X-ray powder diffractometer X-ray powder diffractometer.

Similar presentations


Presentation on theme: "X-ray powder diffractometer X-ray powder diffractometer."— Presentation transcript:

1 X-ray powder diffractometer X-ray powder diffractometer

2 X-ray powder diffractometer radiation counter radiation counter

3 X-ray powder diffractometer

4 Powder diffractometer (see Jenkins & Snyder, chap. 7)

5 Bragg-Brentano geometry

6 X-ray powder diffractometer

7

8

9

10

11

12

13

14

15

16

17

18 To maintain the focusing geometry, when counter is rotated at rate of 2  /min, specimen is also rotated at rate of  /min  /min  /min

19 Powder diffractometer Bragg-Brentano instrument - errors (see Klug & Alexander) Axial divergence - axis, here, is main rotation axis - asymmetric broadening toward low 2  - negative 2  error up to 90° - positive 2  error beyond 90° Bragg-Brentano instrument - errors (see Klug & Alexander) Axial divergence - axis, here, is main rotation axis - asymmetric broadening toward low 2  - negative 2  error up to 90° - positive 2  error beyond 90°

20 Powder diffractometer Bragg-Brentano instrument - errors (see Klug & Alexander) Axial divergence - axis, here, is main rotation axis Bragg-Brentano instrument - errors (see Klug & Alexander) Axial divergence - axis, here, is main rotation axis

21 Powder diffractometer Bragg-Brentano instrument - errors (see Klug & Alexander) Axial divergence - axis, here is main rotation axis each section of Soller slit produces its own Debye ring from a specific region in the specimen divergence angle typically ±2.5° Bragg-Brentano instrument - errors (see Klug & Alexander) Axial divergence - axis, here is main rotation axis each section of Soller slit produces its own Debye ring from a specific region in the specimen divergence angle typically ±2.5°

22 Powder diffractometer Bragg-Brentano instrument - errors (see Klug & Alexander) Axial divergence - axis, here is main rotation axis each section of Soller slit produces its own Debye ring from a specific region in the specimen divergence angle typically ±2.5° Bragg-Brentano instrument - errors (see Klug & Alexander) Axial divergence - axis, here is main rotation axis each section of Soller slit produces its own Debye ring from a specific region in the specimen divergence angle typically ±2.5°

23 Powder diffractometer Bragg-Brentano instrument - errors (see Klug & Alexander) Flat specimen peak shifts to lower 2  & asymmetric broadening depends on beam divergence & 2 , plus specimen shape Bragg-Brentano instrument - errors (see Klug & Alexander) Flat specimen peak shifts to lower 2  & asymmetric broadening depends on beam divergence & 2 , plus specimen shape

24 Powder diffractometer Bragg-Brentano instrument - errors (see Klug & Alexander) Specimen transparency Bragg-Brentano instrument - errors (see Klug & Alexander) Specimen transparency

25 Powder diffractometer Bragg-Brentano instrument - errors (see Klug & Alexander) Specimen displacement Usually largest error in expt'l data Neg. peak shift - ~0.01° 2  /15  displacement Asymmetric broadening toward lower 2  Bragg-Brentano instrument - errors (see Klug & Alexander) Specimen displacement Usually largest error in expt'l data Neg. peak shift - ~0.01° 2  /15  displacement Asymmetric broadening toward lower 2 

26 Powder diffractometer Bragg-Brentano instrument Divergence slit choice Bragg-Brentano instrument Divergence slit choice

27 Powder diffractometer Bragg-Brentano instrument Divergence slit choice Or, theta-compensating slit - divergence slit opens as theta increases Bragg-Brentano instrument Divergence slit choice Or, theta-compensating slit - divergence slit opens as theta increases

28 Powder diffractometer Bragg-Brentano instrument Receiving slit choice Bragg-Brentano instrument Receiving slit choice

29 Powder diffractometer Bragg-Brentano instrument Receiving slit choice lowers intensity - little improvement in resolution poor resolution - little improvement in intensity Bragg-Brentano instrument Receiving slit choice lowers intensity - little improvement in resolution poor resolution - little improvement in intensity

30 Powder diffractometer Bragg-Brentano instrument Receiving slit choice correct slit size ~0.15° Bragg-Brentano instrument Receiving slit choice correct slit size ~0.15°

31 Powder diffractometer Alignment steps

32 Powder diffractometer Alignment steps Align to focal spot Set take-off angle Alignment steps Align to focal spot Set take-off angle

33 Powder diffractometer Alignment steps Align to focal spot Set take-off angle Zero alignment 2:1 alignment Alignment steps Align to focal spot Set take-off angle Zero alignment 2:1 alignment

34 Powder diffractometer Alignment steps Align to focal spot Set take-off angle Zero alignment 2:1 alignment Alignment steps Align to focal spot Set take-off angle Zero alignment 2:1 alignment

35 Powder diffractometer Alignment steps Align to focal spot Set take-off angle Zero alignment 2:1 alignment Alignment steps Align to focal spot Set take-off angle Zero alignment 2:1 alignment

36 Powder diffractometer Alignment checks Run calibration pattern - use suitable standard NIST 640c (Si) Alignment checks Run calibration pattern - use suitable standard NIST 640c (Si)

37 Powder diffractometer Alignment checks Run calibration pattern - use suitable standard Alignment checks Run calibration pattern - use suitable standard

38 Powder diffractometer Alignment checks Run calibration pattern - use suitable standard Watch for: Intensity loss Angle shifts Resolution loss Alignment checks Run calibration pattern - use suitable standard Watch for: Intensity loss Angle shifts Resolution loss

39 Powder diffractometer Alignment checks

40 Powder diffractometer Alignment check "Five fingers of quartz" Alignment check "Five fingers of quartz"

41 Powder diffractometer Alignment check "Five fingers of quartz" Measure: position & intensity of (203)  1 H 1 /H 2 Alignment check "Five fingers of quartz" Measure: position & intensity of (203)  1 H 1 /H 2

42 Powder diffractometer Alignment checks Run calibration pattern - use suitable standard NIST 1976 (Al 2 O 3 ) Alignment checks Run calibration pattern - use suitable standard NIST 1976 (Al 2 O 3 )

43 Powder diffractometer Alignment checks Run calibration pattern - use suitable standard Alignment checks Run calibration pattern - use suitable standard

44 Powder diffractometer Alignment checks Run calibration pattern - use suitable standard Alignment checks Run calibration pattern - use suitable standard