Secondary Ion Mass Spectrometry Professor Paul K Chu
2
Secondary Ion Mass Spectrometry (SIMS)
3
Sputtering by Elastic Collisions Single knock-on < 1keV All secondary ions virtually originate from the uppermost atomic layers Linear cascade 1 keV – 1 MeV sputtering yield proportional to beam energy Spike > 1 MeV High density of recoil atoms
4
Ion – Solid Interactions
5
(a) Sputtering event, T=0 Predicted trajectories (b) Sputtering event, T 10 -13 s Post trajectories - indicated (c) Sputtering event, T 10 -10 s Sputtering Events with Time
6
Simulated Trajectories Computer simulation: Displacement of Cu atoms due to the impact of 4 keV argon ions (a) Trajectories within the entire volume of collision cascade for 10 incident particles (b,c) Transport of target atoms out of and into the designated layer (20 incident particles) (d) Trajectories of sputtered atoms (50 incident particles)
7
Sputtering Yield Sputtering yield is the average number of sputtered particles per incident ion. In the linear cascade regime, the sputtering yield is proportional to ion beam energy. Sputtering yield depends on a) atomic number, b) Displacement energy, c) Matrix of solid. Ion sputtering yield is the average number of ions emitted per incident primary ion. Many factors affect the ion yield. The most obvious are Intrinsic tendency to be ionized Positive ion : Ionization potential (IP) Negative ion: Electron affinity (EA) Matrix effects Al + from Al 2 O 3 versus Al + from Al metal Secondary ion cluster spectrum from Ar ion bombardment of Al. Note that the ordinate is in a log scale. Predominant species are Al + ions; Al 2 + and Al 3 + are also abundant
8
Matrix Effects Absolute secondary ion yields as a function of atomic number, under high vacuum conditions (a) and under oxygen saturation (b): 3keV Ar +, incident angle 60 o, beam density 10 -3 A/cm -2, pressure 10 -10 Torr I = I - I CLEAN
9
Ion Yield Enhancement Relative positive ion yield for 13.5 13.5 keV normal incident O - ; - compound was used, B.D; Barely detectable Relative negative ion yield for 16.5 keV Cs +, normal incidence Enhancement by O - Enhancement by Cs +
10
Ion Yield versus Ionization Potential and Electron Affinity (a) Positive relative ion yield of various certified elements (M + /Fe + ) in NBS 661 stainless steel reference material versus ionization potential (b) Negative relative ion yields of various certified elements (M - /Fe - ) in NBS 661 stainless steel reference material versus electron affinity
12
Secondary Ion Yield The variability in ionization efficiency leads to different analysis conditions for elements as indicated on the periodic table.
13
Selection of Primary Ions
14
Positive and Negative Ion Spectra Al alloy Positive ion spectrum Negative ion spectrum GaAs
15
Positive and Negative Ion Spectra Positive and negative spectra are complementary and useful in searching for traces of chemical elements and their complexes.
16
Instrumentation Ion Sources Ion sources with electron impact ionization - Duoplasmatron: Ar +, O 2 +, O - Ion sources with surface ionization - Cs + ion sources Ion sources with field emission - Ga + liquid metal ion sources Mass Analyzers Magnetic sector analyzer Quadrupole mass analyzer Time of flight analyzer Ion Detectors Faraday cup Dynode electron multiplier
17
SIMS CAMECA 6F
18
Cameca SIMS 1.Cs ion source 2.Duoplasmatron ion source 3.Primary beam mass filter 4.Immersion lens 5.Sample 6.Dynamic emittance matching 7.Transfer lens system 8.Liquid metal source 9.Entrance slit S 1 10.90 o electrostatic analyzer 11.Energy slit S 2 12.Intermediate lens 1 13.90 o magnetic sector 14.Exit slit S3 15.Projection lenses 16.Projection deflector 17.Channelplate 18.Fluorescence screen 19.Electron multiplier 20.Faraday cup
19
Magnetic Sector Analyzer High transmission efficiency High mass resolution Imaging Capability R 2000 Capable: R ~ 10 5
20
Ion Detectors Faraday Cup Secondary electron Multiplier 20 dynodes Current gain 10 7
21
Quadrupole SIMS
22
Energy Distribution of Sputtered Particles Energy distribution of neutral particles of some elemental polycrystalline targets emitted in the direction of the surface normal: Ar + ions, E P = 900 eV, incident angle = 0 O
23
Voltage Offset Technique
25
Mass Resolution Several definitions of mass resolution (R). R - capability of a mass spectrometer to differentiate between masses. M - mass difference between two adjacent peaks that are just resolved M - nominal mass of the first peak or mean mass of two peaks. Resolution is also defined as the full width at half maximum (FWHM) of a peak. C 2 H 4 + 28.0313 CH 2 N + 28.0187 M = 0.0126 N 2 + 28.0061 CO + 27.9949
26
Common Mass Interferences InterferingAnalyticalRequired M ionionresolution 28 Si + 32 S + 9600.0146 Matrix 16 O 2 + 32 S + 18000.0178 ionsSi 2 +56 Fe + 29600.0189 46 Ti 28 Si+ 75 As + 109400.0069 46 Ti 29 Si+ 75 As + 105000.0091 Matrix+ 29 Si 30 Si 16 O +75 As + 31900.0235 primary Hydrates 30 Si 1 H 31 P + 39500.0078 27 Al 1 H- 28 Si - 23000.0120 54 Fe 1 H+ 55 Mn + 62900.0087 120 Sn 1 H+ 121 Sb + 192500.0062 Hydrocarbons 12 C 2 H 3 +27 Al + 6400.0420 12 C 5 H 3 +63 Cu + 6700.0939
27
Boron Implanted Silicon Wafer
28
Quantitative Analysis I A+ T = j p A Y A+ T f A+ T C A+ T Primary ion current density Area of analysis Instrumental transmission factor for A + Measured secondary ion current of A+ in the matrix T Secondary ion yield in the matrix T Atomic concentration of A in the matrix T I A+ T = S A+ T C A+ T Sensitivity factor for A in the matrix T Very difficult to calculate SA+T. It depends on the 1. Element and matrix 2. SIMS instrument 3. System parameters S A+ T Standards are normally used Standard the same matrix Sample Measure I A+ T Use S A+ T from standard Find C A T Measure I A+ T From known C A T Find S A+ T
29
Quantitative Analysis using Relative Sensitivity Factors
31
Detection Limits (Sensitivity)
32
Inherent SIMS Sensitivity Silicon with an atom density of 5 10 22 Si atoms/cm 3 Bombarded area of (100 m) 2 = (10 -2 cm) 2 = 10 -4 cm 2 Sputtering rate of 1.0 nm/sec = 10 -7 cm/sec Then, silicon volume removed per second by sputtering is V = 10 -4 cm 2 10 -7 cm/sec = 5 10 -11 cm 3 /sec Hence, a number of the removed atoms per second by sputtering is N = 5 10 22 cm -3 10 -11 cm -3 /sec = 5 10 11 /sec Assume 1% Secondary ion yield 10% Ion transmission Then, ions detected 5 10 11 /sec 10 -3 ions = 5 10 8 ions/sec If 5 ions/sec is a threshold, then (5 ion/sec)/(5 10 8 ion/sec) = 10 -8 = 10 10 ppb The detection limit is 5 10 14 atoms/cm 3
33
Typical Detection Limits in Silicon Primary Ion Beam O 2 + or Cs + Element Detection Limit Element Detected Ion atom/cm 3 B 11 B + <10 13 P 31 P - /31P + <5 10 14 As 28 Si 75 As - <10 14 Sb 121 Sb + <5 10 13 C 12 C - <5 10 15 O 16 O - <5 10 16 NSiN - <5 10 15 HH - <5 10 17
34
Common Modes of Analysis The bulk analysis mode is used to detect trace-level components, while sacrificing both depth and lateral resolution. The mass scan mode is used to survey the entire mass spectrum within a certain volume of the specimen. The depth profiling mode is use to measure the concentration of pre- selected elements as a function of depth from the surface. The imaging mode is used to determine the lateral distribution of pre- selected elements. In certain circumstances, an imaging depth profile combining both depth profiling and imaging can be obtained.
35
Mass Spectrum
36
Fingerprint of polymers Positive mass spectrum from polyethylene, 0 - 200 amu Positive mass spectrum from polystyrene, 0 - 200 amu
37
Sometimes both positive and negative spectra are needed Positive mass spectrum from polyphenylene sulfide, 0 – 200 ammu No indication of S It looks like polyethylene Negative mass spectrum from polyphenylene sulfide, 0 –250 amu
38
Dynamic SIMS – Depth Profiling
39
Factors Affecting Depth Resolution
40
CRATER EFFECT The shape of the depth profile can be affected by a) Redeposition by sputtering from the crater wall onto the analysis area b) Direct sputtering from the crater wall
41
Crater Effect (a) (b) (a) Ions sputtered from a selected central area (using a physical aperture or electronic gating) of the crater are passed into the mass spectrometer. (b) The beam is usually swept over a large area of the sample and signal detected from the central portion of the sweep. This avoids crater edge effects. The analyzed area is usually required to be at least a factor of 3 3 smaller than the scanned area.
42
Crater Side-Wall Contribution
43
Crater Bottom Flatness
44
Effects of Reducing Primary Ion Energy
46
Effects of Primary Angle of Incidence
47
Crater Bottom Roughening
48
Sample Rotation
49
Stable Primary Ion Gun Mass Analyzer with High Stability Low Noise Electronics and Highly Stable Detector Consistent Secondary Ion Extraction Requirements for High Precision SIMS Analysis
50
High Precision Depth Profiling
51
Typical Applications in Semiconductor Industry
52
Energy Contamination in Ion Implanted Materials
53
P-N Junction in Silicon
54
Gate Oxide Breakdown
55
Imaging The example (microbeam) images show a pyrite (FeS 2 ) grain from a sample of gold ore with gold located in the rims of the pyrite grains. The image numerical scales and associated colors represent different ranges of secondary ion intensities per pixel. Some instruments simultaneously produce high mass resolution and high lateral resolution. However, the SIMS analyst must trade high sensitivity for high lateral resolution because focusing the primary beam to smaller diameters also reduces beam intensity. High lateral resolution is required for mapping chemical elements. 34 S 197 AU
56
Cross-Sectional Imaging Cross-sectional 27 Al - Image depth profile of SiO 2 capped GaAs/AlGaAs superlattice with a 4 micrometer laser melt strip
57
Dynamic SIMS versus Static SIMS
58
Time-of-Flight (TOF) SIMS
59
TOF-SIMS Analysis of Polymers
60
Surface Analysis of Silicon Wafers
61
Characterization of Hard Disk Lubricants
63
Sample Tutorial Questions What are matrix effects? What is the difference between ion yield and sputtering yield? When are oxygen and cesium ions used as primary ions? Why is the primary ion rastered when acquiring a depth profile? How can depth resolution be improved? How are mass interferences separated?