
26
Question Match the exposure or dose with the appropriate item:
A. 15 mR B. 40 mR C. 5 R D. 10 R E. 50 mrem 1. CT head scan ESE – 4 to 6 R typical 2. Lateral chest ESE – mR for PA. 2 to 3 times for Lateral 3. 10 min fluoro (thin patient) – 1-2 R/min for thin patient 4. Monthly limit for a pregnant technologist – 0.5 mSv or 50 mrem
Similar presentations
© 2024 SlidePlayer.com Inc.
All rights reserved.