1 DICOM WG21 MULTI-ENERGY IMAGING SUPPLEMENT OVERVIEW 16-June-2015

2 Agenda Definitions, Use Cases, Objectives Aspects of Multi-Energy (ME) technologies New types of ME images Proposed Approach Risk and Concerns

3 ME-Definition Imaging techniques, including scanning, reconstruction, processing, when the scanner utilizes multiple energies from the X-Ray beam spectrum, as opposed to the conventional CT imaging, when a single (accumulated) X- Ray spectrum is used. The existing CT and Enhanced CT (eCT) IODs do not adequately describe the new CT multi-energy imaging. Although different vendors apply different scanning and detection techniques to achieve multi-energy images, there is large commonality in the generated diagnostic images.

4 Examples of Multi-Energy CT

5 Physics background

6 Clinical Use Cases The primary potential applications this supplement intends to focus on include: Allowing better differentiation of materials that look similar on conventional CT images, e.g., to differentiate Iodine and Calcium in vascular structures Eliminating acquisitions such as non-contrast acquisition, when the “virtual/artificial non-contrast” image is generated from the contrast image

7 Objectives When defining this supplement, the following objectives / goals have been considered: 1. Making multi-energy information (acquisition, reconstruction and processing attributes) available to rendering or processing applications 2. Facilitating fast and easy adoption of this supplement across the imaging community, both modalities and PACS/Displays. 3. Eliminating (or at least minimizing) the risk of mis- interpretation when the ME images are displayed by a non-compliant display, including incorrect measurements

8 New aspects of ME technologies Different vendors apply different technologies for: Scanning Detection Reconstruction Processing Material decomposition This results in a variety of image types calling for standardization of parameters and definitions

9 Virtual Mono- chromatic Image (VMI) Material- Specific Image Material- Subtracted Image Color Overlay Image Color Blending Image Discrete Labeling Image Proportiona l Map Image Iodine Map; Bone Density Virt. Non-C; Virt. Non-Ca Effective AN (Z) Image Electron Density Image Probability Map Image Color Map Image Multi-Energy Imaging Material Quantification (Decomposition) Material Classification (Labeling) Material Visualization (Color) Material- Modified Image Highlighted; Partially- Suppressed Color Eff. AN Gout crystals on top of CT image 4 categories many flavors

10 Proposed approach For VMI and Quantification we reuse basic and enhanced CT IOD‘s for ME images For Classification we consider CT IOD, Segmentation IOD or Parametric Maps IOD For Visualization we consider reusing SC, Presentation States, Blending and may be Parametric map or adding RGB palette information to CT IOD Include the new attributes (add macros) in the existing IOD‘s Add new definitions to the existing standard (like Image Type, defined terms) Adapt some descriptions to multi-energy related cases  Close to current implementations. Better chance to be widely/fast adopted

11 Virtual Mono-chromatic Image (VMI) 40 keV167 keV Essentially analogous to a CT image that would be generated by a monochromatic (of a specific keV value) X-Ray beam

12 Material-Specific Image Iodine Map Image presenting a physical scale of specific material. Pixel values can be in HU or in equivalent material concentration (e.g., mg/ml).

13 Material-Subtracted Image Virtual Non-Contrast Image Image with one or more materials subtracted. Pixel values may have been corrected for displacement of one material by another material.

14 Material-Modified Image Image where pixel values have been modified to highlight a certain target material (either by partially suppressing the background or by enhancing the target material), or to partially suppress the target material. Unlike Material-Specific and Material- Subtracted images (that can allow accurate measurements), the Material-Modified image is primary used for better visualization of the target materials Bone Marrow Image

15 Electron Density Image An image where each pixel represents a number of electrons per unit volume. Widely used in radiotherapy.

16 Effective AN (Z) Image An image where each pixel represents Effective Atomic Number (aka “Effective Z”) of that pixel.

17 Probability Map Image An image where pixels describe the probability that this pixel is classified as one or more of the multiple defined materials Gout Material Map

18 Material Visualization (Color) Users are asking to visualize ME-content in certain ways: color maps, color overlays, blending, etc.

19 How we intend to extend C.8.2.1 CT Image Module CT Additional X-Ray Source Sequence (0018,9360) (modified) Table X-1 “Optional CT Multi-Energy Macro Attributes" (added) CT Multi Energy Acquisition Sequence (1C) >Table X-2 “CT Multi-Energy Acquisition Macro Attributes” CT Multi Energy Material Decomposition Sequence (1C) >Table X-3 “CT Multi-Energy Material Decomposition Macro Attributes” CT Multi Energy Image Sequence (1C) >Table X-4 “CT Multi-Energy Image Macro Attributes” C.8.15.3.9 CT X-Ray Details Macro (for Enhanced CT IOD only) CT X-Ray Details Sequence

20 Risk and Concerns Enhancements can be mis-interpreted in keV image  need for correct display label including keV image type and keV value 100 kV (conventional)50keV (virtual monochomatic)

21 Risk and Concerns High keV (>150) images can be mis-interpreted that no contrast was applied (missing contrast information) Potential risk for VNC images (e.g. changing size of small structures) – currently not broadly validated that VNC is truly equivalent to TrueNonContrast Quantitative measurements are different in non-contrast vs. contrast use case (e.g. renal mass and renal cyst) Missing labeling can cause confusion which can slow down the workflow  delay the diagnosis

22 Questions To guarantee that the created ME images contain the necessary information, a number of new attributes will be defined. These attributes will be added to the ME images, either as an extension to the existing IODs, or as part of new, ME-specific, IODs. However, we cannot mandate using any new specific attribute for existing IOD. Is there any workaround: Can we say that specific attribute is Type 1 if the system supports ME imaging? Yes only for new attributes; What about standard attributes, like Image Type? Ok to extend with new values and make them mandatory if the image is ME? Can we say that specific attribute is Type 1C if the units for pixels are not in HU? Not for existing attributes; ok for new Shall we rather define an optional SQ with mandatory ME attributes? Image Type of VMI Images. It is defined use ORIGINAL unless there is a specific case requiring it to be DERIVED. WG6 recommends leaving it to the vendor to decide if the image is ORIGINAL or DERIVED.

23 Questions To add Real-World Value Mapping to CT IOD to accurately describe the non-HU values. There are several concerns with this approach: RWVM is not specified today for CT images or widely implemented in the field (is this correct?). As a result, the units can be misinterpreted by the display application Rescale Slope and Rescale Intercept are Type-1 for CT image (if it is ORIGINAL); there is potential conflict between rescale attributes and RWVM. If we define image type as DERIVED, we avoid Type-1 requirement for Rescale Slope/Intercept, so they can be omitted. As a consequence, Rescale attributes shall be used if possible; for specific cases (to be identified) RWVM may be considered.

24 Questions To describe the need and recommendations for good labeling in the informative section (e.g., to display keV value for VMI images; to display material + concentration for measurements, etc.). DICOM alone cannot enforce PACS/Workstation to present specific attributes therefore there is little chance new important attributes we introduce here will be presented to the users. Should we work with IHE? Can we “orchestrate” the object in such a way that naïve display will either present the image adequately or fail to present anything? For instance, we can put rescale attributes to zero for non-HU ME images

25 END OF PRESENTATION

26 Virtual Mono-chromatic Image (VMI) 43 HU (160 keV) 307 HU (50 keV) 31 HU (≈ 120kV) - = Contrast VMI Images 12 HU 206 HU Renal Mass use case Non-Contrast Image

27 DETAILS OF IMPLEMENTATION

28 ME Acquisition Techniques X-Ray Sources SINGLE_SOURCE MULTI_SOURCE KV Switching NONE, FAST, SLOW Multi-Energy Acquisition SINGLE_SCAN MULTI_SCAN Multi-Energy Detection CONVENTIONAL MULTILAYER PHOTON_COUNTING Technique-Specific Parameters: What do we want to record for Dual-Layer and PhC?

29 ME Material Decomposition Decomposition Method SINOGRAM_BASED IMAGE_BASED Decomposition Base Materials (sequence) Decomposition Description Vendor-specific label/description Material Attenuation Curves (opt) Decomposition Parameters e.g., dual-energy ratio

30 Pixel Value Units Method 1: using Rescale Rescale Slope Rescale Intercept Rescale Type Measurement Units as standard Coded Values e.g., UCUM: "mg/cm^3“ Method 2 – using Real World Value Mapping Sequence First/Last Values Mapped Value Slope/Intercept LUT (optional) Measurement Units (Coded Value)

31 Material Classification (Labeling) 2 0: Unknown 1: Material A 2: Material B 3: Material C Discrete Labeling (most-probable material): A B C D 1.5 Proportional/Density Map: % or mg/ml 0.2 Probability/Confidence Map:

32 Spectral Imaging Challenges Monochromatic Images Capture keV value How to differentiate from “legacy” CT? Incl. query, display annotations Material Density Images Non-HU values: how to avoid confusion? Several alternatives for solution Segmentation IOD Real-World Mapping PTE-like new IOD Rely on specific Rescale Intercept/Slope/Type Effective Atomic Number Images Similar challenges as for Material Density Using Color mapping prohibits measurements and analysis

33 Open Issues (some) Assess risks ME images been misinterpreted as the conventional ones on a PACS/Workstation AI: work through Mark Armstrong (ACR) to get radiologist to enumerate the risks How shall we model KV Switching? (including duration and gaps, proportion of High/Low KV) Alternative 1: Describe as two different sources Alternative 2: Single dynamic source KV-Switching specific parameters How shall we “model” Photon Counting detector? Shall we better describe the full “Data Path”? Too complicated and vendor-specific? Each vendor to provide a list of potential public attributes specific for the each ME Acquisition/Recon techniques