1 HF Laser/LED Calibration Studies L. Almeida, M. Baarmand, L. Caraway, M. Hohlmann, T. Qureshi, I. Vodopianov FLORIDA TECH 1->9 light splitter w/ pulsed.

Similar presentations


Presentation on theme: "1 HF Laser/LED Calibration Studies L. Almeida, M. Baarmand, L. Caraway, M. Hohlmann, T. Qureshi, I. Vodopianov FLORIDA TECH 1->9 light splitter w/ pulsed."— Presentation transcript:

1 1 HF Laser/LED Calibration Studies L. Almeida, M. Baarmand, L. Caraway, M. Hohlmann, T. Qureshi, I. Vodopianov FLORIDA TECH 1->9 light splitter w/ pulsed LEDs, input for UV laser pulses, and PIN diode light monitoring 1->4 light splitter Electronic pulser for ~100 ns LED pulses w/ single-shot and external trigger capability 10 m jacketed optical fiber

2 2 N 2 Laser 300  QQ 75m CLI ROBox 3 CALIB Box 1  4 Counting Room HF Calibration Optical Path Test Beam 2002 200  J 337 nm 9 600  QP 18m 300  QQ 9m

3 3

4 4 LED 2 MIXER 50mm  15mm  15mm Polystyrene PIN3 ADC 9 600  QP 300  QQ Blue Scintilator 8mm  4mm  8mm Calibration Box - 1 x 9 splitter To CLI in PMT Box PIN Diode 4mm  4mm Pulser PIN1 PIN2 LED 1

5 5 Calibration Light Injector (CLI) Empty PMT TUBE Light Guide Non-reflective coating Light Guide 600  QP PMT BOX 5 120 SHIELDING 600  QP 40cm Light Guide PMT Some PMTs have more or fewer than 5 fibers !!!

6 6 Calibration Light Injector (CLI)

7 7 SMA Connector Fiber 600  QP Jacket SMA Connectors Holographic Diffuser Plate 2Plate 3 SMA-SMA connector Fiber Ferrule Calibration Light Injector in R O Box 3 Calibration Tube 9cm Aspheric Lens 11mm Focal Length Collimation Assembly ±4% Non uniformity

8 8 HF Laser/LED Calibration Issues Test Beam 2002 Insufficient light intensity –low laser output power  laser used at test beam misaligned or badly focused? –attenuation of UV in fibers  move laser downstairs / use DYE (blue) laser –small efficiency in 1  4 splitter, CALIB Box, CLI  improvements under investigation – see below –low LED light  use brighter LEDs and drive at higher bias voltage? Non-uniformity as seen by PMTs (~20%) –mostly due to light guides and calibration fibers downstream of CLI

9 9 PMT 154 8-10 fibers!!! R O Box 3

10 10 N 2 Laser 75m/150m/20m 300  QQ CALIB Box 1  4/8 Counting Room 300  J 337 nm 0m/20m/20m 300  QQ Attenuation of UV in 300  QQ Atten. Length = 60m length (m)attenuation Test beam 75 + 0 3.5 CMS UV Laser upstairs 150 + 20 17 CMS UV Laser downstairs 20 + 20 2 CMS DYE Laser upstairs 150 + 20 2 without splitter attenuation

11 11 300  QQ  QQ (Silica) 300  QQ SMA connectors Splitter 1 × 4 (1 × 8) efficiency 75 mm cross section ratio  11 (9%) Lab, red/blue 4.5% Test beam, UV 2.5% 1 × 8 (calc. UV) 1.3%

12 12 laser Mixer 50  15  15 mm 3 PIN 10 × 600  fiber bundle Ø3mm (7mm 2 ) Light collection: 0.1% by bundle, 0.004% by each fiber Scintillator 8  4  8 mm 3 ε = 10% Total efficiency = 4  10  6 per output fiber Present Calibration Box ) 22° 10 × 600  fiber bundle Ø3mm (7mm 2 )

13 13 20  7  4 mm 3 Max light collected within 22° = 3.6%  max efficiency for this mixer+bundle = 1.4  10  5 per output fiber LED ~10 times better efficiency ~ 4  10  5 Calibration Box with smaller mixer to be tested laser PIN 10 × 600  fiber bundle Ø3mm (7mm 2 )

14 14 SMA Connector Holographic Diffuser Present C L I → Calibration Bundle Light guide Lens (F=11mm Ø=6mm) Efficiency = 1.9 × 10  4 per PMT CALIB Bundle 120 × 600  lens  = 16% 5 fibers  = 0.12% with non-uniformity of 4%

15 15 Modified Calibration Light Injector (to be tested) SMA fiber 60°diffuser 10°diffusers Lens Ø = 16mm F = 20mm current Estimated efficiency ≈ 1.7 × 10  3 per PMT ~ 10 times better !

16 16 Results from 2002 test beam Estimated total efficiency = 2.2  10  10 –attenuation of UV in fibers = 3.5 –CALIB Box efficiency = 4  10  6 –CLI → bundle efficiency = 1.9  10  4 Attenuation (bundle → PMT) ≈ 2 * Hamamatsu: QE(PMT) = 27% Test beam: (laser → PMT) = 74 PE (wish 750 PE!)  2.5  10 12 laser photons 200  J is 3.3  10 14 photons (~100 times higher!) * Obtained by combining LEDs power output and test beam measurements – in agreement with measured light guide attenuation

17 17 Outlook (laser calibration) Improvements: –UV laser downstairs or DYE laser   1.8 –Better CALIB Box efficiency   10 –Better CLI efficiency   10 Efficiency of 1  8 splitter = 1.3%  overall efficiency increased by ~ 2 Possible laser power recovered up to ~10 times (to be confirmed with measurements)

18 18 Outlook (LED calibration) LED subsystem is independent of splitter efficiency and UV attenuation Improvements: –use more powerful LEDs assume this to be compensated by gate 100ns → 25ns –improve CALIB Box and CLI as above  overall efficiency increased by ~ 100