INSTRUMENTAL ANALYSIS CHEM 4811

Similar presentations


Presentation on theme: "INSTRUMENTAL ANALYSIS CHEM 4811"— Presentation transcript:

1 INSTRUMENTAL ANALYSIS CHEM 4811
CHAPTER 15 DR. AUGUSTINE OFORI AGYEMAN Assistant professor of chemistry Department of natural sciences Clayton state university

2 ELECTROANALYTICAL CHEMISTRY
CHAPTER 15 ELECTROANALYTICAL CHEMISTRY

3 ELECTROCHEMISTRY - The study of the relations between chemical reactions and electricity - The study of the interconversion of chemical energy and electrical energy - The study of redox reactions - Electrochemical processes involve the transfer of electrons from one substance to another

4 Electroactive Species
ELECTROCHEMISTRY Electroactive Species - Species that undergoes an oxidation or a reduction during reaction - Species may be complexed, solvated, molecule, or ion - Species may be in aqueous or nonaqueous solution

5 ELECTROANALYTICAL CHEMISTRY
- The use of electrochemical techniques to characterize a sample - Deals with the relationship between electricity and chemistry - Analytical calculations are based on the measurement of electrical quantities (current, potential, charge, or resistance) and their relationship to chemical parameters

6 ELECTROANALYTICAL CHEMISTRY
Advantages - Measurements are easy to automate as they are electrical signals - Low concentrations of analytes are determined without difficulty - Far less expensive equipment than spectroscopy instruments

7 FUNDAMENTAL CONCENPTS
Redox Reaction - Oxidation-reduction reaction - Reactions in which electrons are transferred from one substance to another Oxidation - Loss of electrons Reduction - Gain of electrons

8 FUNDAMENTAL CONCENPTS
Oxidized Species - The species that loses electrons - The reducing agent (reductant) - Causes reduction Fe(s) ↔ Fe2+(aq) + 2e-

9 FUNDAMENTAL CONCENPTS
Reduced Species - The species that gains electrons - Oxidizing agent (oxidant) - Causes oxidation Cu2+(aq) + 2e- ↔ Cu(s)

10 FUNDAMENTAL CONCENPTS
Half Reactions - Just the oxidation or the reduction is given - The transferred electrons are shown Oxidation Half-Reaction - Electrons are on the product side of the equation Reduction Half-Reaction - Electrons are on the reactant side of the equation

11 FUNDAMENTAL CONCENPTS
Half Reactions Oxidation half reaction Fe(s) ↔ Fe2+(aq) + 2e- Fe2+ ↔ Fe e- Reduction half reaction Cu2+(aq) + 2e- ↔ Cu(s) Cl2(g) + 2e- ↔ 2Cl-

12 FUNDAMENTAL CONCENPTS
- Many redox reactions are reversible - Reduction reaction becomes oxidation reaction when it is reversed and vice versa - Sum of oxidation and reduction half-reactions gives the net redox reaction or the overall reaction - No electrons appear in the overall reaction

13 FUNDAMENTAL CONCENPTS
The Overall Reaction - Both an oxidation and a reduction must occur in a redox reaction - The oxidizing agent accepts electrons from the reducing agent Cu2+(aq) + Fe(s) ↔ Cu(s) + Fe2+(aq) - Reducing agent - Oxidized species - Electron loss - Oxidizing agent - Reduced species - Electron gain

14 FUNDAMENTAL CONCENPTS
Charge (q) Charge (q) of an electron = x C Charge (q) of a proton = x C C = coulombs Charge of one mole of electrons = (1.602 x C)(6.022 x 1023/mol) = 96,485 C/mol = Faraday constant (F) - The charge (q) transferred in a redox reaction is given by q = n x F

15 FUNDAMENTAL CONCENPTS
Current (i) - The quantity of charge flowing past a point in an electric circuit per second i = q/time Units Ampere (A) = coulomb per second (C/s) 1A = 1C/s

16 FUNDAMENTAL CONCENPTS
Voltage or Potential Difference (E) - The amount of energy required to move charged electrons between two points - Work done by or on electrons when they move from one point to another w = E x q or E = w/q Units: volts (V or J/C) 1V = 1J/C

17 FUNDAMENTAL CONCENPTS
Ohm’s Law i = E/R R = resistance Units Ω (ohm) or V/A

18 FUNDAMENTAL CONCENPTS
Electrode - Conducts electrons into or out of a redox reaction system - The electrode surface serves as a junction between an ionic conductor and an electronic conductor Examples platinum wire carbon (glassy or graphite) Gold Silver

19 FUNDAMENTAL CONCENPTS
Electroactive Species Donate or accept electrons at an electrode - Can be made to oxidize or reduce Electrochemical Measurements - Occur at the electrode – solution interface Chemical Measurements - Involve homogeneous bulk solutions

20 ELECTROCHEMICAL CELL - Made up of the electrodes and the contacting sample solution - Electrical conductor is immersed in a solution of its own ions - A potential difference (voltage) is created between the conductor and the solution - The system is a half-cell - The metal conductor is an electrode and the solution is an electrolyte

21 ELECTROCHEMICAL CELL Electrode Potential
- A measure of the ability of the half-cell to do work (the driving cell for the half-cell reaction) Anode - Electrode where oxidation occurs Mo → Mn+ + ne- - Metal loses electrons and dissolves (enters solution) Cd(s) → Cd e- Ag(s) → Ag+ + e-

22 ELECTROCHEMICAL CELL Cathode - Electrode where reduction occurs
Mn+ + ne- → Mo - Positively charged metal ion gains electrons - Neutral atoms are deposited on the electrode - The process is called electrodeposition Cd e- → Cd(s) Ag+ + e- → Ag(s)

23 ELECTROLYSIS - Voltage is applied to drive a redox reaction that
would not otherwise occur Examples - Production of aluminum metal from Al3+ - Production of Cl2 from Cl-

24 ELECTROLYTIC CELL - Nonspontaneous reaction
- Requires electrical energy to occur - Consumes electricity from an external source

25 GALVANIC CELL - Spontaneous reaction - Produces electrical energy
- Can be reversed electrolytically for reversible cells Example Rechargeable batteries Conditions for Non-reversibility - If one or more of the species decomposes - If a gas is produced and escapes

26 GALVANIC CELL - Also known as voltaic cell
- A spontaneous redox reaction generates electricity - One reagent is oxidized and the other is reduced - The two reagents must be separated (cannot be in contact) - Electrons flow through a wire (external circuit)

27 Oxidation Half-Reaction Reduction Half-Reaction
GALVANIC CELL Oxidation Half-Reaction - Loss of electrons - Occurs at anode (negative electrode) - The left half-cell by convention Reduction Half-Reaction - Gain of electrons - Occurs at cathode (positive electrode) - The right half-cell by convention

28 GALVANIC CELL Salt Bridge
- Connects the two half-cells (anode and cathode) - Filled with gel containing saturated aqueous salt solution (KCl) - Ions migrate through to maintain electroneutrality (charge balance) - Prevents charge buildup that may cease the reaction process

29 GALVANIC CELL - For the overall reaction
Cu2+(aq) + Zn(s) → Cu(s) + Zn2+(aq) e- Voltmeter e- - + Cu electrode Zn electrode Cl- K+ Zn2+ Salt bridge (KCl) Cu2+ Anode Oxidation Zn(s) → Zn2+(aq) + 2e- Cathode Reduction Cu2+(aq) + 2e- → Cu(s)

30 GALVANIC CELL Line Notation
Phase boundary: represented by one vertical line Salt bridge: represented by two vertical lines Fe(s) FeCl2(aq) CuSO4(aq) Cu(s)

31 Standard Reduction Potential (Eo)
STANDARD POTENTIALS Standard Reduction Potential (Eo) - Used to predict the voltage when different cells are connected - Potential of a cell as cathode compared to standard hydrogen electrode - Species are solids or liquids - Activities = 1 - We will use concentrations for simplicity Concentrations = 1 M Pressures = 1 bar

32 Standard Hydrogen Electrode (SHE)
STANDARD POTENTIALS Standard Hydrogen Electrode (SHE) - Reference electrode half-cell - Used to measure Eo for half-reactions (half-cells) - Connected to negative terminal (anode) Assigned Eo = under standard state conditions (T = 25 oC, concentration = 1M, pressure = 1 bar, pure solid or liquid)

33 STANDARD POTENTIALS Standard Hydrogen Electrode (SHE) Consists of
- Platinized Pt electrode immersed in a solution of 1M HCl - H2 gas (1 bar) is bubbled over the Pt electrode 2H+(aq, 1 M) + 2e- ↔ 2H2 (g, 1 bar)

34 CELL POTENTIALS - The potential for a cell containing a specified concentration of reagent other than 1 M Standard Cell Potential Eocell = Eocathode – Eoanode Cell Potential Ecell = Ecathode – Eanode - Ecell is positive for spontaneous reactions - Half-reaction is more favorable for more positive Eo

35 CELL POTENTIALS Junction Potential
- Is produced when there is a difference in concentration or types of ions of the two half-cells - Is created at the junction of the salt bridge and the solution - Is a source of error - Minimized in KCl salt bridge due to similar mobilities of K+ and Cl-

36 CELL POTENTIALS Eo (V) Half Reaction 2.890 F2 + 2e- ↔ 2F- 1.507
1.280 1.229 0.799 0.339 0.000 -0.402 -0.440 -0.763 -1.659 -2.936 -3.040 Half Reaction F e- ↔ 2F- MnO e- ↔ Mn2+ Ce e- ↔ Ce3+ (in HCl) O H e- ↔ 2H2O Ag+ + e- ↔ Ag(s) Cu e- ↔ Cu(s) 2H+ + 2e- ↔ H2(g) Cd e- ↔ Cd(s) Fe e- ↔ Fe(s) Zn e- ↔ Zn(s) Al e- ↔ Al(s) K+ + e- ↔ K(s) Li+ + e- ↔ Li(s) Oxidizing agents Reducing agents Increasing oxidizing power Increasing reducing power

37 CELL POTENTIALS - Elements that are more powerful reducing agents than hydrogen show negative potentials - Elements that are less powerful reducing agents than hydrogen show positive potentials - Metals with more negative Eo are more active - More active metals displace less active metals from solution Fe will displace Cu2+ out of solution Zn dissolves in HCl but Cu does not

38 NERNST EQUATION Gives relationship between the potential of an electrochemical cell and the concentration of reactants and products O + ne- ↔ R E = electrode potential Eo = standard potential for the redox reaction R = gas constant = J/K-mol T = absolute temperature in Kelvin F = Faraday’s constant = 96,485 C/mol n = number of electrons transferred

39 The half-cell potential (at 25 oC), E, is given by
NERNST EQUATION For the half reaction aA + ne- ↔ bB The half-cell potential (at 25 oC), E, is given by

40 NERNST EQUATION For the overall reaction aA + bB ↔ cC + dD
The potential at 25 oC is given by

41 NERNST EQUATION - E = Eo when [O] = [R] = 1M
- Concentration for gases are expressed as pressures in bars or atm - Concentrations for pure solids, liquids, and solvents are omitted (activity = 1) - Reduction is more favorable on the negative side of Eo - When a half reaction is multiplied by a factor Eo remains the same

42 REFERENCE ELECTRODES - An ideal reference electrode
- Has a fixed potential over time and temperature - Long term stability - Ability to return to the initial potential after exposure to small currents (reversible) - Obey the Nernst equation

43 Standard Hydrogen Electrode (SHE) Saturated Calomel Electrode (SCE)
REFERENCE ELECTRODES Standard Hydrogen Electrode (SHE) E = V Saturated Calomel Electrode (SCE) - Composed of metallic mercury in contact with saturated solution of mercurous chloride (calomel, Hg2Cl2) - Pt wire is in contact with the metallic mercury - Calomel is in contact with saturated KCl solution E = V at 25 oC

44 Silver/Silver Chloride Reference Electrode (Ag/AgCl)
REFERENCE ELECTRODES Silver/Silver Chloride Reference Electrode (Ag/AgCl) - Consists of silver metal coated with silver chloride paste - Immersed in saturated KCl and AgCl solution E = V at 25 oC

45 ELECTROANALYTICAL METHODS
Two main types - Potentiometric and Potentiostatic - The type of technique reflects the type of electrical signal used for quantitation - Techniques require at least two electrodes and an electrolyte (containing solution) Electrodes Working (indicator) electrode, reference electrode, counter electrode

46 ELECTROANALYTICAL METHODS Potentiometric Technique
- Based on a static (zero-current) situations - Based on measurement of the potential established across a membrane - Used for direct monitoring of ionic species (Ca2+, Cl-, K+, H+)

47 ELECTROANALYTICAL METHODS Potentiostatic Technique
- Controlled-potential technique - Based on dynamic (non-zero-current) situations - Deals with the study of charge transfer processes at the electrode-solution interface - Chemical species are forced to gain or lose electrons

48 ELECTROANALYTICAL METHODS
- Potentiometry - Coulometry - Voltammetry - Polarography - Methods are classified according to the variable being measured - One variable (current, voltage, charge) is measured and the others are controlled

49 POTENTIOMETRY - Based on static (zero-current) measurements
- Involves measurement of potential (voltage) of an electrochemical cell - Used to obtain information on the composition of an analyte - Potential between two electrodes is measured (indicator electrode and reference electrode) - Indicator (sensing) electrode responds to the concentration of the analyte species

50 POTENTIOMETRY - The analyte concentration is related to the potential difference between the indicator electrode and the reference electrode (by applying the Nernst equation) - Indicator electrode is connected to a reference electrode (SCE, Ag/AgCl) to form a complete cell - Implies Etotal = Eindicator – Ereference - Reference electrode is connected to the negative terminal of the readout device (potentiometer)

51 POTENTIOMETRY Applications - Environmental monitoring
- Clinical diagnostics (blood testing, electrolytes in blood) - Control of reaction processes

52 INDICATOR ELECTRODE - Electrode that responds to change in analyte activity - Generally show high degree of selectivity Types of indicator electrodes - Metallic electrodes (metal wire, mesh, or strip) - Metal coated with its sparingly soluble salt (Ag/AgCl) - Electrode whose equilibrium reaction responds to nalyte cation - Redox indicator electrode (measures redox reactions)

53 ION-SELECTIVE ELECTRODES (ISE)
- Are indicator electrodes - Respond directly to the analyte - Used for direct potentiometric measurements - Selectively binds and measures the activity of one ion (no redox chemistry) Examples pH electrode Calcium (Ca2+) electrode Chloride (Cl-) electrode

54 ION-SELECTIVE ELECTRODES (ISE)
Advanteages - Exhibit wide response - Exhibit wide linear range - Low cost - Color or turbidity of analyte does not affect results - Come in different shapes and sizes

55 ION-SELECTIVE ELECTRODES (ISE)
- Made from a permselective ion-conducting membrane (ion-exchange material that allows ions of one electrical sign to pass through) - Reference electrode is inbuilt - Internal solution (solution inside electrode) contains ion of interest with constant activity - Ion of interest is also mixed with membrane - Membrane is nonporous and water insoluble

56 ION-SELECTIVE ELECTRODES (ISE)
- Responds preferentially to one species in solution Internal reference electrode Internal (filling) solution Ion-selective membrane

57 ION-SELECTIVE ELECTRODES (ISE)
- If C+ is the preferential ion - [C+] inside the electrode ≠ [C+] outside the electrode - Results in a potential difference across the membrane Generally (at 25 oC) - 10-fold change in activity implies 59/zi mV change in E - zi is the charge on the selective ion (negative for anions) - zi = +1 for K+, zi = +2 for Ca2+, zi = -2 for CO32-

58 ION-SELECTIVE ELECTRODES (ISE)
- Let ci = molarity of C+ - Activity (ai) rather than molarity is measured by ISEs - Activity is the effective (active) concentration of analyte (effective concentration decreases due to ionic interactions) ai = γici where γi = activity coefficient (between 0 and 1)

59 ION-SELECTIVE ELECTRODES (ISE) Selectivity Coefficient (k)
- A measure of the ability of ISE to discriminate against an interfering ion - It is assumed that ISEs respond only to ion of interest - In practice, no electrode responds to only one specific ion - The lower the value of k the more selective is the electrode - k = 0 for an ideal electrode (implies no interference)

60 ION-SELECTIVE ELECTRODES (ISE) Selectivity Coefficient (k)
For k > 1 - ISE responds better to the interfering ion than to the target ion For k = 1 - ISE responds similarly to both ions For k < 1 - ISE responds more selectively to ion of interest

61 ION-SELECTIVE ELECTRODES (ISE)
Empirical Calibration Plot Slope = 59/zi mV zi = charge of ion Called Nernstian slope Potential (mV) p[C+] - Used to determine the unknown concentration of analytes - Departure from linearity is observed at low concentrations

62 ION-SELECTIVE ELECTRODES (ISE)
Three groups of ISEs - Glass electrodes - Liquid electrodes - Solid electrodes

63 GLASS ELECTRODES - Responsive to univalent cations
- Employs thin ion-selective glass membrane

64 pH GLASS ELECTRODE - The most widely used
- For pH measurements (selective ion is H+) - Response is fast, stable, and has broad range - pH changes by 1 when [H+] changes by a factor of 10 - Potential difference is V when [H+] changes by a factor of 10 For a change in pH from 3.00 to 6.00 (3.00 units) Potential difference = 3.00 x V = 0.177

65 pH GLASS ELECTRODE - Thin glass membrane (bulb) consists of SiO4
- Most common composition is SiO2, Na2O, and CaO Glass membrane contains - dilute HCl solution saturated in AgCl - inbuilt reference electrode (Ag wire coated with AgCl)

66 pH GLASS ELECTRODE Glass Electrode Response at 25 oC
(potential across membrane with respect to H+) ΔpH = pH difference between inside and outside of glass bulb β ≈ 1 (typically ~ 0.98) (measured by calibrating electrode in solutions of known pH) K = assymetry potential (system constant, varies with electrodes)

67 pH GLASS ELECTRODE - Equilibrium establishes across the glass membrane with respect to H+ in inner and outer solutions - This produces the potential, E - Linearity between pH and potential - Calibration plot yields slope = 59 mV/pH units - Electrode is prevented from drying out by storing in aqueous solution when not in use

68 pH GLASS ELECTRODE Sources of Error - Standards used for calibration
- Junction potential - Equilibration time - Alkaline (sodium error) - Temperature - Strong acids - Response to H+ (hydration effect)

69 OTHEER GLASS ELECTRODES
Glass Electrodes For Other Cations K+ -, NH4+-, Na+-selective electrodes - Mechanism is complex - Employs aluminosilicate glasses (Na2O, Al2O3, SiO2) - Minimizes interference from H+ when solution pH > 5 pH Nonglass Electrodes - Quinhydrone electrode (quinone – hydroquinone couple) - Antimony electrode

70 SOLID-STATE ELECTRODES
- Solid membranes that are selective primarily to anions Solid-state membrane may be - single crystals (most common) - polycrystalline pellets or - mixed crystals

71 SOLID-STATE ELECTRODES
- Ionic solid contains the target ion - Solid is sealed to the end of a polymer tube - Contains internal reference electrode and filling solution - Concentration difference across the membrane causes migration of charged species across the membrane - Can measure concentrations as low as 10-6 M

72 SOLID-STATE ELECTRODES
Examples - Most common is fluoride-ion-selective electrode (limited pH range of 0-8.5) (OH- is the only interfering ion due to similar size and charge) - Iodide electrode (high selectivity over Br- and Cl-) Chloride electrode (suffers interference from Br- and I-) Thiocynate (SCN-) and cyanide (CN-) electrodes

73 LIQUID MEMBRANE ELECTRODES
- Employs water-immiscible substances impregnated in a polymeric membrane (PVC) - For direct measurement of polyvalent cations and some anions - The inner solution is a saturated solution of the target ion - Hydrophilic complexing agents (e.g. EDTA) are added to inner solutions to improve detection limits - Inner wire is Ag/AgCl

74 LIQUID MEMBRANE ELECTRODES Ion-Exchange Electrodes
- The basis is the ability of phosphate ions to form stable complexes with calcium ions - Selective towards calcium - Employs cation-exchanger that has high affinity for calcium ions (diester of phosphoric acid) - Inner solution is a saturated solution of calcium chloride - Cell potential is given by

75 LIQUID MEMBRANE ELECTRODES Other Ion-Exchange Electrodes
- Have poor selectivity and are limited to pharmaceutical formulations Examples - IEE for polycationic species (polyarginine, protamine) - IEE for polyanionic species (DNA) - IEE for detection of commonly abused drugs (large organic species)

76 LIQUID MEMBRANE ELECTRODES Anion-Selective Electrodes
- For sensing organic and inorganic anions Examples of Anions - Phosphate - Salicylate - Thiocyanate - Carbonate

77 OTHER ELECTRODES - Coated-wire electrodes (CWE)
- Solid-state electrodes without inner solutions - Made up of metallic wire or disk conductor (Cu, Ag, Pt) - Mechanism is not well understood due to lack of internal reference - Usually not reproducible For detection of amino acids, cocaine, methadone, sodium

78 GAS SENSING PROBES - For monitoring gases such as CO2, O2, NH3, H2S
- Device is known as compound electrode (probe is usually used in place of electrode) - Highly sensitive and selective for measuring dissolved gases - For environmental monitoring for clinical and industrial applications

79 GAS SENSING PROBES - Gas permeable membrane (teflon, polyethylene) is immobilized on a pH electrode or ion-selective electrode - Thin film of electrolyte solution is placed between electrode and membrane (fixed amount, ~0.1 M) - Inbuilt reference electrode - The target analyte diffuses through the membrane and comes to equilibrium with the internal electrolyte solution

80 GAS SENSING PROBES - The target gas then undergoes chemical reaction and the resulting ion is detected by the ion-selective electrode - Electrode response is directly related to the concentration of gas in the sample - Two types of polymeric materials are used Microporous and Homogeneous - Membrane thickness is ~ 0.01 – 0.10 mm - Membrane is impermeable to water and ions

81 GAS SENSING PROBES CO2 Sensors
- Consists of pH electrode covered by a CO2 selective membrane (silicone) - Electrolyte between electrode and membrane is NaHCO3-NaCl solution - pH of inner solution lowers when CO2 diffuses through membrane - Inner glass electrode senses changes in pH - Overall potential is determined by CO2 concentration in sample

82 GAS SENSING PROBES CO2 Sensors HCO3- solution pH glass electrode
CO2 + H2O ↔ H+ + HCO3- H+ lowers pH pH glass electrode Membrane (silicone)

83 GAS SENSING PROBES NH3 Sensors
- Consists of pH electrode covered by NH3 selective membrane (teflon or polyethylene) - Electrolyte between electrode and membrane is NH4+-KCl solution - NH3 goes through membrane and raises pH - Inner glass electrode senses changes in pH - Increase in pH is proportional to amount of NH3 in sample

84 Other Gas Sensing Devices
GAS SENSING PROBES Other Gas Sensing Devices NO2 and SO2 - Makes use of modified pH electrode H2S - Makes use of S2- ISE or modified pH electrode HF - Makes use of F- ISE or modified pH electrode

85 IMMOBILIZED ENZYME MEMBRANE
ELECTRODES - Enzymes are proteins that catalyze chemical reactions in living things - Based on coupling a layer of an enzyme with an electrode (enzyme is immobilized on an electrode) - Electrode serves as a transducer - Very efficient and extremely selective

86 IMMOBILIZED ENZYME MEMBRANE
ELECTRODES - Enzyme (biocatalytic) layer immobilized on an electrode Electrode Biocatalytic Layer

87 IMMOBILIZED ENZYME MEMBRANE
ELECTRODES Applications - Useful for monitoring clinical, environmental, food samples - For determination of glucose in blood (glucose sensors) For amperometric sensing of ethanol (ethanol electrodes) For sensing urea in the presence of urease enzyme (urea electrodes)

88 SOLID-STATE DEVICES - Known as ion-selective field effect transistors (ISFET) - Are semiconductor devices - Surface of transistor is covered with silicon nitride - Absorbs H+ from solution (results in change of conductivity) - Provides the ability to sense several ions (Na+, Ca2+, K+, pH in blood samples, etc) - For detection of hydrocarbons and NOx in exhaust

89 SOLID-STATE DEVICES - External reference electrode is required
- Does not require hydrating - Has rapid response time Examples Na+ ISFET NH3 ISFET Cl- ISFET

90 POTENTIOMETRY INSTRUMENTATION
- Employs a potential measuring device (handheld device) (high-impedance circuit) - Example is the pH meter (or pIon meter) - Designed to work with various electrodes - Have built-in temperature measurement and compensation - Three-point or more auto calibration - Two-electrode system (auxiliary reference electrode and working electrode)

91 APPLICATIONS OF POTENTIOMETRY
- Used as detectors for automated flow analyzers (flow injection systems) - High-speed determination of blood electrolytes in hospitals (H+, K+, Cl-, Ca2+, Na+) - For measuring soil samples (NO3-, Cl-, Li+, Ca2+, Mg2+) - Coupling ion chromatography with potentiometric detection - Micro ISEs as probe tips for SECM - Column detectors for capillary-zone electrophoresis

92 APPLICATIONS OF POTENTIOMETRY
- For studying chemical reactions (kinetics, equilibria, mechanism, solubility product constant, stability constant of complexes) - For characterization of materials - Quality control of raw materials and finished products - Pharmaceutical and biological studies - Elemental and molecular analysis - Environmental monitoring

93 APPLICATIONS OF POTENTIOMETRY
- Electronics - Electrochemical sensors Advantages of controlled potential processes - High sensitivity and selectivity - Very low detection limits - Wide range of electrode types - Wide range of linearity - Portable and low cost instrumentation

94 CONTROLLED POTENTIAL TECHNIQUES
- Electrostatic technique - Measurement of the current response to an applied potential - Various combinations of potential excitations exist (step, ramp, sine wave, pulse strain, etc)

95 CONTROLLED POTENTIAL TECHNIQUES
Instrumentation - Potentiostat (Voltammetric Analyzer) - Electrochemical cell with a three-electrode system Working Electrode (WE) Reference Electrode (RE) Counter/Auxiliary Electrode (CE/AE) - Plotter - Other components may be required depending on the type of experiment

96 CONTROLLED POTENTIAL TECHNIQUES
Potentiostat - Instrument that controls the potential at a working electrode - Connects the three electrodes Electrochemical Cell - Covered glass container of 5 – 50 mL volume - Contains the three electrodes immersed in the sample solution - Electrodes are inserted through holes in the cell cover - N2 gas used as deoxygenated gas

97 CONTROLLED POTENTIAL TECHNIQUES
Working Electrode (WE) - Electrode at which the reaction of interest occurs (Pt, Au, Ag, C) Reference Electrode (RE) - Provides a stable and reproducible potential - Independent of the sample composition (Ag/AgCl, SCE) Counter/Auxiliary Electrode (CE/AE) - Current-carrying electrode made of inert conducting metal (Pt wire, Graphite rod)

98 CONTROLLED POTENTIAL TECHNIQUES
- RE is placed as close as possible to WE to minimize potential drop caused by the cell resistance (iR) - Flow cannot occur through RE hence the need for CE to complete the current path - Current flows through solution between WE and CE - Voltage is measured between WE and RE

99 CONTROLLED POTENTIAL TECHNIQUES
Electrochemical Cell RE CE WE Opening N2 Teflon cap Glass container

100 MASS TRANSPORT - Three modes of mass transport Diffusion
- Spontaneous movement as a result of concentration gradient - Movement from regions of high concentration to regions of low concentration

101 MASS TRANSPORT - Three modes of mass transport Convection
Transport to the electrode by gross physical movement Forced Convection - Driving force is an external mechanical energy - Solution stirring or flowing - Electrode rotation or vibration Natural Convection - Physical movement as a result of density gradient

102 MASS TRANSPORT - Three modes of mass transport Migration
- Movement of charged particles along an electric field - Charge is carried through the solution as a result of movement of ions

103 SUPPORTING ELECTROLYTE
- Inert - Decreases the resistance of the solution - Eliminates electromigration effects - Maintains a constant ionic strength - Concentration range in usually 0.1 M – 1.0 M - Should be in large excess of analyte concentration

104 SOLVENTS - Medium for electrochemical measurements
- Contains a supporting electrolyte - Choice of solvent depends on the solubility and the redox activity of the analyte Solvent Properties - Electrical conductivity - Electrochemical activity - Chemical reactivity

105 OXYGEN REMOVAL - Purging with an inert gas for about 10 minutes
- Nitrogen gas is usually used - Purging is done just before voltammetric measurements - Necessary as oxygen complicates interpretation Other Methods - Formation of peroxides followed by reduction of peroxides Reduction by addition of sodium sulfite or ascorbic acid

106 COULOMETRY - Method in which charge is measured
- Species being measured is converted quantitatively to a new species The Methods Based on Electrolysis - Electrogravimetry - Constant-potential coulometry - Constant-current coulometry (coulometric titrimetry) Electrolysis - A process causing a thermodynamically nonspontaneous oxidation or reduction reaction to occur by application of potential or current

107 COULOMETRY Electrogravimetry
- Product of electrolysis is plated on a pre-weighed electrode - Electrode is weighed again after process and the amount plated is determined by difference - Metal dissolves from the anode and deposits on the cathode (electroplating, electrowinning, or electrorefining) Examples of metals commonly determined Cd, Bi, Co, Cu, Sb, Zn, Ni, In, Ag

108 Controlled Potential Coulometry
- Three electrode system - Permits applied potential pulse or ramp at the working electrode - Metal elements are deposited as potential is increased which increases charge passing through cell - The instrument is the coulometer which measures q

109 Controlled Potential Coulometry
Applications - Used to eliminate interferences from other reactions that take place at different potentials - Used to determine the number of electrons involved in a reaction - Used for coulometric titrations

110 Conductometric Analysis
COULOMETRY Conductometric Analysis - Measures electrical conductivity between two electrodes by ions in solution Applications - To determine the ionic content of drinking water, deionized water, solvents, beverages - Used as a detector for ion chromatography, HPLC - Used for conductometric titrations (end point determination)

111 COULOMETRY Instrumentation Apparatus comprises of
- Potentiostat with DC output voltage - Inert cathode and anode - Stirring rod set-up - Solution may be heated - Working electrode can be either anode or cathode - Controlled potential conditions

112 VOLTAMMETRY - Voltage between two electrodes is varied as current is measured - Solid working electrodes are used - Oxidation-reduction takes place at or near the surface of the working electrode - Graph of current versus potential is obtained - Peak current is proportinal to concentration of analyte

113 VOLTAMMOGRAM - Current versus potential plot
- Current on vertical axis and excitation potential on horizontal axis - Electrode reactions involve several steps and can be complicated - The rate is determined by the slowest step and depends on the potential range

114 CYCLIC VOLTAMMETRY - Involves linear scanning of potential of a stationary electrode using a triangular waveform - Solution is unstirred - The most widely used technique for quantitative analysis of redox reactions Provides information on - the thermodynamics of redox processes - the kinetics of heterogeneous electron transfer reactions - the kinetics of coupled reactions

115 CYCLIC VOLTAMMETRY - Is a three electrode system
- Pretreatment (polishing) of working electrode is necessary - The current resulting from an applied potential is measured during a potential sweep - Current-potential plot results and is known as cyclic voltammogram (CV)

116 CYCLIC VOLTAMMOGRAM (CV)
Triangular waveform (left) and CV (right) of ferricyanide

117 CYCLIC VOLTAMMETRY O + ne- ↔ R - Assume only O is present initially
- A negative potential sweep results in the reduction of O to R (starting from a value where no reduction of O initially occurs) - As potential approaches Eo for the redox process, a cathodic current is observed until a peak is reached - The direction of potential sweep is reversed after going beyond the region where reduction is observed

118 CYCLIC VOLTAMMETRY - This region is at least 90/n mV beyond the peak
- R molecules generated and near the electrode surface are reoxidized to O during the reverse (positive) scan - Results in an anodic peak current - The characteristic peak is a result of the formation of a diffusion layer near the electrode surface - The forward and reverse currents have the same shape

119 CYCLIC VOLTAMMETRY - Increase in peak current corresponds to achievement of diffusion control Characteristic Parameters - Anodic peak current (ipa) - Cathodic peak current (ipc) - Anodic peak potential (Epa) - Cathodic peak potential (Epc)

120 CYCLIC VOLTAMMETRY Reversible Systems
- Peak current for a reversible couple is given by the Randles-Sevcik equation (at 25 oC) n = number of electrons A = electrode area (cm2) C = concentration (mol/cm3) D = diffusion coefficient (cm2/s) ν = potential scan rate (V/s)

121 CYCLIC VOLTAMMETRY Reversible Systems ip is proportional to C
- Implies electrode reaction is controlled by mass transport ip/ic ≈ 1 for simple reversible couple - For a redox couple

122 CYCLIC VOLTAMMETRY Reversible Systems
- The separation between peak potentials - Used to determine the number of electrons transferred - For a fast one electron transfer ∆Ep = 59 mV - Epa and Epc are independent of the scan rate

123 Quasi-reversible Systems
CYCLIC VOLTAMMETRY Irreversible Systems - Systems with sluggish electron transfer - Individual peaks are reduced in size and are widely separated - Characterized by shift of the peak potential with scan rate Quasi-reversible Systems - Current is controlled by both charge transfer and mass transport - Voltammograms are more drawn out - Exhibit larger separation in peak potentials compared to reversible systems

124 CYCLIC VOLTAMMETRY Applications For analyzing - drugs - herbicides
- insecticieds - foodstuff additives - pollutants

125 POLAROGRAPHY - Voltammetry in which the working electrode is
dropping mercury - Makes use of potential ramp - Conventional DC - Wide cathodic potential range and a renewable surface - Hence widely used for the determination of many reducible species

126 POLAROGRAPHY - Initial potential is selected such that the reaction of interest does not take place - Cathodic potential scan is applied and current is measured - Current is directly proportional to the concentration-distance profile - Reduction begins at sufficiently negative potential [concentration gradient increases and current rises rapidly to its limiting value (iL)]

127 POLAROGRAPHY - Diffusion current is obtained by subtracting response due to supporting electrolyte (background current) - Analyte species entering region close to the electrode surface undergo instantaneous electron transfer reaction - Maximum rate of diffusion is achieved - Current-potential plot provides polarographic wave (polarogram)

128 DC POLAROGRAPHY - Three electrode system
WE = dropping mercury electrode (DME) CE = Pt wire or foil RE = SCE

129 DC POLAROGRAPHY The Ilkovic Equation
D = cm2/s C = mol/cm3 m = g/s t = s iL is current at the end of drop life (the limiting current) iL is a measure of the species concentration

130 Half Wave Potential (E1/2)
DC POLAROGRAPHY Half Wave Potential (E1/2) - Potential at which the current is one-half its limiting value - E1/2 is independent of concentration of species DR = diffusion coefficient of reduced species DO = diffusion coefficient of oxidized species - Experimental E1/2 is compared to literature values to identify unknown analyte

131 Half Wave Potential (E1/2)
DC POLAROGRAPHY Half Wave Potential (E1/2) At 25 oC - A graph of E versus log[(iL-i)/i] is linear if reaction is reversible (Nernstian behavior) - Slope = /n and intercept = E1/2 E = E1/2 when [Ox] = [Red]

132 STRIPPING ANALYSIS Two step technique
1. Deposition Step (Preconcentration step) - Involves preconcentration of analyte species by reduction (anodic stripping) or oxidation (cathodic stripping) into a mercury electrode 2. Stripping Step - Measurement step - Rapid oxidation or reduction to strip the products back into the electrolyte

133 STRIPPING ANALYSIS - Very sensitive for trace analysis of heavy metals
- Favorable signal to background ratio - About four to six metals can be measured simultaneously at levels as low as M - Low cost instrumentation - There are different versions of stripping analysis depending on the nature of the deposition and stripping steps

134 Anodic Stripping Voltammetry (ASV)
STRIPPING ANALYSIS Anodic Stripping Voltammetry (ASV) - The most widely used stripping analysis - Preconcentration is done by cathodic deposition at controlled potential and time - Metals are preconcentrated by electrodeposition into a small-volume Hg electrode - Deposition potential is usually 0.3 – 0.5 V more negative than Eo for the analyte metal ion

135 Anodic Stripping Voltammetry (ASV)
STRIPPING ANALYSIS Anodic Stripping Voltammetry (ASV) - Metal ions reach the Hg electrode surface by diffusion and convection - Electrode rotation or solution stirring is employed to achieve convection - Metal ions are reduced and concentrated as amalgams Mn+ + ne- + Hg → M(Hg) - Hg film electrodes or Hg drop electrodes may be used

136 Anodic Stripping Voltammetry (ASV)
STRIPPING ANALYSIS Anodic Stripping Voltammetry (ASV) Following preselected deposition period: - Forced convection is stopped - Anodic potential scan is employed (may be linear or pulse) - Amalgamated metals are reoxidized (stripped off electrode) - An oxidation (stripping) current then flows M(Hg) → Mn+ + ne- + Hg

137 Cathodic Stripping Voltammetry (CSV)
STRIPPING ANALYSIS Cathodic Stripping Voltammetry (CSV) - Mirror image of ASV - Involves anodic deposition of analyte and subsequent stripping by a potential scan in the negative direction An- + Hg ↔ HgA + ne- (Deposition to the right and stripping to the left) - Useful for measuring organic and inorganic compounds that form insoluble salts with Hg (thiols, penicillin, halides, cyanides)