1 TUBES, CATHETERS and DEVICES …and when they go BAD

2 A dr Z Lecture On the placement (and misplacement) of monitoring and therapeutic devices in the critically ill patient

3 Radiography It is mandatory to check for position and complications after placing ANY device within a patient! Radiography is definitive! Clinical evaluation is NOT sufficient!

4 Devices MOVE! In critically ill patients, you must RECONFIRM the position of ALL devices at least every day.

5 Complications HAPPEN! Another reason to recheck critically ill patients is to detect complications and correct them. The complications can be device-related or not, but they are frequent and can be serious or life threatening.

6 ICU PATIENTS It IS necessary to re-check the position of ALL devices and to look for complications EVERY 24 hours in all ICU patients, by getting a Portable Chest Radiograph.

7 How Frequent? In recent studies, 25% of ICU portable chest radiographs showed an adverse change in position of a device, or a complication that needed intervention!

8 The Devices Nasogastric (NGT) and oral gastric tubes Endotracheal tubes (ETT) Vascular catheters Pacemakers, AICDs, Swan-Ganz catheters, chest tubes, etc.

9 The Complications Pneumothorax Pneumomediastinum Obstructive atelectasis Pleural and mediastinal fluid Pulmonary infarction Pulmonary edema Aspiration and pneumonia

10 ENDOTRACHEAL TUBES ETT

11 Endotracheal Tubes: optimally positioned Tip about 5 cm above the carina Tip at top 1/3rd of aortic arch

12 Endotracheal Tube: optimal position

13 Endotracheal Tubes: mal- positioned Too high: Can damage larynx. Patient can extubate if neck extended

14 Endotracheal tube: mal- positioned Too low: If patient’s head is flexed, ETT can enter right mainstem bronchus

15 ETT: malpositioned Too low: The ETT can easily enter the right main stem bronchus. It likes to go there-don’t let it!

16 ETT: too low ETT has entered right main stem bronchus ETT has obstructed the left mainstem bronchus and collapse the left lung If mechanically ventilated, can cause a right pneumothorax also

17 Endotracheal Tube: mal- positioned Esophageal intubation An ETT in the esophagus does not ventilate the patient Hypoxia results, with serious or fatal consequences

18 Esophageal Intubation: signs ETT tip below carina Part of ETT outside trachea wall Balloon overlaps trachea walls Trachea visible outside of ETT

19 Esophageal Intubation

20 Nasogastric Tubes NGT

21 Nasogastric tubes Tip of NGT must be at least 10 cm distal to the gastroesophageal junction There is a side hole at 7 cm. If above the ge junction, can lead to aspiration

22 NGT: good position

23 NGT: the ge junction

24 NGT: the side hole

25 NGT: too high

26 NGT: coiled in pharynx

27 NGT: in right bronchus

28 Vascular Catheters and Devices

29 Catheters and Devices Venous access catheters Central venous catheters Swan-Ganz catheters Pacemakers

30 Vascular Catheters Placement and Landmarks

31 Venous Catheter placement Ideally, in the superior vena cava Acceptable, in the brachio-cephlic veins Marginal, in the right atrium

32 Venous Landmarks Subclavian vein: thoracic margin to head of clavicle, where it joins Internal Jugular vein, to become the Brachio-cephalic vein

33 Venous Landmarks, upper To find the junction of the two brachio- cepahlic veins and so origin of Superior Vena Cava, Follow the curve of the lower margin of the right First Rib to the right paramidline

34 Venous Landmarks, upper

35 Venous Landmarks, lower To find the termination of the Superior vena Cava at the Right Atrium, look for the convex lateral curve of the heart

36 Venous Landmarks, lower

37 Review: Venous Landmarks

38 Venous Catheter placement: ideal

39 Venous catheter placement: marginal

40 Misplaced catheters Venous Aterial Extra-vascular

41 Misplaced catheter: venous In addition to too far or not far enough, places to avoid are: Internal jugular vein Azygos vein Internal mammary vein

42 Misplaced catheter: Internal Jugular vein

43 Misplaced catheter: Azygos vein

44 Venous catheter: subclavian artery to aorta

45 Extra-vascular catheter placement IV fluid infuses into mediastinum, pleural space, or extrapleural space Pneumothorax, pneumomediastinum may occur When in doubt, do CT Chest.

46 Swan-Ganz Catheter Ideal placement is tip in right or left pulmonary artery More peripheral placement can cause an infarct if wedged into a small artery

47 Swan-Ganz Catheter: good placement

48 Swan-Ganz Catheter: too far

49

50 Pacemakers Leads are in the right atrium and right ventricle; some units have a third lead in the coronary sinus. Some are also AICD

51 Pacemaker

52 So….. Don’t ASSUME a device is OK CONFIRM the placement of ALL devices by radiology imaging RECONFIRM the position of ALL devices EVERY DAY in critically ill patients

53 Goodbye… Copyright 2005 Michael Zucker, MD