Tools for Non-Invasively Measuring Brain Compliance By Mark L. Manwaring.

Similar presentations


Presentation on theme: "Tools for Non-Invasively Measuring Brain Compliance By Mark L. Manwaring."— Presentation transcript:

1 Tools for Non-Invasively Measuring Brain Compliance By Mark L. Manwaring

2 What is Engineering? To contrive or plan out usually with more or less subtle skill and craft; to guide the course. S ystematic study of algorithmic processes that describe and transform information: their theory, analysis, design, efficiency, implementation, and application. Encompasses many areas of specialization. Software development, systems development and hardware selection, studies of compatibility between hardware and software, language development and modification, or perhaps a combination of these and any number of other diverse applications and concepts. Must be familiar with the language of the physical sciences, mathematics, and English.

3 Normal fetal brain

4 A Ventriculoperitioneal (VP) Shunt

5 Over drainage and slit ventricles – a statistical “setup” for repeat surgery

6 Detection of shunt failure Standard practice rests on history, clinical exam, and CT scan changes—unreliable on initial assessment ¼ of time ICP measurement is definitive, but invasive and impractical prior to ER

7 Optimizing outcome of hydrocephalus Avoidance of complications at surgery: malposition and infection Avoidance of longterm complications of obstruction and overdrainage Simplify and eliminate shunts where possible (evolving role of third ventriculostomy and related surgeries) Optimize hydrocephalus control for cognitive outcome Wean shunts to ligation and removal—the cure?

8 The Ear is a Speaker -- a recent discovery 1.Internal carotid artery and its distral tributary is the generator of the ICP pulse inside the cranium. 2.Pulse travels through CSF and brain. 3.Tissues change the ICP waveform morphology dependent on brain compliance—increased amplitude and risetime. 4.CSF waveform transmitted backwards across hearing apparatus (cochlear endolymph  ossicles  tympanic membrane). 5. ICP pulsation not perceived due to frequency below hearing threshold (20 Hz to 20 kHz).

9 General Ear Anatomy Middle & Inner Ear— CSF pulse travels into endolymph

10 Measuring the Output of the Ear: Our Prototype Head band with “microphone” for detection of low frequency ICP pulse.

11 Ear microphone allows tracings from both ears simultaneously.

12 Unfiltered signals: Piezo sensor tympanic membrane and ICP sensor

13 The “brain stethoscope”: an apparatus and method for outpatient and home ICP monitoring

14 ICP vs earlobe pulse – standing upright

15 Vertical Position Coray Jotham

16 Horizontal Position CorayJotham

17 Minus 45 degrees—intracranial pressure exceeds 27 cm H2O (20 mm Hg)

18 Completely inverted—ICP greater than +45 cm (31 mmHg); CPP <50 mm Hg; cerebral congestion; retinal petichiae after prolonged period

19 Blue – signal from external ear canal; Green – pulse oximeter

20 Phase relationship between TMD and pulse oximeter at +45°

21 Phase relationship in horizontal position Note increased amplitude but no phase shift

22 Inversion to -45° Marked phase shift

23 Complete inversion = -90° Further phase shift seen

24 27 males, age 18-25. Five positions: +45°, 0°, -15°, -30°, -45°

25 27 males, the average and +/- 1 SD

26 Hysteresis of phase shift in a subject moved from horizontal to -45° and back, pausing 30sec in each position

27 Patients with VP shunts showed three patterns: Poor compliance, normal range, and overdrainage

28 A Prototype Hardware/Software System to Capture and Display Compliance Information

29 Issues to be worked on -- 1.A better method than using the FFT for determining phase angles 2.Measuring ICP with pulse oximeters 3.A long-term, implantable measuring device associated with the shunts. 4.Theory – the head is resonant chamber. Can the problem of hydrocephalus be “cured” by fixing the resonance problem ?