Damaca MRI 508 anaesthesia machine

Similar presentations


Presentation on theme: "Damaca MRI 508 anaesthesia machine"— Presentation transcript:

1 Damaca MRI 508 anaesthesia machine

2 Gauss and Tesla Gauss(G) and Tesla(T) are directly interrelated
1T equals 10,000G 0,0001T equals 1G Tesla represents the magnetic strength at the center of the MR scanner Gauss is typically used when expressing the strength of the magnetic field surrounding the MR scanner

3 High Gauss level equipment
Major advantage of equipment tolerating a high Gauss level is that equipment can be placed close to diagnostic equipment such as scanner and new MR scanners become more and more powerful (=> 7T) Practical Operating the anaesthesia machine close to the scanner Safety No interference with diagnostic images No jamming of ventilator functions (halt) No attraction of anaesthesia machine to the scanner

4 Feature – benefits - gains
3.0 Tesla approved Working together with state-of-the-art scanners Better diagnostics Tested up to 2,000 Gauss Work close to scanner Patient safety Scanner safety Patient observation/monitoring Same user interface throughout Siesta i series Dual use Time saving No specialised persons (education) Minimising risks of mis-use High performance anesthesia ventilator 7 ventilation modes Broad patient range IBS Integrated Breathing System Integrated design Patient safety, no leakage problems Easy cleaning procedure Low life time costs (TCO) Certificate from leading MR scanner manufacturers Tested with leading MRI scanners in an MRI environment Can be used together with leading scanner manufactures

5 Dameca MRI 508 — intuitive and easy-to-use
Feature Benefits 3.0 Tesla approved Working together with state-of-the-art scanners Better diagnostics Tested up to 2,000 Gauss Work close to scanner Patient & Scanner safety Patient observation/monitoring Slim design Small footprint – occupies a minimum of space Same user interface throughout Siesta i series Dual use Time saving No specialised persons (education) Minimising risks of mis-use High performance anesthesia ventilator 8 ventilation modes incl. PRVT Broad patient range Same as the well-known Siesta i Whispa Integrated Breathing System (IBS) Patient safety, no leakage problems Easy cleaning procedure Low total costs of ownership (TCO) Low running costs Certificate from leading MR scanner manufacturers Can be used together with a very broad range of MRI scanners

6 Dameca MRI 508 ventilation modes
Automatic Manual VCV — Volume Controlled Ventilation PCV — Pressure Controlled Ventilation SIMV — Synchronised Intermittent Mandatory Ventilation PSV — Pressure Support Ventilation PRVT — Pressure Regulated Volume Target VSV — Volume Support Ventilation Manual Spontanous

7 Dameca MRI 508 Machine Dimensions HxWxD i mm Patient monitor mounting
1510x760x670 Wheels/Breaks Twins - Interlock Drawers No Patient monitor mounting Brand independent Self test (time) 2-3min Battery operation time 60min AGS system 30-40L/min (Built-in) Screen 8" TFT Colour Patient System Description IBS (integrated breathing system) Mounting Left side Insp/Exp. valves flaps Vertical Disassembly for cleaning In 30sec. Disassembly without tools YES Freshgas introduced after insp. value Autoclavable Absurber capacity (disposable) 900gr. System compatible with standard anaesthesia machines Can be used on Siesta i Breasy, Siesta i Whispa and Siesta i TS Ventilator Type Microprocessor control Ventilation modes 8 VCV, PCV, SIMV, PSV, VSV, PRVT Manual spontaneous Vaporiser System Back bar for two Selectatec (German and Bristish standard) with interlock safety mechanism. Ventilation meassurements Tidal Volume Adult ml Paediatric ml Minut Volume 0.2-60L/m PEEP Electronic 4-20hpa Plateau 0-70% Measured patient compliance VCV, PCV, PRVT, SIMV Measured inspired 02 % Yes Spirometry Loop Respiration rate 4-80bpm I:E ratio 3:1 to 1:99 Inspiratory pressure PCV PEEP hpa PSV PEEP hpa Fresh Gas Flow Flow 0-20 L/min Rotameter Minoxyguard 3 tubes 5 tubes (option)

8 MRI 508 vs. competition Dameca MRI 508 Dräger Fabius MRI Aestiva/5 MRI

9 MRI market comparison & positioning
Features Dameca MRI 508 Dräger Fabius MRI GE Aestiva/5 MRI Gauss Tested up to 2,000 Rated up to 1,000 400 300 Tesla 3.0 Patient system compatibility Same as the Siesta i series Only Fabius series Only Aestiva User interface Integrated colour monitor/display 8” 6.5” External display Ventilation modes 8 adults / infants / neonates 6 adults / infants / neonates 5 adults / infants / neonates PRVT option yes No Gauss alarm Special audible & visual ventilator alarm Yes incorporated in the alarm button Yes no

10 Philips - MRI compatibility statement
Tested at the Philips MR scanner test facility in Best with a good result Tested with 1.0T to 3.0T scanners Dameca has tested the machine in addition with a number of different types of scanners

11 MRI safety space

12 MRI safety space