1.1 Range of molar absorptivity

Similar presentations


Presentation on theme: "1.1 Range of molar absorptivity"— Presentation transcript:

1 1.1 Range of molar absorptivity
Chapters 13&14 Ultraviolet-Visible Molecular Absorption Spectrometry 1 Absorbing Species 1.1 Range of molar absorptivity Electronic excitation of outer valence (i.e. bonding) electron How probable for this electronic excitation? (allowed transition, or forbidden) judged by the range of   = L mol-1 cm-1, strong absorption  < 103 L mol-1 cm-1, low intensity

2 1.2 Which electron get excited? 1.2.1 Organic molecules
,  (bonding) and n (non-bonding) orbitals *, * (anti-bonding) orbitals   * E large ( < 150 nm, out of range)  = ,000 Lmol-1cm-1 n  * E smaller ( = nm)  = Lmol-1cm-1   * n  * E smallest ( = nm)  = 10-10,000 Lmol-1cm-1 Ideal for UV-Vis spectrometry of organic chromophore

3

4 1.2.2 Inorganic molecules Most transition metal ions are colored (absorption in Vis) due to d  d electronic transition Fig (p.370) Fig (p.371)

5 1.2.3 Charge-transfer absorption
A: electron donor, metal ions D: electron acceptor, ligand  > 10,000 Fig (p.371)

6 2 Beer’s Law 2.1 Approximation of T and A
A = -log T = log (P0/P) = ··b·c : molar absorptivity at one particular wavelength (L·mol-1cm-1) b: path length of absorption (cm) c: molar concentration (mol·L-1) Fig (p.158) Light loss due to reflection (17.3%), scattering, … Fig (p.337)

7

8 = 1bc1 + 2bc2 A2 = 1,2·b·c1 + 2,2·b·c2
2.2 Application of Beer’s law to mixtures Absorbance is additive Atotal = A1 + A2 + … = 1bc1 + 2bc2 For a 2-component mixture, we measure the absorption at two different wavelength, respectively A1 = 1,1·b·c1 + 2,1·b·c2 A2 = 1,2·b·c1 + 2,2·b·c2

9 A = -log T = log (P0/P) = ··b·c
2.3 Limitations of Beer’s law 2.3.1 Real deviations At low concentration A = -log T = log (P0/P) = ··b·c At c > 0.01 M solute-solute interaction, hydrogen-bond, … can alter the electronic absorption at a given wavelength  dilute the solution

10 2.3.2 Chemical effects analyte associates, dissociates or reacts with a solvent to give molecule with different  Example: acid-base equilibrium of an indicator 430 570 HIn 6.30 x x (measured in HCl solution) In x x102 (measured in NaOH solution) What’s the absorbance of unbuffered solution at c = 2 x 10-5M?

11 Fig (p.340) =[HIn] + [In-]

12 2.3.3 Instrumental deviations due to polychromatic radiation
Beer’s law applies for monochromatic absorption only. If a band of radiation consisting of two wavelength ’, and ” Assuming Beer’s law applies to each wavelength Fig (p.176)

13 2.3.3 Instrumental deviations due to polychromatic radiation
Non-linear calibration curve Fig (p.341)

14 How to avoid : Select a wavelength band near its maximum absorption where the absorptivity changes little with wavelength Fig (p.341)

15 2.3.4 Other physical effects
stray light – the scattering, reflection radiation from the instrument, outside the nominal wavelength band chosen for measurement mismatched cell for the sample and the blank

16 3 Noise of Spectrophotometric analyses
3.1 Standard deviation of c

17 3.2 Sources of instrumental noise
Case I Limited readout resolution (31/2-digit displays  0.1% uncertainty from 0%T -100% T) Thermal noise in thermal detector, etc (particularly for IR and neat IR spectrophotometer) Case II Shot noise in photon detector (random emission of photon from the light source or random emission of electrons from the cathode in a detector) Case III Flicker noise, Fail to position sample and blank cells reproducibly in replicate measurements (as a result, different sections of cell window are exposed to radiation, and reflection and scattering losses change)

18 3.2 Sources of instrumental noise
Fig (p.344)

19 4 Instrumentation 4.1 Designs a. Single beam b. Double-beam-in-space
c. Double-beam-in-time Advantage of double beam configuration Compensate for fluctuation in the radiant output, drift in transducer, etc. Continuous recording of spectra Fig (p.352)

20 Shimadzu UV-2450 Spectrophotometer

21 Wavelength Range 190 to 900nm (performance guaranteed range). Extendable to 1,100nm through the use of the optional photomultiplier. (The measurable range maybe restricted in the shorter wavelength side depending on the type of photomultipler used.) Monochromator System UV-2450: Single monochromator with a high-performance blazed holographic grating in the aberration corrected Czerny-Turner mounting. Resolution 0.1nm Spectral Bandwidth 0.1, 0.2, 0.5, 1, 2 and 5nm Wavelength Accuracy ±0.3nm Wavelength Repeatability ±0.01nm Wavelength Scanning Speed FAST, MEDIUM, SLOW, and SUPER SLOW Light Source 50W halogen lamp (2,000 hours of life) and D2lamp (500 hours of life) Light Source lamp switching Selectable between 282nm and 393nm Stray Light UV-2450: Less than 0.015% at 220nm and 340nm Detector Photomultiplier R-928 Photometric System Double beam, direct ratio system with dynode feedback Photometric Mode Absorbance (Abs.), transmittance (%), reflectance (%) and energy (E). Photometric Range Abosrbance: -4~5 Abs. (0.001 Abs. increments) Transmittance: 0~999.9% (0.01 increments) Reflectance: 0~999.9% (0.01% increments) Photometric Accuracy ±0.002Abs(0~0.5Abs), ±0.004Abs(0.5~1Abs),±0.3T (0~100%T) (all determined with NIST 930D standard filter) Photometric Repeatability 0.001Abs (0~0.5Abs), ±0.1%T Baseline Correction Selectable with storage in firmware Baseline Flatness Within ±0.001Abs. (excluding noise, 2nm slit width and SLOW wavelength scanning speed) Drift Less than Abs. per hour (after 2 hours warm-up) Dimensions 570 (W) x 660 (D) x 275 (H) (mm) Weight 36kg Power Supply AC 100V/120V/220V/240V, 50/60Hz 250VA (swithc-selectable)

22 Shimadzu UV-2450 Spectrophotometer

23 d. Effects of monchromator exit slit width on spectra
Narrow exit slit width improves the spectrum resolution but it also significantly reduce the radiant power Trade-off between resolution and S/N ratio

24 e. Multichannel spectrometer
No monochromator, but disperses transmitted light and measures “all wavelength at once” No Scanning-simple and fast More expensive Limited resolution Fig (p.353)

25 5.1 Quanlitative spectra 5 Applications
Solvent effects on the UV-Vis spectra Polar solvents “blur” vibrational features Polar solvents shift absorption maxima n  * blue shift   * red shift - UV-Vis not reliable for qualitative but excellent for quantitative analysis

26 5.2 Quantitative analysis
- Determining the relationship between A and c External Standards Standard-Addition

27 5.3 Spectrophotometric kinetics
Fig (p.382)

28 Stopped-flow mixing Fig (p.384)