Presentation is loading. Please wait.

Presentation is loading. Please wait.

Fundamentals of Digital Radiology

Similar presentations


Presentation on theme: "Fundamentals of Digital Radiology"— Presentation transcript:

1 Fundamentals of Digital Radiology
George David Medical College of Georgia

2 So what is “Digital”?

3 Filmless Department What we mean by Digital Digital Radiographs PACS
Picture Archival & Communication Systems Reading from Monitors

4 What we really mean by Digital
No more File Room!!!

5 Digital Image Formation
Place mesh over image

6 Digital Image Formation
Assign each square (pixel) a number based on density Numbers form the digital image 194 73 22

7 Digital Image Formation
The finer the mesh, the better the digital rendering

8 What is this? 12 X 9 Matrix

9 Same object, smaller squares
24 X 18 Matrix

10 Same object, smaller squares
48 X 36 Matrix

11 Same object, smaller squares
96 X 72 Matrix

12 Same object, smaller squares
192 X 144 Matrix

13 Numbers / Gray Shades Each number of a digital image corresponds to a gray shade for one picture element or pixel

14 So what is a digital image?
Image stored as 2D array of #’s representing some image attribute such as optical density x-ray attenuation echo intensity magnetization 125 25 311 111 182 222 176 199 192 85 69 133 149 112 77 103 118 139 154 120 145 301 256 223 287 225 178 322 325 299 353 333 300

15 Computer Storage 125 25 311 111 182 222 176 199 192 85 69 133 149 112 77 103 118 139 154 120 145 301 256 223 287 225 178 322 325 299 353 333 300 125, 25, 311, 111, 182, 222, 176, 199, 192, 85, 69, 133, 149, 112, 77, 103, 118, 139, 154, 125, 120, 145, 301, 256, 223, 287, 256, 225, 178, 322, 325, 299, 353, 333, 300

16 Digital Copies = If you’ve got the same numbers ...
125, 25, 311, 111, 182, 222, 176, 199, 192, 85, 69, 133, 149, 112, 77, 103, 118, 139, 154, 125, 120, 145, 301, 256, 223, 287, 256, 225, 178, 322, 325, 299, 353, 333, 300 125, 25, 311, 111, 182, 222, 176, 199, 192, 85, 69, 133, 149, 112, 77, 103, 118, 139, 154, 125, 120, 145, 301, 256, 223, 287, 256, 225, 178, 322, 325, 299, 353, 333, 300 = If you’ve got the same numbers ...

17 Digital Copies then you have an identical copy =

18 = Digital Copies Digital copies are identical
All digital images are originals =

19 Image Matrix Doubling the matrix dimension quadruples the # pixels
125 25 311 111 199 192 85 69 77 103 118 139 145 301 256 223 111 87 118 155 2 X 2 Matrix 4 pixels 4 X 4 Matrix 16 pixels

20 Doubling the matrix dimension quadruples # pixels
Image Matrix Doubling the matrix dimension quadruples # pixels A matrix compared to a 5122 matrix quadruples disk storage requirements image transmission time digital image manipulation Matrix # Pixels 512 X => ,144 1024 X1024 => 1,048,576 2048 X2048 => 4,194,304

21 Matrix Size & Resolution
More pixels = better spatial resolution

22 The Bit Fundamental unit of computer storage Only 2 allowable values
1 Computers do all operations with 0’s & 1’s BUT Computers group bits together

23 Special Binary Digit Grouping Terms
Nibble 4 binary bits (0101) Byte 8 binary bits ( ) Word 16 binary bits ( ) Double Word 32 binary bits ( )

24 Abbreviations Review Bit (binary digit) Byte Kilobyte Megabyte
Smallest binary unit; has value 0 or 1 only Byte 8 bits Kilobyte 210 or 1024 bytes sometimes rounded to 1000 bytes Megabyte 213 or 1,048,576 bytes or 1024 kilobytes sometimes rounded to 1,000,000 bytes or 1,000 kilobytes

25 # of unique values which can be represented by 1 bit
2 unique combinations / values 1 2

26 # of unique values which can be represented by 2 bits
1 2 4 unique combinations / values 3 4

27 # of unique values which can be represented by 3 bits
5 1 6 2 7 3 8 4 8 unique combinations / values

28 Digital Image Bit Depth
the number of computer bits (1’s or 0’s) available to store each pixel value Values Bits # Values 1 2 3 . 8 0, 1 00, 01, 10, 11 000, 001, 010, 011, 100, 101, 110, 111 . , , 2 1 = 2 2 2 = 4 2 3 = 8 . 2 8 = 256

29 Digital Image Bit Depth
bit depth indicates # of possible brightness levels for a pixel presentation of brightness levels pixel values assigned brightness levels brightness levels can be manipulated without affecting image data window level

30 Bit Depth & Contrast Resolution
The more bits per pixel the more possible gray shades and the better contrast resolution. 2 bit; 4 grade shades 8 bits; 256 grade shades

31 Computer Storage Storage = # Pixels X # Bytes/Pixel
Example: 512 X 512 pixels; 1 Byte / Pixel 512 X 512 pixel array # pixels = 512 X 512 = 262,144 pixels Storage = 262,144 pixels X 1 byte / pixel = 262,144 bytes = 256 KBytes = .25 MBytes

32 Image Size Related to both matrix size & bit depth
higher (finer) matrix requires more storage doubling matrix size quadruples image size higher bit depth requires more storage doubling bit depth theoretically doubles image size Computer may require storage in multiples of 8 bits (bytes) 10 or 12 bits stored in 16 bit slot alters image size requirements 1 2 3 4 5 6 7 8 9

33 Image Compression jpg gif (20) 37’s
reduction of digital image storage size by application of algorithm for example, repetitive data could be represented by data value and # repetitions rather than by repeating value jpg 37, 37, 37, 37, 37, 37, 37, 37, 37, 37, 37, 37, 37, 37, 37, 37, 37, 37, 37, 37 gif (20) 37’s

34 Image Compression Image Decompression Compression Ratio
calculating original digital image from previously compressed data Compression Ratio original image size compressed image size ratio depends upon data to be compressed algorithm

35 Compression Types Reversible Compression Non-reversable Compression
Image decompresses to original pixel values Low compression ratios only Non-reversable Compression Decompressed image’s pixel values not necessarily identical to original much higher compression ratios possible variation from original image may or may not be visible or clinically significant

36 Non-Reversable Compression
variation from original image generally increases with increasing compression ratio but a higher compression ratio means less storage requirements variation less noticeable for dynamic (moving) images than for still images such as radiographs

37 Computed Radiography (CR)
Re-usable metal imaging plates replace film & cassette Uses conventional bucky & x-ray equipment                                                                  

38 CR Exposure & Readout

39 CR Readout

40 Another View: CR Operation

41 Computer Radiography (CR)
plate is photostimulable phosphor radiation traps electrons in high energy states higher states form latent image H i g h e r E n e r g y - E l e c t r o n S t a t e P h o t o n p u m p s e l e c t r o n t o X - R a y h i g h e r e n e r g y s t a t e P h o t o n - - - L o w e r E n e r g y - - - - - - E l e c t r o n - - - - - - - - - S t a t e - - - - - - - - -

42 Reading Imaging Plate reader scans plate with laser
laser releases electrons trapped in high energy states electrons fall to low energy states electrons give up energy as visible light light intensity is measure of incident radiation Lower Energy Electron State

43 Reading Imaging Plate Reader scans plate with laser light using rotating mirror Film pulled through scanner by rollers Light given off by plate measured by PM tube & recorded by computer

44 Laser & Emitted Light are Different Colors
Phosphor stimulated by laser light Intensity of emitted light indicates amount of radiation incident on phosphor at each location Only color of light emitted by phosphor measured by PMT

45 CR Operation after read-out, plate erased using a bright light
plate can be erased virtually without limit Plate life defined not by erasure cycles but by physical wear

46 CR Resolution Small cassettes have better spatial resolution
Smaller pixels More pixels / mm

47 CR Throughput Generally slower than film processing
CR reader must finish reading one plate before starting to read the next Film processors can run films back to back                

48 CR Latitude Much greater latitude than screen/film
Plate responds to many decades of input exposure under / overexposures unlikely Computer scale inputs exposure to viewable densities Unlike film, receptor separate from viewer                

49 Film Screen vs. CR Latitude
CR Latitude: .01 – 100 mR 100

50 Digital Radiography (DR)
Digital bucky Incorporated into x-ray equipment

51 Digital Radiography (DR)
Receptor provides direct digital output No processor / reader required Images available in < 15 seconds Much less work for technologist

52 TFT = THIN-FILM TRANSISTOR ARRAY
Direct vs. Indirect TFT = THIN-FILM TRANSISTOR ARRAY

53 Digital Radiography (DR)
Potentially lower patient dose than CR High latitude as for CR Digital bucky fragile First DR portables coming to market

54 Raw Data Image Unprocessed image as read from receptor
CR Intensity data from PMT’s as a result of scanning plate with laser DR Raw Data read directly from TFT array Not a readable diagnostic image Requires computer post-processing Specific software algorithms must be applied to image prior to presenting it as finished radiograph

55 Enhancing Raw Image (Image Segmentation)
* Identify collimated image border Separate raw radiation from anatomy Apply appropriate tone-scale to image Done with look-up table (LUT) This process is specific to a particular body part and projection

56 Image Segmentation Computer then defines anatomic region
Computer must establish location of collimated border of image Computer then defines anatomic region Finished image produced by tone scaling Requires histogram analysis of anatomic region

57 Histogram Graph showing how much of image is exposed at various levels

58 Tone Scaling Post-Processing
Body part & projection-specific algorithms determine average exposure Must correctly identify anatomical region LUT computed to display image with proper Density Contrast

59 Film/Screen Limited Latitude
Film use has little ambiguity about proper radiation exposure

60 Should I Worry? In CR & DR, image density is no longer a reliable indicator of exposure factor control.

61

62 CR / DR Latitude DANGER Will Robinson!!!
Almost impossible to under or overexpose CR / DR Underexposures look noisy Overexposures look GOOD!!!

63 So how do I know if exposure is optimum by looking at my image?

64 Exposure Index Each manufacturer provides feedback to technologist on exposure to digital receptor Displayed on CR reader monitor Displayed on workstations

65 Calculated Exposure Index Affected by
X-Ray technique selection Improper centering of image on cassette Improper selection of study or projection Placing two or more views on same cassette Can cause image to appear dark

66 Phototimed Phantom Image
75 kVp 88 mAs 2460 EI

67 Let’s Approximately Double mAs
75 kVp 88 mAs 2460 EI 75 kVp 160 mAs 2680 EI

68 Let’s Go Crazy 75 kVp 88 mAs 2460 EI 75 kVp 640 mAs 3300 EI

69 How Low Can You Go? Cut mAs in Half!
75 kVp 88 mAs 2460 EI 75 kVp 40 mAs 2060 EI

70 Let’s Go Crazy Low 75 kVp 8 mAs 1380 EI 75 kVp 1 mAs 550 EI

71 CR Artifacts Physical damage to imaging plates Dirt in reader
Cracks, scuffs, scratches Contamination Dust / dirt Dirt in reader Highly sensitive to scatter radiation

72 DR Artifacts Dead detector elements
Spatial variations in background signal & gain Grid interference Software can help correct for above

73 Shifting Gears: Fluoroscopy Issues

74 Digital Video Sources DR type image receptor
Conventional Image Intensifier with Video Signal Digitized (“Frame Grabber”) I m a g e T u b X-Ray Input Image Tube TV Amplfier Analog to Digital Converter Memory (Computer) Lens System

75 Digital Spot Film Frame grabber digitizes image
Digital image saved by computer Radiographic Technique used required to control quantum noise

76 Last Image Hold Computer displays last fluoro image before radiation shut off. Image noisier than for digital spot Image made at fluoroscopic technique / intensity Allows operator to review static processes without keeping beam on ideal for teaching environments ideal for orthopedic applications such as hip pinning Less radiation than digital spot

77 Fluoro Frame Averaging
Conventional fluoro only displays current frame Frame averaging allows computer to average current with user-selectable number of previous frames Averages current frame & history

78 Fluoro Frame Averaging Tradeoff
Advantage: Reduces quantum noise Disadvantage Because history frames are averaged with current frame, any motion can result in lag

79 Other Fluoro Features Real-time Edge Enhancement / Image Filtering
Option of using lower frame rates (15, 7.5, 3.75 fps rather than 30) computer displays last frame until next one reduces flicker Lowers patient and scatter exposure Exposure proportional to frame rate dynamic studies may be jumpy

80 Digital Subtraction Immediate replay of run Free selection of mask
before or after bolus >1 frame may be averaged for mask Note subtraction adds noise

81 Digital Image Manipulations
on-screen measurements distances angles volumes/areas stenosis image annotation peak opacification / roadmapping peak opacification displays vessels after a test injection allows visualization of live catheter on top to saved image of test injection

82 Digital Possibilities
Multi-modality imaging / Image fusion PET/CT

83 DR & Energy Subtraction
2 images taken milliseconds apart at 2 kVp’s Combine / subtract images Soft Tissue Image Bone Image

84 DR Mobile Units GE Definium AMX 700 See image immediately
Wireless transmission of images GE Definium AMX 700

85 Other Possibilities Tomosynthesis Histogram Equalization
Multi-slice linear tomography from one exposure series Histogram Equalization Use computer to provide approximately equal density to various areas of image.

86 The End ?


Download ppt "Fundamentals of Digital Radiology"

Similar presentations


Ads by Google