Presentation is loading. Please wait.

Presentation is loading. Please wait.

pNeuton Transport Ventilator

Similar presentations


Presentation on theme: "pNeuton Transport Ventilator"— Presentation transcript:

1 pNeuton Transport Ventilator
mini Ventilator Self Guided Tour

2 Program Objectives This program is a self guided tour of the
pNeuton mini Ventilator. At the end of this tour you will be able to: Describe the main operating features of the ventilator Identify the controls and connections Describe the patient circuit Set-up the ventilator for patient operation Define various clinical environments for ventilator application Clean and maintain the ventilator

3 Features and Controls – Side Panel
Driving Gas Inputs Connections for high pressure oxygen and medical grade compressed air source gas Always use a high flow gas source that can provide psi ( bar) pressure at a minimum of 20 L/min Insure that the compressed air is clean and free of water

4 Features and Controls – Side Panel
Remote Alarm Output Remote Alarm connection A properly configured Remote Alarm will indicate an active ventilator alarm condition in a remote location

5 Features and Controls – Side Panel
Patient Connection Connection for the large bore hose of the patient circuit

6 Features and Controls – Side Panel
Prox Pressure Patient circuit medium hose connection Connects to the blue tube from the proximal patient pressure port on the patient circuit

7 Features and Controls – Side Panel
Exp Valve Patient circuit small hose connection Connects to the clear tube for the expiratory valve on the patient circuit

8 Features and Controls – Side Panel
Audible Alarm Audible alarm sound generator Used for patient disconnect, high pressure and low gas supply alarms Will not turn off during low gas supply alarm Do NOT put tape over the holes, push sharp objects into the holes or interfere in any way with the alarm The alarm is for patient protection. If you suspect the alarm is not operating properly have the unit serviced immediately

9 Features and Controls – Front Panel
PEEP / CPAP Control Adjusts the PEEP / CPAP from 0 to 20 cm H2O Operates in either mode Automatic sensitivity Oxygen concentration is maintained at set % Oxygen control

10 Features and Controls – Front Panel
Peak Pressure Control Limits peak airway pressure from 15 to 60 cm H2O Operates only in the IMV + CPAP mode Used to set the pressure limit during mandatory breath delivery

11 Features and Controls – Front Panel
Continuous Flow Control Sets baseline continuous flow for 6, 8, 10, 15 or 20 L/min Operates in either mode Used to set the flow used for both spontaneous breaths and mandatory breath delivery

12 Features and Controls – Front Panel
Inspiratory Time Control Sets inspiratory time for mandatory breaths from 0.25 to 2.0 seconds Operates only in the IMV + CPAP mode Can be used to preset tidal volume using the tidal volume chart on the top of the ventilator

13 Features and Controls – Front Panel
Expiratory Time Control Sets expiratory time of mandatory breaths from 0.25 to 20 seconds Operates only in the IMV + CPAP mode Use with the respiratory rate chart on the top of the ventilator to determine mandatory breath breath rate

14 Features and Controls – Front Panel
% Oxygen Control Delivered oxygen choice from 21 to 100% Operates in either mode Sets oxygen concentration of the continuous flow

15 Features and Controls – Front Panel
Visual Alarm Indicator Illuminates for an active alarm condition Alarm warns of patient disconnect after 10 seconds Alarms for high pressure and low gas supply Will turn off when alarm is corrected or Reset / Silence Button is pressed Will not turn off during low gas supply alarm

16 Features and Controls – Front Panel
Alarm Reset / Silence Button Press button to silence the alarm system for 25 seconds Every time the button is pressed the alarm silence period will reset for 25 seconds

17 Features and Controls – Front Panel
Mode Control Activates IMV + CPAP mode for delivery of ventilator breaths Mode is the same as CMV + PEEP CPAP mode provides continuous flow and CPAP for spontaneous breathing patients No ventilator breaths in the CPAP mode

18 Features and Controls – Front Panel
Pressure Gauge Represents pressure in patient circuit at the patient wye of the circuit All patient alarms are sensed at the patient wye

19 Features and Controls – Front Panel
High Pressure Alarm Control Sets the high pressure alarm from 15 to 70 cm H2O Independent audible and visual alarm from the Peak Pressure Control

20 Patient Circuit Patient circuit is a disposable, dual limb system
All hoses attach to the side of the ventilator Patient end connects for clinical intervention Directly to an endotracheal tube Use NIV kit to provide nasal prong or nasal mask application in neonates and infants Use full face CPAP mask (non-vented)

21 Patient Circuit Which size circuit - Neonatal or Pediatric?
The neonatal circuit uses 10 mm hoses and is designed for patients up to 3 kg The pediatric circuit uses 15 mm hoses and is designed for patients above 3 kg If your patient needs 10 L/min or more continuous flow to meet spontaneous flow demand, you should use the pediatric circuit to minimize expiratory flow resistance

22 Patient Circuit Use only Airon Corporation patient circuits
Calibration of the PEEP / CPAP and Peak Pressure controls is based on the Airon circuit and exhalation valve Using other manufacturers circuits / exhalation valve may cause: errors in pressure delivery high expiratory flow resistance interference with CPAP setting

23 Program Objectives pNeuton mini Ventilator self guided tour,
objectives completed so far Describe the main operating features of the ventilator Identify the controls and connections Describe the patient circuit With an understanding of pNeuton’s capabilities, let’s look at patient application.

24 Patient Application Set Main Controls
PEEP / CPAP can be set to minimum Determine peak ventilating pressure Set desired continuous flow to meet patient needs Set Inspiratory Time Set Expiratory Time Choose % Oxygen Adjust High Pressure Alarm for safety

25 Patient Application Initial Set-up
Attach Airon patient circuit to ventilator Attach a test lung to the patient connection on the circuit Attach high pressure oxygen and air hoses to the side of the ventilator Attach these hoses to the gas supplies

26 Patient Application Begin Ventilation Turn ON gas supplies
Alarm system will immediately activate Press Alarm Silence button if desired For IMV or CMV, set Mode control to IMV+CPAP Ventilator will start to provide breaths at the settings on the controls

27 Patient Application Set PEEP / CPAP
Adjust the PEEP / CPAP control to the level desired Adjust continuous flow to meet patient demand – not too high to create expiratory flow resistance – not to low to cause pressure swings and high patient effort Use 6 – 8 for neonates Use for infants Use 10 – 20 for children

28 Ventilator Operation Alarm System Functionality
Alarm preset to activate within 10 seconds Pressing the Alarm Silence button equals 25 second delay Disconnect Alarm is looking for a specific minimum pressure of at least 3 cm H2O at all times If there is not at least 3 cm H2O CPAP or mandatory breaths pressure of 3 cm H2O, the alarm will sound High pressure alarm activates whenever the pressure in the circuit exceeds the alarm control setting. Input Gas Pressure below 40 psi (2.8 bar) will activate the alarm

29 Ventilator Operation Peak Pressure Limit
Operates to limit the peak airway pressure during mandatory breaths For volume ventilation, set the Peak Pressure 10 cm H2O above the pressure needed to deliver the set Tidal Volume Peak Pressure control is not an alarm. Use the High Pressure Alarm control for this function

30 Ventilator Operation % Oxygen Control
Set % Oxygen control to the desired oxygen level Accuracy is plus / minus 3% The ventilator uses the oxygen gas supply for internal operation so high oxygen concentrations may deplete tanks quicker

31 Clinical Issues Neonatal Ventilation
Inspiratory time setting 0.25 to 2.0 sec Use low continuous flow for proper breath delivery and minimize expiratory flow resistance Inspiratory time, flow and pressure (along with lung compliance / resistance and ET-Tube leak) will determine tidal volume to the patient Use the respiratory rate chart on top of the ventilator to estimate ventilator rate Use NIV kit for nCPAP Monitor closely for patient synchrony

32 Clinical Issues Non-invasive Ventilation (NIV)
Use Airon NIV kit with nasal prongs or masks Proximal pressure measured will display on pressure gauge and be used by alarm system Effective patient interface seal will insure proper alarms Rotate from nasal masks to prongs every few hours for patient comfort and proper skin care

33 Clinical Issues Pediatric Volume Ventilation
Set the Peak Pressure to a safe pressure level Refer to the Tidal volume chart on the top of the ventilator for I-Time / Flow combinations to determine the proper settings for the tidal volume desired Adjust Inspiratory Time control / flow to increase or decrease tidal volume - more flow or time equals larger tidal volume Adjust PEEP / CPAP to established baseline pressure

34 Clinical Issues MRI Operation
pNeuton mini is compatible with MRI scanners up to 3 Tesla in strength While tests show that pNeuton functions at the bore of the MRI unit, a minimum proximity of 12 inches (0.3 meter) from the bore is recommended Safe MRI practice calls for all devices used in the proximity of a MRI scanner, including pNeuton mini, be anchored to prevent inadvertent movement

35 Clinical Issues Altitude Operation
The pNeuton mini may be used for air transports in unpressurized cabins up to 15,000 feet Inspiratory time will increase and expiratory time will decrease as altitude increases Pressure settings will not change as altitude changes Always use external monitors and pulse oximeters to monitor your patient

36 Patient Safety pNeuton - Patient Safety Features
Audible and Visual Alarms for patient disconnect, high pressure and low supply gas pressure High Pressure Release – pressure is relieved at the Peak Pressure control setting Internal Safety Pressure Release – limits circuit pressure to 70 cm H2O, regardless of the setting of the Peak Pressure control Anti-Suffocation System – in the event of ventilator malfunction an internal valve will allow ambient air to enter the patient circuit

37 Cleaning and Maintenance
Clean ventilator between patient use applications using a soft cloth and a mild detergent / disinfectant Do not immerse the ventilator in water Patient circuit is disposable, single use application and should not be cleaned, disinfected or reused Ventilator should receive preventative maintenance every 2 years

38 Program Objectives pNeuton mini Ventilator self guided tour
completed; you should now be able to: Describe the main operating features of the ventilator Identify the controls and connections Describe the patient circuit Set-up the ventilator for patient operation Define various clinical environments for ventilator application Clean and maintain the ventilator Next Step – consider reviewing the pNeuton Case Study and completing the competency evaluation

39 Contact Information Thank you for you interest in the pNeuton mini Ventilator For more information visit or call Airon Corporation at USA + Canada toll free International


Download ppt "pNeuton Transport Ventilator"

Similar presentations


Ads by Google