M. Bonesini - MICE CM19 RAL 07/10/071 M. Bonesini INFN Milano PMTs tests in Milano (M. Bonesini, M. Rayner, S. Terzo)

Similar presentations


Presentation on theme: "M. Bonesini - MICE CM19 RAL 07/10/071 M. Bonesini INFN Milano PMTs tests in Milano (M. Bonesini, M. Rayner, S. Terzo)"— Presentation transcript:

1 M. Bonesini - MICE CM19 RAL 07/10/071 M. Bonesini INFN Milano PMTs tests in Milano (M. Bonesini, M. Rayner, S. Terzo)

2 M. Bonesini - MICE CM19 RAL 07/10/072 Outline Hot topics TOF2 PMTs shielding TOF/KL PMTs magnetic tests: new power supply (up to 600 G), more PMTs under test TOF2 frame with local PMTs shielding Conclusions

3 M. Bonesini - MICE CM19 RAL 07/10/073 Problems: cabling (HV, signal, calibration fibers …) checks with 3-D calculations: TOF PMTs with mu-metal shielding only can bare B // up to ~ 60 G cost issue ($$$) Ghislain “TOF1/2 cage”

4 M. Bonesini - MICE CM19 RAL 07/10/074 PMTs tests: conventional PMTs shielding  TOF0 PMTs requires no shielding, aside 1 mm  -metal due to small B field (B << 50 Gauss, Kevin private communication)  TOF1/2 even after the 100 mm global iron shielding needs some additional shielding to use conventional PMTs (as B // ~ 200-300 G, B _|_ ~ 1000 G)

5 M. Bonesini - MICE CM19 RAL 07/10/075 Problem: R4998 can bare B perp but not B // B // < 50 Gauss fine; B _|_ < 100-120 Gauss fine eventually additional  -metal or soft iron

6 M. Bonesini - MICE CM19 RAL 07/10/076 B field is ~ 200-300 G //; ~ 1000 G _|_ at r=30-35 cm (PMTs position) It seems that single PMTs may be individually shielded preserving good timing properties the design of the final shielding may be optimized, using our test facility at MIB Background infos MIB test magnet: coil with 5 windings for fields up to 600-650 G

7 M. Bonesini - MICE CM19 RAL 07/10/077 Advocated solutions: Solution A (“Ghislain cage”): extra shield connected with joints to first shield (100 mm) Solution B: no-extra shield but heavy local shielding of PMTs a la D0 (test magnet at Milano Bicocca)

8 M. Bonesini - MICE CM19 RAL 07/10/078 D0 tests (note # 2706) Problems: It will work for R4998 1” PMTs? Tests to be done for Gain + timing See some results later

9 M. Bonesini - MICE CM19 RAL 07/10/079 R. Bertoni, M. Bonesini, M. Rayner, S.Trezzo INFN MIB Changed with a big 200A (30 V) power supply from Eutron (on loan) -> B up to 600 G

10 M. Bonesini - MICE CM19 RAL 07/10/0710 Laser source (Nichia Blue laser diode + up to 1MHz Avtec fast pulser: signals 100 ps- 2 ns) field was up to 470 G (40A power), now up to 600 G with new power supply (on loan) signal comparable to BTF (~ 200-700 mV) Use of different shieldings: mu-metal only, additional Fe shielding, …

11 M. Bonesini - MICE CM19 RAL 07/10/0711 B field calibration B is uniform to some % in a region +- 5 cm from solenoid center calibration with a Hirst GM404 Hall probe

12 M. Bonesini - MICE CM19 RAL 07/10/0712 KL PMTs tests More than fine !!! and in addition Ludovico has put thicker shielding (7.6 cm x 7.6 cm) and connected shields

13 M. Bonesini - MICE CM19 RAL 07/10/0713 TOF PMTs: only 1 mm mu metal

14 M. Bonesini - MICE CM19 RAL 07/10/0714 Timing measurements: 1.START laser sync out 2.STOP PMT signal after LE PLS 707 discriminator + PLS inverter MCA + Silena 7422 QVT Measure  t=start-stop START STOP LE PLS 707 discr Silena QVT PMT Delay+inverter Laser sync AVTECH pulser MCA

15 M. Bonesini - MICE CM19 RAL 07/10/0715 FE360 shielding 5x5 cm 2 + 1mm mu-metal : 0 cm extension

16 M. Bonesini - MICE CM19 RAL 07/10/0716

17 M. Bonesini - MICE CM19 RAL 07/10/0717 Fe360 5x5 cm2 shielding + 1mm mu- metal : 1cm extension

18 M. Bonesini - MICE CM19 RAL 07/10/0718

19 M. Bonesini - MICE CM19 RAL 07/10/0719 Fe360 5x5 cm2 shielding + 1mm mu- metal: 2 cm extension

20 M. Bonesini - MICE CM19 RAL 07/10/0720

21 M. Bonesini - MICE CM19 RAL 07/10/0721 0 cm Extending the Fe box beyond the mu-metal shielding makes it much more effective

22 M. Bonesini - MICE CM19 RAL 07/10/0722 Preliminary conclusions: System tests have been extended from 400 G to 600 G (thanks loan of a new power supply of CNR IFP “Thorello”) more statistics needed, some syst. must be reduced, the best design of local shielding must be worked out BUT it seems that local shielding may work well for R4998 PMTs (we may also go from 5x5 cm 2 to 6x6 cm 2 in a bar-connected structure : more shielding !!!)

23 M. Bonesini - MICE CM19 RAL 07/10/0723 Local shielding of TOF2 Assuming local PMT shield will work for TOF2 CAD design for TOF2 frame+local bar PMT shielding+ support structure is not much more complicate and expensive than not shielded TOF2 ~30 Kg

24 M. Bonesini - MICE CM19 RAL 07/10/0724 Local shielding of TOF2 Some details: Inventor CAD for hanging of metal shielding blocks If funded, all may be done at MIB

25 M. Bonesini - MICE CM19 RAL 07/10/0725 Conclusions TOF2 local shielding is a valuable (cost-effective) solution requires some extra work from MIB mechanics shop (no extra funding required, if what promised for TOF0/1/2 will arrive) some tests to be finalized, to be completely sure (but I am confident) Preferred solution (if feasible) by TOF+KL teams Proposal: NOT decide now, taking also into account delays arising from experiment (6 months beamline delay), NTA INFN funding still pending [up to now we are at 0 E or $ as you like]: continue with tests to optimize local shielding design)


Download ppt "M. Bonesini - MICE CM19 RAL 07/10/071 M. Bonesini INFN Milano PMTs tests in Milano (M. Bonesini, M. Rayner, S. Terzo)"