Statistics & graphics for the laboratory 54 Metrological basis of IQC Introduction The error concept chosen here The total error concept Total error – calculations Instability of the analytical process Instability and analytical process specifications Instability – how much can be tolerated? Analytical process specifications - TEa IQC and TEa The error model for IQC Basic formula Critical errors Graphical presentation of critical errors Calculation of critical errors Special topic: The TEa problematic Control rules based on TEa Automatic selection of rules based on TEa: The Validator OPSpecs®-Charts Critical error graphs Selection of a control rule based on TEa with the Validator: an example Other selection tools: the TEa/CV a,tot ratio; the IQC decision tool Summary and Conclusion
2
Statistics & graphics for the laboratory 55 Metrological basis of IQC Introduction FOREWORD Currently, there is confusion about metrological terms. While the term error is used in the metrological “bible” (VIM [1]), it is deprecated in one of the recent ISO concepts (GUM [2]). This concept, even, recommends to avoid the distinction between random and systematic effects (= uncertainty concept: GUM). Moreover, ISO uses the concept of accuracy [3]: Accuracy = Combined trueness & precision Inaccuracy = Combined ([untrueness]: note: this term is not defined by ISO) and imprecision. Note: Some traditions link accuracy to systematic error. This book uses the [total] error concept (Adapted from: Stöckl D. Scand J Clin Lab Invest 1996;56:193-7): Total-, systematic-, and random error [In]Accuracy: combined [“un“]trueness and [im]precision Experimental estimates: Total error, bias, SD Note: all experimental estimates have a statistical uncertainty! References [1] ISO VIM. Vocabulaire international des terms fondamentaux et généraux de métrologie. [2] ISO GUM. Guide to the expression of uncertainty in measurement. [3] ISO 5725-1. Accuracy (trueness and precision) of measurement methods and results. Classical metrological error concept: distinction of random and systematic error. Inaccuracy, “untrueness“ and imprecision: pendents for total-, systematic-, and random error. Note: All experimental values (bias, s) have a certain statistical level of probability («±uncertainty»).
3
Statistics & graphics for the laboratory 56 Metrological basis of IQC The total error (TE) concept The term "total error" (TE) (#) is applied to measurement results that are influenced by random (RE) and systematic (SE) components of error. Calculation of TE for single measurements: TE = SE + z RE, or TE = SE + z s –(z is usually set to 1.96 or 2.58, encompassing 95% or 99% [two-tailed] of a gaussian distributed population) Calculation of TE for multiple measurements: TE = SE + t n-1 [s/ n] –s = the standard deviation, n = number of measurements, t n-1 = student's t- value for (n-1) degrees of freedom (e.g., at the 95 % confidence level) Note: Random error can be reduced by increasing the number of measurements, systematic error cannot. Instability of the analytical processes -Due to occurrence of systematic error (1) -Due to increased random error (2) References #Random, systematic, and total error of measurement [adapted from: Westgard JO, Barry PL. Cost-effective quality control. 4th printing. Washington: AACC Press, 1995]
4
Statistics & graphics for the laboratory 57 Instability – how much can be tolerated? The answer is: Poor analytical quality must not invalidate the medical decision! The conclusion is: We have to know the medical requirements for analytical quality («analytical process specifications») Allowable total error (TEa). Analytical process specifications - TEa Concepts and numerical examples can be found in the literature below (#). Note: Optimally, TEa should be medically justified! IQC and TEa Observation The statistically defined "stop-limit" of IQC (TE IQC ; e.g., 3 s) may be much lower than medically relevant (e.g., triacylglycerides). The idea Let's stop the process at a medically relevant (or otherwise defined) allowable total error: TEa (TEa: also called "quality specification"). Advantages If TEa >> TE Stable, "very loose" IQC-rules can be selected. Prerequisite for IQC TE stable < TEa There must be some room left for the IQC-procedure Note: IQC cannot transform an intrinsically poor method into a good method #References Strategies to set global analytical quality specifications in laboratory medicine. Consensus Statement, Stockholm 1999. Scand J Clin Lab Invest 1999;59:585 Stöckl D, Baadenhuijsen H, Fraser CG, Libeer J-C, Hyltoft Petersen P, Ricós C. Desirable routine analytical goals for quantities assayed in serum. Eur J Clin Chem Clin Biochem 1995;33:157-69. www.westgard.com Metrological basis of IQC Instability and analytical process specifications How much instability? TE IQC, max < TEa work within specifications TE IQC,max : Maximum total error guaranteed by the IQC-procedure (95% of results should be within ±TEa)
5
Statistics & graphics for the laboratory 58 Metrological basis of IQC The error model for IQC Basic formula (#) TE = Bias + [ SE cont s meas ] + z [ RE cont s meas ] Bias: method bias, usually assumed to be zero SE cont : change in systematic error that can be detected by the quality control procedure s meas : measured, intrinsic standard deviation of the test RE cont : change in random error that can be detected by the quality control procedure z: statistical multiplier related to the portion of a distribution exceeding a quality requirement ("defect rate") Note: mostly, a z-value of 1.65 is selected to obtain a 5% defect rate (one-tailed). When pure random error is addressed, a z-value of 1.96 is selected to obtain a 5% defect rate (two-tailed). Critical random and systematic error definition: The amount of error that places 5% of results outside TEa. The role of IQC is to detect process deterioration before >5% of the results exceed TEa. Graphical presentation of critical errors References # Westgard JO, Barry PL. Cost-effective quality control. 4th printing. Washington: AACC Press, 1995] Introduction of an SE that places 5% of results outside TEa: SE crit = 2 Increase of an RE that places 5% of results outside TEa: RE crit = 2 (RE i = intrinsic or stable imprecision)
6
Statistics & graphics for the laboratory 59 Metrological basis of IQC Calculation of critical errors Assumption of imprecision and bias Critical random error ( REc) and critical systematic error ( SEc) can be calculated that cause measurements to exceed TEa in 5% of the cases (= defect rate) due to either increased imprecision or increased systematic error. Formulae for a 5% defect rate: Case with SEc = 0 REc = (TEa - bias)/1.96 s meas ; Case with REc = 1 SEc = [(TEa - bias)/s meas ] - 1.65 Assumption of zero bias For the purpose of IQC, the bias of the method often is assumed to be zero. This is justified if a laboratory establishes its own IQC target values or if it can verify the system-specific target values of commercial IQC materials. Formulae for REc and SEc with bias = 0 REc = TEa/1.96 s meas ; SEc = (TEa/s meas ) - 1.65 Note: For simplification, the QC-Validator® program applies the default factor 2 (for RE and SE) for the stable process (elder versions: 1.65!). The factor may be changed by the user. Checklist – Metrology and rule selection Metrology, knowledge of The total error concept (TE, SE, RE) Concept of maximum allowable total error (TEa) (process specifications) Critical error Rule selection Statistical basis: A rule is chosen based on P fr and P ed. SD-limits are taken from stable performance. TEa basis: From a specification for TEa, critical error values can be calculated. From the critical error values, adequate IQC rules can be selected, naturally, on statistical basis. Selection tools: power functions, OPSpecs, the TEa/CV a,tot ratio; the IQC decision tool
7
Statistics & graphics for the laboratory 60 Special topic: the TEa problematic Introduction Specifications for TEa, the concepts Clinical Biological –Notes on IQC rules based on biological TEa –IQC rules from biological TEa and reality Expert Summary Checklist – TEa values
8
Statistics & graphics for the laboratory 61 Special topic: the TEa problematic Introduction BEWARE There are a lot of misunderstandings when selecting a TEa for IQC purposes. We look into TEa in more detail! Questions What is a responsible choice for TEa? Is there a consensus about TEa? Investigate current concepts for establishing values for TEa. Specifications for TEa Current concepts for TEa The Table below lists, in hierarchical order, current concepts for establishing values for TEa. Sources (in hierarchical order#) Clinical concepts (few analytes) Concepts based on biological variation Expert opinion Regulations "State-of-the-art" #Consensus Statement (Stockholm 1999). Scand J Clin Lab Invest 1999;59:585. See also: ISO 15196
9
Statistics & graphics for the laboratory 62 Special topic: the TEa problematic Specifications for TEa Inspection of proposed numbers for TEa EXAMPLE: Serum-sodium When we compare the currently proposed numbers, we can conclude that: There is no consensus about TEa! Additionally: TEa may depend on concentration/disease
10
Statistics & graphics for the laboratory 63 Special topic: the TEa problematic TE from biology BUT: Some are too stringent for current technology Biology gives extreme low values for some analytes: TEa sodium: Ricos: 0.9% (http://www.westgard.com/biodatabase1.htm; but note: CV g on that site = CV b )http://www.westgard.com/biodatabase1.htm Apply a bottom-line# #Stöckl D. Desirable Performance criteria … based on biological analyte variation - hindrances to reaching some and reasons to surpass some. Clin Chem 1993;39:913-4. Bottom-line (for stable process!) CV: 1% Bias: 1,5% TE: 3,2% Consider to surpass some Think about using more stringent goals than derived from biology the goals are a compromise; account for IQC; some goals are surpassed, by far, by current technique. Stöckl D. Desirable Performance criteria … based on biological analyte variation - hindrances to reaching some and reasons to surpass some. Clin Chem 1993;39:913-4. Tonks: 10% upper limit for TE
11
Statistics & graphics for the laboratory 64 Special topic: the TEa problematic Proposed numbers – How realistic are they? The numbers proposed by clinical and biological concepts, usually, are desirable numbers (= goals), which would allow medical decisions to be made without compromise of analytical quality. However, the laboratory has to work with “real numbers” = “state-of-the-art” quality. Comparison of reality with “goals” from biology The Figure below compares “ state-of-the-art” quality with desired quality: 1/4 of the reference interval. [1] TEa = ¼ of the reference interval [2] State-of-the-art total CVa. *Rules calculated with the above TEa and CV values by the Validator (90% assurance; 2 materials; bias: Na 0.5%, Trigl 2%, others: 1%). Observation Some state-of-the-art CVs are too big to allow calculation of IQC-rules from biology-derived TEa values, for example, for sodium. Expert TEa values Sodium: other numbers Koch, Westgard et al. Clin Chem 1990;36:230-3. We have seen that sodium was a problem for IQC with TEa values derived from biology. For others, however, sodium seems to be very easy for IQC (see below). They propose a 3.5 s rule. Note: TEa was chosen by experts and was the US EQA-limit (2.9%). Additional, their “state-of-the-art” CV (0.5%) seems to be quite optimistic! CONCLUSION Don’t cheat yourself by selecting the most liberal TEa values and by underestimating your actual CV.
12
Statistics & graphics for the laboratory 65 More on expert TEa values Below, are shown more examples of IQC rule setting by use of expert TEa values (upper example) and by general expert opinion (lower example). Koch, Westgard, et al. Clin Chem 1990;36:230-33 (see also : Westgard & Stein. Clin Chem 1997;43:400-3). Based on “expert” TEa values, following rules were proposed: 1 3.5s (n = 2) for: Sodium (TEa = 2.9%, CV = 0.5%), potassium, urea nitrogen, creatinine, phosphate, uric acid, cholesterol, total protein, total bilirubin, GGT, ALP, AST, and LD. 1 2.5s (n = 2) for: Chloride, total CO2. Calcium was a problem: Compared to TEa, precision was considered as too poor. Mugan, Carlson, Westgard. J Clin Immunoassay 1994;17:216-22. 1 2.5s (n = 3) for: Prolactin, ß-hCG, CEA, FSH, LH, TSH, ß2-microglobulin. Experts tend to simple rules; choice is not directly related to medical requirements Rules selected by general opinion Steindel SJ, Tetrault G. Arch Pathol Lab Med 1998;122:401-8. 2.5 – 2.7s rule: generally Tetrault GA. CAP Today 1995 (April):60-1. 3.5 s rule: generally Experts tend to simple rules; choice is often more related to considerations about P fr than to medical requirements. Sometimes, experts change their minds. Observations Experts tend to simple rules; choice is not directly related to medical requirements Choice is often more related to considerations about P fr than to medical requirements. Sometimes, experts change their minds. Special topic: the TEa problematic
13
Statistics & graphics for the laboratory 66 Special topic: the TEa problematic TEa and IQC Summary Different models give different TEa values Additionally: TEa may depend on concentration/disease Many proposals do not account for the needs of IQC Some values that are recommended are beyond the state-of-the-art of routine methods, or give IQC rules with high P fr Experts tend to simple rules; choice is often more related to considerations about P fr than to medical requirements. Sometimes, experts change their minds. Don’t choose the most convenient TEa. The most useful general purpose TEa numbers result from the concept of biological variation. –But, apply bottom-line values Checklist – TEa values Apply TEa values from the following hierarchy 1. Clinical models (e.g., cholesterol; glucose) 2. Biological variation (obtain the database) Bottom-line values No numbers from 1-2: 3. Expert models No numbers from 1-3: 4. Regulation No numbers from 1-4: 5. Better state-of-the-art Note: Critically review the proposed numbers.
14
Statistics & graphics for the laboratory 67 Control rules based on TEa Automatic selection of rules based on TEa: The Validator OPSpecs®-Charts Critical error graphs Selection of a control rule based on TEa with the Validator: an example Other selection tools: the IQC decision tool; the TEa/CV a,tot ratio; Summary and Conclusion
15
Statistics & graphics for the laboratory 68 Control rules based on TEa Selection of control rules based on TEa Tools for IQC rule decision based on TEa The Westgard software tools (OPSpecs® chart) The STT IQC decision tool The ratio of TEa/total-CVa: should be >4 Automatic selection of rules based on TEa: The Validator® Based on the TEa concept and the calculation of critical errors, the Validator software allows an automatic selection of the most appropriate control rule. OPSpecs-charts The Validator makes use of so-called OPSpecs-charts (“Charts of Operational Process Specifications”) to relate TEa and stable method performance (bias and imprecision) with IQC-rule selection. Based on a preselected TEa, these charts allow to select IQC rules for different combinations of bias/imprecision (see Figure). The charts are specific for a specific total error requirement, for example, a chart for a TEa of 5% will differ from one requiring 2%. A selection between QC procedures for systematic error or random error has to be made and the probability level for error detection has to be specified, for example, 90% (80% for RE), 50%, etc (OPSpecs®-Charts for RE not in EZ rules!). Imprecision is plotted on the x-axis and bias is plotted on the y-axis. Stable imprecision and bias are plotted as operating point (Figure: 2%/0%). A selection of IQC-rules is presented.
16
Statistics & graphics for the laboratory 69 Control rules based on TEa Critical error graphs The Validator® Additionally, it shows the power functions of the rules with indication of the critical error (= critical error graphs). Critical error graphs are made for systematic and random error, separately. They show the power functions for the IQC rules that the Validator proposes. Additionally, a box is presented that indicates P fr and P ed of the respective rules. The critical error is indicated by a vertical line. Usually, it should intersect the power functions at P ed values >0.9 (=90% for error detection). Note: for the RE graph, the critical error line intersects most power functions at values <0.9. This indicates the generally weaker power of these rules to detect random error! Oper. Point: RE: 2% SE: 0%
17
Statistics & graphics for the laboratory 70 Control rules based on TEa Selection of a control rule based on TEa EXAMPLE Analyte Cholesterol TEa 13%: From German EQA (RILIBÄK) Assumed stable performance Bias:0% CV a,tot :2% Apply the Validator See OPSpecs-chart The Validator selects a very convenient rule: 1 3.5s. Note: powerful control rules are at the right in the OPSpecs! Conclusion If TEa is high: «Easy» rules are selected Attractivity of the approach WHAT, if TEa is low? Will be discussed later under the topic IQC-POLICY
18
Statistics & graphics for the laboratory 71 The Figure shows a simplified version of an OPSpecs® chart. Construction of the lines: TE = SE + k RE; at RE = 0: SE = TE; all lines start at {RE=0;SE=TE}; they stop at {RE=TE/k, SE=0} According to the Westgard approach, the operating point of your test (= stable performance) should be left of the TE line with k = 4. If the operating points is located on the right, the TE cannot be controlled in 90% of the cases by IQC. REMEMBER TEa (preselected!) = SE + k RE (SE & RE are calculated from TEa) For the stable process: k = 1.65 or 1.96 For the IQC controlled process: k = 4 Graphical comparison of the process at TEa and of your actual, IQC controlled process (see figure above, the two X-es) The Figure shows the room for method instability (bias and imprecision) of a process that is controlled with the Westgard approach. Note, the process at TEa, exceeds the TEa in 5% of the cases. REMARK: TE of the IQC controlled process is ~50% of the stable process! Control rules based on TEa The OPSpecs® chart: a more detailed look
19
Statistics & graphics for the laboratory 72 Control rules based on TEa Remark on the location of the operating point The Figures below show the expected measurement populations (for different k values) for an operating point with a low ratio SE/RE (upper Figure) an operating point with a high ratio SE/RE (lower Figure). Observations For low SE/RE ratios, the IQC controlled process is in a robust situation: additional SE does not move it out of TEa. For high SE/RE ratios, the IQC controlled process is in a labile situation: additional SE quickly moves it out of TEa. CONCLUSION The ratio SE/RE that contributes to a TEa is important!
20
Statistics & graphics for the laboratory 73 As already addressed before, different Validator® versions may use different default k values for the stable process: - Either 1,96: pure RE - Or 1.65: pure SE Usually, both components are present. In practice, there is a gradual move from 1.96 to 1.65 when SE moves from 0 to SE = RE. If SE > RE, the multiplier 1.65 is justified. This has also implications for the IQC controlled process. For this reason, the STT variant was developed. IQC decision tool – the STT variant The variant is based on the OPSpecs® charts. From RE = 0, the lines with k = 1.65 (k = 4) are used up to SE = RE. However, from that point on, a line is drawn to the RE point that corresponds to TE/2 (TE/4.85). Control rules based on TEa Other selection tools Problem of the OPSpecs® charts When RE becomes > SE, one moves from the 1-sided case (k = 1,65; k = 4) to the 2-sided case (k = 2; k = 4,85)
21
Statistics & graphics for the laboratory 74 Control rules based on TEa The ratio of TEa/total-CVa (EXCEL-file) The ratio of TEa/total-CVa allows a quick estimate of test performance. This ratio should be >4. However, the ratio depends strongly on the TEa chosen (see also discussion later: special topic TEa problematic). REMARK: bias is not considered by that tool! The EXCEL-file contains a nearly complete list of TEa and state-of-the-art CVa values for the following analyte groups: Ion selective electrode (ISE) Substrates Enzymes Specific proteins Therapeutic drug monitoring (TDM) TEa criteria are: Ricos et al., or 1/6th of the reference (therapeutic) interval CVa data State-of-the-art Summary IQC rule selection can proceed via Statistics TEa and statistics Development of an IQC policy has to consider both possibilities. Conclusion Because regulation and manufacturer recommendations describe minimum IQC, only, the laboratory has to develop an IQC policy itself. The IQC-policy of the laboratory has to respect statistical and medical aspects and should be developed according to the needs of the laboratory, naturally, respecting the guiding rules of regulation. Development of IQC policy Laboratory task
22
Statistics & graphics for the laboratory 75 IQC policy Introduction Software Samples Frequency (& location) of IQC measurements Performance (State-of-the-art) IQC rule selection –Patient release –Process control –Examples IQC rules for state-of-the-art performance –Screening with TEa/CV a,tot –STT IQC decision tool –EZ rules/Validator Special topics –Calculation of an actual TE[a] –Rule “n” and patient release –Dealing with a bias –IQC rules with wide limits (e.g. 6s) and lot variations –“Fine-tuning” of IQC according to instability Remedial actions & –P fr of the IQC rule and frequency of remedial actions –CV a,tot /CV a,w ratio –Inspecting IQC charts Summary
23
Statistics & graphics for the laboratory 76 IQC policy Introduction The IQC policy will be developed on the basis of the Belgian situation: RECALL: Koninklijk Besluit Art. 34. §1. The laboratory director has to organize IQC in all disciplines. §3. IQC consists of several procedures which allow, before the release of patient results, to detect all significant within- or between-day variations: Remedial actions policy/rule selection Art. 35. §1. The frequency of control measurements has to be such that it can guarantee a clinically acceptable imprecision. This frequency depends on the characteristics of the method and/or the instrument: Additional: Praktijkrichtlijn Control rules used for start and for acceptance of a run at least 2 IQC events. Rule selection/frequency §2. The control material, … must be stable within a defined period of time. Different aliquots of the same lot must be homogeneous; Concentrations: Praktijkrichtlijn. Sample requirements §3. For each new lot, the mean and the SD have to be determined. … IQC materials may, at the same time, not be used as calibrator and control material: Establish stable performance: own targets!
24
Statistics & graphics for the laboratory 77 IQC policy Introduction On the basis of the Belgian regulation, we will apply all input elements that we have seen to the analytes ISE; substrates; enzymes, specific proteins; TDM. The elements we have seen (see also “ckecklists”) were: “Knowledge-base” -Statistics -Rule selection -Metrology/Error concept for IQC -TEa data, biological variation State-of-the-art performance data -CV a,tot, CV a,w -Experience from "Peer-IQC" -Laboratory experience (also: Questionnaire to participants) Software tools -Laboratory IQC software -EZ-rules -IQC decision tool/Ratio TEa/CV a,tot From established rules, we look at the magnitude of undetected errors (e.g., in 10% of the cases) Additionally, we will address the topic -Remedial actions. We will develop the policy in an interactive way. We start with the Laboratory IQC software (see also checklist).
25
Statistics & graphics for the laboratory 78 IQC policy IQC-software Discussion ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… …………………………………………………………………………………….………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………
26
Statistics & graphics for the laboratory 79 IQC policy The IQC sample Samples: §2. The control material, … must be stable within a defined period of time. Different aliquots of the same lot must be homogeneous. Concentrations: Praktijkrichtlijn See also checklist – Samples Select commercial one Note: In principle, their target and range cannot be used by the laboratory (Art. 35. §3.) (§3. For each new lot, the mean and the SD have to be determined). Consider participation in an IQC "Peer"-system Sample advantages due to the high number of participants System specific target means and SDs Low target uncertainty Control of sample stability Discussion ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… …………………………………………………………………………………….………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… …………………………………………………………………………………………
27
Statistics & graphics for the laboratory 80 IQC policy Frequency of measurement and location Frequency: Art. 35. §1. The frequency of control measurements has to be such that it can guarantee a clinically acceptable imprecision. This frequency depends on the characteristics of the method and/or the instrument. Praktijkrichtlijn Control rules used for start and for acceptance of a run. Frequency: at least 2 IQC events Location: at least at start and end See also checklist – Frequency and location Discussion ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… …………………………………………………………………………………….………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… …………………
28
Statistics & graphics for the laboratory 81 IQC policy Stable performance §3. For each new lot, the mean and the SD have to be determined. The laboratory has to establish its own stable performance data. REMEMBER The target SD (= stable imprecision) is the cornerstone of IQC. It deserves special attention. All instabilities (random and systematic) are compared relative to the stable imprecision. Recommendation 1 Compare your performance with your colleagues (IQC-"Peer") Advantages due to the high number of participants Better IQC-sample Easier set-up of IQC (more reliable estimates of stable performance) Easier troubleshooting by direct comparison with “peer” Recommendation 2 Compare your performance with the data from the manufacturer! See EXCEL-file for an overview about “state-of-the-art” performance data for the Analyte groups Ion selective electrode (ISE) Substrates/Enzymes Specific proteins Therapeutic drug monitoring (TDM) A list of total and within-run CV data for analyte concentrations in the reference interval (RI) can be found in the EXCEL-file (sorted by biological CVw). See also Checklist “stable” imprecision Note: CV data from concentrations in the RI have been chosen because they will be compared with TEa goals derived from biological variation, which apply to concentrations within the RI.
29
Statistics & graphics for the laboratory 82 IQC policy Stable performance Discussion ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… …………………………………………………………………………………….………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………
30
Statistics & graphics for the laboratory 83 IQC policy IQC rule selection Generally, IQC rule selection can be done on: Statistical basis: A rule is chosen based on P fr and P ed. SD-limits are taken from stable performance. TEa basis: From a specification for TEa, critical error values can be calculated. From the critical error values, adequate IQC rules can be selected, naturally, on statistical basis. Selection tools are: power functions, OPSpecs, critical error graphs, the TEa/CV a,tot ratio; the IQC decision tool See also checklists basic statistics; power of control rules; metrology. Belgian regulation, however … §3. IQC consists of several procedures which allow, before the release of patient results, to detect all significant within- or between-day variations Art. 35. §1. … guarantee clinically acceptable imprecision? What is significant; what is clinically acceptable? We need a goal for TEa! … requires, in principle, IQC rule selection on the TEa basis. see also TEa checklist
31
Statistics & graphics for the laboratory 84 Selection of TEa Discussion ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… …………………………………………………………………………………….………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… …………………………………………………………………………………… IQC policy
32
Statistics & graphics for the laboratory 85 IQC policy Example Selection of the TEa We select TEa from biology TEa from the Ricos concept, and The most stringent TEa, with bottom-line We apply TEa to manufacturer’s performance Example Serum Glucose Ricos TEa: 6.3% CV a,tot : 2% Bias: 0% Exercise STT IQC decision tool EZ Rules selection: NONE with 90%: see OPSpec (note TEa/CV a,tot = 3.2) Conclusion from the example: It did not work! We need a “modified TEa approach”
33
Statistics & graphics for the laboratory 86 Apply the TEa approach (TEa/CV a,tot ratio; the Validator®/EZ rules; IQC decision tool) as evaluation tool for current quality. Analytes that cannot fulfill the TE specifications Decide on the most stringent rule you want to apply (P fr ; n) Analytes that fulfill TE very easy Decide on the most loose rule you want to apply All others: decide whether you want to use individually optimized rules: –Different rules –Same rule, but movement –Adapt the frequency (less/more measurements) Problem Optimization always should consider individual test stability! Example 1: most stringent rule (left) Westgard multirule: 1 3s / 2 2s / R 4s / 4 1s n = 4 (required for full power) P fr : 3% Relatively high! P ed for a 2.5s-shift ~90% P ed for 3 RE ~90% Example: the most loose rule (right) 1 3.5s n = 4 P fr : 0,2%; P ed for a 2.5s-shift ~50% P ed for 3 RE ~70% NOTE For what concerns n, the rule can be applied flexible (e.g., n = 1). IQC policy The modified TEa approach Mean 3.5 s
34
Statistics & graphics for the laboratory 87 Example 2 for the most stringent rule (left) Westgard multirule: 1 3s / 2 2s n = 4 P fr : 1% P ed for a 2.5s-shift ~83% P ed for 3 RE ~83% Example for the most loose rule (right) 1 3.5s n = 4 P fr : 0,2%; P ed for a 2.5s-shift ~50% P ed for 3 RE ~70% NOTE For what concerns n, the rule can be applied flexible (e.g., n = 1). Note (most stringent rule): The Westgard multirules have been chosen because of their widespread availability in IQC software programmes. When possible, choose others: Think of control rules with variable limits, or mean and variance rules! We will go further with The modified TEa approach IQC policy The modified TEa approach 1 3s /2 2s 3.5 s 1 3s /2 2s 3.5 s
35
Statistics & graphics for the laboratory 88 IQC policy IQC rule selection – Process control Basic idea Independent of using IQC for patient release, we may want to use it for process control! For process control, each test could be controlled by the same IQC procedure. For process control, higher P fr values can be accepted. However, P ed should be considerably higher than for patient release. Examples Westgard multirules with high n P ed (90%) ~1.8 SE (see Figure); for the logic of Westgard multirules: lower Figure Mean and variance rules with high n Logic scheme for applying the Westgard multirule X Control Data 1 2s 1 3s 2 2s R 4s 4 1s 10 In-controlAccept run Out-of-controlReject run No Yes No Yes No 1 2s /1 3s /2 2s /R 4s /4 1s /10 x(mean) Pfr: ~3%, Ped for a 2.5s shift: 90% (for n = 4, without the mean rule)
36
Statistics & graphics for the laboratory 89 IQC policy Rule selection Discussion ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… …………………………………………………………………………………….………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………
37
Statistics & graphics for the laboratory 90 IQC policy Selection of IQC rules – Tools & Examples Tools The ratio of TEa/total-CVa: should be 4 The STT IQC decision tool The Westgard software tools (Validator/EZ Rules) EXERCISES Sodium We apply the modified TEa approach, the “most stringent TE” (with bottom-line), and the state-of-the-art CV data 1. Select TEa: Bottom-line: 3.2% 2. Apply the IQC decision tool - Set stable Bias = 0 - Stable CV (total) = 1% 3. “Sort of analyte?” 4. Apply Validator® for individual rule selection OUTCOME The IQC tool and the Validator® show that sodium, with the values chosen, cannot be controlled according to the TEa concept. The ratio of TEa/total-CVa is 3.2. Triglycerides Strategy as before (see Figures on the right) TEa: 13.8 (most stringent) Bias: 0% CV a,tot : 1.7% Ratio TEa/CV a,tot : 8.1 OUTCOME The IQC tool and the Validator® show that, even with the most stringent TEa, triglycerides can be controlled easily with the TEa concept: The Validator® selects the 1 3.5s rule (n = 2) NOTE In the EXCEL-file, you find for ISE, substrates, enzymes, specific proteins, TDM a list of CVw, CVb TE for IQC decision - Ricos et al. State-of-the-art CV data
38
Statistics & graphics for the laboratory 91 IQC policy IQC rule selection – Applications First, we screen the analytes with the Ricos-TEa/CV a,tot ratio The tables show analytes with ratios <4 (full list in EXCEL-file). In principle, the most stringent IQC rule should be applied for all these analytes. Example: 1 3s /2 2s (n = 4) Note on low TEa/CV a,tot ratios Low TEa/CV a,tot ratios are usually associated with analytes that have narrow biological variation. Current state-of-the-art technology is unable to reach the goals for these analytes. This holds true for all companies. TEa/CV a,tot ratio
39
Statistics & graphics for the laboratory 92 IQC policy IQC rule selection – Applications The tables below show analytes with ratios 4 (full list in EXCEL-file). In principle, individually optimized IQC rules can be applied for all of these analytes (note: most loose = 1 3.5s ). Now, we will apply the STT IQC decision tool for some analytes with ratios from 4-7. APO A1 (4,3):TEa: 9.1CV a,tot : 2.1 -1-Antitrypsin (5,1)TEa: 9.2CV a,tot : 1.8 IgA (6,8)TEa: 13.5CV a,tot : 2.0 We observe: from ratios >~5-6, IQC should be relatively easy! We verify this with the EZ rules (note: fill in all fields; if appropriate, use dummy values). TEa/CV a,tot ratio
40
Statistics & graphics for the laboratory 93 IQC policy IQC rule selection – Applications Excercises with the EZ rules, using the specific protein examples (ratios from ~4-7). Note: all fields must be filled in (may be dummy values)! Selection criteria: 2 materials, SE detection 90% APO A1 (4.3)Rule: 1 2.5s n=4 P fr : 4% -1-Antitrypsin (5.1)Rule: 1 3s / 2 2s n=2 P fr : 1% IgA (6.8)Rule: 1 3.5s n=2 P fr : <0.1% Example (6)Rule: 1 3.5s n=2 P fr : <0.1% Observation: from ratios >6, the process can be controlled very easy! 6 sigma process We take the 1 3.5s (n=4) rule for all analytes with a TEa/CV a,tot ratio 6 TEa/CV a,tot ratio
41
Statistics & graphics for the laboratory 94 IQC policy IQC rule selection – Applications Screening with the Ricos-TEa/CV a,tot ratio: all 4 - <6 The table shows examples where optimized rules could be used. However, Consider: Is it worth? Should one use the most stringent IQC rule? REMARK Remember process control. One stringent rule could be used for all analytes. TEa/CV a,tot ratio
42
Statistics & graphics for the laboratory 95 IQC policy VARIA Calculation of an actual TE Rule “n” and patient release Dealing with a bias Rules with wide limits (e.g. 6s) & lot variations “Fine-tuning” of IQC according to instability Calculation of an actual TE When actual performance does not satisfy the TEa requirement, and one selects the most stringent rule, one can calculate (or extract with EZ rules) the real TE for the actual performance and the rule chosen. Examples, IQC rule 1 3s /2 2s (n = 4) CVaRicos TEaReal TE (%)(%)(%) Chloride:1.01.5~4.3 Albumin:1.23.9~5.2 Calcium:2.02.4~8.6 Note: with this rule (without bias), actual TE = 4.3 CVa Rule “n” and patient release When IQC rules with higher number of n are selected, and IQC measurements are done continuous, patient release has to be postponed until the full power of the rule is reached (cumulation needed). However, at the end of the day, cumulation is not possible anymore. Consequently, it may be required to increase the frequency of measurement per IQC event (1 = usual, 2 = last but two, 3 = last but one, 4 = final event). This is not necessary when block measurement is performed. Dealing with a bias It is not recommended to work with the bias in the QC chart: Increases P fr Bias should be addressed during rule selection via TEa. A bias will require more stringent IQC rules.
43
Statistics & graphics for the laboratory 96 IQC policy VARIA IQC rules with wide limits (e.g. 6s) and lot variations When rules with wide limits are used, significant lot-to-lot variations may be recognized in the chart, but may not cause rule violations (see Figure below). Decide about the medical relevance of the variations whether you want to pick them up by the process control whether you address them with quality assurance means. Note: Similar may happen when you overestimated CV a (for example, high CV a,tot /CV a,w ratios). “Fine-tuning” of IQC according to type of instability Steadily increasing Abrupt changes Fluctuating Intermittent (sporadic) IQC is strong in detecting steadily increasing instability or abrupt changes. This, in particular, when patient release can be postponed until the IQC procedure reaches its full power. Namely, after error detection and correction, patient samples can be remeasured. IQC is weak in detecting fluctuating or intermittent (sporadic) instability. This the more, the less IQC measurements are performed. “Fine-tuning” of IQC according to instability requires in-depth knowledge about the weak points of the analytical process (error frequency; expected error types). “Fine tuning” may consider: IQC sample concentration Rule selection Frequency of measurement Location (block or continuous)
44
Statistics & graphics for the laboratory 97 IQC policy Varia Discussion ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… …………………………………………………………………………………….………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………
45
Statistics & graphics for the laboratory 98 IQC policy Remedial actions EXCERCISES Inspecting IQC charts Note in advance: «When results of controls fail to meet the laboratory's established criteria for acceptability, all patient test results obtained in the unacceptable run must be evaluated to determine if patient results have been adversely affected and the laboratory must take remedial actions to ensure the reporting of accurate and reliable patient test results» “Westgard philosophy” (Basic QC Practices) Don’t simply rerun the control Don’t simply “try a new bottle” Identify the problem (inspect the chart) and correct it Which type of error (concentration dependent, SE, RE) Relate the error to a potential cause Consider common factors (temperature; pipetting: volumes; kind of test: kinetic/endpoint; wavelength) Relate to recent changes (operator, calibration, reagent, IQC-material, maintenance, etc.) Demonstrate and document that the error was fixed Run controls after fixing the problem Acceptability limits Duplicates (Note: s = “within day”) z 2 s, for 95%: 1.96 2 s : 2.77 s New control samples, recalibration For n measurements, decrease the limit with a factor of 1/ n Case n = 3, limit = 2.5s, new limit = 2.5 [1/ 3] = 1.44, Because of practical considerations one may choose 1.5s New batches (immunoassays), Control of target values Basic formula: xt s n (,) Example: n = 10, 95%-level: ±0.715 s, simplified:±0.7 s
46
Statistics & graphics for the laboratory 99 IQC policy P fr of the IQC rule and frequency of remedial actions When you compare your actual short-to-medium term frequency of remedial actions with the P fr you expect from the IQC rule, consider the nature of the CV a,tot you chose. The more variations you included, the lower the frequency of remedial actions will be. If actual remedial actions frequency << rule P fr Consider to pick up certain variations by IQC Consequence: reduces your CV a,tot that you use for IQC May allow a different rule when the TEa concept was used. Consider that you used a wrong CV a,tot The “state-of-the-art” example We have ~40 analytes with the rule 1 3s /2 2s, P fr : 1% (n = 4), ~30 analytes with the rule 1 3.5s, P fr : 0,2% (n = 4), and assume 8 IQC measurements per day. Then, we expect ~80% chance for a false rejection Under stable conditions, we expect at maximum 1 remedial action per day, More, if system is unstable Less, if Cv a,tot > CV a short-to-medium term. CV a,tot /CV a,w ratio Note: Troubleshooting is facilitated by increased CV a,tot /CV a,w ratios.
47
Statistics & graphics for the laboratory 100 IQC policy Remedial actions Note: Either the measurement procedure or the control procedure can be faulty. Release of patient results You suspect the QC sample was the reason for failure Take a new bottle. Remeasure with n = 4, for example. If the IQC rule is not violated, assume the IQC sample caused the problem. However, realize the uncertainty of your estimate! Keep an eye on the process for a while. You suspect a calibration problem Recalibrate the system, however, document the post-calibration status by a sufficient number of control measurements (for example: 4). The system is under control again when the IQC rule is not violated. However, realize the uncertainty of your estimate! Remeasure the patient samples between the last in-control event and the out-of- control event. Compare the mean of the new patient results with that of the old results. When there was a calibration problem, this should be reflected in the difference of both means. You suspect other problems Look at patient data Investigate all IQC levels (= IQC across materials), check other analytes (= IQC across analytes). Investigate related analytes (e.g., all enzymes: could be temperature related). Compare the actual CV with the expected one, and compare the IQC results with appropriate IQC-"Peer" (EQA) data. Look at “electronic” QC data. Check instrument (maintenance), reagents, operator change. Locate and fix the problem and document that the problem was solved (perform IQC). Remeasure the patient samples between the last in-control event and the out-of- control event. Compare the mean of the new patient results with that of the old results. When there was an SE problem, this should be reflected in the difference of both means.
48
Statistics & graphics for the laboratory 101 IQC policy Remedial actions Process control Rules with higher P fr (e.g. 2.5s, Westgard multirules, or others) can be used for closely monitoring the analytical process. In case that these rules are violated, patient results can still be reported. However, when time, the analytical process (calibration, reagent, instrument) or the IQC material should be investigated. Possibilities Look at patient data Check “electronic” QC data Investigate all IQC levels (= IQC across materials), check other analytes (= IQC across analytes). Invstigate related analytes (e.g., all enzymes: could be temperature related). Compare the actual CV with the expected one, and compare the IQC results with appropriate IQC-"Peer" (EQA) data. When necessary, recalibrate the system and check the success of recalibration as described above. Check instrument (maintenance), reagents, operator change. In general, trouble-shooting should be done with an increased number of control measurements, however, higher CV a,tot /CV a,w ratios facilitate troubleshooting. Over the time, check the success of your remedial actions. CAUTION The described strategy is an extremely simplified one. In practice, problems have a great variety of origins and require different trouble- shooting strategies (“mans phantasy is too limited to imagine beforehand what the problem might be”). Only an experienced analyst will be able to solve serious problems in a reasonable time. System experience is invaluable!
49
Statistics & graphics for the laboratory 102 IQC policy Remedial actions Discussion ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… …………………………………………………………………………………….………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………
50
Statistics & graphics for the laboratory 103 IQC policy Internal Quality Control General discussion ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… …………………………………………………………………………………….………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………………… ……………………………………………………………………………………
51
Statistics & graphics for the laboratory 104 IQC policy Reminders Reminder 1 Troubleshooting should be done with replicates! Reminder 2 Don’t forget proactive quality assurance A new lot (bottle) of IQC material is introduced The operator changed Instrument was calibrated or underwent major maintenance New lots (bottles) of reagents or calibrators Overall summary The guiding rule (regulation) Knowledge –Basic statistics –Power functions –TE error concept (metrology) –TEa (critical errors, specifications) –Selection of TEa values (and problems) IQC Software –Additional: rule selection tools Adequate samples (sort, targets, concentration) Adequate frequency (and location) policy –Cost/benefit calculations Instrument stability data & test stability & reproducibility data –Stable performance: Compared to manufacturer/peers Integrated rule selection (TEa; statistics; costs) –Distinguish: Release of patient data/Process control Remedial actions policy –Distinguish: Release of patient data/Process control –Compare expected P fr with actual rejection rate –Investigate success rate Integration of IQC in the overall quality management Dedicated personnel