Presentation is loading. Please wait.

Presentation is loading. Please wait.

Session 2: Fundamentals and Classical methods of quantitative elemental analysis http://bcs.whfreeman.com/qca7e http://www.good-weighing-practice.com/gwp/proper-weighing.

Similar presentations


Presentation on theme: "Session 2: Fundamentals and Classical methods of quantitative elemental analysis http://bcs.whfreeman.com/qca7e http://www.good-weighing-practice.com/gwp/proper-weighing."— Presentation transcript:

1 Session 2: Fundamentals and Classical methods of quantitative elemental analysis

2 Measurement of Mass and Volume: Recognising random and systematic errors

3 Recap: Quantitative Analysis - Principles
Define sample amount (mass or volume) Measure quantity proportional to analyte concentration Measured property must vary in a defined way: calibration with known standards necessary Analysis must be specific: Interferences must be known and if possible be eliminated Accuracy: Proximity of measured value to accepted (or "true") value: must be determined Precision: Closeness of measured values to one another: must be defined and reported

4 Classical two-pan balance
Measuring Mass Classical two-pan balance Modern electronic analytical balance F = m x g Electroma-gnetic force to counter From: Harris, 6th edition

5 Random errors in weight measurements: tolerances of analytical balances and weights for calibration
Analytical balances need to be calibrated regularly Typically use stainless steel weights (d = 8.0 g/ml) From: Harris, 6th edition

6 Specifications for balances
Capacity Readability Repeatability (standard deviation); larger than readability Capacity x readability: Analytical balances: tens to hundreds of g x mg Ultra-micro balance: e.g. 6 g x mg

7 Avoiding systematic error in weight measurements: buoyancy
Any object displaces a certain amount of air This reduces the apparent mass that a balance measures If density of the object being weighed is significantly different from calibration weights, buoyancy correction is necessary: m = true mass; m’ = measured mass; da = density of air ( g/ml); dw = density of calibration weights (8.0 g/ml); d = density of weighed object

8 Correcting buoyancy errors
Buoyancy correction in dependence on density of weighed object

9 Exercise: A bottle weighed g empty and g after introduction of an organic liquid with a density of 0.92 g cm-3. The balance was equipped with stainless steel weights having a density of 8.0 g cm-3. Correct the weight of the sample for the effects of buoyancy.

10 Avoiding systematic errors in weight measurements
Temperature effects Convection air currents Warm air in balance weighs less Measured mass of object appears lower Essential to weigh at room temperature Prevent object from picking up moisture: Do not touch with bare fingers Let cool in desiccator If weighing substances that are kept in fridge or freezer, let warm up before weighing Be aware of hygroscopic substances Absolute error in weight as a function of time after object was removed from a 110°C oven (A: porcelain filtering crucible, B: weighing bottle containing 7.5 g of KCl.)

11 Measuring Volume Apparatus for volume measurement
Pipettes, Burettes, Volumetric flasks Calibrated either for containment (flasks) or delivery (pipettes, burettes) of specified volume Typical pipettes: (a) volumetric (transfer pipette); (b) Mohr; (c) serological; (d) Eppendorf micropipette

12 Characteristics of pipettes
Name Function Capacity (cm3) Type of drainage Volumetric Delivery of fixed volume 1-200 Free Mohr Delivery of variable volume 1-25 To lower calibration line Serological 0.1-10 Blow out last drop Eppendorf Delivery of fixed or variable vol. Empty tip by air displacement

13 Random errors in volume measurements: Tolerances of Class A Pipettes
Capacity (cm3) Tolerances (cm3) 0.5  0.006 1 2 5  0.01 10  0.02 20  0.03 25 50  0.05 100  0.08

14 Random errors in volume measurement: Range and precision of typical Eppendorf micropipettes
Volume range (mL) Standard dev. (mL) 1-20 2 mL 20 mL 10-100 15 mL 100 mL 20-200 25 mL 200 mL 250 mL mL mL mL

15 Tolerances of Class A burettes and volumetric flasks
Burette Vol. (mL) Tolerance (mL) Volumetric flask vol. (mL) 5  0.01  0.02 10 25  0.03 50  0.05 100  0.20  0.08 250  0.12 500 1000  0.30 2000  0.50

16 Avoiding systematic errors in volume measurements: Temperature effects
Volume occupied by a given mass of liquid, as well as the device that holds the liquid, varies with temperature For dilute aqueous solution: Coefficient of expansion = 0.025% / °C 1°C increase in temp. yields 0.025% increase in volume. Refer volumetric measurements to temperature at which they were made (standard temperature is 20 °C). Exercise: A mL sample is taken from an aqueous solution at 5°C. What volume does it occupy at 20 °C?

17 Avoiding systematic errors in volume measurement: Calibration of Volumetric Ware
Measure mass of liquid of known density and temperature contained in or delivered by a stated volume Buoyancy correction must be made (see Table) Divide corrected mass by density of liquid Express results at standard temperature (20°C).

18 Volume occupied by 1.0000g water weighed in air against stainless steel weights
Volume (mL/g) at T Temperature T (°C) Corrected for buoyancy Corrected for buoyancy and change in container volume = 20°C 10 1.0013 1.0016 12 1.0015 1.0017 14 1.0018 1.0019 16 1.0021 1.0022 18 1.0024 1.0025 20 1.0028 22 1.0033 1.0032 25 1.0040 1.0036 26 1.0043 1.0041 28 1.0048 1.0046 30 1.0054 1.0052 Exercise: A 25 mL pipette has been measured to deliver g of water weighed against stainless steel mass at 25°C. Use the data in the Table to determine the volume delivered by this pipette at 25°C and 20 °C.

19 Treatment of glassware
Need to ensure that containers are clean and not contaminated Important that liquids interact in defined way with glass surfaces: use detergents For trace analysis, it is common to use an “acid wash” If possible, use polypropylene or teflon rather than glass

20 Summary All measurements carry errors/uncertainty
Systematic errors can be corrected Accuracy of methods can be improved Random errors cannot be corrected Precision of method can be determined and must be known All quantitative data must be reported with error Methods to solve a given analytical question can be selected according to their performance characteristics Analysts must be aware of the performance characteristics of their tools

21 Classical analytical methods: Gravimetric and volumetric analyses

22 Gravimetric analysis Analyte is converted to a solid product of known (pure) composition and weighed Conversion of the analyte can be accomplished in several ways: Reduction of an ion to its elemental form (e.g. by electrolysis) Roasting (hydrolysis/oxidation) of a compound Precipitation of an ion with a counterion Precipitation of an organic molecule Methods exist for most inorganic anions and cations, H2O, SO2, CO2, and iodine Organic compounds can also be quantified

23 Examples Precipitate analyte using precipitating agent
Convert analyte (usually ions) to its elemental form using reducing agents Reducing agent Analyte SO2 Se, Au SO2 + H2NOH Te H2NOH Se H2C2O4 Au H2 Re, Ir HCOOH Pt NaNO2 SnCl2 Hg Precipitate analyte using precipitating agent Precipitating agent Ion precipitated (Precipitate) AgNO3 Cl- (AgCl), Br- (AgBr), I- (AgI) HCl Ag+ (AgCl), Hg+ (Hg2Cl2), Na+ (NaCl) HNO3 Sn4+ (SnO2) H2SO4 Li+, Mn2+, Sr2+, Cd2+, Pb2+, Ba2+ (all as sulfates) H2C2O4 (oxalate) Ca2+, Sr2+, Th4+ (as oxalates or oxides)

24 Gravimetric analysis: precipitation of insoluble salts or complexes
Involves precipitation, filtration, drying, weighing e.g.: Sulfate with BaCl2 Ni(II) with dimethylglyoxime 8-hydroxyquinoline (oxine): range of metal ions. Forms sparingly soluble complexes For accuracy, certain conditions must be fulfilled: The ion of interest must precipitate completely (=quantitatively). The formed salt must have a very low solubility product Precipitate must be a pure compound (avoid co-precipitation) Precipitate must be easy to filter

25 Why gravimetry is still in use, although time-consuming and challenging:
Accurate and precise (if done properly) Absolute method: No calibration required Apparatus required is relatively inexpensive

26 Exercise: Lead (as Pb2+) can be determined by precipitation with sodium iodide Write down the stoichiometric reaction formula What mass of NaI is needed to convert 1.00 g of Pb(NO3)2 to PbI2? What mass of PbI2 will be formed?

27 Exercise: A sample of metallic tin (2.00 g) was reacted with iodine (8.80 g) in a refluxing organic solvent, and an orange-yellow solid (A) (8.62 g) was isolated. A qualitative elemental analysis of A showed it contained only tin and iodide. A sample of A ( g ) was accurately weighed into a pre-weighed silica crucible and roasted in air (A reacts with H2O) to produce SnO2 ( g). A second sample of A ( g) was dissolved in a small excess of nitric acid, and excess silver nitrate added dropwise to precipitate silver iodide, which was collected in a weighed sinter crucible, dried in an oven at 110°C, and then cooled and weighed (mass of AgI obtained= g). When a sample of A was exposed to air for several months, it became hydrated as shown by a second analysis of the impure product, B, which showed Sn=17.92%, I=76.64%, and H=0.61%. (Sn=118.69, I=126.90, O=16.00, Ag=107.87, H=1.008)

28 Exercise (continued):
From the amount of SnO2 obtained in the original analysis, calculate the percentage of tin in A From the amount of AgI obtained, calculate the percentage of iodide in A. Assuming a molecular formula for A of SnaIb, the molecular weight of A is therefore = x a x b. Percentage tin X (1) Percentage iodide Y (2) Use equations (1) and (2) and your calculated values for X and Y to estimate the ratio b/a and determine the empirical formula of A. A mass spectrum of A showed a cluster of peaks centred at a charge/mass ratio m/z=626, and no other peaks at higher m/z. Assuming the observed m/z corresponds to the approximate molar mass, M, what is the molecular formula of A? What is the percentage purity of the exposed sample, B, compared with A (regard A as pure), and how many water molecules are there in B?

29 Volumetric analysis Amount of analyte determined by measurement of volume of a reagent needed to react with analyte Titrimetry: Determining the quantity of a reagent of known concentration that is required to react completely with the analyte Titration: Adding standard solution (titrant) to solution of the analyte until reaction is complete. Solution dispensed from burette to determine volume of reagent required for reaction Requires that Reaction has a large equilibrium constant Reaction proceeds rapidly

30 Titrimetric methods Acid-base titrations Precipitation titrations
Volhard (Ag+ directly or Cl- via back titration) Complexometric titrations Cations with EDTA Redox titrations Manganometry, iodometry Spectrophotometric titrations Measures changes in UV-Vis spectra Potentiometric titrations Measures changes in potential (e.g. with pH electrodes or Ion-selective electrodes)

31 General terminology Equivalence point: Point in a titration when quantity of added titrant is the exact amount necessary for stoichiometric reaction with analyte. This is the “ideal” point sought in a titration. In reality, we find the End point: Point reached when a (ideally sudden) physical change in the solution occurs, which indicates the absence of unreacted analyte. End points are often detected through an indicator Ideally, there is very little difference between the volumes for the equivalence and end points. This difference is the titration error Can be determined with a blank titration Back titration: Excess of a standard solution added to consume analyte is determined by addition of second standard. Required when direct reaction is slow or unstable

32 Typical titration curve
Decrease in concentration of analyte Note: semi-log plot log c Equivalence point Volume/ amount of titrant added

33 General terminology: Standardisation
Titrations require standard solutions: Reagent of known concentration used to carry out titration Primary standard: Solution of a highly purified compound (>99.9%) that can be accurately weighed Serves as a reference material in a given volumetric titration method. The accuracy of such methods is critically dependent on the properties of this compound Must be stable (not decomposed during storage) Must be a compound that can be dried to remove residue water Standard reference materials commercially available (SRMs) Secondary standard: Solution of titrant that has been standardised by titrating a known amount of primary standard (also commercially available)

34 Precipitation Titrations
Based on reactions that give products of low solubility One of the oldest analytical techniques (mid 1800s) E.g. Volhard method for silver(I) titrations For direct analysis of silver ions or indirect detn. of halides Titrant: NaSCN Fe(III) acts as the indicator Red colour observed at [Fe(SCN)2+] = 6.4×10-6 M Reaction 1: Ag+ + SCN¯ ⇌ AgSCN(s) Ksp=[Ag+][SCN¯]=1.1 ×10-12 Reaction 2: Fe3+ + SCN¯ ⇌ [Fe(SCN)]2+ (red) Kf= [Fe(SCN)]2+ =1.4 ×102 [Fe3+][SCN¯]

35 Volhard method for Ag+ Exercise:
Titrate 50 mL of 0.05 M Ag+ with 0.1 M KSCN What concentration of Fe3+ should be used to reduce titration error to zero? Note: For zero titration error, the Fe(SCN)2+ colour should appear when [Ag+] = [SCN-]

36 Effect of solubility product
Log [Ag+] Ksp≈10-12 Ksp≈10-18  The higher the solubility, the more difficult becomes the end point recognition

37 Compleximetric titration
Metal ion determination Metal ion reacts with ligand to form complex Can form soluble complexes or precipitates Equivalence point determined by indicator EDTA: Ethylenediamine tetraacetic acid; is a hexadentate ligand pK1 = 0.0 pK2 = 1.5 pK3 = 2.0 pK4 = pK5 = pK6 = 10.24 (n-4)+ [M(H2O)6]n+ + [H2(EDTA)]2- + 6 H2O + 2 H+ ISO 6059: Determination of Hardness in water

38 Stoichiometric formation constants for EDTA complexes
[MY(n-4)+] [Mn+][Y4-] =KMY (Kf) Mn++Y4- ⇌ MY(n-4)+ Cation KMY log KMY Ag+ 2.1×107 7.32 Cu2+ 6.3×1018 18.80 Mg2+ 4.9×108 8.69 Zn2+ 3.2×1016 16.50 Ca2+ 5.0×1010 10.70 Cd2+ 2.9×1016 16.46 Sr2+ 4.3×108 8.63 Hg2+ 6.3×1021 21.80 Ba2+ 5.8×107 7.76 Pb2+ 1.1×1018 18.04 Mn2+ 6.2×1013 13.79 Al3+ 1.3×1016 16.13 Fe2+ 2.1×1014 14.33 Fe3+ 1.3×1025 25.10 Co2+ 2.0×1016 16.31 V3+ 7.9×1025 25.90 Ni2+ 4.2×1018 18.62 Th4+ 1.6×1023 23.20

39 Titration curve shape depends on formation constant
[MY(n-4)+] [Mn+][Y4-] Mn++Y4- ⇌ MY(n-4)+ =KMY (or Kf) Titration curves for 50 mL of 0.01 mol/L cation solutions at pH 6.0. Ca2+ has smallest formation constant (weakest EDTA complex) Fe3+ has largest formation constant (strongest EDTA complex)

40 Effect of pH Depending on pH, only a certain portion of EDTA is present as Y4-: [Y4-] = aY4- [EDTA]total The value of aY4- decreases with pH pH aY4- at 20°C 1.3×10-23 1 1.9×10-18 2 3.3×10-14 3 2.6×10-11 4 3.8×10-9 5 3.7×10-7 6 2.3×10-5 7 5.0×10-4 8 5.6×10-3 9 5.4×10-2 10 0.36 11 0.85 12 0.98 13 1.00 14 Y4- HY3- H2Y2- H3Y- Speciation curve

41 Effect of pH This leads to an apparent reduction in stability:
Significant for complexes with small K values: Influence of pH on the titration of 0.01 mol/L Ca2+ (50 mL) with 0.01 mol/L EDTA.

42 Minimum pH needed for the satisfactory titration of various cations with EDTA

43 Endpoint recognition in Titrations with EDTA
Indicator for EDTA titrations: Eriochrome Black T Different forms of indicator ( -, 2-, 3- ) have different colours n-3 M + Mn+ + 2H+ (red) H2O + H2In- ⇌ HIn2- + H3O Ka1 = 5×10-7; pKa = 6.3 (red) (blue) H2O + HIn2- ⇌ In3- + H3O+ Ka2 = 2.8×10-12; pKa = 11.6 (orange) Kf for M(In) < Kf for M(EDTA): Solution stays red until no more M is left for complexation with ET pH must be > 6.3 to see colour change to blue

44 Summary Both gravimetric and volumetric methods require an understanding of the underlying Chemistry Gravimetry: absolute method, no standardisation required (but accuracy of a given method must be tested) Titrimetry: careful standardisation is required to achieve satisfactory accuracy

45 Exercise Find and list gravimetric and/or volumetric methods that may be commonly used in a commercial Analytical Lab


Download ppt "Session 2: Fundamentals and Classical methods of quantitative elemental analysis http://bcs.whfreeman.com/qca7e http://www.good-weighing-practice.com/gwp/proper-weighing."

Similar presentations


Ads by Google