Presentation is loading. Please wait.

DAQ Calibration for Heat Exchanger Calibration of Cold/Hot Flow Rate and Temperature UTC ENCH/ENEV 435 11/25/03 Greg Kirton Kevin Zitzow Phuong Mai.

Similar presentations


Presentation on theme: "DAQ Calibration for Heat Exchanger Calibration of Cold/Hot Flow Rate and Temperature UTC ENCH/ENEV 435 11/25/03 Greg Kirton Kevin Zitzow Phuong Mai."— Presentation transcript:

1

2 DAQ Calibration for Heat Exchanger Calibration of Cold/Hot Flow Rate and Temperature UTC ENCH/ENEV 435 11/25/03 Greg Kirton Kevin Zitzow Phuong Mai

3 Objectives  Calibrate Cold Water Flow Rates from DAQ with Measured Flow Rates  Calibrate Hot Water Flow Rates from DAQ with Measured Flow Rates  Calibrate Temperatures from DAQ with Measured Temperatures

4 Heat Exchanger in UTC Lab

5 FT=Flow Transmitter S=Solenoid TT= Temp Transmitter CW=Cold Water W/H=Water Heater CWS=Cold Water Supply CWR=Cold Water Return Schematic of Heat Exchanger

6 Procedure for Measurements  A stopwatch was used to keep time  A 1000 ml graduated cylinder was used to measure volume  Measured volume of water obtained in 10 seconds.  Varied % Opening of Cooling Water Valve % Hot Water Pump Speed  4 measurements were made for each setting

7 Data Analysis  Measurements were averaged and error was calculated using student’s T equation. http://chem.engr.utc.edu/engr329/Lab-manual/Students-T.htm  Computer DAQ was averaged and standard deviation was calculated in Excel DAQ error was reported as 2xSTDEV

8 Cold Water Flow Calibration

9 Cold Water Flow Data

10

11 Hot Water Flow Calibration

12 Hot Water Flow Data

13

14 Temperature Calibration

15 Insulation

16 Tube Side Port B Tube Side Port A Shell Side Port B Shell Side Port A

17 Temperature Data

18 Overall Heat Transfer Coefficient of Heat Exchanger

19 Objectives  Compare total heat transfer rates of the hot and cold streams.  Determine the overall heat transfer coefficient.

20 Pictures of the Heat Exchanger Hot water in Cold water in Tube bundle

21 Heat Transfer Properties  55 tubes  Tube Volume = 143 cc  Tube Length = 506 mm  Total heat transfer area = 0.27 m 2

22 Heat Transfer Diagram T hot,in T hot,out T cold,in T cold,out q

23 Log Mean Temperature

24 Heat Transfer Rate Equations

25 1 st Run  Input = 75% Hot water mass flow rate (HWS) of 0.4124kg/s  CW valve open = 60 % Cold Water mass flow rate (CWS) of 0.080kg/s

26 q and U Results of 1 st Run  q c =5.76 kW  q h = 3.37 kW  U c = 1352 W/m 2 K  U h = 792 W/m 2 K

27 2 nd Run  Input = 40% Hot water mass flow rate (HWS) of 0.0258 kg/s  CW valve open = 40 % Cold water mass flow rate (HWS) of 0.0505 kg/s

28 q and U Results of 2 nd Run  q c =2.33 kW  q h = 2.25 kW  U c = 532 W/m 2 K  U h = 514 W/m 2 K

29 1). HWS = 0.04124kg/s and CWS = 0.080 kg/s 2). HWS = 0.0258 kg/s and CWS = 0.0505 kg/s 1st 2nd 1st 2nd Hot Flow Cold Flow

30 Hot Flow Cold Flow 1st 2nd 1st 1). HWS = 0.04124kg/s and CWS = 0.080 kg/s 2). HWS = 0.0258 kg/s and CWS = 0.0505 kg/s

31 Conclusions  1 st run: q c and q h are not closed.  2 nd run: q c and q h are closed.  1 st run: U c and U h are not closed.  1 st run: U c and U h are closed.  U increases as HWS and CWS increase. 1). HWS = 0.04124kg/s and CWS = 0.080 kg/s 2). HWS = 0.0258 kg/s and CWS = 0.0505 kg/s

32 Questions ?????


Download ppt "DAQ Calibration for Heat Exchanger Calibration of Cold/Hot Flow Rate and Temperature UTC ENCH/ENEV 435 11/25/03 Greg Kirton Kevin Zitzow Phuong Mai."

Similar presentations