Presentation is loading. Please wait.

Presentation is loading. Please wait.

OnSite Rapid Test Procedure Training

Similar presentations


Presentation on theme: "OnSite Rapid Test Procedure Training"— Presentation transcript:

1 OnSite Rapid Test Procedure Training
Part 4 CTK-MK-PPT-R0001 (4)

2 Disclaimer This is a general training presentation based on the OnSite Rapid Test procedure. Trainees should take precautions when performing a specific assay, and should strictly follow the procedure provided by the test kit.

3 Training Contents Introduction Perform Assay Before, During, After 1
2 Perform Assay Before, During, After 3 Assaying with Accuracy 4 Troubleshooting 5 Product Evaluation Additional Information Storage Temperature & Shelf Life 6

4 4 Tro uble sho otin g

5 Limitation of Rapid Test
The Assay Procedure and the Interpretation of Assay Result sections in package insert provided with kits must be followed strictly. The test band intensity does not have a linear correlation with the analyte titer in the specimen. A negative result indicates absence of detectable analyte, which does not preclude the possible of exposure or infection. Is a screening test, results should only be interpreted in conjunction with: Clinical findings Confirmatory test

6 Limitation of Rapid Test
Is a qualitative or semi-quantitative device Has detection limit Analyte has genetic variation or strain variation limitation Example Troponin I Detects 0.5 ng/mL tropnoin I PSA Detects 4 ng/mL and 10 ng/mL PSA FOB Detects 25 ng/ml or 50 ng/mL hHb Variation Example Isotype or serotype variation Dengue has 4 serotypes. A good test has to be able to detect all of the serotypes Genetic variation Few strains of P.f lack of HRP-II protein, leading to false-negative results on the P.f HRP-II antigen test

7 Limitation of Rapid Test
False Negative It means a test result indicates no presence of analyte (the result is negative), when compared to reference test. Causes Explanation Limit of detection (LOD) Analyte in the specimen is below the LOD Isotype or serotype variation Different Isotype and serotype of analyte in the specimen can not be detected Timing to sample Analyte is not present during very early or very late stage of disease Hook effect Analyte concentration is too high; for example, 10 mg/mL hemoglobin in fecal specimen masks the binding of anti-Hb Ab, leads to false negative result Interference by other substances Anticoagulants, or medicines present in the sample can inhibit the reaction, leads to false negative

8 Limitation of Rapid Test
False Positive It means a test result indicates presence of analyte (the result is positive), when compared to reference test. Causes Explanation Cross reaction A reaction which occurs when surface antigenic determinants on different molecules of quite different sources are identical, so that antibody directed against one antigen also reacts with another. Interference by other substances Some specimens containing unusually high titers of heterophile antibodies, rheumatoid factor (RF), HAMA or ANA may affect expected results Medicine or chemicals may affect expected results. For example, in Filariasis, if the patient would have taken DEC or albendazole, IgG4 level will increase and it may interfere with the test Read result exceeds the defined time window Increase of reaction time leads to false positive

9 Troubleshooting Internal control line (C line) Problem
Potential Causes Action No control line or significant decrease of the intensity of control line Damaged device Retest using a new device Improper procedure Follow procedures in the package insert provided Recheck buffer and/or specimen volume Read the result at defined time Device expired or stored improperly Check expiration date of the device Do not use beyond expiration date stated Check storage temperature records T and C line antibody are both from mouse. If the T line is too strong, the conjugate is used up by the T line, so C line is weaker Use different C line system from the T line, such as rabbit antibody control line system, while the T line is mouse antibody system Repeatedly invalid results Defect devices Check device for defective packaging Verify device quality by external control Inform supervisor and CTK

10 Troubleshooting Interpretation of assay result Problem
Potential Causes Action False positive Wrong specimen type Use correct specimen type Specimen contains unusually high titers of lipids, RF, human anti-mouse Ab, anti-nuclear Ab or other inference Spin specimen and get rid of lipid Check for the presence of interference Incorrect volume of specimens or buffer Strictly follow IFU procedure Using components from different kit Don’t use component from different kit Read results exceeded defined in IFU Using distilled water or other inappropriate solution as external control specimen Use clinical specimens or provided control as external control False negative Read result less or exceed defined in IFU Specimen is collected too early Collect specimen few hours, days or weeks late Analyte present below the LOD Use more sensitive detection methods IFU: Instruction for Use

11 Troubleshooting External Control Storage condition Problem
Potential Causes Action External control fails to show accurate results Wrong procedure to reconstitute control Strictly follow control IFU Control is not stored as required Store control properly. Return control to its storage condition immediate after use, if the control is for multiple use Control is expired Use only valid control Storage condition Problem Potential Causes Action Storage temperature exceeds recommended range Storage does not have A/C system to control temperature Run external control to verify kit quality Kit is frozen External temperature is too cold during transportation or accidently keep in freezer Bring up the kit into room temperature before testing, Verify the kit qualify with external control

12 Troubleshooting Flow Chart
Unexpected Results Review Procedure and Repeat Assay with a New Device Repeated Unexpected Results Storage & Test Environment Specimen Collection, Extraction & Storage Procedure Limitation of Test Review Manufacture Instruction Test with Alternative Methods External Control Retest with Different Lot/Device Report to CTK

13 Trouble shooting – Follow the Correct Procedure!
More than 50% of complaints received are caused by incorrect procedure Potential reasons Assay is performed by new, untrained lab technician The procedure from other product insert is used The insert from old revision is used The buffer from a different product is used Wrong buffer volume is added Wrong specimen transfer device is used Wrong read time Suggestions When a complaint is received from a lab, immediately: Go through the procedure point by point with the complainant Ask if correct components have been used Therefore 50% of complaints can be solved instantly!

14 Complaint Handling Procedure
Customer Complaints Recorded and Classified the Complaints by Customer Service & QA General Complaints (Not Products Related) Product Complaints Check Procedure Evaluated and Processed by Customer Service Investigated by QA Record, Review, Inspection, Reference Corrective and Preventive Action if Necessary Corrective and Preventive Action if Necessary Customer Service Reply to Customer QA & Product Manager Reply to Customer Reviewed and Closed by QA

15 5 Pro duc t Eval uati on

16 Evaluation of Rapid Test
For sensitivity and Specificity, generally in situation: Increased desire to detect all truly positive patients for treatment or the treatment is not necessary High Sensitivity High Specificity Increased desire to avoid false positives if the illness is rare or treatment is burdensome to the patient

17 Evaluation of Rapid Test Collection, Extraction, Storage, Dilution
Evaluation protocol Specimen Collection, Extraction, Storage, Dilution Compare “Apple to Apple” OnSite Rapid Test Competitor Rapid Test Results Results Evaluation Agree Disagree: resolve discrepant results Compare to 3rd party test Conclusion

18 Evaluation of Rapid Test
Performance on BBI Mixed Titer Performance Panel 08454 RDT OnSite competitor ELISA IgM IgG Positive Negative Step1 Comparison with competitor RDT Step2 Resolve difference By ELISA Conclusion: OnSite RDT is more accurate than competitor’s test, by using ELISA as confirmatory test It could mislead if ELISA test was not used to verify the discrepancy specimens

19 Evaluation of Rapid Test
Specimen selection Should represent the target population A appropriate number of positive and negative samples required Should be collected, extracted and stored appropriately For example: For malaria Abs or Ags test, prefer specimens from population at epidemic area Recommend: For sensitivity, to select at least 20 positive and 2 negative specimens For specificity, to select at least 20 negative and 2 positive specimens Correct specimens handling and storage is critical for reliable evaluation

20 Evaluation of Rapid Test
Proper Specimen dilution Specimen Diluent Don’t (to avoid matrix effect) Serum, plasma, whole blood Diluted with negative specimen Don’t dilute with assay diluent provided in the kit Fecal extraction specimen Diluted with the extraction buffer Don’t dilute with water or other buffer Nasal extraction specimens Genital extraction specimen Diluted with mixed extraction buffer at different ratio defined in assay procedure Don’t dilute with water or individual extraction buffer Urine specimen Diluted with negative urine Use of proper specimen diluents are critical for reliable evaluation.

21 Evaluation of Rapid Test
About discrepancy results A 3rd party test with more sensitive detection technique should be used to verify the discrepancy specimens Check if the specimen contains common interference factors: ELISA, chemiluminescence, or western blot test may be used to verify the discrepancy by two rapid tests RF, HAMA, ANA or others may interfere the results

22 Evaluation of Rapid Test
Reference test Compare “Apple to Apple” - Use the same technology Compare with the market leader Reference kit has high false positive, the test kit becomes “false” negative Reference kit has high false negative, the test kit becomes “false” positive Use of gold standard Gold standard was usually developed many years ago It may be lack of good sensitivity or specificity Reference standard RDTs for evaluation of RDTs; ELISA for evaluation of ELISA Errors of evaluation will arise if the reference test itself does not have good sensitivity and specificity. Commercial standards (such as NIBSC, BBI) are designed for the specific technology (e.g. RDT or ELISA). A standard intended to be used with ELISA may not be suitable with Rapid Test.

23 Standardization Concentration of analyte, such as PSA and Troponin I in specimens is determined with different kits or machines The values from different kits or machines may be different e.g. Bio-Rad Liquicheck Cardiac Markers control level 1) contains 0.25ng/ml Troponin I. The concentration of this control determined by different assays varies between ng/ml When evaluating rapid tests with these specimens, one should take this into consideration regarding the method to be used to determine the value of the analyte. Standardization against an international standard (WHO , NIST) is preferred

24 Common Questions in Product Evaluation
Potential Issues Suggestion T line is faint in comparison with the competitor’s T line Each company uses their own colloid gold formulation , so the gold color varies As long as the line is present, one should be confident in the test result. Can ask 2nd person to verify the weak color line The test result of Rapid Test is different from that of competitor rapid test Specimen size is too small to represent the true product performance. Test with at least 20 specimens, ideally a panel of minimal 30 positives and 70 negatives Use competitor’s kit component in the comparison study Only use the component provided by the test kit, not to switch or substitute The LOD of the Rapid Test is different from that of competitor’s Compare the product with the same LOD Send feedback to CTK for the test with the same LOD The result of Rapid Test is negative on Bio-Rad or other control Bio-Rad control or other control is designed for quantitative test, such as ELISA, CLIA, might not suit for rapid test. Test the control with the Rapid Test as well as with the market leader rapid test. “Compare Apple to Apple”

25 Analysis of Sample Evaluation
ID Company A Company B Leading brand Company C ELISA 1 Negative N/A 2 3 Positive 4 6 11 12 Weak positive 16 17 18 19 24 30 32 5 Strong positive 7 15 positive 23 27 20 21 31 If we compare “Apple to tomato”, then Company A kit has false negative . Company C kit has inferior quality. If using company C as reference, company A kit has false negative or false positive. . Tested against 22 specimens, company A kit is similar as the leading kit If only #31 specimen is tested, company A kit has false positive. However the positive result is verified by ELISA kit .

26 Add itio nal Info rma tion Storag e Temp eratur e & Shelf Life
6 Add itio nal Info rma tion Storag e Temp eratur e & Shelf Life Reference: Clinical and Laboratory Standard Institute. EP25-A. Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline

27 Stability of an IVD Product
Stability describes the ability of an IVD reagent to retain its properties and performance over a specified time interval when stored under specific conditions. The product stability is defined on the basis of ensuring that key performance metrics are met within predefined acceptance criteria throughout the claimed duration. IVD product stability is affected by external and internal variables, such as storage conditions, handling, final product container system, formulation. Reference CLSI. EP25-A

28 Shelf Life of an IVD Product
The Shelf life is the period of time until the expiration (expiry) date, during which an IVD reagent in the packaging configuration provided to the user maintains its stability under the storage conditions specified by the manufacturer. Reference CLSI. EP25-A

29 Temperature Outside the Specified Range Negatively Affects the Shelf Life of an IVD Product
Temperatures outside the range specified by the manufacturer damage antibodies and antigens used in the IVD product, and therefore reduce product performance. Based on the Arrhenius equation, the chemical reaction rate (e.g. antibody degradation) may double for every 10˚C increase in temperature. Therefore, stability of an IVD product may decrease 2 times faster if temperature is elevated 10˚C above recommended storage temperature Temperature Stability Reference CLSI. EP25-A

30 Temperature Outside the Specified Range Negatively Affects the Shelf Life of an IVD Product

31 Rapid Test Shelf Life in Different Storage Condition
Rapid Test stable for 18 months ≈ 9 months ≈ 3 months WHO qualifies a good rapid test that shows stability for 2 months at 45⁰C storage (quote: WHO malaria rapid test evaluation program)

32 To Ensure IVD Product Quality and Reliability
Store all components of an IVD product under conditions specified by the manufacturer. Monitor and report temperature of storage areas daily. If product was exposed to temperatures outside the specified range: Contact CTK technical service Verify performance of the product Temperature monitor report

33 We Appreciate Your Feedback Technical Support Product Feedback
Contact Info.: 10110 Mesa Rim Rd. San Diego, CA USA Telephone: 1 (858) Fax: 1 (858) Technical Support Product Feedback

34 QUALITY . INNOVATION . SIMPLICITY
THANK YOU


Download ppt "OnSite Rapid Test Procedure Training"

Similar presentations


Ads by Google