Presentation is loading. Please wait.

Presentation is loading. Please wait.

SELECTING AND MANAGING CONTRACT STERILIZERS AND TESTING LABORATORIES

Similar presentations


Presentation on theme: "SELECTING AND MANAGING CONTRACT STERILIZERS AND TESTING LABORATORIES"— Presentation transcript:

1 SELECTING AND MANAGING CONTRACT STERILIZERS AND TESTING LABORATORIES
Pacific BioLabs Inc. (510)

2 CONTRACT STERILIZERS Written agreement between the contract sterilizer and the device manufacturer is required by 21 CFR (e), Medical Devices, processing, labeling or repacking

3 WRITTEN AGREEMENT MUST INCLUDE:
Instructions for maintaining proper records Acknowledgement that the device is non-sterile and awaiting processing Detailed specifications for the sterilization process Division of responsibilities

4 DIVISION OF RESPONSIBILITY
The device manufacturer is ultimately responsible for the sterility of its products The contract sterilizer is responsible for a quality control system for steps under its control Therefore, ……..

5 RESPONSIBILITIES OF MANUFACTURER
Ensure devices are sterile by confirming quality control system of contractor Validation of the process Ensure process controls are appropriate Maintain records of specification & evidence they were followed Product release

6 RESPONSIBILITIES OF CONTRACT STERILIZER
All requirements of the quality system regulation that apply to its operations Equipment installation, maintenance, and equipment calibration Installation and operation qualifications Building adequacy Personnel training Environmental controls

7 RESPONSIBILITIES OF CONTRACT STERILIZER (cont.)
Quality system requirements: Process controls Documentation and record keeping Acceptance of manufacturing materials Procedures for handling process deviations Any other requirement of the written agreement

8 COMMON PROBLEMS FOUND BY FDA
Manufacturer does not assume responsibility for sterilization performed by the contract sterilizer Communication between the two parties of changes Product is released following sterilization without approval of device manufacturer

9 TESTING LABORATORY Manufacturing Facility Contract Laboratory
Contract Sterilizer

10 LAB ATTRIBUTES Quality System FDA Track Record Experienced Personnel
SOP Calibrated equipment Training FDA Track Record Experienced Personnel Industry Reputation Committed Client Services

11 STERILIZATION QUALITY TESTS
Bioburden Bioburden Recovery Validation Sterility Bacteriostasis and Fungistasis Biological Indicators Pyrogen or Bacterial Endotoxin Inhibition and Enhancement

12 STERILIZATION QUALITY TESTS (cont.)
Package Integrity Chemical Residuals Environmental Monitoring Microbial Identification

13 BIOBURDEN Webster’s Medical Device bio (Greek) – life, living
burden – that which is carried, load Medical Device Population (number and types) of viable microorganisms on a product and/or a package

14 BIOBURDEN TEST Three steps Bioburden Recovery Validation
Removal of microbes from a product Culturing on a medium appropriate for growth Enumeration Bioburden Recovery Validation Demonstrates the adequacy of the steps above

15 BIOBURDEN METHODS Pour plating Spread plating Membrane filtration
Automated microbiology systems

16 STERILITY TESTS Product Sterility Biological Indicators (BI) Sterility
Membrane filtration Direct transfer Biological Indicators (BI) Sterility Testing the BI by placing it in microbiological growth medium

17 STERILITY TEST VALIDATION
Bacteriostasis and Fungistasis (B & F) Test using known organisms to demonstrate whether the combination of the product and sterilization method do or do not inhibit microbial growth

18 STERILITY TEST USP vs. AAMI/ISO
Two microbiological media SCDM and FTM B&F (6 organisms) AAMI/ISO One microbiological medium SCDM B&F (3 organisms)

19 BIOLOGICAL INDICATOR TESTS
Three quality attributes may be measured: Species identification Population Resistance D-Value Z-Value

20 BACTERIAL ENDOTOXIN TEST
Endotoxin = Pyrogen Bacterial Endotoxin Test = LAL Test Gel clot method Photometric techniques Photochromogenic Turbidimetric

21 INHIBITION AND ENHANCEMENT TEST
Validation of the LAL test 3 lots of product must be tested

22 DEVICE PACKAGING 21 CFR “Each manufacturer shall ensure that device packaging and shipping containers are designed and constructed to protect the device from alteration or damage during the customary conditions of processing, handling, and distribution”

23 DEVICE PACKAGING 21 CFR Package integrity is required to maintain the microbial barrier and sterility of the product Expiration dating Required for IVD’s and certain other devices that may degrade over time If label includes expiration, you must have data to back it up

24 PACKAGE INTEGRITY TESTS
Dye Penetration Microbial Challenge Burst Creep Seal peal

25 CHEMICAL RESIDUALS (EO ONLY)
EO (ethylene oxide) ECH (ethylene chlorohydrin) EG (ethylene glycol) No longer required

26 ENVIRONMENTAL MONITORING
Conducted in-house Laboratory field technicians conduct sampling at manufacturing facility Samples collected by sponsor’s personnel and incubated/counted at testing laboratory

27 ENVIRONMENTAL MONITORING TESTS
Non-viable particles Viable particles Bacteria Soybean Casein Digest Medium (SCDM) 30 – 35 °C Yeast and Mold Sbouraud Dextrose Agar (SDA) 20 – 25 °C Growth promotion required

28 MICROBIAL IDENTIFICATION
Morphological characterization Colony description, Gram stain Identification to genus and species level Phenotypic Genotypic

29 MICROBIAL IDENTIFICATION
Biochemical Tests Vitek Fatty Acid Analysis MIDI Carbon Utilization Biolog Genetic Based Technology (preferred) RiboPrinter

30 MICROBIAL WATER TESTS Total Heterotrophic Counts Yeast and Mold Counts
Coliform Determinations Thermophiles Counts Anaerobes counts Pseudomonas screening

31 WATER TESTS Choice of Media Choice of Incubation Temperature
Plate Count Agar Soybean Casein Digest Agar R2A Tryptone Glucose Extract Choice of Incubation Temperature 30-35°C, 20-25°C Choice of Incubation Times 2, 3, 5 or 7 days

32 THANK YOU Q & A


Download ppt "SELECTING AND MANAGING CONTRACT STERILIZERS AND TESTING LABORATORIES"

Similar presentations


Ads by Google